Doran 7000 User manual

Digital Scale
INSTRUCTION MANUAL
DORAN SCALES, INC.
1315 PARAMOUNT PKWY.
1-800-262-6844
FAX: (630) 879-0073
http://www.doranscales.com
MANUAL REVISION: 3.0
MAN0190 8/26/2002
Model 7000
Model 7000XL
Model 8000XL
Made in USA

1
Table of Contents
Table of Contents ......................................................................................... 1
Introduction ................................................................................................... 4
Quick Start User's Guide ............................................................................. 5
Fig. 1: Model 7000 Front Panel Layout .............................................................................. 5
Model 7000:................................................................................................................................5
Fig. 2: Model 7000XL Front Panel Layout......................................................................... 5
Model 7000XL:....................................................................................................................... 6
Fig. 3: Model 8000XL Front Panel Layout......................................................................... 6
Model 8000XL:........................................................................................................................... 6
Power Up:................................................................................................................................... 6
Basic Weighing Operation:...................................................................................................... 7
Turn Off: (8000XL) .................................................................................................................... 7
Low Battery Indication: (8000XL) ........................................................................................... 7
Electrical Connections:............................................................................................................. 7
Zero: ............................................................................................................................................ 7
Units: ........................................................................................................................................... 8
Print: ............................................................................................................................................ 8
Battery Operation: (8000XL) ................................................................................................... 8
Quick Setup Guide........................................................................................ 9
Fig. 4: Load Cell connections............................................................................................... 9
Load Cell and Power Connections: ........................................................................................ 9
Option Connections: ................................................................................................................. 9
Fig. 5: Serial Communications and Remote Push Button Connections......................10
Re-assembling Rear Cover:..................................................................................................10
Entering Calibrations and Parameter setup Mode:...........................................................10
Capacity: ...................................................................................................................................11
Calibration: ...............................................................................................................................11
A/D Ranging:............................................................................................................................12
Table 1: Calibration requirements in raw counts ............................................................12
Parameter Setup ......................................................................................... 13
Entering and Exiting the Calibration and Parameter Setup Mode:..................................13
Stepping through the menu parameters:.............................................................................14
Changing a Parameter:..........................................................................................................14
Quick Review of Setup Parameters:....................................................................................14
Legal for Trade Restrictions: .................................................................................................14
Audit Counters: ........................................................................................................................14
Setup Menus Explained ............................................................................. 15
Capacity Setup Menu..........................................................................................................16
Calibration Menu..................................................................................................................17
Digital Filter Setup Menu ....................................................................................................18
Automatic Zero Tracking Setup Menu..............................................................................18
Motion Aperture Setup Menu.............................................................................................19
Start Up Zero Setup Menu .................................................................................................19
Latching Zero Request Setup Menu.................................................................................19

2
Latching Print Request Setup Menu.................................................................................19
Printer Data Output Setup Menu.......................................................................................20
Output Formats ....................................................................................................................20
Baud Rate Setup Menu.......................................................................................................20
Serial Data Handshaking Setup Menu.............................................................................21
Units Conversion Setup Menu...........................................................................................21
Start Up Units Selection Menu...........................................................................................21
Push-button Function Setup Menu ...................................................................................22
Remote Push-button Configuration Menu .......................................................................22
Legal For Trade Setup Menu.............................................................................................22
Scale Resolution Setup Menu ...........................................................................................22
Unit On Timer (8000XL only) .............................................................................................23
Default all Scale Parameter settings ................................................................................23
Raw Counts Display Mode.................................................................................................23
Calibration and Parameter Menu Exit...............................................................................23
Data Communications................................................................................ 24
Introduction to data communications:..................................................................................24
Printer Modes:..........................................................................................................................25
Transmit on demand (tod):....................................................................................................25
Continuous print (CP): ............................................................................................................25
Auto Print 1 (AP1):....................................................................................................................25
Auto Print 2 (AP2): ...................................................................................................................25
Data output format: .................................................................................................................26
"F0" Format: ............................................................................................................................26
"2d" Format: ............................................................................................................................26
"SSP" format:.............................................................................................................................26
"Fg" Format: ............................................................................................................................26
Table 2: Doran serial protocol............................................................................................27
Specifications and Interconnect Data ..................................................... 28
Specifications:..........................................................................................................................28
Table 3: Scale Specifications.............................................................................................28
Interconnect Data:...................................................................................................................28
Table 4: TB1 Load Cell Connections ................................................................................28
Table 5: P2 Options Connections .....................................................................................29
Table 6: J1 Power Connections.........................................................................................29
P3 Keyboard Connections..................................................................................................29
Table 7: P3 Keyboard Connections ..................................................................................29
Fig. 6: Connector J2, Remote switch cable assembly...................................................29
Fig. 7: Connector J2, Serial cable assembly...................................................................30
Table 8: Serial Output pin description ..............................................................................30
Fig. 8: RS232 Output DB9 Connector (optional) ............................................................30
Fig. 9: Installation of EMI / RFI / ESD protection devices. ............................................31
Fig. 10: Jumpers and Connector Locations.....................................................................32
Table 9: Board Jumper settings ........................................................................................32
4-20mA Analog Output (optional)............................................................. 33
Introduction...............................................................................................................................33
Setup .........................................................................................................................................33
Operation..................................................................................................................................33
Fig. 11: 4-20mA analog option board ...............................................................................33

3
Troubleshooting.......................................................................................... 34
General problem resolution:..................................................................................................34
Resetting the scale parameters: ...........................................................................................35
Resetting the scale:................................................................................................................35
Scale Messages: .....................................................................................................................36
Error messages: ......................................................................................................................36
Replacement Part List................................................................................ 37
Warranty Statement.................................................................................... 41

4
Introduction
Introducing the Doran Scales, Inc. Model 7000, 7000XL, 8000XL Digital Scale Indicator.
This scale uses state of the art technology to provide you with a low cost solution to the
most demanding weighing applications. With ease of use and setup in mind, the scale is
simple to set up and ready to use. The Model 7000, 7000XL, 8000XL offers many
features. A few of these features are listed below:
NTEP certification for Class III installations to 10,000d (CoC 99-129).
A six digit, 0.56" red LED display for easy reading.
lb, kg, oz, g, lb-oz display units supported.
Fully configurable duplex printer port with RS232 support.
EEPROM nonvolatile data storage of all calibration and setup information.
Microprocessor monitoring system to prevent scale failure under severe fault
conditions.
Support for up to four 350 ohm load cells.
115/230 VAC 50/60 Hz (jumper selectable) operation.
Field selectable digital filtering.
Software configurable remote push-button support (Optional).
Non NTEP parameters are user configurable.
Password protected, Front Panel Calibration Access Feature (Optional on
7000).
60 hour of battery operation, with built in charger (8000XL).
Six digit, 0.56" red LED remote display (Optional).
4-20mA analog output (Optional).
Please be sure to read the entire manual to ensure obtaining all the benefits that the
Model 7000, 7000XL, 8000XL can provide. If any questions arise, please feel free to
contact the Doran Scales Technical Support Department at 1-800-262-6844.
Unpacking Your Scale
Before unpacking your Doran scale, please read the instructions in this section. Your
new scale is a durable industrial product, but it is also a sensitive weighing instrument.
Normal care should be taken when handling and using this product. Improper handling
or abuse can damage the scale and result in costly repairs that will not be covered by
the warranty. If you notice any shipping damage, notify the shipper immediately. Please
observe the following precautions to insure years of trouble free service from your new
scale.
DO NOT drop the scale.
DO NOT immerse the scale.
DO NOT drop objects on the platform.
DO NOT pick up the scale by the "spider."
Carefully remove the scale from the shipping carton.

5
Quick Start User's Guide
Fig. 1: Model 7000 Front Panel Layout
Model 7000:
The Model 7000 (Fig. 1) controls consist of the ZERO and optional UNITS buttons
located under the main display. A six-digit LED display is used to provide weight
indications, negative polarity (except lb/oz) and operator messages describing scale
operation. A status annunciator, "NEG," is used to display negative polarity as well as
the minus sign (except in lb & oz). The "NEG" LED is located in the lower left corner of
the display area. Four LED's located at the bottom of the display indicate the current
weighing units. Motion is indicated when the units LED's are off. The LED's are on
when stable.
Fig. 2: Model 7000XL Front Panel Layout

6
Model 7000XL:
The operational controls for the Model 7000XL (Fig. 2) consist of the ZERO,PRINT and
UNITS buttons. A six digit LED display is used to provide weight indications and
operator messages describing scale operation. Scale status such as motion (MOT),
polarity (NEG) and center of zero (ZERO) is displayed on annunciators located to the
left of the display area. Scale units are displayed on four annunciators located to the
right of the main display.
Fig. 3: Model 8000XL Front Panel Layout
Model 8000XL:
The operational controls for the Model 8000XL (Fig. 3) consist of the ZERO (on/off),
PRINT and UNITS buttons. A six digit LED display is used to provide weight indications
and operator messages describing scale operation. Scale status such as motion (MOT),
polarity (NEG), center of zero (ZERO), and low battery (BATT) are displayed on
annunciators located to the left of the display area. Scale units are displayed on four
annunciators located to the right of the main display.
Power Up:
Connect the Model 7000, 7000XL to a compatible power source. With the 8000XL,
press and release ZERO to turn on scale.

7
Basic Weighing Operation:
1) Remove all items from the scale platter.
2) Press ZERO to zero the scale. The weight display should now read zero.
3) Place an item on scale platter and wait for the motion (MOT) indicator to go out,
indicating a stable weight.
4) Read the weight on the scale display.
Turn Off: (8000XL)
1) To turn off manually, press and hold the ZERO push button until the display shows
"rEL PbrEL Pb." Then release, the ZERO button and, the indicator will turn off, or…
2) Indicator will turn off automatically at the end of the unit on timer setting when that
mode is selected (see page 23).
Low Battery Indication: (8000XL)
The "BATT" indicator indicates that the battery is in need of recharging. Once the
"BATT" indicator turns on, there will be approximately one more hour of battery life
before the scale shuts down. When the battery is too low to run the scale, the Model
8000XL simply turns off and will not operate again until the battery is recharged. The
Model 8000XL remains accurate and useable even with the "BATT" indicator on.
Note:Battery life can vary depending on the following:
The operating temperature.
If optional remote display is installed.
Whether or not the battery is fully charged after each low battery event.
Load cell input impedance.
Electrical Connections:
Prior to connecting your Model 7000, 7000XL, 8000XL to power, check the serial
number tag on the scale for the correct operating voltage. Verify that the power matches
the rated voltage.
Be sure the AC power is not excessively noisy - this can occur if large inductive loads,
such as solenoids or motors, are on the same power line. The Model 7000, 7000XL,
8000XL has a filtered power supply to reduce the effects of normal line noise, but they
cannot limit severe fluctuations. If problems occur, noise producing devices may have to
be suppressed to minimize their effect.
Zero:
The ZERO push button is used to zero the scale prior to making a reading. The zero
button can function over the full range of the scale or it can be limited to a zero band
equal to 4% of scale capacity for Canadian Legal for Trade applications. To zero the
scale, wait until the scale is stable and press the ZERO button. The scale will zero
immediately. The scale will not "zero" if the scale is in motion. The scale is equipped

8
with an optional "Zero on Demand" feature which holds "zero" requests until motion
stops. This option may be activated during the scale setup procedure.
Units:
The UNITS button permits the operator to change the 7000XL, 8000XL, and
7000(optional) units by pressing a button. After pressing UNITS, the next display
update will indicate the correct weight in the new units. The UNITS button has several
configuration parameters, which can disable the UNITS button or limit the display units
available.
Print:
The PRINT button permits the operator to send data to a printer or other external
devices. Like the ZERO button, the user must wait for motion to stop before pressing
the PRINT button. The current weight will then be transmitted to the printer. The
7000XL, 8000XL, and 7000(optional) has a "Print on Demand" feature which stores a
PRINT request until the scale is stable. Once stable, the scale transmits the current
weight. These models also have several automatic print options, which may be used to
simplify printer operation.
Battery Operation: (8000XL)
The 8000XL is equipped with a self-contained rechargeable, sealed, gelled-electrolyte
battery and charging circuit, both internal. The 8000XL is designed to run continuously
for 60 hours (with one 350 ohm loadcell) on a fully charged battery. The charging circuit
will fully charge the battery in approximately four hours. To charge the battery, simply
plug the line cord into a standard 115V (230V optional) wall outlet.
The 8000XL can be used while recharging the battery, in fact, the 8000XL can be used
with the AC charger cord plugged in on a continuous basis. A full recharge takes four
hours, whether the 8000XL is on or off.
If an AC power failure occurs with the charger plugged in, the 8000XL battery
immediately takes over to provide uninterrupted scale operation for up to 60 hours.
The 8000XL's charging circuit is a two-stage, current limited type charger. The 8000XL
will sense the charge condition of the battery and charge at a high rate when the battery
is depleted. When the battery comes up to a fully charged state, the charger will switch
to a "float" or "trickle" mode which maintains the battery at a fully charged state without
overcharging.
The "BATT" indicator indicates that the battery is in need of recharging. The 8000XL
will continue to operate accurately for approximately four hours (with one 350 ohm
loadcell) after "BATT" is lit. When the battery is too low to run the scale, the 8000XL
simply turns off and will not operate again until the battery is recharged. At this point,
when ZERO (ON/OFF) is pressed, the "BATT" indicator will be lit as the 8000XL
performs its display test and then the scale will shut down immediately. This eliminates
any potential scale errors.

9
Quick Setup Guide
Fig. 4: Load Cell connections
Load Cell and Power Connections:
Load cell connections are made through a terminal block located at the bottom center of
the main PC. Board. The power cord connects to a connector adjacent to the
transformer. These connections are accessible by removing the rear cover screws and
laying the rear cover and main board on the work surface. Locate the ferrite core kit and
read "load cell cable grounding and ferrite assembly" instruction sheet. Thread loadcell
wires through the ferrite core as shown in Fig. 9. Connect load cell per Fig. 4.
Option Connections:
The Remote Switch and Serial Communications connector is found on the topside of
the main PC. Board. This connection is accessed by removing the rear cover.
Connections are made by either crimping (or soldering) a connector contact onto each
lead of the option cable. After crimping (or soldering), the contact is pressed into a
connector housing. The completed option connector is then snapped onto the option
connector found on the main board. Like the load cell cable and power cord, the option
cables are passed through watertight fittings mounted on the back of the indicator.
ZERO
CAL
S1
UNITS
S2
S/N
UNITSCAL
S3 S4
+ Signal
- Signal
+ Excitation
- Excitation
+ Sense
- Sense
Remove JU7 and JU8 when connecting a 6-
wire load cell. (JU7 and JU8 must be present
when connecting to a 4-wire loadcell)
Remove factory installed
shipping jumper.
ZERO P3B
CAL S3

10
Fig. 5: Serial Communications and Remote Push Button Connections
Re-assembling Rear Cover:
The rear cover must be pressed solidly against the front cover to ensure a good seal.
To accomplish this, torque the four rear cover attachment screws to 20 in-lb.
Entering Calibrations and Parameter setup Mode:
The Calibration and Parameter Setup Mode can be entered by two different ways.
1) The Front Panel CAL Access Feature is only available when scale is not in Legal
For Trade mode (“oP” = “Std”). This feature is not available on 7000 without the units
push button option. To enter the calibration mode, power up scale while pressing the
ZERO and the UNITS button at the same time. When the “rEL Pb” is shown, release
both buttons. The display will momentary show “Ent Cd” then go blank. Press the
ZERO button five times, the display will indicate the number of times the button has
been pressed. When “ 5“ is displayed then press and release the units button and
wait a few seconds. Note: If the code is not entered before the timer is finished, the
scale will bypass code entry and enter the normal run mode.
2) The CAL switch S1 is a momentary push button located in the lower left corner of
main PC. Board (see Fig. 4). On the 7000XL and 8000XL, the calibration switch can
be accessed by removing the meter’s back cover. For the 7000 remove the Cal
access panel located on bottom of indicator.

11
Capacity:
Capacity selection is performed by entering in the Calibration mode, then press and
release UNITS until "CAPCAP XX" appears on the display (where "XX" can be any valid
capacity). Once the CAP prompt appears press and release ZERO until the desired
capacity is displayed. When finished, return to the run mode by pressing the CAL
switch again or scroll through the menu with the UNITS button to "donEdonE nn" and select
"donEdonE 44" with the ZERO button, then press the UNITS button.
Once the desired capacity has been selected, place the correct capacity label on the
front panel, to the right of the display.
NOTE: On scales with factory installed platforms, the capacity is set at the factory. It
will not be necessary to set the scale capacity.
Calibration:
1) Enter Calibration mode, then press and release the UNITS until "CALCAL 00" appears on
the display.
2) Remove all weight from the scale platter and wait for about 10 seconds. Press
ZERO and wait for the display to count down to 0. If the calibration zero was in
range, the display will return with "CALCAL FSFS". If the display returns with a "CALCAL 00".
Repeat the process.
NOTE: If "rgrg ErrErr" appears on the display, the calibration zero is out of range. Press
ZERO to clear error. Refer to the A/D Ranging section for additional information.
3) The scale can be calibrated using (a) full capacity weight, (b) half capacity, (c)
quarter capacity, (d) one tenth capacity, or (e) any capacity weight between 2% and
100% of full scale (7000XL and 8000XL only).
To select the weight to calibrate the scale, press the UNITS button and select one of the
four following capacities.
CALCAL FSFS: Full load calibration.
CALCAL .50.50: Half load calibration. (50% of full load)
CALCAL .25.25: Quarter load calibration. (25% of full load)
CALCAL .10.10: 1/10th load calibration. (10% of full load)
If you want to use a weight between 10% and 100% of capacity, not 25% or 50%, select
the capacity above the closest to the desired weight (do not use CAL FS). Then scroll to
the exact weight by pressing CAPACITY to go higher or PRINT to go lower. This is not
available on the Model 7000.

12
4) To complete the calibration process, place the correct weight on the platter and
press ZERO, wait for the display to count down to 0. If the span calibration was in
range, the display will return with "donEdonE". If the display returns with a "CALCAL 00", repeat
steps 2 and 3.
NOTE: If " SPAnSPAn EE" appears on the display, the calibration span is out of range. Press
ZERO to clear error. Refer to the A/D Ranging section for additional information.
NOTE: Calibration at 10% of capacity has been provided as a convenience to
customers with high capacity scales in remote or inaccessible locations. Scales
calibrated at 10% of capacity are more likely to have significant errors at full capacity
than are scales calibrated at 25% or 50%. Doran Scales recommends that all scales be
calibrated at full capacity whenever possible. 10% calibration should not be used when
calibrating scales for legal for trade applications. It is the responsibility of the
installer/user to ensure that NTEP accuracy is achieved after any calibration.
A/D Ranging:
(Refer to this section only if you encounter a calibration problem)
NOTE: On scales with factory installed platforms, the zero and span will lie within
permissible limits.
1) Enter Calibration mode by using Front Panel CAL Access feature or by
pressing the CAL switch (S1).
2) Press and release UNITS until in the raw counts mode.
3) Return the scale platform to "No Load" by removing all items from the platform.
4) Record the "No Load" counts. The "No Load" or dead load raw counts must be
between 130,000 and 393,000 counts. If the readings are outside of the limits
specified, change dead load until you meet the requirements.
5) Place "Full Load" on the platform and record the "Full Load" counts. Subtract
the "No Load" counts from the "Full Load" counts to calculate "span". Refer to
Table 1 and verify that the span falls within the limits specified range. The "Full
Load" raw counts (span + dead load) should not exceed 900,000 counts.
6) When using 50%, 25% or 10% of full load to calibrate, refer to Table 1 50%,
25%, 10% span ranges.
Calibration requirements in raw counts
Platform load Minimum span Maximum span
Full 50,000 600,000
50% 25,000 300,000
25% 12,500 150,000
10% 5,000 60,000
Table 1: Calibration requirements in raw counts

13
Parameter Setup
The Models 7000/7000XL/8000XL have 22 calibration and parameters setup menu
items, which can be accessed two different methods, CAL switch and the Front Panel
CAL Access Feature. The Front Panel CAL Access Feature allows the user to change
the scale’s calibration and parameter settings by way of entering in a password code.
This Feature is only available when scale is not configured for legal for trade mode (see
Operating mode on page 22).
Entering and Exiting the Calibration and Parameter Setup Mode:
The Calibration and Parameter Setup Mode can be entered by two different methods.
1) The Front Panel CAL Access Feature is only available when scale is not in Legal
For Trade mode (“oP” = “Std”). This feature is not available on 7000 without the units
push button option. To enter the calibration mode, power up scale while pressing the
ZERO and the UNITS button at the same time. When the “rEL Pb” is shown, release
both buttons. The display will momentary show “Ent Cd” then go blank. Press the
ZERO button five times, the display will indicate the number of times the button has
been pressed. When “ 5“ is displayed then press and release the UNITS button and
wait a few seconds.
Note: If the code is not entered before the timer is finished, the scale will bypass
code entry and enter the normal run mode.
2) Apply power to the Scale. Then momentarily press the CAL switch S1 located in the
lower left corner of main PC. Board (see Fig. 4). The calibration switch can be
accessed by removing the meter’s back cover. Warning: do not press CAL switch
while powering up scale, this will cause the scale to reset all parameter settings.
The scale will Momentarily display a Parameter and Calibration audit counters. The
indicator will then display the first menu item, "CALCAL 2525" (Where 25 can be any valid
capacity). Press the UNITS button to access the next menu item.
To exit the Calibration and Parameter Setup Menu, momentarily press the CAL switch
or scroll through the menu options, by pressing the UNITS button, until "donEdonE nn" appears.
Press ZERO button until "donEdonE 44" appears and then press the UNITS button. The
indicator will return to the normal weighing mode. If any menu selections were changed,
The new values will be saved
NOTE: No new setup information is saved until the scale displays "SAVEdSAVEd" and returns to
the RUN mode. In the event of a power failure while in the Calibration and Parameter
Mode, any changes that have been made will be lost.

14
Stepping through the menu parameters:
Once the Calibration and Parameter Setup Mode has been entered, you may step
through the menu by pressing and releasing UNITS. A different display prompt will
appear for each parameter in the menu.
The parameter list on the following pages corresponds to the parameters available in
the Calibration Setup Menu.
Changing a Parameter:
After finding the desired menu item, the parameters for that item may be changed.
Press and release ZERO to step through the parameter list for that item. The list of
choices will repeat if you keep pressing and releasing ZERO. When you have found the
desired setting, press UNITS to go to the next menu item.
Quick Review of Setup Parameters:
The Setup parameters for the Model 7000, 7000XL, 8000XL may be quickly reviewed
without entering in the Calibration and Parameter Setup Mode. Remove power and
press and hold the ZERO button while you apply power. Hold the button until the scale
begins to scroll through the setup parameters. The button may be released anytime
after the review has begun. After parameters are displayed, scale will then go to the
normal weighing mode automatically.
Legal for Trade Restrictions:
When the Legal for Trade mode is enabled, it automatically disables some menus and
parameter options. This is done to comply with NTEP requirements. The menus and
parameter sections are shown on the following pages. Those menus and/or parameters
not available when in the Legal for Trade mode are marked by an asterisk.
Audit Counters:
When entering calibration mode, the Parameter audit counter and the Calibration audit
counter will momentarily be displayed. The Parameter audit counter only increments
when A2t,nn.A.,SU0,oP values are changed (ZERO pb is pressed). The Calibration audit
counter only increments when CAL 0, and Span are performed or the CAP,CtS, values are
changed (ZERO pb is pressed).

15
Setup Menus Explained
(In order of occurrence)
CAPCAP CALCAL AVgAVg A2tA2t
2
Calibration Setu
p
Menu
A0
A9
A7
A6
A5
A4
A3
C4
C8
C16C16
C32
C64
oFF
0.50.5
1
3
FS
00
nnnn.A..A.
0.5
0.25
0.1
xxxxxx
SU0SU0
onon
CL0
PB0
2od2od
on
oFFoFF
PodPod
on
oFFoFF
HSHS
oFF
SFSF
CtS
CSLCSL
CACA
Lgo
Lh
UnitSUnitS
IbIb
hg
P.b.P.b.
CPCP
P
Lo
go
O2
g
Ib o2
C
non
r.P.b.r.P.b.
oFFoFF
P
C
0
oPoP
StdStd
44
CAn
CtSCtS
3
PP
SP
td4td4
on
0.50.5
1
1.5
2
3
5
10
30
xxxxxx donEdonE
4
nn
d.o.d.o.
t.o.d.t.o.d.
C.P.
A.P.1
A.P.2
For.For.
F0F0
2d
SSP
Fg
11
3
5
br.br.
12
24
48
9696
14.4
CC PP
100.
dEFtdEFt
nn
4
4
nn

16
Capacity Setup Menu
CAPCAP Capacity Select Menu
Allows the selection of scale capacity.
22 pounds 3
55 pounds
66 pounds
10 10 pounds
15 15 pounds 2,3
20 20 pounds 3
2525 25 pounds 3
30 30 pounds 3
50 50 pounds
60 60 pounds
100 100 pounds
150 150 pounds 2,3
200 200 pounds 3
250 250 pounds 3
300 300 pounds 3,4
500 500 pounds 2,3,4
600 600 pounds 2,3,4
1.1000 pounds 1,2,3,4
1.51500 pounds 1,2,3,4
2.2000 pounds 1,2,3,4
2.52500 pounds 1,2,3,4
3.3000 pounds 1,2,3,4
5.5000 pounds 1,2,3,4
6.6000 pounds 1,2,3,4
10.10,000 pounds 5
15.15,000 pounds 5
20.20,000 pounds 5
25.25,000 pounds 5
30.30,000 pounds 5
50.50,000 pounds 5
60.60,000 pounds 5
100.100,000 pounds 5
1) No lb-oz display for this capacity in standard resolution.
2) No lb-oz display for this capacity in precision resolution.
3) No lb-oz display for this capacity in super precision resolution.
4) No grams display for this capacity.
5) Pound - kilograms display only at this capacity.

17
Calibration Menu
CALCAL Zero Calibration Mode.
00
Calibration Zero
Press ZERO to perform calibration of the scale zero.
Successful calibration is indicated by " donEdonE"
NOTE: The scale will automatically adjust the offset and gain to
compensate for dead load and span. When making these adjustments,
the scale may ask you to repeat zero calibration immediately after
performing a zero calibration or after a span calibration. Successful
calibration is indicated by "donEdonE"
CALCAL Span Calibration Mode.
(Does not appear if CAL 0 is not activated.)
FSFS Full load calibration.
.50 Half capacity calibration.
.25 Quarter capacity calibration.
.10 1/10th of capacity calibration.
XXXXXX
By pressing the optional print push button weight
value can be entered in. (note: XXXXXX will be the
dialed in weight value. This feature is not available
on 7000.)
NOTE: For maximum accuracy, Doran Scales recommends that all
scales be calibrated at full capacity. When location or installation make
it difficult to bring full capacity weights to the scale, calibration with as
little 10% of capacity is possible.

18
Digital Filter Setup Menu
AvgAvg Averaging mode
Determines the number of samples to average
A0 Stabil-izer©auto averaging. All readings are
averaged. Display updates 10 times a second.
A9 Stabil-izer©auto averaging. All readings are
averaged. Display updates 9 times a second.
A7 Stabil-izer©auto averaging. All readings are
averaged. Display updates 7 times a second.
A6 Stabil-izer©auto averaging. All readings are
averaged. Display updates 6 times a second.
A5 Stabil-izer©auto averaging. All readings are
averaged. Display updates 5 times a second.
A4 Stabil-izer©auto averaging. All readings are
averaged. Display updates 4 times a second.
A3 Stabil-izer©auto averaging. All readings are
averaged. Display updates 3 times a second.
C4 Fixed averaging 4 readings are averaged.
Display updates 10 times a second.
C8 Fixed averaging 8 readings are averaged.
Display updates 5 times a second.
C16C16 Fixed averaging 16 readings are averaged.
Display updates 3 times a second.
C32 Fixed averaging 32 readings are averaged.
Display updates 1½ times a second.
C64 Fixed averaging 64 readings are averaged.
Display updates 1 time every 1.4 seconds.
Automatic Zero Tracking Setup Menu
A2tA2t Automatic Zero Tracking Range
Small weights within the specified number of
divisions are automatically zeroed.
oFF Zero tracking is off. No automatic zeroing.
0.50.5 Zero tracking to within 0.5 division.
1* Zero tracking to within 1.0 division.
3*Zero tracking to within 3.0 divisions.
* NOTE: The Legal for Trade mode disables some options and selections
listed above. These items have been indicated by an asterisk.

19
Motion Aperture Setup Menu
nn.A.nn.A.*
Motion aperture *
Determines how many divisions consecutive
readings must change before the scale is
considered in motion.
11 1 division change must be seen to enter motion.
33 division change must be seen to enter motion.
55 division change must be seen to enter motion.
Start Up Zero Setup Menu
SU0SU0*Start Up Zero
Controls the start up zero status.
onon Zeros on the first stable reading on power up.
CL0 Loads the calibration zero for zero reference
PB0* Loads the last pushbutton zero. (8000XL option only)
Latching Zero Request Setup Menu
2od2od Zero on Demand
Enables or disable zero latching.
on If ZERO is pressed, it is saved until the scale
becomes stable.
oFFoFF If the scale is in motion, the zero request is
discarded.
Latching Print Request Setup Menu
PodPod Print on Demand
Enables or disables print latching.
on If PRINT is pressed, the print request is saved until
the scale becomes stable.
oFFoFF If the scale is in motion, the print request is discarded.
* NOTE: The Legal for Trade mode disables some options and selections
listed above. These items have been indicated by an asterisk.