GE Healthcare Portascan User manual

Portascan
Portable Bladder Scanner
Service Manual
GE Healthcare
Portascan Service and Assembly Manual
MAN-004-049
Issue 1

Table of Contents
Mediwatch Service Department
Service Documentation
© Mediwatch UK Ltd
Cosford Lane, Rugby CV21 1QN
Phone +44 1788 547888 Fax +44 1788 536434
Section 1
Service Information
Chapter 1. Introduction
About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Chapter 2. Warnings and Precautions
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Chapter 3. Technical Specifications of the Scanner
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Acoustic Output information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Chapter 4. Connector overview
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Main board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
ETX Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Chapter 5. Planned Maintenance
Handling and Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Handling and Care probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Agents and procedures that may damage the probes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Chapter 6. Unit Disassembly
Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Chapter 7. Faults and Solutions
Screen Related . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Probe Related . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Keypad Related . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Software Related . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Chapter 8. Parts list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Chapter 9. ESD
What is ESD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Preventing ESD damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
ESD safe workshop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
ESD safe field service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Circuit Diagrams
Diagram 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Diagram 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Diagram 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix 1
Testing of Final Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Table of Figures
Mediwatch Service Department
Service Documentation
Figure 1
Portascan Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Figure 2
Mainboard Top View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Figure 3
ETX Board Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Figure 4
Portascan Top View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 5
Portascan complete. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 6
Battery release & removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 7
Unscrew 4 screws ... & remove cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 8
Battery Pack & PCB Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 9
Back Panel with release screws arrowed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 10
Note location of long screw. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 11
Probe connector lock nut. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 12
Use special Lemo wrench to loosen probe connector nut. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Figure 13
Cut or remove security sticker then hinge top away to separate case. . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 14
The bottom case and handle grip can be placed to one side. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 15
Two screws hold the main assembly in place. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 16
Lift main board peeling away earth strap & disconnect plugs/skts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 17
Note location of sounder & fan ... and display inverter board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 18
Carefully turn over main panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 19
Shielding removed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Table of Figures
Mediwatch Service Department
Service Documentation
Figure 20
Probe interface panel set. ... Remove screws to separate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 21
Separate ribbon connector, or carefully hinge probe interface boards . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Figure 22
... to reveal ETX Heatsink screws. Retain Heatsink from other side. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Figure 23
The compact flash can now be removed, reprogrammed & replaced. . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Figure 24
RAM can also be accessed if necessary. (Observe E.S.D. precautions) . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Figure 25
Thermal Printer PCB … and location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Figure 26
Paper roll installed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Dual Frequency prove & connector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Rear panel connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Bottom case label. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Keypad membrane. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Section 1
1Mediwatch Service Department
Service Documentation
About this manual
Chapter 1. Introduction
This service manual can be used to service the Portascan at board level. The manual explains the functioning of the boards by
means of functional block diagrams and photographs while connections can be checked at the connector overview.
Preventative and corrective maintenance is also included as are circuit diagrams.
Warnings
Chapter 2. Warnings and Precautions
A probe may only be connected to or disconnected from the scanner while the instrument is switched off. Ignoring
this may cause severe damage to your scanner and/or probe.
To avoid a risk of explosion the equipment must not be operated in the presence of flammable anaesthetics.
To avoid a risk of electric shock do not open the equipment. Refer servicing to qualified personnel only.
Be careful not to place the patient into contact with the ultrasound equipment or other devices. If the ultrasound
equipment or other devices are defective, there is a risk of electrical shock.
For continued protection against fire hazard, replace fuses only with the same type and rating.
The use of non-Mediwatch Plc components with this scanner may result in damage to Mediwatch Plc components.
To prevent hazards; refer to your local requirements for adequate electrical installation in case of class 1 type BF
equipment.
Do not subject the equipment to excessive shock, for example, when moving the equipment. If the equipment is
repeatedly subjected to excessive shock, mechanical parts may be damaged.
Assembly operations, extensions, re-adjustments, modifications or repairs must be carried out by authorized persons.
The electrical installation of the relevant room must comply with the IEC requirements.
The product must be used in accordance with the instructions for use.
Precautions
Cleaning the probe is done by first removing the ultrasound coupling gel with a soft tissue and then gently wipe the
probe dry using a new tissue or dry cloth.
When more cleaning is required only a mild detergent or hand-soap may be used together with some water and a
soft tissue cloth.
To avoid possible damage, the probe cable must not be coiled to a diameter of less than 9 cm (3.5 inch).
Although there is no danger to a patient with an implantable pulse generator (IPG); ultrasonic scanning equipment
could cause mechanical damage to the IPG if used directly over the device's implant site.
Do not use the equipment in locations subject to intense electric or magnetic fields (near transformers, for
example). If the equipment is used in such locations, the monitor will be adversely affected.
Do not use the equipment near devices generating high frequencies (such as medical telemeters and cordless
telephones). If used near such devices, the equipment may malfunction or adversely affect such devices.
To guarantee proper unit operation; do not operate the scanner in an environment with a temperature in excess of
35 °C. If the equipment is used in a small room, the room temperature may rise. Proper ventilation must be
provided.
Avoid installation near a heater or in direct sunlight.
For correct image geometry, only monitors properly adjusted by the manufacturer may be used on the scanner.
Inspect the probe carefully after a drop. A dangerous situation may arise due to damaged insulation of the probe
surface.
To prevent damage to mechanical probes due to excessive heat, a warning mechanism has been built into the system.

Section 1
2Mediwatch Service Department
Service Documentation
Chapter 3. Technical Specifications of the Scanner
General Specifications:
Power Input 10 - 13.5 Vdc
Power Consumption 12 VA
Scanning Method Mechanical Sector
Display Modes B - mode
Probe Frequency 3.5/5.0 MHz
Max Image Depth 20 cm
Scan Converter Full Digital 512 x 512 x 8
Bladder Volume 0 -1500 ml
Accuracy 0 - 699 ml ± 20%, ± 20ml
(whichever is700 ml - 1500ml ± 25% ± 25ml greater)
Accuracy is based upon usage as per instructions and scanning a tissue equivalent phantom
Display Format 8 inch colour TFT LCD 640 x 480 pixels
Battery 12 Volt rechargeable at least 1.5 Hrs continuous operation
from full charge.
Dimensions WxDxH = 16 x13 x 13 cm
Weight 2540g without battery and probe3220g with battery and probe
Enclosure Leakage and Earth Leakage Current Within specification for Class II type BF according to
EN 60601 - 1
Environmental Operating Conditions Temperature 8 - 40ºC Humidity up to 90% (20ºC)
General Storage Conditions Temperature 0 - 50ºC
Standards EN 60601 - 1 & Class II, Type BF handheld equipment
Acoustic Output information for the 3.5/5.0 MHz Mechanical sector (410047):
Parameter: Mode B W o max.
P_ (MPa) 1.2
1spt (mW.cm2) 2.8
System settings 5.0HMz
Min. scan angle
Min. scan depth
1p 53.8
(mm)
Wpb6 (_1_) (mm) 2.4
Prr (KHz) 1708
Srr
(Hz) 2.8
Output Beam Dimensions 520.6
Area
(mm²)
Ø,< (mm), 19.35
(º)
Fawf (MHz) 3.2
Acoustic power up fraction 0
(%)
Maximum Power 2.6
(mW)
lob 0.6
(mW/cm²)
Power up mode B
Initialisation mode B
Acoustic output freeze YES
L tt 7.29
(mm)
Lts CONTACT
(mm)
Inclusive modes B

Section 1
3Mediwatch Service Department
Service Documentation
Backlight
Probe
Board 1 Probe Board 2
Composit
e
USB
Converter
Motherboard
ETX Board With Dimm &
Flash Memory 7 Way Keypa
d
LCD Displa
y
Backlight
Inverte
r
Printe
r
Batter
y
Battery Charge
Control
Uni ver sal DC S upply
(Applied Only When
Charging Removed
Battery Pack
)
Removable Battery Pac
k
Ultrasonic Probe
A
.C. Mains
External Peripherals
Repairable Parts
Serviceable by Module Only
Figure 1. Portascan Block Diagram
Figure 2. Mainboard Top View
Figure 3. ETX Board Block Diagram
Chapter 4. Connector overview
Connections:
Main Board:
See Main Board Schematic for further details
of connector Pin assignments.
EXT Board:
CPU
Geode GX1
1028aseT
Ethernet
Controller
Companion
Chip
CS5530A
DRAM
BIOS
flash
memory
USB0
USB1
Sound
Codec
Hard Disk
Interface
CF Socket
sec. master
Setup
Data
LCD
JLI
CRT
NDA
Floppy
LTP
Mouse
Key-
COM1
COM2
Converter
12C
12C
ISA BUS
PCI BUS
Connector X1
Connector X4Connector X3
Connector X2

Section 1
4Mediwatch Service Department
Service Documentation
Chapter 5. Planned Maintenance
Handling and Care
Although the scanner is produced with the utmost care and only the highest quality components are used, maintenance will be
necessary from time to time to ensure trouble-free operation.
Remove loose dust from the exterior with a soft cloth or a dry brush. A solution of water with a mild detergent may
be used, but not freely applied. Avoid aggressive cleaners.
Check the ventilation fans of the unit for proper operation.
Handling and Care probes
Basic precautions probes
Always follow these basic precautions:
Inspect the probe daily for cracks and other damage.
DO NOT use a probe that has been cracked or damaged.
DO NOT use a probe that has been dropped or struck against another object until it is inspected by a service
engineer.
Avoid pinching or kinking the probe cable. If the probe housing becomes cracked or broken or if there are cuts or
openings in the probe cable the electrical safety of the probe could be compromised.
Disconnect the probe from the scanner before cleaning and disinfection. Make sure to switch the system off before
disconnecting or connecting the probe.
To prevent hazards; refer to your local requirements for adequate electrical installation in case of class 1 type BF
Agents and procedures that may damage the probes.
Some agents and procedures damage probes. Use of any of the following procedures or products WILL VOID your probe
warranty.
Agents that contain the following chemicals are known to damage the probe:
Acetone
Methanol
Denatured ethyl alcohol
Mineral oil
Iodine
Any lotions or gels containing perfume. Check with the ultrasound gel manufacturer regarding gel contents. If you
have additional questions, please contact your representative.
The following procedures are known to damage probes:
Autoclaving
Soaking the probe in chlorine bleach
General probe cleaning
Cleaning the probe is done by first removing the ultrasound coupling gel with a soft tissue and then gently wiping the probe
dry using a new tissue or dry cloth. When more cleaning is required only a mild detergent or hand-soap may be used
together with some water and a soft tissue cloth.

Section 1
5Mediwatch Service Department
Service Documentation
Figure 4. Portascan Top View Figure 5. Portascan complete
Figure 6. Battery release & removal
Figure 7. Unscrew 4 screws & remove cover Figure 8. Battery Pack &
PCB Board
Figure 9. Back Panel with
release screws arrowed
Figure 10. Note location of
long screw
Figure 11. Probe connector
lock nut
Chapter 6. Unit Disassembly
Figure 12. Use special Lemo wrench to loosen probe
connector nut
Figure 13. Cut or remove security sticker then hinge top away
to separate case
Figure 14. The bottom case and
handle grip can be placed to one
side
Figure 15. Two screws hold the
main assembly in place
Figure 16. Lift main board peeling away earth strap &
disconnect plugs/skts.
Figure 17. Note location of sounder & fan … and display
inverter board
Figure 18. Carefully turn
over main panel.
Figure 19. Shielding removed Figure 20. Probe interface panel set. Remove screws to
separate

Section 1
6Mediwatch Service Department
Service Documentation
Chapter 6. Unit Disassembly - continued
Figure 21. Separate ribbon connector, or carefully hinge probe
interface boards
Figure 22. … to reveal ETX Retain Heatsink from other side
Heatsink screws
Figure 23. The compact flash can now be removed,
reprogrammed & replaced
Figure 24. 24 RAM can also be accessed if necessary
(Observe E.S.D. precautions)
Figure 25. The compact flash can now be removed,
reprogrammed & replaced
Figure 26. 24 RAM can also
be accessed if necessary
(Observe E.S.D. precautions)
LCD mounting brackets &
screws
Dual frequency probe &
connector
Rear panel connections Bottom case label Keypad membrane

Section 1
7Mediwatch Service Department
Service Documentation
Chapter 7. Faults and Solutions
Screen Related
LCD blank, can be caused by loose connector at rear of display panel or faulty display.
LCD poor contrast or dark due to failure of inverter board P/N C-00049.
Probe Related
Probe knocking: generally caused by one or more broken wires at the back of probe socket.
Probe not scanning, error message probe too hot: Probe or probe interface board
Keypad Related
Led failure, internal led becomes detached possibly due to operator, replace keypad.
No response from keypad. Due either to connector failure or to short caused by poor etching of the keypad ribbon cable.
Reconnect or replace keypad
Software Related
On the rare occasion that Windows 98SE crashes, the display will freeze and keypad will be inactive. In this situation the
system will automatically shut down after 15 seconds. DO NOT remove the battery pack until this has happened. Removing
the battery pack before complete shutdown could cause the following problems:
Unit boots into windows safe mode, connect external keyboard and follow screen prompts, if problem is not resolved after
two or three reboot attempts then unit requires the compact flash to be reprogrammed.
Fatal exception error, run scandisk to establish integrity of compact flash using the ETX test assembly. If flash disk fails then
replace and reprogram with latest version.
Boot error message writing to drive c, treat as per fatal exception error.

Section 1
8Mediwatch Service Department
Service Documentation
Chapter 8. Parts List
Parts Description Mediwatch Code
Battery Cells PSA00020
Battery Gold Contacts PSA00021
Battery Pack PA00141B
Battery Pack Case (Lower) C00052A
Battery Pack Case (Upper) C00051A
Battery PCB PSA00052
Fan C00068
Flash Memory 256Mb C00169
Flexi-Links (Interface PCB-pair) PSA00065
Frame Grabber PCB C00165
Motherboard PCB PSA 00042
ETX PCB C00148
SDRAM 256Mb C00168B
Interface PCB PA00143A
Inverter Cable C00049
Screening Cans PSA00062
Keyboard Membrane PSA00182
LCD Backlight Inverter C00048
LCD Bracket(Bottom) PSA00015
LCD Bracket(Top) PSA00014
LCD Data Cable PSA00017
LCD Screen C00047
Lid PSA00006A
Magnet Contact PSA00061
Magnet Support Plates PSA00060
Handle Soft Grip PSA00011
Magnet(Print Lid) C00216
Main Case(Lower) C00037A
Main Case(Upper) C00036A
Main Printer Lid C00040A
Printer Bowl C00041A
Printer Pcb Assy. PSA00045
Probe 3.5/5.0Mhz PA00143
Probe Connector Location Washer PSA00072
Probe Connector Mounting Ring PSA00064A
Battery Charger PA00142B
Speaker C00067
Thermal Print Unit C00057

Chapter 9. ESD
What is ESD
Electro Static Discharge (ESD) is the transmission of electro static charges between two bodies with a potential difference. This
transmission can be achieved by direct contact or by an inducted electrostatic field. Electro static discharges are not always
noticed by operators, because ESD is:
sensed at 3,500 Volt
audible at 4,500 Volt
visible at 5,000 Volt
Components and integrated circuits are more sensitive than operators and might be damaged by lower voltage discharges.
Component Type ESD sensitivity in Volts
V-MOS 30-1800
mosFET 100-200
EPROM 100
JFET 140-7000
OP-AMP 190-2500
C-MOS 250-3000
Schottky Diodes 300-2000
Bipolar Transistors 380-7000
ECL 500-1500
SCR 680-1000
Schottky TTL 1000-2500
Preventing ESD damage
Use electro static sensitive parts only in an ESD safe workshop.
Use ESD safe packing material.
Remove all static chargeable materials (plastics) from the workshop.
Wear ESD safe clothing.
Always check wrist strap and other ESD equipment before use.
Be grounded.
BE AWARE.
ESD safe workshop
To be sure that there is no potential difference between body, boards and work surface, the work surface should be made of
electrostatic dissipative rubber. The work surface should be connected to a central ground point. The engineer, handling the
boards should also be connected to the central ground point, either through wrist strap or heal strap (and conductive floor).
ESD safe field service
To enable ESD field safe service an ESD field service kit has to be used whenhandling boards.
This ESD field service kit should be connected to a central ground point and the service engineer should be connected to the
ESD field service kit through a wrist strap.
Handle boards only on this ESD field service kit and pack in ESD safe packing material.
More information about ESD precautions
All precautions against ESD damage are described in the CECC 00015/I regulations, composed by the CECC - Cenelec
electronic Components Committee.
Section 1
9Mediwatch Service Department
Service Documentation

Section 1 Circuit Diagrams
10 Mediwatch Service Department
Service Documentation

Section 1 Circuit Diagrams
11 Mediwatch Service Department
Service Documentation

Section 1
12 Mediwatch Service Department
Service Documentation
Appendix 1
Testing of Final Assembly
Scanner Serial No:
Interface Board Serial No:
Probe Serial No:
Instructions: Pass / fail:
Insert a charged 'test' battery pack and check that it fits securely.
Ensure that the printer door opens and closes smoothly and that the magnetic latch holds the door
securely when the door is closed.
Restrain paper and shake unit and ensure that there are no loose parts inside.
Unplug the probe from the unit by pulling on the release ring on the plug and ensure that
there is no damage to the connector on the unit or the probe plug. Plug the probe in and check
that it clicks securely in place.
Plug a standard PC keyboard into the appropriate socket on the back of the unit. Switch the unit
on by pressing the red power button & wait for the unit to start up.
Hold down 'Alt'& 'Shift'& 'Control' then press 'W' very briefly to disable the Interface board
watchdog timer. Switch off by pressing the red button and wait for the unit to power down.
Remove the keyboard and switch the unit back on by pressing the red button. Wait for the unit
to start up.
Press the display brightness button several times and ensure that the display brightness toggles
between its high and low setting.
Press the menu button and check that the menu options are displayed.
Using the "UP", "DOWN" menu buttons select the "Date/Time" option and then press the
"SELECT" button. Press the "SELECT" button again to select the "Set Date" option. Check that the
correct date is displayed. If the incorrect date is displayed then set the correct date by using the
"UP", "DOWN" and "TAB" menu buttons and lock the setting by pressing the "SELECT" button.
Press the menu button. Using the "UP", "DOWN" menu buttons select the "Date/Time" option and
then press the "SELECT" button. Press the "DOWN" button and then the "SELECT" button to select
the "Set Time" option. Check that the correct time is displayed. If the incorrect time is displayed
then set the correct time by using the "UP", "DOWN" and "TAB" menu buttons and lock the setting
by pressing the "SELECT" button.
Press the menu button. Using the "UP", "DOWN" menu buttons select the "Set Probe Frequency"
option and then press the "SELECT" button. Press the "DOWN" button and then the "SELECT"
button to select the "5Mhz" option. The unit will switch to 5Mhz operation and "5Mhz" will be
displayed at the top of the display.
Press the green button to start scanning.
Align the probe on a phantom bladder so the largest section of the phantom can be seen
on the display. Press the magnification button repeatedly so that the unit cycles through all
levels of magnification then set the magnification so that the phantom fills the display but does
not overlap the edges of the scan sector.
Press the green scan button to record the reading. Check that the reading is within tolerance
(+ or - 10% of indicated phantom volume).
Press the print button to obtain a hard copy of the result.
Press the sagittal scan button and rotate the probe 90 degrees from the position of the first
scan and obtain the largest section and press the sagittal scan button again to obtain the reading.

Section 1
13 Mediwatch Service Department
Service Documentation
Press the print button to obtain a hard copy of the result.
Press the menu button. Using the "UP", "DOWN" menu buttons select the "Set Probe Frequency"
option and then press the "SELECT" button. Press the "DOWN" button and then the "SELECT"
button to select the "3.5Mhz" option. The unit will switch to 3.5Mhz operation and "3.5Mhz" will be
displayed at the top of the display.
Press the green button to start scanning.
Align the probe on a phantom bladder so the largest section of the phantom can be seen
on the display. Press the magnification button repeatedly so that the unit cycles through all levels
of magnification then set the magnification so that the phantom fills the display but dose not
overlap the edges of the scan sector.
Press the print button to obtain a hard copy of the result.
Press the sagittal scan button and rotate the probe 90 degrees from the position of the first
scan and obtain the largest section and press the sagittal scan button again to obtain the reading.
Press the print button to obtain a hard copy of the result.
On the printed results record in the field "Patient Name" the serial number of the unit that is
being inspected.
Open the printer door and raise the print head and remove the paper. Feed the paper back into
the printer until the end appears from the front paper slot and re-engage the print head.
Press the print button. The last recorded image should be printed.
Leave the unit sitting for 5 minutes in which time the display should have shut down. (If the unit
is left for longer than 7 minutes the unit will switch off and the test will need to be repeated in
order to proceed to the next step).
Press any function button to reactivate the display.
Press the red power button. The unit will now shutdown and switch off. Remove the battery pack.
Fit the top cover with protective polythene.
Ensure that the top cover opens and closes smoothly and that the latch holds the cover securely
in place when closed. Ensure that the friction latch holds it from falling forward when fully opened.
Check that the Power drawn by the Battery when the unit is powered down is less than 10mA.
Appendix 1
Testing of Final Assembly - continued
Instructions: Pass / fail:
This Bladder scanner has PASSED/FAILED the above Test Procedure.
(Delete as appropriate)
Signed: ________________. Date: ____/____/________

GE and GE Monogram are
trademarksof General Electric
Company.
GE Medical Systems, a General
Electric company, going to
market as GE Healthcare.
Windows CE™ is either a
registered Trade Mark or a Trade
Mark of Microsoft Corporation in
the United States and/or other
countries.
Portascan manufactured by:
Mediwatch (UK) Ltd.
Swift House
Cosford Lane
Rugby
Warwickshire
CV21 1QN
General Electric Company
reserves the right to make
changes in specifications and
features shown herein, or
discontinue the product
described at any time without
notice or obligation.
Visit us online at:
www.gehealthcare.com
©2005 General Electric
Company
All Rights Reserved
For more than 100 years, scientists and industry leaders
have relied on General Electric for technology, services and
productivity solutions. So no matter what challenges your
healthcare system faces – you can always count on GE
to help deliver the highest quality services and support.
For details, please contact your GE Service Sales Team.
GE Healthcare
Building 4B, 21 South Street
Rydalmere, NSW 2116
Australia
Phone: 1800 223 615
www.gehealthcare.com
Table of contents
Other GE Medical Equipment manuals

GE
GE CARESCAPE R860 User manual

GE
GE DINAMAP Compact TS User manual

GE
GE Modular MAC Trolley User manual

GE
GE SIGNA Voyager User manual

GE
GE LOGIQ C5 Premium User manual

GE
GE LOGIQ P3 PRO User manual

GE
GE Dash 3000 User manual

GE
GE Vivid P3 Installation and user guide

GE
GE Optima XR200amx User manual

GE
GE Senographe SecondLook User manual

GE
GE Datex-Ohmeda S/5 Product manual

GE
GE Precision RXi User manual

GE
GE SENOGRAPHE DMR+ User manual

GE
GE MAC 1600 User manual

GE
GE Voluson 730Pro Quick manual

GE
GE Senographe DS User manual

GE
GE LOGIQ C3 Premium User manual

GE
GE SENOGRAPHE DMR+ User manual

GE
GE auisys cs2 Assembly instructions

GE
GE Marquette Responder 3000 User manual