GE CTF878 Owner's manual

GE
Sensing
Model CTF878
Clamp-On Tag Flowmeter
Programming Manual

GE
Sensing
Model CTF878
Clamp-On Tag Flowmeter
Programming Manual
910-254PA4
February 2009

iii
February 2009
Warranty Each instrument manufactured by GE Sensing, Inc. is warranted to be
free from defects in material and workmanship. Liability under this
warranty is limited to restoring the instrument to normal operation or
replacing the instrument, at the sole discretion of GE. Fuses and
batteries are specifically excluded from any liability. This warranty is
effective from the date of delivery to the original purchaser. If GE
determines that the equipment was defective, the warranty period is:
•one year for general electronic failures of the instrument
•one year for mechanical failures of the sensor
If GE determines that the equipment was damaged by misuse,
improper installation, the use of unauthorized replacement parts, or
operating conditions outside the guidelines specified by GE, the
repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu of
all other warranties whether statutory, express or implied
(including warranties of merchantability and fitness for a
particular purpose, and warranties arising from course of
dealing or usage or trade).
Return Policy If a GE Sensing, Inc. instrument malfunctions within the warranty
period, the following procedure must be completed:
1. Notify GE, giving full details of the problem, and provide the model
number and serial number of the instrument. If the nature of the
problem indicates the need for factory service, GE will issue a
RETURN AUTHORIZATION number (RA), and shipping instructions
for the return of the instrument to a service center will be
provided.
2. If GE instructs you to send your instrument to a service center, it
must be shipped prepaid to the authorized repair station indicated
in the shipping instructions.
3. Upon receipt, GE will evaluate the instrument to determine the
cause of the malfunction.
Then, one of the following courses of action will then be taken:
•If the damage is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and returned.
•If GE determines that the damage is not covered under the terms
of the warranty, or if the warranty has expired, an estimate for the
cost of the repairs at standard rates will be provided. Upon receipt
of the owner’s approval to proceed, the instrument will be repaired
and returned.

v
February 2009
Table of Contents
Chapter 1: Programming Site Data
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Powering UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Using the Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Using the Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Entering the Program Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Entering Transducer Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Parameters for Typical Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
Parameters for Special Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
Entry Confirmation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
Entering Pipe Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
Entering Fluid Types and Speeds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14
Entering Path Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-15
Path Parameters for Clamp-On Transducers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-16
Entering Analog Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-17
Entering Analog Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-20
Entering Digital Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-23
Entering Alarm Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-25
Entering Standard Volume Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-27
Entering Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-27
Entering Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-28
Entering User Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-30
Entering a Calibration Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-33
Chapter 2: Displaying and Configuring Data
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Format Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
View Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Limits Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Measurement Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Customizing the Display Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Specifying the Number of Displayed Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Customizing Softkeys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Accessing Meter Data —The About Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10

February 2009
vi
Table of Contents (cont.)
Chapter 3: Creating and Managing Sites
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Site Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Creating a New Site. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Opening an Existing Site. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Saving a Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Saving a Site with a Different Name. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Refreshing a Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Renaming a Site. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Deleting a Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Listing Files by Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Listing Files in Chronological Order. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Chapter 4: Programming Meter Settings
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Entering the Meter Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Selecting Measurement Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Entering Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Changing Date and Time Appearance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Adjusting Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Setting Backlight Timeout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Changing Communications Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Resetting Forward and Reverse Totals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Setting Up User Tables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11

vii
February 2009
Table of Contents (cont.)
Chapter 5: Logging Data
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Entering the Logging Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Log Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
File Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Setting up a New Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
Copying (Cloning) a Selected Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
Renaming a Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
Deleting a Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Deleting All Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Refreshing the Log Manager Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Log Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
Stopping (Pausing) a Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
Restarting a Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
Ending a Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Stopping All Logs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Restarting All Logs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Ending All Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
View All Sites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
View Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-13
Displaying Log Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-13
Displaying Log Data in Graphical Form. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-14
Displaying Log Data in Spreadsheet Form. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-17
Sort Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-18
Listing Logs by Name. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-18
Listing Logs in Chronological Order . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-18

Chapter 1

Programming Site Data
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Powering UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Using the Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Using the Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Entering the Program Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Entering Transducer Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Entering Pipe Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Entering Fluid Types and Speeds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
Entering Path Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Entering Analog Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
Entering Analog Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
Entering Digital Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-23
Entering Alarm Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-25
Entering Standard Volume Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-27
Entering User Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30
Entering a Calibration Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-33

Programming Site Data 1-1
February 2009
Introduction This chapter provides instructions for entering the programming data
required to place the CTF878 flowmeter into operation. Before the
CTF878 can begin taking measurements and displaying valid data,
the current system and pipe parameters must be entered. Additional
programming options provide access to the more advanced features of
the CTF878, but this information is not required to begin taking
measurements.
Powering UP Immediately upon power up the CTF878 emits a short beep and
displays a “PCI Loader” message. It then validates the instrument
programming, and displays the GE Sensing logo and the software
version. If the meter fails any of these tests, contact the factory.
After the meter conducts all the self checks, the screen then appears
similar to the one shown in Figure 1-1 below.
Figure 1-1: Screen After Powering ON
050100
DEFAULT 2003/07/17 03:35 PM

February 2009
1-2 Programming Site Data
Using the Screen The essential features for operating the CTF878 are the screen and
keypad. Although these features are common, the CTF878 design
offers particular features to simplify and speed operation. The new
menu system is very user-friendly; program information is now
located on tabbed forms instead of on a long, linear menu.
The primary function of the screen is to display information in order
for you to take measurements accurately and easily. The CTF878
screen consists of those parts shown in Figure 1-2 below.
Figure 1-2: CTF878 Screen in Operate Mode
The top line of the screen is the status bar, which normally displays
the time and date. However, when you press [MENU] (the menu key),
the menu bar replaces the status line.
The middle of the screen is the work area, which displays the
measured parameters, numeric measurements, and both bar and line
graphs. When you enter a selection on the menu bar, this area displays
menu prompts. A line at the bottom of the area also displays error
code messages, which are described in more detail in the Service
Manual.
E1: Receiver
0 50 100
DEFAULT 2003/07/17 03:35 PM
5
-12.2 0 12.2
Function Keys System Tray
Status BarMeasurementsCurrent Site
Example
Error Message

Programming Site Data 1-3
February 2009
Using the Screen (cont.) The system tray displays icons that indicate meter operations not
otherwise shown. Table 1-1 below lists the icons and their meanings.
The bottom of the screen displays the three function key options: F1,
F2 and F3. These keys have different functions, depending on the task
being performed. The initial, factory setup, configuration for these
keys is as follows:
•F1 = Transducer
•F2 = Diagnostics
•F3 = Save Now
Table 1-1: Icons in the System Tray
Icon Function Meaning
Alert The meter encountered an error
in operation.
Log A log is running (marks).
Log A log is pending (no marks).

February 2009
1-4 Programming Site Data
Using the Keypad The CTF878 keypad contains 24 keys that are labeled with their
functions. The complete keypad is illustrated in Figure 1-3 below and
a detailed description of the functions for each of the 24 keys is listed
in Table 1-2 on page 1-5.
Figure 1-3: CTF878 Keypad
ENTERMENU
?ESC SEL
.
7
4
1
0
8
5
2
9
6
3
F1 F2 F3
CTF878 Flowmeter

Programming Site Data 1-5
February 2009
Using the Keypad (cont.) Table 1-2: CTF878 Key Functions
Key Function
Software Function Keys: Press to select the
special functions displayed directly above them
at the bottom of the screen.
12 Numeric Keys: Use to enter numeric data.
Arrow Keys: Press to scroll through function
choices in measurement mode or to scroll
through menu choices in programming mode.
Press [<] as a backspace key in programming
mode.
Help Key: Use to access on-line help. See the
next section for details.
Escape Key: Press to exit menus or menu
options at any time, or cancel a numeric entry.
Select Key: Use to move between data
measurements on the screen.
Menu Key: Press to access the Main Menu.
Enter Key: Use to enter a particular menu or to
enter selected values into CTF878 memory.
F1 F2
F3
12
34
5 6
78
9 0
.–
?
ESC
SEL
MENU
ENTER

February 2009
1-6 Programming Site Data
Entering the Program
Menu To enter the Program Menu, press the [MENU] key at the lower right of
the CTF878 keypad. The Menu Bar replaces the Status Bar at the top
of the screen. Press the [X] arrow key once to scroll from the Site
Menu to the Program Menu. At the Program Menu, press [ENTER].
The screen appears similar to Figure 1-4 below. While following the
programming instructions, see the Program Menu Map, Figure 1-34
on page 1-37.
Figure 1-4: Program Menu
To scroll to a particular option, press the [T]or [S] arrow keys until
you reach the option. Then press [ENTER] to open the option window.
When entering parameters in an option, press:
•The [T] key to step through the available parameters
•The [S] key to scroll back to a previously entered parameter
•The [F2] key (Cancel) or the [ESC] key to exit an option at any time
and return to Operate Mode without changing the parameters.
Note: The Transducer, Pipe, Fluid and Path parameters are
categorized under the same Transducer/Pipe menu, as shown
in Figure 1-5 below. You can be more time-efficient by
programming the parameters in all four entries under
Transducer/Pipe before exiting the menu.
When you are highlighting a parameter in one of the four entries, and
you wish to go to another entry:
1. Use the [S] key to scroll up to the current tab (the entry that you
are under).
2. Press the [W] or [X] key to select the parameter you want to
access, then press [ENTER].
Site Meter Logging Service
Pipe
Fluid
Path
IO
Standard Volume
User Functions
Correction Factors
0 50 100
Program
Transducer elocity, m/s
Cross Corr

Programming Site Data 1-7
February 2009
Entering Transducer
Parameters To enter the Transducer option, scroll to the Transducer tab on the
Program Menu and press [ENTER]. The screen appears similar to
Figure 1-5 below. To step through each parameter, press the [T] key.
Refer to the Program Menu Map, Figure 1-34 on page 1-37.
Note: If the CTF878 does not scroll to a particular parameter, the
parameter is not necessary for that transducer type.
.
Figure 1-5: Transducer Option Window
The next prompt asks you to enter the transducer number (printed on
the transducer itself), or to specify that you are using a special
application transducer.
1. From the Type prompt, press the [T] key to reach the Transducer
prompt, and press [ENTER].
2. A drop-down list of transducer numbers opens (see Figure 1-6
below). Press the [T] or [S] keys to scroll to the appropriate
number, or scroll to “Special” for a special application transducer.
3. Press [ENTER] to confirm your selection.
.
Figure 1-6: Transducer Type Options
r
ansducer
Transducer/Pipe
Pipe Fluid Path
#304 (Shear)
0.500 0.471
Transducer
Frq Up/Dn MHz
OKCancel
°35Wedge Ang
Wedge Tmp 45 °C
Wedge SS 2480 m/s
Transducer/Pipe
Transducer
a
nsducer
p
Transducer/Pipe
T
ransducer
Transducer Pipe Fluid Path
SPECIAL
Transducer
Frq Up/Dn Hz
OKCancel
#304 (Shear)
#305 (Shear)
#306 (Shear)
Transducer/Pipe
#305 (Shear)

February 2009
1-8 Programming Site Data
Entering Transducer
Parameters (cont.)
Upon your selection, a warning screen appears similar to Figure 1-7
below. In three out of four warnings, the Receiver Board option
number appears as part of the text.
IMPORTANT: If SPECIAL is selected as the transducer type, a
number does not appear, however, the warning still
states that the transducer selection must match the
Receiver Board.
Figure 1-7: Transducer Selection Warning Screen
The Receiver Board option number corresponds to the frequency
range selected with the original order (see Table 1-3 below).
4. Compare the Receiver Board option number listed on the
WARNING screen with the one assigned to your CTF878.
Note: The Receiver Board option number is the last digit of the
CTF878’s model number. That model number can be found on
the label attached to the inside of the front cover. In
Figure 1-8 below the Receiver Board option number is 3.
Figure 1-8: Transducer Type Options
Table 1-3: Frequency Options
Option No. Frequency Range
3 1 MHz / 888 KHz
4 500 KHz / 471 KHz
5 250 KHz / 222 KHz
T
z
OK
This transducer selection requires
a matching Option 3 (1000/888
kHz) Receiver Board. See manual
for details.
Transducer/Pipe
* * * WARNING * * *
GE Infrastructure Sensing
Model No:
Serial No:
Volts:
CTF878-1-11-XXXX2X
XXX
85-264 VAC
Made in USA
20Watts:
3

Programming Site Data 1-9
February 2009
Entering Transducer
Parameters (cont.)
The program now varies, depending on whether you have selected
typical or special transducers.
•If you have selected a typical transducer, the CTF878 comes
programmed with the needed parameters. Proceed to
Parameters for Typical Transducers below.
•However, if your application requires entering a special
transducer / receiver option, go to the next section, Parameters
for Special Transducers on page 1-9.
Parameters for Typical
Transducers If a typical transducer has been selected, the user must:
1. Enter the transducer wedge temperature (Wedge Temp) text box
and press [ENTER].
2. Type in the value (provided by the factory) and press [ENTER].
3. Proceed to Entry Confirmation on page 1-10.
Parameters for Special
Transducers
Note: The factory will supply the information required below with a
special transducer / receiver option.
The user must enter the appropriate transducer / receiver operating
frequency range.
IMPORTANT: The CTF878 flowmeter and the transducers have the
same fixed frequency band of operation defined by
the customer's application. The frequency printed on
the transducers represents a frequency band of
operation and does not always match the up or down
frequencies.
1. From the Transducer prompt, press the [T] key to reach the
Frequency Up (Frq Up) prompt, and press [ENTER].
2. A drop-down list of thirty-six frequencies opens, ranging from
0.222 to 8.00 MHz. Scroll to the correct frequency, and press
[ENTER].
3. Repeat the same selection process for the Frequency Down
(Frq Dn) prompt.
Next, you must enter the transducer wedge angle (Wedge Ang) and
wedge soundspeed (Wedge SS).
4. Scroll to the appropriate text box and press [ENTER].
5. Type in the value (provided by the factory).
Note: If you enter an incorrect numeric value, press the [W] key to
erase the last digit entered.
6. Once the value has been entered, press [ENTER] and proceed to
Entry Confirmation below.

February 2009
1-10 Programming Site Data
Entry Confirmation •To confirm the entries and return to Operate mode, press [F3] (OK).
•To leave the window without confirming the entries, press [F2]
(Cancel) or the [ESC] key.
In either case, the CTF878 returns to Operate Mode.
Entering Pipe
Parameters To enter the Pipe option, scroll to the Pipe tab on the Program Menu
and press [ENTER]. (From the Transducer window, you can scroll back
up to the Transducer tab and press the [X] arrow key to reach the
Pipe window, and press [ENTER].) The screen appears similar to
Figure 1-9 below. To step through each parameter, press the [T] key.
See the Program Menu Map, Figure 1-34 on page 1-37.
Note: Refer to the installation reference Sound Speeds and Pipe Size
Data (914-004) for additional information about pipe sizes
and sound speeds.
Figure 1-9: Pipe Option Window
Pipe
Transducer Fluid Path
Other
5
Material
OKCancel
OD, mm 15.71
ODxPI, mm Wall, mm 2
Sound Speed 600 m/s
Transducer/Pipe
Pipe

Programming Site Data 1-11
February 2009
Entering Pipe
Parameters (cont.)
1. The first prompt asks you to select the pipe material.
a. Press [ENTER] to enter the material prompt.
b. A drop-down list of materials opens as shown in Figure 1-10
below. Press the [T] or [S] keys to scroll to the appropriate
material, or scroll to “Other” for a material not on the list. Press
the [X] key to scroll down by a page, or the [W] key to scroll
up by a page.
c. Press [ENTER] to confirm the choice.
Figure 1-10: Pipe Material Options
Transduc
Mate
OD, mm
Nominal
Steel (1% Carbon)
Steel (1% Carbon, hardened)
Steel (Carbon)
Steel (Mild)
Steel (Stainless 302)
Steel (Stainless 303)
Steel (Stainless 304)
Transduc
Mate
OD, mm
Nominal
Copper
Aluminum (rolled)
Brass
Copper (annealed)
Copper (rolled)
Copper / Nickel (10%)
Copper / Nickel (30%)
Glass (Pyrex)
Transduc
Mate
OD, mm
Nominal
Gold (hard-drawn)
Inconel
Iron (Armco)
Iron (Cast)
Iron (Ductile)
Iron (electrolytic)
Monel
Nickel
Plastic (Acrylic)
Steel (Stainless 347)
Steel (Stainless 316)
Steel (Stainless 304L)
Glass (Flint)
Plastic (Polyethylene)
Titanium
Plastic (CPVC)
OD, mm
Tungsten, annealed
Tungsten, carbide
Tin, rolled
Tungsten, drawn
Nominal
Plastic (Polypropylene)
Plastic (PVC)
Transduc
Mate
Zinc, rolled
Other
Transducer/Pipe
Aluminum
Transducer/Pipe
Transducer/Pipe
Other Transducer/Pipe
Plastic (Nylon)
Glass (Crown)

February 2009
1-12 Programming Site Data
Entering Pipe
Parameters (cont.)
Figure 1-11: Pipe Option Window
2. For pipe diameter, two alternatives are available. The meter asks
for the pipe outside diameter (OD) and wall thickness (Wall). But if
you have selected certain pipe materials (carbon or stainless steel,
cast iron, PVC and CPVC), you have the option of entering the
pipe dimensions by a standardized schedule. Once you enter the
nominal pipe size (Nominal) and identification (Schedule), the
CTF878 determines the OD and wall thickness from an internal
table.
If you want to use the diameter alternative (available for all
materials):
a. You must select from two choices, outside diameter or
circumference.
•The prompt moves to the outside diameter (OD)text box.
Type the OD (in mm or in., whichever is displayed) into the
box, and press [ENTER] to confirm your choice,
or
•Move the prompt to the circumference (ODxPI)text box. Type
the circumference (in mm or in., whichever is displayed) into
the text box, and press [ENTER] to confirm the choice.
b. In either case, the next prompt asks for the wall thickness. Type
the value (in mm or in., whichever is displayed) into the box,
and press [ENTER] to confirm your choice.
Note: The measurement units shown depend on the choices made in
the English/Metric window or the Meter Settings menu. You
can, however, change the measurement units in any text box by
moving to that box and pressing the F1 key.
Pipe
Transducer/Pipe
Pipe
Transducer Fluid Path
OtherMaterial
OKCancel
OD, mm 15.71
ODxPI, mm Wall, mm 2
Pipe
English
Sound Speed 600 m/s
Transducer/Pipe
5
Other manuals for CTF878
1
Table of contents
Other GE Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Endress+Hauser
Endress+Hauser Deltabar FMD71 Description of Device Parameters

PCB Piezotronics
PCB Piezotronics 602D61 Installation and operating manual

Teledyne Lecroy
Teledyne Lecroy Frontline NFC-F user manual

PCB Piezotronics
PCB Piezotronics 356M57 Installation and operating manual

Perel
Perel ABA12 user manual

Kurz
Kurz MFT-B Series manual