Prepare cleaning solution.
Immerse flow sensor housing
and sensor insert bubble-free
in the cleaning solution.
Ensure that the solution flushes
all surfaces and interior spaces.
At the beginning and end of the
entire reprocessing procedure
vigorously swirl both parts for at
least one minute.
... until no cleaning agent
residues can be recognized.
For disinfection, repeat step 1 to 4 with disinfection solution.
For disinfection, repeat step 2a to 2c with disinfection solution.
Sufficiently rinse the housing and
the insert in water (preferably
drinking-water quality) ...
* The Neonatal Flow Sensor is also working with Evita® V300 and Evita Infinity® V500.
Reprocessing of ISO 15 Neonatal Flow Sensor
and Y-piece Neonatal Flow Sensor*
D-12377-2016
At the beginning of immersion
(step 1 and 2) flush each opening
of the housing with cleaning
solution using a 20 mL syringe.
At the end of immersion flush
each opening of the housing
once more with cleaning solution
using a 20 mL syringe.
Clean the housings with a lint-free
brush: Insert the brush ten times
each vertically into both ports of
Y-piece and pull it out again and
insert it ten times each diagonally
into both angles of the opening
for the sensor insert and pull it
out again.
When using Y-piece Neonatal Flow Sensor please add the steps 2a-2c to the cleaning and disinfection immersion step
(step 1 and 2), always using the corresponding cleaning or disinfection solution.
Check parts for visible soiling or
damage. Check the sensor insert
for damage to the measuring
wires and their pegs. Thoroughly
shake out residual water.
Allow the flow sensor to air dry for
at least 30 minutes.
Final check and drying for
both Flow Sensors
!
!
2a 2b 2c
Dräger Babylog® VN500
and Dräger Neonatal Flow Sensor
Reprocessing instructions
Manual cleaning and disinfection
This guideline does not replace or amend the instructions for use for Infinity Acute Care System Workstation Neonatal Care.
The instructions for use and the hospital guidelines and policies always have to be observed.