DRIVE PUREFILL™ OXYGEN COMPRESSOR MODEL PF1100
....................................................................................................
....................................................................................................
DRIVE PUREFILL™ OXYGEN COMPRESSOR MODEL PF1100
.................................................................................................... 3
WARNINGS AND CAUTIONS
• To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth
• The PureFill™ Oxygen Compressor is equipped with a detachable cord. Do not position
the equipment so that it is dicult to operate this disconnection device.
• If while using the system you experience any discomfort or signs of any unusual physical
symptoms, contact your physician immediately
• Oxygen can promote rapid burning when exposed to an open flame. Do not allow any
smoking objects or open flames near this unit.
• Do not use any oil, grease, or petroleum-based products on or near this unit.
• Do not smoke while the device is in use.
• Protect the device against the spillage of fluids on or around it. Do not place drinks on it.
Do not allow fluids to come in contact with the device. If fluids do come in contact with
the device, contact your service representative.
• Do not set any objects on, or allow to be draped over, the PureFill™ system as they may
block air flow into the device. Place the unit in a ventilated location with the sides and
rear two feet away from any object.
• Do not allow minors or children to operate the PureFill™ system. Do not allow unsuper-
vised children to play on or near the device.
• To prevent an accidental tip over, do not attempt to move the PureFill™ when it is con-
nected to a cylinder.
• Do not use the PureFill™ System if either the plug or power cord is damaged. Do not use
an extension cord with the PureFill™ System.
• Do not connect the PureFill™ System to an electrical outlet controlled by a wall switch or
ground fault interrupter.
• Make sure there is no other high power device such as a TV, stereo, computer; radio
transmitter; space heater, hair dryer, refrigerator, microwave oven, electrical range, or
other large appliance connected to this outlet.
• Electrical shock hazard: Do not remove the cabinet of the PureFill™ Station. The removal
of the cabinet should only be performed by Drive Medical or a trained representative.
• Disconnect the power cord from the electrical outlet before cleaning the housing of the
PureFill™ Station.
• This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical
electrical equipment and/or systems. This standard is designed to provide reasonable
protection against harmful interference in a typical medical installation. However, be-
cause of the proliferation of radio-frequency transmitting equipment and other sources
of electrical noise in health care and other environments, it is possible that high levels of
such interference due to close proximity or strength of a source might disrupt the perfor-
mance of this device. Medical electrical equipment needs special precautions regarding
EMC, and all equipment must be installed and put into service according to the EMC
information specified in this manual.
• Take precautions to prevent electrostatic discharge (ESD). Precautionary procedures
include methods to prevent buildup of electrostatic discharge (e.g., air conditioning,
humidification, conductive floor coverings, and non-synthetic clothing), discharging one’s
body to the frame of the equipment or system or to earth or a large metal object, and