ClINICal SETUP MENU
FIRST - Make sure the device is
plugged in and turned off.
Simultaneously press the DOWN
arrow button and the DELAY
button.
SECOND - While holding the
DOWN arrow button and the
DELAY button, press the ON/
OFF button. This will turn on the
device and take you to the
"Clinical Menu". You can now
access the Clinical Setup Menu.
(reference chart below)
Use the LEFT and RIGHT
arrow buttons to navigate
through each category of
the menu.
Menu Item Description Factory
Default
Clinical Menu
Language
Selection
Select device language. English
Set Pressure Set to CPAP pressure prescribed by doctor. Range: 3-20 cmH2O 10.0
cmH20
Delay Time Select period of time to stay at delay pressure. This delay is patient activated.
Range: 0-45 minutes, increments of 5
20 min
Delay Pressure If prescribed, set to a level that is comfortable for the patient to fall asleep. This level of pres-
sure is maintained for the first portion of the delay time and rises to the set pressure in the
last 10 minutes of the delay time. Range: 3 cmH2O to CPAP, increments of 0.5
4.0 cmH20
Pressure Unit The device operating pressure may be displayed in one of three pressure units: cmH2O, hPa
or mbar.
cmH20
Compliance
Meter
Cumulative hours of “while breathing” time. N/A
Enable Menu Press the button to enter. Reference the Enable Submenu section for details. N/A
Lockout Menu Press the button to enter. Reference the Lockout Submenu section for details. N/A
Enable Submenu
Auto-ON This option controls the Auto-ON feature, which automatically starts the flow of air after
breathing once or twice into the mask
Enabled
Auto-OFF This option controls the Auto-OFF feature, which automatically stops the flow of air after
taking off the mask.
Disabled
Notifications Setting to enable mask/filter check notification. Enabled
Exit Submenu Press the button to exit and return to Clinical Menu. N/A
Lockout Submenu
Delay Time
Lockout
Used to allow patient access to CPAP delay time setting.
Range: Locked, Unlocked
Unlocked
Exit Submenu Press the button to exit and return to Clinical Menu. N/A
ClINICal SETUP MENU ChaRT
Use the UP and DOWN
arrow buttons to adjust
settings for the individual
patient.
Follow these instructions to access the device settings and prepare the device for patient use. All settings have
pre-set values as shown below.