Fluke 713 User manual

713 English Instruction Sheet
Page
1
®
713
Pressure Calibrator
Instruction Sheet
Introduction
Fluke Model 713 30G and 713 100G Pressure Calibrators
(hereafter called “calibrator”) calibrate P/I (pressure to current)
transmitters, or measure pressure or current. The calibrator
simultaneously displays pressure and current measurements.
Full scale pressure input is as follows:
•Model 713 30G: 30 psi (206.85 kPa, 2.0685 bar)
•Model 713 100G: 100 psi (689.5 kPa, 6.895 bar)
The calibrator measures and displays current up to 24 mA, and
the pressure applied to the 1/8-inch NPT pressure fitting. The
calibrator measures pressure in the units shown in the next table.
Your calibrator is supplied with a Flex-Standholster, an
installed 9 V alkaline battery, one set of TL75 test leads, one set
of AC70A alligator clips, and this instruction sheet.
If the calibrator is damaged or something is missing, contact the
place of purchase immediately. Contact your Fluke distributor for
information about accessories. See “How to Contact Fluke.” To
order replacement parts or spares, see “Replacement Parts.”
PN 650298 August 1997 Rev. 1, 10/97
1997 Fluke Corporation. All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.

713 English Instruction Sheet
Page
2
Pressure Range and Resolution
Displayed
Pressure Units Model 713 30G
Range and
Resolution
Model 713 100G
Range and
Resolution
psi 30.000 psi 100.00 psi
inH2O at 4°C 830.4 inH2O 2768.0 inH2O
inH2O at 20°C 831.9 inH2O 2772.9 inH2O
cmH2O at 4°C, 2109.0 cmH2O 7030 cmH2O
cmH2O at 20°C 2113.0 cmH2O 7043 cmH2O
bar 2.0685 bar 6.895 bar
mbar 2068.5 mbar 6895 mbar
kPa 206.85 kPa 689.5 kPa
inHg 61.080 inHg 203.6 inHg
mmHg 1551.3 mmHg 5171 mmHg
kg/cm22.1090 kg/cm27.030 kg/cm2
Safety Information
Use the calibrator only as specified in this instruction sheet,
otherwise the protection provided by the calibrator may be
impaired.
A Warning identifies conditions and actions that pose hazard(s)
to the user; a Caution identifies conditions and actions that may
damage the calibrator or the equipment under test.
WWarning
To avoid possible electric shock or personal injury:
•Never apply more than 30 V between the mA
terminals, or between either of the mA terminals
and earth ground.

713 English Instruction Sheet
Page
3
•Remove the test leads from the calibrator
before you open the battery door.
•Make sure the battery door is closed and
latched before you operate the calibrator.
•Do not operate the calibrator if it is damaged.
•Do not operate the calibrator around explosive
gas, vapor, or dust.
•To avoid a violent release of pressure in a
pressurized system, shut off the valve and
slowly bleed off the pressure before you attach
or detach the calibrator pressure fitting to the
pressure line.
•When servicing the calibrator, use only
specified replacement parts
Caution
To avoid overpressure damage, do not apply
pressure that exceeds the following:
•Model 713 30G: 3X top of range (90 psi, 620 kPa,
6.2 bar)
•Model 713 100G: 2X top of range (200 psi, 1380
kPa, 13.8 bar)
To avoid corrosion in the pressure sensor, use the
calibrator only with media compatible with glass,
ceramic, silicon, RTV, type 316 stainless steel, and
nickel.
To avoid mechanically damaging the calibrator, do
not apply torque between the pressure fitting and
the calibrator case. See the figure for the proper
use of tools.

713 English Instruction Sheet
Page
4
Hold in
fixed
position
kf01f.eps
Explanation of International Symbols
The symbols in the table below are used on the calibrator or in
this instruction sheet.
International Symbols
Symbol Meaning
JEarth ground
IFuse
MBattery
WRefer to this instruction sheet for information about
this feature.
TDouble insulated
Conforms to relevant Canadian Standards
Association directives.
PConforms to European Union directives

713 English Instruction Sheet
Page
5
Getting Acquainted with the Calibrator
Press the green Opushbutton to turn the calibrator on and off.
The calibrator displays pressure and current measurements
simultaneously. See the figure below.
The upper part of the display shows the gage pressure applied to
the pressure fitting. Press Uto select a different pressure unit.
When you cycle the power off and on, the calibrator retains the
pressure unit you last used.
The lower part of the display shows the current (up to 24 mA)
applied to the current (mA) inputs. The current inputs are fused
with a 0.125 A, 250 V fast fuse (Littelfuse type 2AG).
Current mA
measurement
Pressure
measurement
HOLD
0-100psi 0-689kPa
RANGE
200psi 1378kPa
MAX
713 100G
PRESSURE
CALIBRATOR
DAMPUNITS
MIN
MAX
ZERO
CLR
HOLD
Pressure
fitting Current input
kf05f.eps

713 English Instruction Sheet
Page
6
Pushbutton Functions
Pushbutton Function
USelects a different pressure unit.
DTurns pressure reading damping on and off. With
damping on, the display does not update as
quickly.
ZPress this to zero the pressure display. Vent the
pressure fitting on the calibrator to atmosphere
before you press this pushbutton.
mPress and hold to read the minimum pressure and
current readings since the power was turned on or
cwas pressed.
cPress to clear the MIN and MAX memories.
MPress and hold to read the maximum pressure and
current readings the power was turned on or c
was pressed.
hPress hto freeze the display. The
gsymbol appears on the display. Press h
again to resume normal operation.
Calibrating a P/I Transmitter
To calibrate a P/I (pressure to current) transmitter, you apply a
pressure to the transmitter and measure the transmitter’s current
loop output. Connect the calibrator to the transmitter as shown in
the figure.

713 English Instruction Sheet
Page
7
SIGNAL
TEST
+–
713 100G
PRESSURE
CALIBRATOR
DAMPUNITS
MIN
MAX
ZERO CLR
HOLD
kf02f.eps

713 English Instruction Sheet
Page
8
Maintenance
For maintenance procedures not described in this instruction
sheet, or if the calibrator needs repair, contact a Fluke Service
Center.
In Case of Difficulty
•Check the battery, test leads, and pressure tubing. Replace
as necessary.
•Review this instruction sheet to make sure you are using the
calibrator correctly.
If the calibrator needs repair, and the calibrator is under
warranty, see the warranty statement for terms. If the warranty
has lapsed, the calibrator will be repaired and returned for a fixed
fee.
Cleaning
Periodically wipe the case with a damp cloth and detergent; do
not use abrasives or solvents.
Calibration
Calibrate your calibrator once a year to ensure that it performs
according to its specifications. A calibration manual is available
(PN 686540). Call 1-800-526-4731 from the USA and Canada. In
other countries, contact a Fluke Service Center.

713 English Instruction Sheet
Page
9
Replacing the Battery
When the Bsymbol appears on the display, replace the battery
with a 9 V alkaline battery.
SIGNAL
TEST
+–
713 100G
PRESSURE
CALIBRATOR
DAMPUNITS
MIN
MAX
ZERO CLR
HOLD
it07f.eps

713 English Instruction Sheet
Page
10
Replacing the Fuse
!Warning
To avoid personal injury or damage to the
calibrator, use only a 0.125A 250V fast fuse,
Littelfuse2AG.
Fuse F1 is probably blown if the mA measurement display
reading does not respond to current applied to the current (mA)
inputs.
Replace the fuse as follows:
1. Remove the test leads and turn the calibrator off.
2. Remove the battery door.
3. Remove the three Phillips-head screws from the case
bottom and turn the case over.
4. Gently lift the top cover from the end nearest the current
(mA) inputs until it unsnaps from the bottom cover.
5. Replace the fuse with a 0.125 A 250 V fast fuse, Littelfuse
type 2AG.
6. Carefully fit the case top and circuit board assembly
together, making sure that the O-ring is properly seated
between the pressure sensor and the pressure fitting on the
case top. Fit the case bottom onto the case top, engaging
the two snaps near the display end of the case. Reinstall the
three screws.
7. Replace the battery door.

713 English Instruction Sheet
Page
11
DAMP
UNITS
MIN
MAX
ZERO
CLR
HOLD
F1
Snaps
kf03f.eps

713 English Instruction Sheet
Page
12
Replacement Parts and Accessories
Replacement Parts
Item Description Part or
model
no. Qty.
BT1 9V battery, ANSI/NEDA 1604A or
IEC 6LR61 614487 1
CG81Y Holster, Yellow CG81Y 1
!
F1 Fuse, 125 mA, 250V fast 686527 1
MP85 713 30G case top 620218 1
MP85 713 100G case top 620226 1
MP86 Case bottom 620168 1
H2, 3, 4 Case screw 832246 3
MP89, 90 Non-skid foot 824466 2
OR1 O-ring for pressure input 146688 1
SP1 Spacer for pressure input 687449
MP92 Battery door 619947 1
H5, 6 Battery door fasteners 948609 2
S1 Keypad 687068 1
TL75 Test lead set TL75 1
AC70A Alligator clips AC70A 1
−713 Instruction Sheet 650298 1
TL20 Industrial test lead set TL20 Option
−71X Series Calibration Manual 686540 Option

713 English Instruction Sheet
Page
13
H2, 3, 4
AC70A
Alligator Clips
BT1
H5, 6
MP85
OR1
SP1
F1
MP86
Holster
S1
MP92
Instruction
Sheet
TL75
Test Lead Set
MP89, 90
kf04c.eps

713 English Instruction Sheet
Page
14
Specifications
Specifications are based on a one year calibration cycle and
apply for ambient temperature from +18 °C to +28 °C unless
stated otherwise. “Counts” are the number of increments or
decrements of the least significant digit.
Pressure Input, 713 30G
Range Accuracy
30 psi (206.85 kPa) ±0.05% of range
Maximum nondestructive pressure: 3X top of range (90 psi, 620 kPa,
6.2 bar)
Temperature coefficient: 0.01% of range per
°
C for temperature ranges
-10
°
C to 18
°
C and 28
°
C to 55
°
C
Pressure Input, 713 100G
Range Accuracy
100 psi (689.5 kPa) ±0.05% of range
Maximum nondestructive pressure: 2X top of range (200 psi, 1380 kPa,
13.8 bar)
Temperature coefficient: 0.01% of range per
°
C for temperature ranges
-10
°
C to 18
°
C and 28
°
C to 55
°
C
DC mA Input, 713 30G and 713 100G
Range Resolution Accuracy,
±(% of Reading + Counts)
24 mA 0.001 mA 0.025 + 1
Overload protection: 125 mA, 250V fast acting fuse
Temperature coefficient: 0.005% of range per
°
C for temperature
ranges -10
°
C to 18
°
C and 28
°
C to 55
°
C

713 English Instruction Sheet
Page
15
General Specifications
Maximum voltage applied between either mA terminal and
earth ground or between the mA terminals: 30 V
Storage temperature: -40°C to 60°C
Operating temperature: -10°C to 55°C
Operating altitude: 3000 meters maximum
Relative humidity: 95% up to 30°C, 75% up to 40°C, 45% up to
50°C, and 35% up to 55°C
Vibration: Random 2 g, 5 Hz to 500 Hz
Shock: 1 meter drop test
Safety: Certified as compliant to CAN/CSA C22.2 No.
1010.1:1992. Complies with ANSI/ISA S82.01-1994.
Power requirements: Single 9 V battery (ANSI/NEDA 1604A or
IEC 6LR61)
Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x
7.35 in L);
With holster and Flex-Stand: 52 mm H x 98 mm W x 201 mm L
(2.06 in H x 3.86 in W x 7.93 in L)
Weight: 369 g (13 oz);
With holster and Flex-Stand: 624 g (22 oz)

713 English Instruction Sheet
Page
16
How to Contact Fluke
To order accessories, receive operating assistance, or get the
location of the nearest Fluke distributor or Service Center, call:
1-800-44FLUKE (1-800-443-5853) in U.S.A. and Canada
+31-402-678-200 in Europe
+1-425-356-5500 from other countries
Address correspondence to:
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090, P.O. Box 1186,
Everett, WA 98206-9090 5602 BD Eindhoven
USA The Netherlands
Visit us on the World Wide Web at: www.fluke.com
LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship
for three years from the date of purchase. This warranty does not cover
fuses, disposable batteries or damage from accident, neglect, misuse or
abnormal conditions of operation or handling. Resellers are not
authorized to extend any other warranty on Fluke’s behalf. To obtain
service during the warranty period, send your defective calibrator to the
nearest Fluke Authorized Service Center with a description of the
problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE,
ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES
OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some
states or countries do not allow the exclusion or limitation of an implied
warranty or of incidental or consequential damages, this limitation of
liability may not apply to you.
Table of contents
Other Fluke Test Equipment manuals

Fluke
Fluke DTX-CLT CertiFiber User manual

Fluke
Fluke 820-2 User manual

Fluke
Fluke i6000s Flex-24 User manual

Fluke
Fluke 712 User manual

Fluke
Fluke 9100 Series Instruction Manual

Fluke
Fluke DTX Series User manual

Fluke
Fluke 1550C Instruction manual

Fluke
Fluke TL1000-KIT Original instructions

Fluke
Fluke TS 30 User manual

Fluke
Fluke 729 User manual

Fluke
Fluke 975 AirMeter User manual

Fluke
Fluke AIRCHECK Instruction manual

Fluke
Fluke 5730A User manual

Fluke
Fluke 5730A Parts list manual

Fluke
Fluke T5-1000 Technical manual

Fluke
Fluke 522A/6 User manual

Fluke
Fluke MicroScanner MS-POE User manual

Fluke
Fluke 5560A User manual

Fluke
Fluke 74X Series Use and care manual

Fluke
Fluke TL2X4W-TWZ User manual