ENGLISH • 4
4.1 Infection Control Information
Reference Card
For your convenience, an Infection Control Information
reference booklet, has been included with your Cavitron™
PROPHY-JET®Air Polishing Prophylaxis System. If your book-
let is missing or misplaced, you can request a replacement in
the U.S. by calling Customer Service at 1-800-989-8826 Monday
through Friday, 8:00 AM to 5:00 PM (Eastern Time) For other
areas, contact your local DENTSPLY representative.
4.2 General Infection Control
Recommendations
• Aswithalldentalprocedures,usestandardpersonal
protection equipment (i.e., wear face mask, eyewear
or face shield, gloves and protective gown).
• Aswithhighspeedhandpieces,andotherdentaldevices,
the combination of water, air and powder from your
Cavitron™PROPHY-JET®Air Polishing Prophylaxis
System will create aerosols. With proper technique,
much of the Cavitron™PROPHY-JET®System’s aerosol
dispersion can be effectively controlled and minimized.
Please carefully follow the procedural guide lines in this
manual regarding the use of your System.
• Formaximaloperatorandpatientsafety,carefully
follow the Infection Control Information procedures
detailed on the reference card accompanying your
System.
4.3 Water Supply Recommendations
It is highly recommended that all dental water supply systems
conform to applicable CDC (U.S. Centers for Disease Control
and Prevention) and ADA (American Dental Association)
standards, and that all recommendations be followed in terms
of flushing, chemical flushing, and general infection control
procedures. See Sections 5.2 and 9.
Section 4: Infection Control
Section 5: Installation Instructions
5.1 General Information
If the installation of your Cavitron™PROPHY-JET®Air
Polishing Prophylaxis System is performed by someone other
than trained DENTSPLY®Dealer personnel, care should
be taken to observe the following requirements and
recommendations.
5.2 Water Line Requirements
• Awatersupplylinewithuser-replaceablefilteris
factory installed on your Prophy JET Prophylaxis
system. Do not disconnect the water supply line. See
9.0 System Care for filter replacement instructions.
• IncomingwatersupplylinepressuretotheSystemmust
be 25 psi (172 kPa) to 60 psi (414 kPa). If your dental
water system’s supply line pressure is above 60 psi
(414 kPa), install a water pressure regulator on the
supply line to your Cavitron™PROPHY-JET®Air
Polishing Prophylaxis System.
• Amanualshut-offvalveonthedentalwatersystem
supply line should be used so that the water can be
completely shut-off when the office is unoccupied.
• InadditiontothewaterfiltersuppliedwithyourSystem,
it is recommended that a filter in the dental water system
supply line be installed so that any particles in the water
supply will be trapped before reaching the System.
• Aftertheaboveinstallationsarecompletedonthedental
water supply system, the dental office water line should
be thoroughly flushed prior to connection to the System.
5.3 Air Line Requirements &
Recommendations
• Anairsupplylinewithauser-replaceablefilter
assembly is factory installed on your PROPHY-JET
Prophylaxis system. Do not disconnect the air supply
line. A filter-mounting bracket is included for hanging
the air filter. The clear bowl should hang downward
allowing for moisture separation and drainage of water
from the air filter. See 9.0 System Care for replacement
instructions.
• IncomingairsupplylinepressuretotheSystemmustbe
65 psig (448 kPa) to 100 psig (690 kPa). If your office air
line pressure is above 100 psig (690 kPa), install an air
pressure regulator on the supply line to your Cavitron™
PROPHY-JET®Air Polishing Prophylaxis System.
• Amanualshut-offvalveontheofficeairsupplyline
should be used so that the air line can be completely
shut-off, and the line pressure relieved when the office
is unoccupied.
• YourSystemmustbesuppliedwithclean,dryairto
help prevent water condensation from forming in the air
supply line which may cause the system to malfunction.
In addition to the air filter supplied with your system, it
is strongly recommended that an air dryer be used on
the compressor line supplying the System.
5.4 Electrical Requirements
Refer to Section 11: Specifications.