Contec CMS-50f User manual

After
setting
,
short
press
the
button
to
select
“
Exit
”
,
then
long
press
the
button
to
exit
the
system
menu
and
return
to
main
menu
.
5.3
.
5
Bluetooth
setting
Under
main
menu
,
short
press
the
button
to
select
“
Bluetooth
”
,
then
long
press
the
button
to
enter
its
selection
interface
as
shown
in
Figure
10
and
Figure
11
.
When
the
Bluetooth
is
“
ON
’
,
if
there
is
no
data
to
be
transmitted
,
the
Bluelooih
will
turn
off
automatically
after
a
few
seconds
.
Note
:
under
data
transmitting
state
by
Bluetooth
,
the
Bluetooth
can
not
be
turned
off
.
Table
4
:
cathode
.
)
Guidance
and
manufacturer
'
s
declaration
-
electromagnetic
Immunity
Battery
life
Charge
and
discharge
:
no
less
than
500
times
Recycling
garbage
WEEE
(
2012
/
19
/
EU
)
The
[
Code
SI
]
is
intended
for
use
in
the
electromagnetic
environment
specified
below
.
The
customer
or
the
USB
Output
voltage
:
DC
5
V
Output
current
:
1000
mA
user
of
the
Pulse
Oximeter
should
assure
that
i
(
is
used
in
such
an
environment
Adapter
specification
I
MM
UN
I
Test
Freque
Band
Modul
alien
b
)
TY
Dimensions
and
Weight
Modulation
Distance
Battery
anode
Battery
cathode
a
)
Service
a
)
TEST
LEVEL
(
V
/
m
)
Dimension
61
mm
(
L
)
>
:
56
mni
(
W
)
x
24
mm
(
H
)
b
)
(
m
)
ncy
(
MHz
)
(
W
)
(
MHz
)
Weight
About
60
g
(
including
a
lithium
battery
)
It
means
this
pulse
is
protected
against
harmful
effects
of
dripping
water
when
tilled
at
15
°
.
Pulse
modulation
Turn
Off
81
?
Yes
E
3
Turn
On
BT
?
IP
22
RST
RESET
hole
Yi
'
h
Q
380
Note
1
:
the
claims
of
SpC
>
2
accuracy
shall
be
supported
by
clinical
study
measurements
taken
over
the
full
range
.
By
artificial
inducing
,
get
the
stable
oxygen
level
to
the
range
of
70
%
to
100
%
SpO
:
.
compare
the
Sp
(
)
:
values
collected
by
the
secondary
standard
pulse
oximeter
equipment
and
the
tested
equipment
al
the
used
for
(
he
accuracy
analysis
.
(
li
is
applicable
for
the
probes
385
TETRA
400
K
8
0
,
3
27
—
390
b
)
18
Hz
FMc
)
±
5
kHz
Humidity
limitation
Temperature
limitation
time
,
to
form
paired
data
,
which
430
GMRS
460
,
FRS
460
equipped
.
450
2
0
,
3
28
-
470
deviation
1
kHz
sine
II
There
are
12
healthy
volunteers
(
male
:
6
.
female
:
6
;
age
:
18
-
50
;
skin
color
:
black
:
2
,
light
:
8
,
while
:
2
)
data
in
the
clinical
report
.
Note
2
:
because
pulse
oximeter
equipment
measurements
are
statistically
distributed
,
only
about
hvo
-
thirds
of
pulse
oximeter
equipment
measurements
can
be
expected
to
fall
within
土
Amis
of
the
value
measured
by
a
CO
-
OXIMETER
.
Figure
10
Bluetooth
“
ON
”
interface
Figure
11
Bluetooth
“
OFF
”
interface
Atmospheric
pressure
limitation
0
Radiated
This
way
up
5.3
.
6
Shutdown
Under
the
main
menu
interface
,
short
press
the
button
to
select
“
Power
off
1
’
,
then
long
press
the
button
to
turn
off
the
device
.
710
Pulse
modulation
RF
704
LTE
Band
IEC
610
00
-
4
-
3
(
Test
sped
flea
tions
for
ENCLO
SURE
PORT
1
MMUN
ITYto
745
02
0
,
3
9
r
-
787
13
,
17
b
)
5.3
.
7
E
\
il
main
menu
780
Under
main
menu
,
short
press
the
button
to
select
“
Exit
”
,
then
long
press
the
button
to
exit
the
main
menu
and
return
to
the
measurement
interface
.
5.4
Data
upload
5.4
.
1
Wired
transmission
Connect
the
device
to
computer
by
the
USB
cable
,
upload
the
data
after
connecting
with
the
PC
software
properly
,
refer
to
“
Software
operating
instruction
”
for
details
.
Note
:
Ihe
PC
software
can
be
downloaded
from
our
official
website
.
5.4
.
2
Bluetooth
wired
transmission
Turn
for
d
5.5
Charging
Power
adapter
can
be
selected
to
charge
for
Ihe
device
.
It
indicates
that
the
device
is
charging
when
the
indicator
is
orange
,
the
charging
is
finished
when
the
indicator
turns
to
green
.
5.6
Reset
Use
a
pointed
and
hard
object
(
for
example
,
a
paper
clip
)
to
press
Ihe
reset
button
inside
of
the
RESET
hole
,
to
reset
the
device
.
217
Hz
Note
3
:
Patient
simulator
has
been
used
to
verily
the
pulse
rate
accuracy
,
it
is
stated
as
the
root
-
mean
-
square
difference
between
the
PR
measurement
value
and
the
value
set
by
simulator
.
Note
4
:
percentage
modulation
of
infrared
signal
as
the
indication
of
pulsating
signal
strength
,
patieut
simulator
has
been
used
to
verify
ils
accuracy
under
conditions
of
low
perfusion
.
SpO
?
and
PR
values
are
different
due
to
low
signal
conditions
,
compare
Ihcm
with
the
known
SpO
:
and
PR
values
of
input
signal
.
Note
5
:
optical
sensors
as
the
light
-
emitting
components
,
will
affect
other
medical
devices
applied
the
wavelength
range
.
The
information
may
be
useful
for
the
clinicians
who
carry
out
the
optical
treatment
.
For
example
,
photodynamic
therapy
operated
by
clinician
.
810
GSM
800
/
900
,
TETRA
870
Close
the
sound
prompt
Pulse
modulation
800
800
,
2
0
,
3
28
b
)
O
iDEN
820
,
CDMA
850
,
LTE
Band
5
930
Open
the
sound
prompt
Pause
Ihe
sound
prompt
18
Hz
ou
the
device
Bluetooth
and
the
PC
software
to
upload
data
,
refer
to
“
Software
operating
instruction
'
d x
etails
.
0
-
山
Close
ihe
PR
sound
Menu
/
Power
button
RF
1720
GSM
1800
;
CDMA
1900
;
GSM
1900
;
DECT
;
LTE
Band
1
.
3
,
4
,
25
;
UMTS
wireless
1845
comm
u
n
ications
equipme
Pulse
Recyclable
Open
the
PR
sound
1700
modulation
2
0
,
3
28
b
)
-
1990
Prompt
condition
delay
Prompt
signal
generation
delay
State
1970
nt
)
Bluetooth
icon
(
Bluetooth
wired
device
)
217
Hz
Manufaclure
Dale
Low
vo
I
tag
cprompt
20
ms
Is
6.1
Cleaning
and
Disinfecting
The
device
must
be
turned
off
before
cleaning
,
and
it
should
not
be
immersed
into
liquid
.
Please
take
out
the
internal
battery
before
cleaning
,
do
not
immerse
it
into
liquid
.
Use
75
%
alcohol
to
wipe
the
device
enclosure
,
and
use
liquid
soap
or
isopropanol
to
wipe
the
watchband
for
disinfection
,
nature
dry
or
clean
it
with
clean
and
soft
cloth
.
Do
noJ
spray
any
liquid
on
the
device
directly
,
and
avoid
liquid
penetrating
into
the
device
.
6.2
IVIainlcnancc
A
.
Check
the
main
unit
and
all
accessories
periodically
to
make
sure
that
there
is
no
visible
damage
that
may
affect
user
’
s
safety
and
monitoring
performance
.
It
is
recommended
lhai
the
device
should
be
inspected
weekly
at
least
.
When
there
is
obvious
damage
,
stop
using
it
.
B
.
Please
clean
and
disinfect
the
device
be
fore
/
after
using
it
according
to
the
User
Manual
(
6.1
)
.
C
.
Plcasc
charge
the
battery
in
lime
when
low
battery
appears
.
D
.
Recharge
the
battery
soon
after
over
-
discharge
.
The
device
should
be
recharged
every
three
months
when
it
is
not
used
for
some
time
.
It
can
extend
the
battery
life
following
this
guidance
.
E
.
The
device
need
not
to
be
calibrated
during
maintenance
.
6.3
Transportation
and
Storage
A
.
The
packed
device
can
be
transported
by
ordinary
conveyance
or
according
io
transport
contract
.
During
traiLsporlation
,
avoid
strong
shock
,
vibration
and
splashing
with
rain
or
snow
,
and
it
can
not
be
material
.
B
.
The
packed
device
should
be
stored
in
room
with
no
corrosive
gases
and
good
vcnlilation
.
Temperature
:
-
40
°
C
~
60
°
C
;
Humidity
:
<
95
SpO
:
prompt
330
ms
20
ms
Sensor
Off
The
probe
is
disconnected
.
Finger
Out
The
finger
is
not
inserted
.
Bluetooth
,
WLAN
,
802.11
b
/
g
/
n
,
RFID
2450
,
LTE
Band
7
Pulse
rate
prompt
330
ms
20
ms
Pulse
modulation
Probe
error
prompt
16
ms
20
ms
Sensor
Fault
2400
-
2570
Recording
Probe
failure
R
*
2450
2
0
,
3
28
b
)
217
Hz
Material
code
Batch
No
.
P
/
N
LOT
EMC
1
.
The
finger
clip
falls
off
Pulse
modulation
5240
Tabic
1
:
WLAN
802.11
5100
5500
0
-
2
0
,
3
9
2
.
Probe
error
3
.
Signal
inadequacy
indicator
Alarm
inhibit
Guidance
and
manufacturer
’
s
declaration
-
electromagnetic
emission
-
5800
b
)
a
/
n
5785
The
Pulse
Oximeter
is
intended
for
use
in
the
electromagnetic
environment
specified
below
.
The
purchaser
or
Ihe
user
of
the
device
should
assure
that
it
is
used
in
such
environment
.
217
Hz
NOTE
If
necessary
to
achieve
the
IMMUNITY
TEST
LEVEL
,
the
distance
between
the
transmitting
antenna
and
the
ME
EQUIPMENT
or
ME
SYSTEM
may
be
reduced
to
I
pcnnitlcd
by
IEC
61000
-
4
-
3
.
Note
:
Your
device
may
not
contain
all
the
following
symbols
.
The
I
m
test
distance
is
sPo
2
Emission
test
Compliance
|
see
note
1
]
RF
emissions
C
1
SPR
11
Group
I
a
)
For
some
services
,
only
tlic
uplink
frequencies
are
included
.
b
)
The
carrier
shall
be
modulated
using
a
50
%
duly
cycle
square
wave
signal
.
c
)
As
an
alternative
to
FM
modulation
,
50
%
pulse
modulation
at
18
Hz
may
be
used
because
while
it
does
not
rq
^
rcscnt
actual
modulation
,
it
would
be
worst
case
.
transported
mixed
with
toxic
,
harmful
,
Display
range
0
%
-
99
%
Table
2
:
Measured
range
0
%
~
100
%
Guidance
and
manufacturer
’
s
dcclaratiou
-
clcctromagnctic
immunity
70
%
-
100
%
:
士
2
%
;
The
Pulse
Oximeter
is
intended
for
use
in
the
electromagnetic
environment
specified
below
.
The
purchaser
or
Ihe
user
of
ihe
Pulse
Oximclcr
should
assure
lhat
it
is
used
in
such
environment
.
Accuracy
|
see
note
2
]
The
MANUFACTURER
should
consider
reducing
the
minimum
separation
distance
,
based
on
RISK
MANAGEMENT
,
and
using
higher
IMMUNITY
TEST
LEVELS
that
reduced
LEVELS
shall
be
ailculated
using
the
following
equation
:
0
%
-
69
%
:
unspecified
.
Trouble
Possible
Reason
Solution
appropriate
for
the
are
1
.
The
finger
is
not
properly
positioned
.
2
.
The
finger
is
shaking
or
the
user
is
moving
.
3
.
The
device
is
not
used
in
environment
required
by
the
manual
.
4
.
Thc
device
works
abnormally
.
Resolution
1
%
.
separation
distance
.
Minimum
separation
distances
for
higher
IMMUNITY
TEST
1
.
Please
insert
the
finger
properly
and
again
.
2
.
Lcl
the
user
keep
calm
3
.
Please
use
the
device
in
normal
Immunity
test
IEC
60601
test
level
Compliance
level
measure
PR
E
=
M
Electrostatic
discharge
(
ESD
)
IEC
61000
-
4
-
2
±
8
kV
contact
±
8
kV
contact
The
values
can
not
be
displayed
normally
or
stably
Display
range
30
bpm
250
bpra
d
土
15
k
V
air
土
15
kV
air
Where
P
is
the
maximum
power
in
W
,
d
is
Ihe
minimum
separation
distance
in
ra
,
and
E
is
the
IMMUNITY
TEST
LEVEL
in
V
/
m
.
Measured
range
30
bpm
~
250
bpm
Power
frequency
(
50
/
60
Hz
)
magnetic
field
IEC
61000
-
4
-
8
4
.
Please
contact
ihe
after
-
sales
.
±
2
bpm
during
the
pulse
rate
range
of
30
bpm
99
bpm
and
±
2
%
during
the
pulse
rate
range
of
100
bpm
-
250
bpm
.
30
A
/
m
30
A
/
m
Accuracy
|
see
note
31
1
.
Low
battery
or
the
battery
is
drained
away
.
2
.
The
device
works
abnormally
.
The
device
can
not
be
1
.
Please
charge
the
battery
.
2
.
Please
contact
the
after
-
sales
.
Resolution
turned
on
arning
Don
'
t
near
active
HFSURGICAL
EQUIPMENT
and
the
KF
shielded
room
of
an
ME
SYSTEM
for
magnetic
resonance
imaging
,
where
the
intensity
of
EM
DISTURBANCES
is
high
.
Use
of
this
equipment
adjacent
to
or
stacked
with
other
equipment
should
be
avoided
because
it
could
result
in
improper
operation
.
If
such
use
is
necessary
,
this
equipment
and
the
other
equipment
should
be
observed
to
verify
that
they
are
operating
normally
.
Use
of
accessories
,
transducers
and
cables
other
than
those
specified
or
provided
by
the
manufacturer
of
this
equipment
could
result
in
increased
electromagnetic
emissions
or
decreased
electromagnetic
immunity
of
this
equipment
and
result
in
improper
operation
.
”
Portable
RF
communications
equipment
(
including
peripherals
such
as
antenna
cables
and
external
antennas
)
should
be
used
no
closer
than
30
cm
(
12
inches
)
lo
any
part
of
the
device
including
cables
specified
by
Che
manufacturer
.
Otherwise
,
degradation
of
Ihe
performance
of
this
equipment
could
result
.
Active
medical
devices
are
subject
to
special
EMC
precautions
and
Ihey
must
be
installed
and
used
in
accordance
、
vilh
these
guidelines
.
Table
3
:
Low
perfusion
0.4
%
:
1
.
The
device
enters
into
the
energy
saving
mode
.
2
.
Low
battery
.
I
.
Please
conlacl
(
he
after
-
sales
.
2
.
Please
charge
the
battery
3
.
Please
conlacl
(
lie
after
-
sales
.
Guidance
and
manufacturer
’
s
declaration
—
electromagnetic
irmnunity
Accuracy
under
low
perfusion
[
see
note
41
The
display
disappears
suddenly
.
The
Pulse
Oximeter
is
intended
for
use
in
the
electromagnetic
environment
specified
below
.
The
customer
the
user
of
the
Pulse
Oximeter
should
assure
that
it
is
used
in
such
environment
.
PR
:
±
2
bpm
during
the
pulse
rate
range
of
30
bpm
-
99
bpm
and
士
2
%
during
the
pulse
rate
range
of
100
bpm
250
bpm
.
3
.
The
device
works
abnormally
.
The
device
can
not
be
used
for
full
time
after
charge
.
1
.
The
battery
is
not
charged
fully
.
2
.
The
device
works
abnormally
.
1
.
Please
charge
the
battery
.
2
.
Please
contact
the
after
-
sales
.
IEC
60601
test
level
Compliance
level
Immunity
test
der
normal
and
ambient
light
conditions
,
the
SpO
?
deviation
Light
interference
%
Radiated
RF
10
V
/
m
80
MHz
-
2.7
GHz
10
V
/
m
80
MHz
-
2.7
GHz
The
battery
can
not
be
full
charged
even
after
10
hours
charging
lime
.
Continuous
bar
-
graph
display
,
the
higher
display
indicates
the
stronger
pulse
.
IEC
61000
-
4
-
3
Pulse
Intensity
The
battery
works
abnormally
.
Please
contact
the
after
-
sales
.
NOTE
1
Al
80
MHz
and
800
MHz
,
the
higher
frequency
range
applies
.
NOTE
2
These
guidelines
may
not
apply
in
all
situations
.
Electromagnetic
propagation
is
affected
by
absorption
and
reflection
from
structures
,
objects
and
people
.
Upper
and
lower
limit
of
measured
values
1
.
The
device
is
not
operated
according
lo
the
manual
.
2
.
The
device
works
abnormally
.
1
.
Please
operate
the
device
according
to
the
manual
.
2
.
Please
contact
the
after
-
sales
.
SpO
:
0
%
100
%
The
data
can
not
be
stored
.
Field
strengths
from
fixed
transmitters
,
such
as
base
stations
for
radio
(
cellular
/
cordless
)
telephones
and
land
mobile
radios
,
amateur
radio
,
AM
and
FM
radio
broadcast
and
TV
broadcast
cannot
be
predicted
theoretically
with
accuracy
.
To
assess
the
electromagnetic
em
ironment
due
to
fixed
RF
transmitters
,
ail
electromagnetic
site
survey
should
be
considered
.
If
the
measured
field
strength
in
the
location
in
which
the
Pulse
Oximeter
is
used
exceeds
the
applicable
RF
compliance
level
above
,
the
Pulse
Oximeter
should
be
observed
to
verify
normal
operation
.
If
abnormal
performance
is
observed
,
additional
measures
may
be
necessary
,
such
as
reorienting
or
relocating
the
Pulse
Oximeter
.
h
Over
the
frequency
range
150
KHz
to
80
MHz
,
field
strengths
should
be
less
than
3
V
/
m
.
PR
0
bpm
-
254
bpm
a
A
Optical
sensor
|
see
note
51
Note
:
Red
light
Wavelength
:
about
660
nm
,
optical
output
power
:
<
6.65
mW
When
the
device
is
disturbed
,
(
he
data
measured
may
fluctuate
,
please
measure
repeatedly
or
in
another
environment
to
ensure
its
accuracy
.
Symbols
Meaning
Symbols
Meaning
Infrared
light
Wavelength
:
about
905
nm
.
optical
oulput
power
:
<
6.75
mW
©
Up
to
99
group
of
data
under
auto
mode
,
total
duration
docs
not
exceed
72
hours
.
Up
to
24
-
liour
data
under
manual
mode
.
Caution
,
consult
accompanying
documents
PRbpm
Pulse
rate
(
bpm
)
Memory
Pulse
oxygen
saturation
(
%
)
Safety
class
Internally
powered
equipment
,
type
BF
applied
pari
International
Protection
1
P
22
i
Fully
charged
Manufacturer
Working
voltage
DC
3.6
V
-
4.2
V
Working
current
<
100
mA
A
rechargeable
lithium
battery
(
3.7
V
)
(
The
red
battery
denotes
anode
,
the
black
wire
on
the
ba
on
the
denotes
Serial
mimhei
.
Use
-
by
date
wire
ncry
Power
supply
Bluetooth
Specification
Working
frequency
:
2402
MHz
2480
MHz
Modulation
mode
:
GFSK
Transmitting
power
:-6
dBm
,
+
4
dBm
Receiving
sensitivity
:
-
93
dBm

FCC Caution
§ 15.19 Labeling requirements.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept
any interference received, including interference that may cause undesired operation.
§ 15.21 Information to user.
Any Changes or modifications not expressly approved by the party responsible for compliance
could void the user's authority to operate the equipment.
§ 15.105 Information to the user.
Note: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates
uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
-Consult the dealer or an experienced radio/TV technician for help.
The device has been evaluated to meet general RF exposure requirement. The device can be used
in portable/mobile exposure condition without restriction
Other manuals for CMS-50f
3
Other Contec Medical Equipment manuals

Contec
Contec CMS-50E User manual

Contec
Contec CMS60C User manual

Contec
Contec CMS-50D Plus User manual

Contec
Contec ABPM 50 Reference guide

Contec
Contec ECG 100G User manual

Contec
Contec CMS50DL User manual

Contec
Contec CMS50DL User manual

Contec
Contec CMS50M User manual

Contec
Contec CMS60C User manual

Contec
Contec PM-60A User manual

Contec
Contec CMS800G User manual

Contec
Contec CMS 50IW User manual

Contec
Contec ECG1200G User manual

Contec
Contec CMS80A Instruction Manual

Contec
Contec CMS60C User manual

Contec
Contec CMS50QB Instruction Manual

Contec
Contec ESA0008 User manual

Contec
Contec CMS8000 User manual

Contec
Contec CMS60D User manual

Contec
Contec CMS50D User manual