INSTRUCnONS
FISHERDRYBATHINCUBATORS
Capacities:
11-718
11-718-2
11-718-4
11-718-6
11-718-8
1Block
2Blocks
3Blocks
4Blocks
6Blocks
CatalogNos.11-718
11-718-2
11-718-4
11-718-8
11-718-8
FisherDryBathIncubatorsprovidecontrolledheatfora
widevarietyofclinicalandgeneralchemistryapplications.
Theyoperatethrougha temperaturerangefromsligjttly
overambientto130C.Thevariotismodelscan
accommodate1,2,3,4,or6 dryblocks.Eachimitis
controlledbytwoadjustablethermostats,onefor
operationuptoabout60C,andtheotherforupto130
C.A fully-insulatedpad-typeheatingelementisbonded
direct^totheentireundersideofthedryblocksupport
platetoprovideevenheatdistribution.Thesupport
plateismadefromaltmiinumforrapidheattransfer.
Eachmodelisdesignedtobeusedwithanyofei|^t
differentaccessorydryblocks.(SeetheAccessories
Sectionfora listing,ofcatalognumbers).Blocksare
madefromsolidaluminumwitha blackanodizedfinish.
Eachblockismarkedsothateverywellcanbeidentified.
Dryblockscanhandlea variety oftubesfrom1.5ml
microcentrifugetubesto25imndiametertubes.Allof
theseaccessorydryblocksareinterchangeable.
Typicalapplicationsforthesedrybathsinclude
incubation,ctiltureinactivations,enzymereactions,
meltingandboilingpoints,andmanyotherprocedures
thatrequirecontroUedheat.
PerfomianceCharacteristics
TemperatureRange:SlightlyAbove
Ambientto130C
Note:Tolerancesontemperaturerangesare
approximately-5to+10C.Maximtunattainable
temperaturemayexceed140C.
Uniformity:* -t-or- 0.5C
*Variation,ata giventime,ofa sampletemperature
fromtheaverageofallsampletemperattu'eswithina
dryblock(at37C).Temperaturevariationbetween
blocksmayexceedthisspec.
PowerRequirements:
11-718115V,50/60Hz,90Watts
11=718-2115V,50/60Hz,180Watts
11-718-4115V,50/60Hz,270Watts
11-718-6115V,60/60Hz,360Watts
11-718-8116V,50/60Hz,540Watts
PhysicalData:
11-7189inX 8inX 3.6in4.6lbs
11-718-29inX8inX3.5in4.7lbs
11-718-49inX llinX 3.5in5.8lbs
11-718-612.5inX8.5inX3.5in8.1lbs
11-718-812.6inX llinX 3.5in7.4lbs,
Unpacking
TheseDryBathIncubatorsareshippedina sin^ecarton,
complete^assembled.Eachissuppliedwiththismantial,
awarrantycard,anda DryBlockHandle(forhandling
ormovingdryblocks).Thewarrantycardshouldbe
completedandretumedassoonaspossible.Incases
whereshippingdamageisobserved,keeptheunitand
cartonintact,includingthepackagingmaterials,andfile
aclaimwiththefinalcarrier.
Note;Becauseofthewidevarietyofapplicationsand
sampleblockdesigns,dryblocksarenotincludedwith
drybaths.Theusermustselectandordertheblocks
thatbestsuittheptirticularapplication.
OPERATION
Efichdrybathisequippedwithtwotemperaturecontrols
(thermostats),a selectorswitch,anda red-lensedlamp
whichindicatesheateroperation.Thelampcomeson
whenthecontrollingthermostat,asselectedbythe
switch,closestosupplypowertotheheater.Oncethe
drybathreachesthennalequilibrium,the(yclingofthe
lamp,
andtheheater,shouldcontinueona regular,
basis.
Incaseofa malfunction,suchasa heaterfailureor
thermostatstuckintheclosedposition,thelampwill
remain
OIL
Shouldthethermostatfailwiththecontacts
open,theli^twiUremainoff.
1.
Placetheimitona flat,levelsurface,inanareafree
ofdrailsandtemperaturechanges.(Note;Draftsand
ambienttemperaturechangeswilladverselyaffect
temperattureconstancyanduniformity.)
i
>
2.
Makesurethatbathcavityisfreeofforeignmatter,
thenplaceemptydryblock(s)intobathcavity.
3.
Connectunittoanappropriate,grotmdedpower
sourceasspecifiedontheunit'sdataplate.
4.
Insertanimmersiontypethermometer,graduatedto
thedesiredtemperaturerange,intothethermometer
wellinthediyblock.Note;Formoreprecise
measurements,thethermometercanbeplacedinto
thesampletubeora similartubeadjacenttothe
sampletube.
5.
Setthethree-positionswitchattheappropriate
setting: liow"settingforambientto60C
"Hii^"
settingfor60Cto130C
6.Ac^usttheappropriatetemperaturecontroltoa
settingthatapproximatesthedesiredtemperature.
Thefollowingcanbeusedasa generalguideline.On
theLowtemperaturecontrola settingof
"0"
willbe
ambient,"6"or'Tabout40C,and"10"about60C
OntheHi^temperaturecontrola settingof
"0"
will
heatsamplestoabout50C,
"5"
willheattoabout90
C,and
"10"
about130C.
7.
Asthedrybathisheating,monitorthetemperature
readingonthethermometerandobservetheindicator
lamp.
Ifthethermometerreadingexceedsthedesired
temperature,reducethetemperaturecontrolsetting.
Whenthelampcyclesonandoffatregularintervals
(or"flickers"),theunithasreachedequilibrium.Ifthe
drybathisnotatthedesiredtemperature,ac|jiistthe
temperaturecontrolintheappropriatedirection.
8.Oncethedrybathhasstabilizedatthedesired
temperature,samplesmaybeinsertedintotheblocks.
9.UsetheDryBlockHandletoremoveblocks.
BesurethattheDryBlockHandleissecurelyengaged
withtheblockbeforeliftingitfromthebath.-
CAUTION:
-Besuredrybathisconnectedtoonly
groundedpowersourcesoftheappropriate
voltage.
-Neverheatsamplesbeyondsafelevels.
-Ti^tlycappedsampletubesmayburstwhen
heated.
-UsetheDryBlockHandletomovethedry
blocks.
-TumtheSelectorSwitchtotheOffposition
andunplugunitwhennotinuse.
FORBESTRESULTS:
-UseDryBathIncubatorsinanareawitha constant
ambienttemperatureandnodrafts.
-Allsamplesshouldbeplacedinidenticaltubes,filled
toequallevelsnohigherthantopofblock.Allsiunple
tubesshouldbeplacedinidenticalblocks.
MAINTENANCE
Thesedrybathsrequirenoperiodicmaintenanceother
thanroutinecleaningwithmildcleaningproducts.Be
sureunitsaretmpluggedpriortocleaning.Neveruse
abrasivesorharshchemicalstocleanunits.Never
immersetinitsinliqtiidsorpourliqtiidsontothedry
blocksupportplate.
SERVICE
Itisrecommendedthatdrybathsbeservicedonlyfay
thosequalifiedininstrumentmaintenance.Serviceis
availableatanyoftheInstnmientServiceDistrictOffices
locatedthrou^outthecountry.Ifyouneedserviceor
serviceinformation,pleasecontacttheofficenearestto
yoiLTherepresentativestherecanissuea retum
authorizationnumber,if
necessary.
Emergency,on-site
servicecallsarenotcowered\mderthewarranty.
UnitDisassembly
WARNING:Unplugunitpriortodisassembly.
1.
Loosensetscrewseuidremovetemperattirecontrol
knobs.
2.
Tumtmitupside-downandremovefourscrews.
3.
Removebasebyfirst liftingatrear,thenmoving
forwardtodeartemperaturecontrolstems.
4.
Notetheroutingofwiresandthelocationofinsulation
sothattheunitcanbepropertyre-assembled.
5.
Carefultyremoveinsulation.
ThermostatReplacement
1.
(Optional)RemovescrewsthatattachHeaterBase
AssembtytotheTopCoverAssembly.
2.
Observewiringconnections,thenremovetwowires
connectedtoappropriatethermostat.
3.
Removefastenerholdingthermostatinplace,then
removethermostat.
4.
Connectwirestonewthermostatandinstall.
5.
Re-attachHeaterBaseAssemblyifithasbeen
removed.
6.Positionbaseontouiutandcheckalignmentof
thermostatshaftwithcenterofholeinfrontofbase.
Removebaseandrepositionthermostatasrequired,
thentightensecurely.
HeaterReplacement
1.
Thebeaterispermanentlybondedtothedryblock
supportplate,andcanonlybereplacedasa pre-
assembledunit.Thisassemblyiscalleda HeaterBase
Assembly.
2.
Removethefastenerswhichattachthethermostatsto
thedryblocksupportplate.
3.
Removethefastenersthatattachthedryblock
supportplatetothedrybathtopcoverassembly.
4.
Locateandremovethesplicesattheendoftheheater
leads.
l-
r\-
5.
Cormectthenewheaterleads,makingsurethatthe
coimectionsaresecureandproperlyinsulated.
6.Attachthenewdryblocksupportplateand
thermostats(seeThermostatReplacementsection).
SelectorSwitchorIndicatorLampReplacement
1.
Notewiringcoimectionsofdevicetobereplaced,then
removewires.
2.
Squeezetabsthatholddeviceinplace,thenpushout
throughfrontofbase.
3.
Installnewdeviceintoappropriateholeinbase.
4Reconnectwires.
Re-Assembly
ACCESSORIES
ItemCapacityCat.no.
DryBlock,24tubes,1.5ml11-718-9
DryBlock,30tubes,6mmdiameter11-718-10
DryBlock,20tubes,10mmdiameter11-718-12
DryBlock,20tubes,13mmdiameter11-718-14
DryBlock,12tubes,16mmdiameter11-718-16
DryBlock,8 tubes,20mmdiameter11-718-18
DryBlock,6 tubes,25mmdiameter11-718-20
DryBlock,Combination11-718-22
'6 tubes,6mmdiameter
5tubes,13mmdiameter
3tubes,25mmdiameter
1.
Positioninsulationoverheater.
2.
Makesurethatwireslu-eroutedsoasnottocomein
contactwithheater.
3.
Properlypositionbase,checkingalignmentof
thermostatsasinThermostatReplacementsection.
4.
Fastenbasetotopcover,andreplaceknobs.
ReplacementParts
ItemPartNundn*
Thermostat,LowTemp83310
Thermostat,Hi^Temp83311
SelectorSwitch83364
IndicatorLamp64057
DryBlockHandle83326
HeaterBaseAssemblyfor11-71883305
HeaterBaseAssemblyfor11-718-283306
HeaterBaseAssemblyfor11-718-483307
HeaterBaseAssemblyfor11-718-683308
HeaterBaseAssemblyfor11-718-883309
PartNo.83315
InstructionsforDryBathIncubators
Rev.A.(12/91)
FisherScientific
711ForbesAve
Pittsburgh,Pa.15219
1-800-388-8355
s-i
REARV IEW
I
H I
32 1
PL1
REARV IEW
SI LOW
T1-W1
'<y^^
SPLICE»3
SPLICE"2
SCHEMATIC

This manual suits for next models

4

Other Fisher Scientific Laboratory Equipment manuals

Fisher Scientific Isotemp 600 Series User manual

Fisher Scientific

Fisher Scientific Isotemp 600 Series User manual

Fisher Scientific 88861021 User manual

Fisher Scientific

Fisher Scientific 88861021 User manual

Fisher Scientific FB-705 User manual

Fisher Scientific

Fisher Scientific FB-705 User manual

Fisher Scientific Isotemp Plus 13-986-120A User manual

Fisher Scientific

Fisher Scientific Isotemp Plus 13-986-120A User manual

Fisher Scientific Fisherbrand 88861027 User manual

Fisher Scientific

Fisher Scientific Fisherbrand 88861027 User manual

Fisher Scientific 15-460-2Q User manual

Fisher Scientific

Fisher Scientific 15-460-2Q User manual

Fisher Scientific 280A User manual

Fisher Scientific

Fisher Scientific 280A User manual

Fisher Scientific Isotemp 642 User manual

Fisher Scientific

Fisher Scientific Isotemp 642 User manual

Fisher Scientific 106146 User manual

Fisher Scientific

Fisher Scientific 106146 User manual

Fisher Scientific Isotemp Plus 13-986-142A User manual

Fisher Scientific

Fisher Scientific Isotemp Plus 13-986-142A User manual

Fisher Scientific 650 Series User manual

Fisher Scientific

Fisher Scientific 650 Series User manual

Fisher Scientific 88861049 User manual

Fisher Scientific

Fisher Scientific 88861049 User manual

Fisher Scientific HPS RT2 Basic User manual

Fisher Scientific

Fisher Scientific HPS RT2 Basic User manual

Fisher Scientific accuSpin 3 User manual

Fisher Scientific

Fisher Scientific accuSpin 3 User manual

Fisher Scientific accumet AR50 User manual

Fisher Scientific

Fisher Scientific accumet AR50 User manual

Fisher Scientific Fisher Isotemp 307C User manual

Fisher Scientific

Fisher Scientific Fisher Isotemp 307C User manual

Fisher Scientific Isotemp User manual

Fisher Scientific

Fisher Scientific Isotemp User manual

Fisher Scientific 150 Homogenizer User manual

Fisher Scientific

Fisher Scientific 150 Homogenizer User manual

Fisher Scientific Isotemp 11-800-49S/.57A User manual

Fisher Scientific

Fisher Scientific Isotemp 11-800-49S/.57A User manual

Fisher Scientific 88861023 User manual

Fisher Scientific

Fisher Scientific 88861023 User manual

Fisher Scientific 15212985 User manual

Fisher Scientific

Fisher Scientific 15212985 User manual

Fisher Scientific Finnpipette User manual

Fisher Scientific

Fisher Scientific Finnpipette User manual

Fisher Scientific 88861041 User manual

Fisher Scientific

Fisher Scientific 88861041 User manual

Fisher Scientific SterilElite24 User manual

Fisher Scientific

Fisher Scientific SterilElite24 User manual

Popular Laboratory Equipment manuals by other brands

Thermo IEC Centra-GP8 Operation manual

Thermo IEC

Thermo IEC Centra-GP8 Operation manual

Comnet CNGE1IPS-95 Installation and operation

Comnet

Comnet CNGE1IPS-95 Installation and operation

Branson Sonifier 250 user manual

Branson

Branson Sonifier 250 user manual

IKA Lab disc operating instructions

IKA

IKA Lab disc operating instructions

Hach Dr Lange LCK 529 Programming instructions

Hach

Hach Dr Lange LCK 529 Programming instructions

Julabo CORIO CP Original operating manual

Julabo

Julabo CORIO CP Original operating manual

Thermo Scientific Dionex UltiMate 3000 Series Standard Applications Guide

Thermo Scientific

Thermo Scientific Dionex UltiMate 3000 Series Standard Applications Guide

Manson Engineering Industrial DPD Series Operation manual

Manson Engineering Industrial

Manson Engineering Industrial DPD Series Operation manual

Bruker BioSpin GRASI-1.0 user manual

Bruker

Bruker BioSpin GRASI-1.0 user manual

InnovaPrep ACD-200 user guide

InnovaPrep

InnovaPrep ACD-200 user guide

Consolidated Mark II operating manual

Consolidated

Consolidated Mark II operating manual

Steenbeck 01 Series user manual

Steenbeck

Steenbeck 01 Series user manual

Biotage Selekt Enkel user manual

Biotage

Biotage Selekt Enkel user manual

Thinky AR-250 Operation manual

Thinky

Thinky AR-250 Operation manual

Thermo Scientific 88882013 quick start guide

Thermo Scientific

Thermo Scientific 88882013 quick start guide

BioLAB BCLT-203 Operation manual

BioLAB

BioLAB BCLT-203 Operation manual

R-Biopharm RIDA qLine autoBlot user manual

R-Biopharm

R-Biopharm RIDA qLine autoBlot user manual

Thermo Scientific Forma 8602 Operating and maintenance manual

Thermo Scientific

Thermo Scientific Forma 8602 Operating and maintenance manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.