manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Contec
  6. •
  7. Medical Equipment
  8. •
  9. Contec CMS50M User manual

Contec CMS50M User manual

tOO
!l{l
85
1m
AV01'l1.llilt
Sa02
and
Sp02
Pules rate Bargraph Dispaly
•
Figure 3Bland-Altman plot
75
10
Figure
5.
''"'
8
...
Of;
I7
~
'a.
6
II'> Ii
((;
g4
'"
3
~
....
2
....
...
1
Q
~
N0
~
-]
I,W
-z
<;,
'3
l>!,;
~i'ii
'4
"
~J.
""5
0
-6
"'~
-1
...
5
·8
<It>
II
,"-----
Figure 4. Front View
•
Pules rate
---
......
~8
88
The display Sp02
------1++++88
~-t+t+!t------
Figure 2
SpOz
regression plot
~
One hanging rope;
~
Two batteries (optional);
~
One User Manual.
5
Accessories.
.
6Installation
6.1 View
of
the
Front
Panel
7
Operating
Guide .
6.2
Battery
Step
1.
Insert the two
AAA
size batteries properly in the right direction.
Step 2. Put the cover.
~lease
take
care
when
you
insert
the
batteries for
the
improper
insertion
may
damage
the
device.
6.3
Mounting
the
Hanging
Rope
Step
1.
Put the end
of
the rope through·the hole.
Step 2. Put another end
of
the rope through the first one and then tighten it.
4Technical Specifications
\
Glow and Infrared-ray
Receipt Tube
instrument is: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning &Recording
Technology~
so
that two beams
of
different wavelength
of
lights can be focused onto human nail tip through perspective clamp finger-type sensor. Then measured signal
can be obtained by aphotosensitive element, information acquired through which will be shown on screen through treatment in electronic circuits and
microprocessor.
7.4 Press the switch button once on front panel.
7.5 Do not shake the fmger and keep the user at ease during the process. Meanwhile, human body is not recommended in movement status.
7.6 Get the information directly from screen display.
7.7 In boot-strap state,press button ,and the device is reset.
~
Fingernails
and
the
luminescent
tube
should
be
on
the
same side.
Glow and Infrared-ray
Emission Tube
3.2
Caution
1.
The finger should be placed properly (see the attached illustration
of
this manual ,Figure 7), or else it may cause inaccurate measurement.
2. The
SpOz
sensor and photoelectric receiving tube should be arranged in away with the subject's arteriole in aposition there between.
3.
The SpOz sensor should not be used at alocation or limb tied with arterial canal or blood pressure cufforreceiving intravenous injection.
4.
Make sure the optical path
is
free from any optical obstacles like rubberized fabric.
5.
Excessive ambient light may affect the measuring result. It includes fluorescent
lamp~
dual ruby
light~
infrared heater, direct sunlight and etc.
6.
Strenuous action
of
the subject or extreme electrosurgical interference may also affect the accuracy.
7.
User can not use enamel or other makeup.
3.3 Clinical Restrictions
1.
As the measure is taken on the basis
of
arteriole pulse, substantial pulsating blood flow
of
subject is required. For asubject with weak pulse due to
shock, low ambientlbody temperature, major bleeding, or use
of
vascular contracting drug, the
SpOz
waveform (PLETH) will decrease. In this case,
the measurement will be more sensitive to interference.
2. For those with asubstantial amount
of
staining dilution drug (such as methylene blue, indigo green and acid indigo blue), or carbon monoxide
hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and some with icterus problem, the SpOz determination by this monitor
may be inaccurate.
3.
The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be amajor factor blamed for serious error
of
SpOz measure.
4. As the SpOz value serves as areference value for judgement
of
anemic anoxia and toxic
anoxia~
some users with serious anemia may also report
good SpOz measurement.
8Re airing
and
Maintenance
7.1
Insert the two batteries properly
to
the direction, and then put the cover.
7.2 Open the
clip.
7.3 Let the user's finger put into the rubber cushions
of
the clip (make sure the finger is in the right position), and
then!
clip the finger. As shown in
Figure 5
1)
Display
Format:
Digital tube Display;
SpOz
Measuring
Range: 0% -100%;
Pulse
Rate
Measuring
Range: 30 bpm -250 bpm;
Pulse Intensity Display: columniation display
2) Power Requirements: 2x1.5V
AAA
alkaline battery, adaptable range: 2.6V-3.6V.
3) Power
Consumption:
Smaller than
25
rnA.
4) Resolution: 1% for
SpOz
and 1bpm for Pulse Rate.
5)
Measurement
Accuracy: ±2% in stage
of
70%-100% SpOz, and meaningless when stage being smaller than 70%. ±2 bpm or±2% (select larger) for
Pulse Rate. Clinical Trial :SpOz regression plot &Bland-Altman plot,Refer to Figure 2&Figure
3.
6)
Measurement
Performance
in
Weak
Filling Condition: SpOz and pulse rate can be shown correctly when pulse-filling ratio is 0.4%.
SpOz
error
is
±4%, pulse rate error is ±2 bpm or ±2% (select larger).
7) Resistance to
surrounding
light: The deviation between the value measured in the condition
of
man-made light, indoor natural light and that
of
darkroom
is
less than
±l
%.
8)
It
is
equipped with aswitch function. The Oxirpeter can be powered offwhen the finger
is
off
the oximeter within 5seconds.
9) Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm, 6.75mW)
Figure
1.
Operating Principle
~
Please change the batteries when the low-voltage displayed on the screen.
~
Please clean the surface
of
the device before using. Wipe the device with medical alcohol first, and then let it dry in air or clean it by dry clean fabric.
~
Using the medical alcohol to disinfect the product after use, prevent from cross infection for next time use.
~
Please take out the batteries
if
the oximeter is not in use for along time.
~
The packed device can be transported by ordinary conveyance or according to transport contract.The device can not be transported mixed with toxic,
harmful, corrosive material.
~
The best storage environment
of
the device is -40°C to
60°C
ambient temperature and not higher than 95% relative humidity.
~
Users are advised to calibrate the device termly (or according to the calibrating program
of
hospital). It also can be performed at the state-appointed
agent orjust contact us for calibration.
&.
High-pressure sterilization
cannot
be
used on
the
device.
&.
Do
not
immerse
the
device
in
liquid.
&.
It
is recommended
that
the
device should
be
kept
in a
dry
environment. Humidity
may
reduce
the
useful life
of
the
device,
or
even
damage
File No.:CMS2.782.395(OCH)ESS/1.0
Release Date: June 2016
1.4.01.01.393
Address: No.112 Qinhuang West Street, Economic &Technical DevelopmentZone,
Qinhuangdao, Hebei Province, PEOPLE'S REPUBLIC OF CHINA
Tel:
0086-335-8015430
Fax: 0086-335-8015588
Technical support: 0086-335-8015431
E-mail:[email protected]
[email protected]
Website: http://www.contecmed.com
CONTEC™
Contec
Medical
Systems
Co.,
Ltd.
Instructions to User
recommended to
be
used
under
this circumstance.
The pulse oxygen saturation is the percentage
of
HbOz in the total Hb in the blood, so-called the
Oz
concentration in the blood. It is an important
bio-parameter for the respiration. For the purpose
of
measuring the
SpOz
more easily and accurately, our company developed the Pulse Oximeter. At the
same time, the device can measure the pulse rate simultaneously.
The Pulse Oximeter features in small volume, low power consumption, convenient operation and being portable. It is only necessary for user to put one
of
his fmgers into afingertip photoelectric sensor for diagnosis, and adisplay screen will directly show measured value
of
Hemoglobin Saturation.
2.1 Features
~
Operation
of
the product is simple and convenient.
~
The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient in carrying.
~
Power consumption
of
the product is low and the two originally equipped AAAbatteries can be operated continuously for 24 hours.
~
The product will automatically be powered
off
when no signal is in the product within 5seconds.
~
Low -battery indicator as battery icon flash manner.
2.2
Major
Applications
and
Scope
of
Application
The Pulse Oximeter can be used in measuring pulse oxygen saturation and pulse rate through finger. The product is suitable for family use(It can be used
before or after doing sports, and it is not recommended to use the device during the process
of
doing sports) .
~
The
problem
of
overrating would emerge when
the
user
is suffering from toxicosis which caused
by
carbon
monoxide,
the
device is
not
CMS50M
2Overview
Dear users, thank you very much for purchasing the Pulse Oximeter.
In case
of
modifications and software upgrades, the information contained in this document is subject to change without notice.
The Manual describes, in accordance with the Pulse Oximeter's features and requirements, main structure, functions, specifications, correct methods for
transportation, installation, usage, operation, repair, maintenance and storage, etc. as well as the safety procedures to protect both the user and equipment.
Refer to the respective chapters for details.
Please read the User Manual carefully before using this product. The User Manual which describes the operating procedures should be followed
strictly.Failure to follow the User Manual may cause measuring abnormality, equipment damage and human injury. The manufacturer is NOT responsible
for the safety, reliability and performance issues and any monitoring abnormality, human injury and equipment damage due to users' negligence
of
the
operation instructions. The manufacturer's warranty service does not cover such faults.
Owing to the forthcoming renovation, the specific products you received may not be totally in accordance with the description
of
this User Manual.
We
would sincerely regret for that.
This product can be used repeatedly. The operating life is 3years.
If
you have any questions regarding to the use
of
this product, please call
us
at 1-847-562-1702 Monday-Friday from 8:00 AM to 5:00 PM Eastern Time.
WARNING:
~
Uncomfortable
or
painful feeling may
appear
if
using
the
device
cea~elessly,
especially for
the
the
microcirculation
barrier
users.
It
is
recommended
that
the
sensor should
not
be
applied to
the
same finger for
over
2hours.
r
For
the
special users,
there
should
be
a
more
prudent
inspecting in
the
placing process.
The
device
can
not
be
clipped on
the
edema
and
tender
tissue.
~
The
light (the
infrared
is invisible)
emitted
from
the
device is
harmful
to
the
eyes, so
the
user
and
the
maintenance
man
should
not
stare
at
the
light.
rUser
can
not
use enamel
or
other
makeup.
r
User's
fingernail can
not
be
too long.
rPlease refer to
the
correlative
literature
about
the
clinical restrictions
and
caution.
rThis device is
not
intended
for
treatment.
The User Manual
is
published by our company. All rights reserved.
Pulse
Oximeter
(UJ~~lr
[tMtl~lJillYl~O
1Safety
1.1 Instructions for Safe
Operations
~
Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect user's safety and monitoring
performance about cables and transducers. It is recommended that the device should be inspected once aweek at least. When there is obvious
damage, stop using the oximeter .
~
Necessary maintenance must be performed by qualified engineers
ONLY.
Users are not permitted to maintain it by themselves.
~
The oximeter cannot be used together with devices not specified
in
User's Manual.Only the accessory that appointed or recommendatory by
manufacture can be used with this device.
~
This product is calibrated before leaving factory.
1.2 Warnings
~
Explosive
hazard-DO
NOT use the oximeter in environment with inflammable gas such
as
some ignitable anesthetic agents.
~
The person who is allergic to rubber can not use this device.
~
The disposal
of
scrap instrument and its accessories and packings(including battery, plastic bags, foams and paper boxes) should follow the local
laws and regulations.
~
Please check the packing before use to make sure the device and accessories are totally in accordance with the packing list, or else the device may
have the possibility
of
working abnormally.
~
Please don't measure this device with function test paper for the device's related information.
~
Parts
of
the device that are not serviced or maintained while in use with the user.
};-
WarninKagainst servicing and maintenance while the me equipment is
in
use.
~
No modification
of
this equipment is allowed.
~
The user
is
an intended operator.
~
The probe
of
the device is the applied part.
1.3 Attentions
GKeep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture.
G
If
the oximeter gets wet, please stop operating it.
GWhen it
is
carried from cold environment to warm or humid environment, please do not use it immediately.
GDO NOT operate keys on front panel with sharp materials.
GHigh temperature or high pressure steam disinfection
of
the oximeter is not permitted. Refer to User Manual in the relative chapter for instructions
of
cleaning and disinfection.
GDo not have the oximeter immerged in liquid. When it needs cleaning, please wipe its surface with medical alcohol by soft material. Do not spray
any liquid on the device directly.
GWhen cleaning the device with water, the temperature should be lower than 60°C.
GAs to the fingers which are too thin or too cold, it would probably affect the normal measure
of
the users'
Sp02
and pulse rate, please clip the thick
finger such as thumb and middle finger deeply enough into the probe.
GDo not use the device on infant or neonatal users.
GThe product
is
suitable for adults(Weight should be between 40kg to 1IOkg).
GThe device may not work for all users.
If
you are unable to achieve stable readings, discontinue use.
GThe update period
of
data is less than 5seconds, which
is
changeable according to different individual pulse rate.
G
If
some abnormal conditions appear on the screen during test process, pull out the finger and reinsert to restore normal use.
GThe hanging rope attached the product is made from Non- allergy material,
if
particular group are sensitive to the hanging rope, stop using it. In
addition, pay attention to the use
of
the hanging
rope,
do not wear it around the neck avoiding cause harm to the users.
GThe instrument dose not have low-voltage alarm function, it only shows the low-voltage,please change the battery when the battery energy is used
out.
GWhen the parameter
is
particularly, The instrument dose not have alarm function.Do not use the device in situations where alarms are required.
GBatteries must be removed
if
the device is going to be stored for more than one month, orelse batteries may leak.
GAflexible circuit connects the two parts
of
the device. Do not twist or pull on the connection.
1.4.Indication for Use
The Pulse Oximeter is anon-invasive device intended for the spot-check
of
saturation
of
arterial hemoglobin(Sp02) and the pulse rate
of
adult in home
use environments.This device is not intended for continuous monitoring.The device can be multi-used.Solely for use with sporting and aviation
activities.Intended to monitor heart rate during exercise.
2.3
Environment
Requirements
Storage Environment
a) Temperature
:-40°C~+60°C
b) Relative humidity
::::s
95%
c) Atmospheric pressure
:500hPa~1060hPa
Operating Environment
a) Temperature:
:10°C~40°C
b) Relative Humidity
::::S75%
c) Atmospheric
pressure:700hPa~
1060hPa
3Principle
and
Caution it.
3.1 Principle
of
Measurement
Principle
of
the Oximeter is
as
follows: An experience formula
of
data process is established taking use
of
Lambert Beer Law according to Spectrum
Absorption Characteristics
of
Reductive Hemoglobin (Hb) and Oxyhemoglobin (HbOz) in glow &near-infrared zones. Operation principle
of
the
11
Function
Secification .
10
Ke
of
Symbols
Portable and mobile RF communications equipment
should be used no closer to any part
of
the CMS50M,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency
of
the transmitter.
Recommendoo
separation
distance
Electromagnetic
environment
-
guidance
d=t
~~
]
v'JT
80 MHzto 800 MHz
WherePis the maximum output power rating
of
the
transmitter in watts (W) according to the transmitter
manufacturer and dis the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined
by
an electromagnetic site survey,a should be
less than the compliance level in each frequency range.b
Interference may occur in the vicinity
of
equipment
marked with the following symbol:
d=[
~J.JP
800
MHz
to 2.5
OHz
3V/m
Compliance
level
3V/m
80 MHz to 2.5 GHz
IEC
60601
test
level
Guidance
and
manufacture's
declaration
-electromagnetic
immunity
Guidance
and
manufacture's
declaration
-
electromagnetic
immunity
-
for
aU
EQUIPMENT
and
SYSTEMS
The CMS50Mis intended for use in the electromagnetic environment specified below. The customer
or
the user
of
CMS50M
should assure that it is used in such an environment.
Recommended
separation
distances between
portable
and
mobile
RF
communications
equipment
and
the
EQUIPMENT
or
SYSTEM-
for
EQUIPMENT
or
SYSTEM
tbat
are
not
LIFE-SUPPORTING
Guidance
and
manufacture's
declaration
-electromagnetic
immunity
-
for
EQUIPMENT
and
SYSTEMS
that
are
not
LIFE-SUPPORTING
Immunity
test
RadiatedRF
IEC 61000-4-3
NOTE 1
At
80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected
by
absorption and
reflection from structures, objects and people.
b
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateurradio, AM and FM radio broadcast and
TV
broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fIxed RF transmitters, an electromagnetic site survey should be considered.
If
the
measured field strength in the location in which the CMS50M is used exceeds the applicable RF compliance level above,
the CMS50M should be observed to verify normal operation.
If
abnormal performance is observed, additional measures
may be necessary, such as reorienting
or
relocating the CMS50M.
Overthe frequency range 150 kHz to 80
MHz~
field strengths should be less than 3 V/m.
RFemission
ClassB
The CMS50M
is
suitable for use in all establishments,
CISPR
11
including domestic establishments and those directly
Harmonic emissions N/A connected to the public low-voltage power supply network
IEC 61000-3-2 that supplies buildings used for domestic purposes.
Voltage f1uctuationsl
flicker emissions N/A
IEC 61000-3-3
Guidance
and
manufacture's
declaration
-
electromagnetic
immunity
The CMS50M is intended for use in the electromagnetic environment specified below. The customer
or
the user
of
CMS50M
should assure that it is used in such an environment.
Immunity
test
IEC
60601
test
level
Compliance
level Electromagnetic
environment
-
guidance
Electrostatic discharge ±6
kV
contact ±6 KV contact Floors should be wood, concrete
or
(ESD) ±8
kVair
±8
kVair
ceramic tile.
If
floor are covered with
IEC 61000-4-2 synthetic material, the relative humidity
should be
at
least 30%.
Electrical fast
±2
kV
for power supply N/A Mains power quality should be that
of
a
transientlburst lines typical commercial
or
hospital
IEC 61000-4-4 environment.
Surge ±1 kV differential mode N/A Mains power quality should be that
of
a
IEC 61000-4-5 typical commercial
or
hospital
environment.
Voltage dips, short
<5%Ur
N/A Mains power quality should be that
of
a
interruptions and (>95% dip in Ur) typical commercial
or
hospital
voltage
varia~ions
on
for 0.5 cycle environment.
If
the user
of
the
power supply input CMS50M requires continued
lines
40%Ur
operation during power mains
IEC 61000-4-11 (60% dip in
Ur)
interruptions, it is recommended that the
for 5cycles
CMS50M
be powered from an
70%Ur
uninterruptible power supply
or
a
(30% dip in
Ur)
battery.
for 25 cycles
<5%Ur
(>95% dip in Ur)
for 5sec
Power frequency
3Nm
3Nm
Power frequency magnetic fields
(50/60Hz) sould be at levels characteristic
of
a
Magnetic field typical location in atypical commercial
IEC-61000-4-8
or
hospital environment.
NOTE
Ur
is the a.c. mains voltage prior to application
of
the test level.
Recommended
separation
distances between
portable
and
mobile RF
communications
equipment
and
tbe
CMS50M
TheCMS50M
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer
or
the user
of
the CMS50M canhelp prevent electromagnetic interference
by
maintaining aminimum distance
between portable and mobile RF communications equipment (transmitters) and the CMS50Mas recommended below, according
to the maximum outputpower
of
the communications equipment.
Separation
distance
according
to
frequency
of
transmitter
Rated
maximum
output
(m)
power
of
transmitter
150
kHz
to
80
MHz
80
MHz
to
800
MHz
800
MHz
to
2.5
GHz
(W)
d=[3~~p
d=[~~J~
d=[~JJP
0.01 0.12 0.12 0.23
0.1 0.39 0.37 0.74
I1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
For transmitters rated at amaximum output power not listed above, the recommended separation distance din metres (m) canbe
estimated using the equation applicable to the frequency
of
the transmitter, where Pis the maximum output powerrating
of
the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At
80
MHz
and 800
MHz,
the separation distance for the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Electromagnetism
Compatibility
Guidance and
manufacture's
declaration -
electromagnetic
emissions-
for
all EQUIPMENT and SYSTEMS
Guidance and
manufacture's
declaration -
electromagnetic
emission
The CMS50M is intended for use
in
the electromagnetic environment specified below. The customer
of
the user
of
the
CMS50Mshould assure that it is used
in
such and environment.
Emission
test
Compliance
Electromagnetic
environment
-
guidance
RF
emissions The CMS50M uses
RF
energy only for its internal function.
CISPR
11
Group 1Therefore, its
RF
emissions are very low and are not likely to
cause any interference
in
nearby electronic equipment.
Trouble
PossibleReason
Solution
The
Sp02
and
Pulse
Rate
1.
The fmger is not properly positioned.
1.
Place the fmger properly and
try
again.
can
not
be
displayed 2. The user's Sp02 is too low to be detected. 2. Try again; Go to ahospital for adiagnosis
if
normally
you are sure the device works all right.
The
Sp02
and
Pulse
Rate
1.
The finger is not placed inside deep enough.
1.
Place the finger properly and
try
again.
are
not
displayed
stably
2.
The fmger is shaking or the user is moving. 2. Let the user keep calm
1.
The batteries are drained or almost drained.
1.
Change batteries.
The
device
can
not
be
2. The batteries are not inserted properly. 2. Reinstall batteries.
turned
on
3. The malfunction
of
the device. 3. Please contact the local service center.
1.
The device will power offautomatically when
1.
Normal.
The
display is
off
suddenly
,it gets no signal within 5seconds. 2. Change batteries.
2.
The batteries are almost drained.
Symbol
Description
00
Type BF
(C Refer to instruction manuallbooklet
Sp02%
The pulse oxygen saturation(%)
PRbpm"
Pulse rate (bpm)
IXJ:I
The battery voltage indication is deficient (change the battery in time avoiding the inexact
measure)
------
1.
no finger inserted
2. An indicator
of
signal inadequacy
+battery positive electrode
battery negative electrode;
dPower switch
SN Serial number
~
Alarm inhibit
)t
WEEE (2002/96/EC)
-
IP22 Ingress
of
liquids rarik
C€
0123
This item is compliant with Medical Device Directive 93/42/EEC
of
June 14, 1993, adirective
of
the European Economic Community.
~REPI
EUROPEAN
REPRESENTATIVE
r
..,
MIl
Manufacturer
L.J
r
..,
dManufacture Date
L.J
r!~
Storage and Transport Temperature limitation
~c
.J
r
95%1
¢Storage and Transport Humidity limitation
~
.J
r
106kP-;)
fJ Storage and TransportAtmospheric pressure limitation
E
kPa
.J
rtf
This side UP
L-.J
r!
..,
Fragile, handle with care
L
.J
r
~'
Keep dry
L
.J
0Recyclable
Display
Information
Display
Mode
The Pulse Oxygen Saturation(Sp02) Digital
Pulse Rate(BPM) Digital
Pulse Intensity (bar-graph) Digital bar-graph display
SpO%
Parameter
Specification
Measuring range 0%-100%, (the resolution
is
1%).
Accuracy 70%-100%:±2% ,Below 70% unspecifIed.
Optical Sensor Red light (wavelength is 660nm)
Infrared (wavelength is 880nm)
Pulse
Parameter
Specification
Measuring range 30bpm-250bpm (the resolution is 1bpm)
Accuracy ±2bpm or +2% select larger
Pulse
Intensity
Range Continuous bar-graph display, the higher display indicate the stronger pulse.
Battery
Requirement
I.5V
(AAA size) alkaline batteries x2or rechargeable battery
Battery
Useful Life
Two batteries can work continually for 24 hours
Dimensions
and
Weight
Dimensions 57(L) X34(W) X31(H)
mm
Weight About 50g (with the batteries)
9
Troubleshooting
Appendix:

Other Contec Medical Equipment manuals

Contec CMS50D User manual

Contec

Contec CMS50D User manual

Contec MDD93EEC Instruction Manual

Contec

Contec MDD93EEC Instruction Manual

Contec CMS8000 User manual

Contec

Contec CMS8000 User manual

Contec CMS 50IW User manual

Contec

Contec CMS 50IW User manual

Contec PM-60A User manual

Contec

Contec PM-60A User manual

Contec CMS-50f User manual

Contec

Contec CMS-50f User manual

Contec CMS-50D Plus User manual

Contec

Contec CMS-50D Plus User manual

Contec ECG1200G User manual

Contec

Contec ECG1200G User manual

Contec CMS50DL1 User manual

Contec

Contec CMS50DL1 User manual

Contec CMS50EW User manual

Contec

Contec CMS50EW User manual

Contec CMS-50E User manual

Contec

Contec CMS-50E User manual

Contec ESA0008 User manual

Contec

Contec ESA0008 User manual

Contec 300G User manual

Contec

Contec 300G User manual

Contec CMS-50f User manual

Contec

Contec CMS-50f User manual

Contec ABPM 50 Reference guide

Contec

Contec ABPM 50 Reference guide

Contec CMS50EW User manual

Contec

Contec CMS50EW User manual

Contec KT88-3200 User manual

Contec

Contec KT88-3200 User manual

Contec CMS50D User manual

Contec

Contec CMS50D User manual

Contec CMS50QB Instruction Manual

Contec

Contec CMS50QB Instruction Manual

Contec CMS60C User manual

Contec

Contec CMS60C User manual

Contec CMS50DL User manual

Contec

Contec CMS50DL User manual

Contec CMS50DL User manual

Contec

Contec CMS50DL User manual

Contec CMS800G User manual

Contec

Contec CMS800G User manual

Contec CMS50DA User manual

Contec

Contec CMS50DA User manual

Popular Medical Equipment manuals by other brands

Sutter Bison Series manual

Sutter

Sutter Bison Series manual

Otivio FlowOx 2.0 quick guide

Otivio

Otivio FlowOx 2.0 quick guide

laerdal SimMom Directions for use

laerdal

laerdal SimMom Directions for use

Sedecal General Purpose Rad System Operation

Sedecal

Sedecal General Purpose Rad System Operation

Merit Medical Achieve Instructions for use

Merit Medical

Merit Medical Achieve Instructions for use

HeartSine samaritan PAD User handbook

HeartSine

HeartSine samaritan PAD User handbook

LINET MULTICARE 2015 Service manual

LINET

LINET MULTICARE 2015 Service manual

Hill Laboratories HA90W Reference manual

Hill Laboratories

Hill Laboratories HA90W Reference manual

Hallowell EMC 000A2787B instruction sheet

Hallowell EMC

Hallowell EMC 000A2787B instruction sheet

GE Vivid 3N Pro Series Service manual

GE

GE Vivid 3N Pro Series Service manual

REH4MAT AKVOLITO user manual

REH4MAT

REH4MAT AKVOLITO user manual

Kodak 104 Service manual

Kodak

Kodak 104 Service manual

Zimmer PremoPortFour Instructions for use

Zimmer

Zimmer PremoPortFour Instructions for use

ResMed Mirage Vista user guide

ResMed

ResMed Mirage Vista user guide

Acteon X-MIND unity Operator's manual

Acteon

Acteon X-MIND unity Operator's manual

ACKERMANN Apollo user manual

ACKERMANN

ACKERMANN Apollo user manual

Oppo GENU X Instructions for use

Oppo

Oppo GENU X Instructions for use

orfit HP Pro 25000/26 Instructions for use

orfit

orfit HP Pro 25000/26 Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.