manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Contec
  6. •
  7. Medical Equipment
  8. •
  9. Contec CMS50D User manual

Contec CMS50D User manual

Instructions to User
Dear Users, thank you very much for purchasing our product.
This Manual is written and compiled in accordance with the council directive MDD93/42/EEC for medical devices and
harmonized standards. The Manual is written for the current Pulse Oximeter. In case of modifications and software upgrades, the
information contained in this document is subject to change without notice.
The Manual describes, in accordance with the Pulse Oximeter's features and requirements, main structure, functions,
specifications, correct methods for transportation, installation, usage, operation, repair, maintenance and storage, etc. as well as
the safety procedures to protect both the user and equipment. Refer to the respective chapters for details.
Please read the Manual very carefully before using this equipment. These instructions describe the operating procedures to be
followed strictly, failure to follow these instructions can cause measuring abnormality, equipment damage and personal injury.
The manufacturer is NOT responsible for the safety, reliability and performance issues and any monitoring abnormality, personal
injury and equipment damage due to user's negligence of the operation instructions. The manufacturer’s warranty service does not
cover such faults.
Owing to the forthcoming renovation, the specific products you received may not be totally in accordance with the description of
this User Manual. We would sincerely regret for that.
This product is medical device, and can be used repeatedly. Its using life is 3 years.
WARNING:
The uncomfortable or painful feeling may appear if using the device ceaselessly, especially for the microcirculation
barrier patients. It is recommended that the sensor should not be applied to the same finger for over 2 hours.
For the individual patients, there should be a more prudent inspecting in the placing process. The device can not be
clipped on the edema and tender tissue.
The light (the infrared is invisible) emitted from the device is harmful to the eyes, so the user and the maintenance
man, can not stare at the light.
Testee can not use enamel or other makeup.
Testee’s fingernail can not be too long.
Please peruse the relative content about the clinical restrictions and caution.
This device is not intended for treatment.
Caution: Federal law restricts this device to sale by or on the order of a physician.
1 Safety
1.1 Instructions for Safe Operations
Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect patient's
safety and monitoring performance about cables and transducers. It is recommended that the device should be inspected
once a week at least. When there is obvious damage, stop using the monitor.
Necessary maintenance must be performed by qualified service engineers ONLY. Users are not permitted to maintain it by
themselves.
The oximeter cannot be used together with devices not specified in User's Manual.Only the accessory that appointed or
recommendatory by manufacture can be used with this device.
This product is calibrated before leaving factory.
1.2 Warnings
Explosive hazard—DO NOT use the oximeter in environment with inflammable gas such as some ignitable anesthetic
agents.
DO NOT use the oximeter while the testee measured by MRI and CT.
The person who is allergic to rubber can not use this device.
The disposal of scrap instrument and its accessories and packings(including battery, plastic bags, foams and paper boxes)
should follow the local laws and regulations.
Please check the packing before use to make sure the device and accessories are totally in accordance with the packing list,
or else the device may have the possibility of working abnormally.
Please don't measure this device with function test paper for the device's related information.
1.3 Attentions
Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture.
If the oximeter gets wet, please stop operating it.
When it is carried from cold environment to warm or humid environment, please do not use it immediately.
DO NOT operate keys on front panel with sharp materials.
High temperature or high pressure steam disinfection of the oximeter is not permitted. Refer to User Manual in the relative
chapter for instructions of cleaning and disinfection.
Do not have the oximeter immerged in liquid. When it needs cleaning, please wipe its surface with medical alcohol by soft
material. Do not spray any liquid on the device directly.
When cleaning the device with water, the temperature should be lower than 60℃.
As to the fingers which are too thin or too cold, it would probably affect the normal measure of the patients' SpO2and pulse
rate, please clip the thick finger such as thumb and middle finger deeply enough into the probe.
Do not use the device on infant or neonatal patients.
The product is suitable for children above four years old and adults (Weight should be between 15kg to 110kg).
The device may not work for all patients. If you are unable to achieve stable readings, discontinue use.
The update period of data is less than 5 seconds, which is changeable according to different individual pulse rate.
The waveform is normalized.Please read the measured value when the waveform on screen is equably and steady-going,
Here this measured value is optimal value. And the waveform at the moment is the standard one.
If some abnormal conditions appear on the screen during test process, pull out the finger and reinsert to restore normal use.
The device has normal useful life for three years since the first electrified use.
The hanging rope attached the product is made from Non- allergy material, if particular group are sensitive to the hanging
rope, stop using it. In addition, pay attention to the use of the hanging rope , do not wear it around the neck avoiding cause
harm to the patient.
The instrument dose not have low-voltage alarm function, it only shows the low-voltage.please change the battery when
the battery energy is used out.
When the parameter is particularly, The instrument dose not have alarm function.Do not use the device in situations where
alarms are required.
Batteries must be removed if the device is going to be stored for more than one month, or else batteries may leak.
A flexible circuit connects the two parts of the device. Do not twist or pull on the connection.
1.4 Indication for Use
The Fingertip Pulse Oximeter is a non-invasive device intended for the spot-check of oxygen saturation of arterial hemoglobin
(SpO2) and the pulse rate of adult and pediatric patients in home and hospital environments (including clinical use in
internist/surgery, anesthesia, intensive care ect.). This device is not intended for continuous monitoring.
2 Overview
The pulse oxygen saturation is the percentage of HbO2in the total Hb in the blood, so-called the O2 concentration in the blood. It
is an important bio-parameter for the respiration. For the purpose of measuring the SpO2more easily and accurately, our company
developed the Pulse Oximeter.At the same time, the device can measure the pulse rate simultaneously.
The Pulse Oximeter features in small volume, low power consumption, convenient operation and being portable. It is only
necessary for patient to put one of his fingers into a fingertip photoelectric sensor for diagnosis, and a display screen will directly
show measured value of Hemoglobin Saturation.
2.1 Classification
Class II b, (MDD93/42/EEC IX Rule 10)
2.2 Features
Operation of the product is simple and convenient.
The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient in carrying.
Power consumption of the product is low and the two originally equipped AAA batteries can be operated continuously for
20 hours.
The product will automatically be powered off when no signal is in the product within 5 seconds.
4 directions display mode without waveform.
Waveform display mode as line drawing or filling manner.
2.3 Major Applications and Scope of Application
The Pulse Oximeter can be used to measure human Hemoglobin Saturation and pulse rate through finger, and indicate the pulse
intensity by the bar-display. The product is suitable for use in family, hospital (Ordinary sickroom), Oxygen Bar, social medical
organizations and also the measure of saturation oxygen and pulse rate.
The product is not suitable for use in continuous supervision for patients.
The problem of overrating would emerge when the patient is suffering from toxicosis which caused by carbon
monoxide, the device is not recommended to be used under this circumstance.
2.4 Environment Requirements
Storage Environment
a) Temperature: -40℃~+60℃
b) Relative humidity: ≤95%
c) Atmospheric pressure: 500hPa~1060hPa
Operating Environment
a) Temperature: 10℃~40℃
b) Relative Humidity: ≤75%
c) Atmospheric pressure: 700hPa~1060hPa
3 Principle and Caution
3.1 Principle of Measurement
Principle of the Oximeter is as follows: An experience formula of data process is established taking use of Lambert Beer Law
according to Spectrum Absorption Characteristics of Reductive Hemoglobin (Hb) and Oxyhemoglobin (HbO2) in glow &
near-infrared zones. Operation principle of the instrument is: Photoelectric Oxyhemoglobin Inspection Technology is adopted in
accordance with Capacity Pulse Scanning & Recording Technology, so that two beams of different wavelength of lights can be
focused onto human nail tip through perspective clamp finger-type sensor. Then measured signal can be obtained by a
photosensitive element, information acquired through which will be shown on screen through treatment in electronic circuits and
microprocessor.
Figure 1 Operating principle
3.2 Caution
1. The finger should be placed properly (see the attached illustration of this manual, Figure 5), or else it may cause inaccurate
measurement.
2. The SpO2sensor and photoelectric receiving tube should be arranged in a way with the subject’s arteriole in a position there
between.
3. The SpO2sensor should not be used at a location or limb tied with arterial canal or blood pressure cuff or receiving
intravenous injection.
4. Make sure the optical path is free from any optical obstacles like rubberized fabric.
5. Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby light, infrared heater, direct
sunlight and etc.
6. Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy.
7. Testee can not use enamel or other makeup.
3.3 Clinical Restrictions
1. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of subject is required. For a subject
with weak pulse due to shock, low ambient/body temperature, major bleeding, or use of vascular contracting drug, the SpO2
waveform (PLETH) will decrease. In this case, the measurement will be more sensitive to interference.
2. For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green and acid indigo blue), or
carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and some with icterus problem,
the SpO2determination by this monitor may be inaccurate.
3. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor blamed for serious error of
SpO2measure.
4. As the SpO2value serves as a reference value for judgement of anemic anoxia and toxic anoxia, some patients with serious
anemia may also report good SpO2measurement.
4 Technical Specifications
1) Display Format: OLED Display;
SpO2Measuring Range: 0% ~ 100%;
Pulse Rate Measuring Range: 30 bpm ~ 250 bpm;
Pulse Wave Display: columniation display and the waveform display.
2) Power Requirements: 2×1.5V AAA alkaline battery (or using the rechargeable battery instead), adaptable range:
2.6V~3.6V.
3) Power Consumption: Smaller than 30mA.
4) Resolution: 1% for SpO2and 1 bpm for Pulse Rate.
5) Measurement Accuracy: ±2% in stage of 70%-100% SpO2, and meaningless when stage being smaller than 70%. ±2 bpm
during the pulse rate range of 30-99 bpm and ±2% during the pulse rate range of 100~250 bpm .
6) Measurement Performance in Weak Filling Condition: SpO2and pulse rate can be shown correctly when pulse-filling
ratio is 0.4%. SpO2error is ±4%, pulse rate error is ± 2 bpm during the pulse rate range of 30~99 bpm and ±2% during the
pulse rate range of 100~250 bpm .
7) Resistance to surrounding light: The deviation between the value measured in the condition of man-made light or indoor
natural light and that of darkroom is less than ±1%.
8) It is equipped with a function switch. The Oximeter can be powered off in case no finger is the Oximeter within 5 seconds.
9) Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm, 6.75mW)
5 Accessories
One hanging rope;
Two batteries(optional)
One User Manual.
6 Installation
6.1 View of the Front Panel
Figure 2 Front view Figure 3 Batteries installation Figure 4 Mounting the hanging rope
6.2 Battery
Step 1. Refer to Figure 3. and insert the two AAAsize batteries properly in the right direction.
Step 2. Replace the cover.
Please take care when you insert the batteries for the improper insertion may damage the device.
6.3 Mounting the Hanging Rope
Step 1. Put the end of the rope through the hole.
Step 2. Put another end of the rope through the first one and then tighten it.
7 Operating Guide
1) Insert the two batteries properly to the direction, and then replace the cover.
2) Open the clip as shown in Figure 5.
Figure 5 Put finger in position
3) Let the patient’s finger put into the rubber cushions of the clip (make sure the finger is in the right position), and then clip
the finger.
4) Press the switch button once on front panel.
5) Do not shake the finger and keep the patient at ease during the process. Meanwhile, human body is not recommended in
movement status.
6) Get the information directly from screen display.
7) The button has three functions.When the device is power off, pressing the button can open it; When the device is
power on, pressing the button shortly can change direction of the screen; When the device is power on, pressing the button
long can change brightness of the screen.
Press the button shortly when the device is power on, the display mode will change, show as follow figure:
Figure 6 Display mode one Figure 7 Display mode two Figure 8 Display mode three
Figure 9 Display mode four Figure 10 Display mode five Figure 11 Display mode six
Fingernails and the luminescent tube should be on the same side.
8 Repairing and Maintenance
Please change the batteries when the low-voltage displayed on the screen.
Please clean the surface of the device before using. Wipe the device with medical alcohol first, and then let it dry in air or
clean it by dry clean fabric.
Using the medical alcohol to disinfect the product after use, prevent from cross infection for next time use.
Please take out the batteries if the oximeter is not in use for a long time.
The best storage environment of the device is - 40ºC to 60ºC ambient temperature and not higher than 95% relative
humidity.
Users are advised to calibrate the device termly (or according to the calibrating program of hospital). It also can be
performed at the state-appointed agent or just contact us for calibration.
High-pressure sterilization cannot be used on the device.
Do not immerse the device in liquid.
It is recommended that the device should be kept in a dry environment. Humidity may reduce the useful life of
the device, or even damage it.
9 Troubleshooting
Trouble
Possible Reason
Solution
The SpO2and Pulse
Rate can not be
displayed normally
1. The finger is not properly positioned.
2. The patient’s SpO2is too low to be detected.
1. Place the finger properly and try again.
2. Try again; Go to a hospital for a
diagnosis if you are sure the device works
all right.
The SpO2and Pulse
Rate are not displayed
stably
1. The finger is not placed inside deep enough.
2. The finger is shaking or the patient is
moving.
1. Place the finger properly and try again.
2. Let the patient keep calm
The device can not be
turned on
1. The batteries are drained or almost drained.
2. The batteries are not inserted properly.
3. The malfunction of the device.
1. Change batteries.
2. Reinstall batteries.
3. Please contact the local service center.
The display is off
suddenly
1. The device will power off automatically
when it gets no signal within 5 seconds.
2. The batteries are almost drained.
1. Normal.
2. Change batteries.
10 Key of Symbols
Symbol
Description
Type BF
Warning –See User Manual
The pulse oxygen saturation(%)
PRbpm
Pulse rate (bpm)
The battery voltage indication is deficient (change the battery in time avoiding the inexact measure)
1. no finger inserted
2. An indicator of signal inadequacy
battery positive electrode
battery cathode
1.Power switch
2.change direction of the screen
3.Change brightness of the screen
SN
Serial number
Alarm inhibit
WEEE (2002/96/EC)
IP22
International Protection
This item is compliant with Medical Device Directive 93/42/EEC of June 14, 1993, a directive of the
European Economic Community.
11 Function Specification
Display Information
Display Mode
The Pulse Oxygen Saturation (SpO2)
OLED
Pulse Rate (PR)
OLED
Pulse Intensity (bar-graph)
OLED bar-graph display
Pulse wave
OLED
SpO2Parameter Specification
Measuring range
0%~100%, (the resolution is 1%).
Accuracy
70%~100%:±2%, Below 70% unspecified.
Optical Sensor
Red light (wavelength is 660nm)
Infrared (wavelength is 880nm)
Pulse Parameter Specification
Measuring range
30bpm~250bpm (the resolution is 1 bpm)
Accuracy
±2bpm or±2% select larger
Pulse Intensity
Range
Continuous bar-graph display, the higher display indicate the stronger
pulse.
Battery Requirement
1.5V (AAA size) alkaline batteries × 2 or rechargeable battery
Battery Useful Life
Two batteries can work continually for 20 hours
Dimensions and Weight
Dimensions
57(L) × 31(W) × 32(H) mm
Weight
About 50g (with the batteries)
Appendix Guidance and manufacture’s declaration-electromagnetic emission
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration –electromagnetic emission
The CMS50D Pulse Oximeter is tended for use in the electromagnetic environment specified below. The customer of the user
of the CMS50D Pulse Oximeter should assure that it isused in such an environment.
Emission test
compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The CMS50D Pulse Oximeter uses RF energy only for their
internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The CMS50D Pulse Oximeter is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
Harmonic emissions
Not applicable
IEC 61000-3-2
network that supplies buildings used for domestic purposes.
Voltage fluctuations/
flicker emission
IEC 61000-3-3
Not applicable
Guidance and manufacture’s declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration-electromagnetic immunity
The CMS50D Pulse Oximeter is intended for use in the electromagnetic environment specified specified below. The the user of
CMS50D Pulse Oximeter should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electromagnetic
environment-guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±6KV contact
±8KV air
±6KV contact
±8KV air
Floors should be wood, concrete or
ceramic tile. If floor are covered with
synthetic material, the relative
humidity should be at least 30%.
Power frequency
(50Hz) magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment
Guidance and manufacture’s declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration-electromagnetic immunity
The CMS50D Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the
user of CMS50D Pulse Oximeter should assure that it is used in such an environment.
Immunity
test
IEC60601 test
level
Compliance
level
Electromagnetic environment -guidance
Radiated RF
ICE
61000-4-3
3V/m
80MHz to
2.5GHz
3V/m
Portable and mobile RF communication equipment should be used no
closer to any part of the CMS50D Pulse Oximeter, including cables, than
the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
recommended separation distance
P
E
d






1
5.3
80MHz to 800MHz
P
E
d






1
7
800MHz to 2.5GHz
Where Pis the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and dis the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcastcannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If
the measured field strength in the location in which The CMS50D Pulse Oximeter is used exceeds the applicable RF
compliance level above, the CMS50D Pulse Oximeter should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the CMS50D Pulse
Oximeter.
b Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM
for EQUIPMENT or SYSTEM that not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the CMS50D Pulse Oximeter
The CMS50D Pulse Oximeter is intended for use in the electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the CMS50D Pulse Oximeter can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the CMS50D
Pulse Oximeter as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150KHz to 80MHz
P
E
d






1
5.3
80MHz to 800MHz
P
E
d






1
5.3
800MHz to 2.5GHz
P
E
d






1
7
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

Other Contec Medical Equipment manuals

Contec CMS50DL User manual

Contec

Contec CMS50DL User manual

Contec ECG 100G User manual

Contec

Contec ECG 100G User manual

Contec CMS-50f User manual

Contec

Contec CMS-50f User manual

Contec CMS-50f User manual

Contec

Contec CMS-50f User manual

Contec CMS-50E User manual

Contec

Contec CMS-50E User manual

Contec CMS800G User manual

Contec

Contec CMS800G User manual

Contec PM-60A User manual

Contec

Contec PM-60A User manual

Contec ESA0008 User manual

Contec

Contec ESA0008 User manual

Contec SP10W User manual

Contec

Contec SP10W User manual

Contec CMS8000 User manual

Contec

Contec CMS8000 User manual

Contec 300G User manual

Contec

Contec 300G User manual

Contec CMS 50IW User manual

Contec

Contec CMS 50IW User manual

Contec CMS50D User manual

Contec

Contec CMS50D User manual

Contec ABPM 50 Reference guide

Contec

Contec ABPM 50 Reference guide

Contec CMS50M User manual

Contec

Contec CMS50M User manual

Contec CMS60D User manual

Contec

Contec CMS60D User manual

Contec CMS80A Instruction Manual

Contec

Contec CMS80A Instruction Manual

Contec CMS60C User manual

Contec

Contec CMS60C User manual

Contec CMS-50f User manual

Contec

Contec CMS-50f User manual

Contec CMS50QB Instruction Manual

Contec

Contec CMS50QB Instruction Manual

Contec MDD93EEC Instruction Manual

Contec

Contec MDD93EEC Instruction Manual

Contec CMS50EW User manual

Contec

Contec CMS50EW User manual

Contec ECG1200G User manual

Contec

Contec ECG1200G User manual

Contec CMS-50f User manual

Contec

Contec CMS-50f User manual

Popular Medical Equipment manuals by other brands

SchureMed Stirrup Dolly Instructions for use

SchureMed

SchureMed Stirrup Dolly Instructions for use

Skytrofa Auto-Injector Quick reference guide

Skytrofa

Skytrofa Auto-Injector Quick reference guide

Advantech POC127 user manual

Advantech

Advantech POC127 user manual

Arthrex Synergy Resection quick start guide

Arthrex

Arthrex Synergy Resection quick start guide

Flaem AirMate Instructions for use manual

Flaem

Flaem AirMate Instructions for use manual

Stryker 1068 Maintenance manual

Stryker

Stryker 1068 Maintenance manual

Dräger Primus SW 4.5n Instructions for use

Dräger

Dräger Primus SW 4.5n Instructions for use

EASTMAN Kodak RP X-Omat Processor M6B manual

EASTMAN

EASTMAN Kodak RP X-Omat Processor M6B manual

TransMedics Organ Care System Technical user guide

TransMedics

TransMedics Organ Care System Technical user guide

Cantel Medical MEDIVATORS ENDO STRATUS EGA-501 instruction manual

Cantel Medical

Cantel Medical MEDIVATORS ENDO STRATUS EGA-501 instruction manual

Ultimate Healthcare UBEE0001-2 user manual

Ultimate Healthcare

Ultimate Healthcare UBEE0001-2 user manual

Diplomat ADEPT DA 370 Product information

Diplomat

Diplomat ADEPT DA 370 Product information

DeVilbiss Healthcare Vacu-Aide 7314 Series Service manual

DeVilbiss Healthcare

DeVilbiss Healthcare Vacu-Aide 7314 Series Service manual

Stryker altrix Operation manual

Stryker

Stryker altrix Operation manual

burmeier Lippe IV slatted frame instruction manual

burmeier

burmeier Lippe IV slatted frame instruction manual

Storz XENON NOVA 3000 instruction manual

Storz

Storz XENON NOVA 3000 instruction manual

Wassermann SG-1 user manual

Wassermann

Wassermann SG-1 user manual

Roscoe Medical RMI-POX4C user manual

Roscoe Medical

Roscoe Medical RMI-POX4C user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.