Installation instructions, DLQA96/144
4189350016B (UK)
5. Technical specifications
Basic accuracy: Class 0.5 (-10...15...30...55°C) according to EN 60051 and IEC 51
Adjustments: None
Instrument sizes: 96 x 96mm and 144 x 144mm according to DIN 43700
Bezel: Slim bezels according to DIN 43718
Scale: Figures, lines and font size according to customer design and
specification
Base material: Scale printed directly on front glass
Design only on black scale base!
Pointer: Rotation disc according to customer specification/design
Design only on black pointer base (disc)!
Glass: Antiglare, thickness 2mm
Mounting angle: May be mounted at any angle between 0…360° to horizontal
without this affecting the linearity according to DIN 16257
Compass safety
distance: To be mounted at least 1m from the compass according to IEC
945 and EN 60945
Analogue inputs: Single or double (for SIN/COS input)
Measuring ranges: -10...0...10V Load: 10kΩ
0…1mA Load: 1.0V
4…20mA Load: 1.0V
Response time: Max. pointer movement: 90°/second, no overshoot
Protection: IP52 from front, IP20 from back (terminals) according to IEC 529
and EN 60529
(IP54 from front is optional with special gasket)
or
IP66 option in special DLQW housing, IP20 from back (terminals)
according to IEC 529 and EN 60529
Climate: Class H U E, short term condensing allowed
Max. 95% RH: Max. 30 days per year
Max. 85% RH: Remaining days
Max. 75% RH: Average per year
According to DIN 40040