
6 7
The Six Scanning Views
Superior
Upper Outer
Quadrant
Inferior
Lateral
Medial
AP
Standard Views
View Type Description Transducer Positioning
AP Central tissue: includes
nipple. Nipple is centered.
Lotion is applied to the
entire breast surface.
Center the scanner assembly on the breast using the nipple-positioning arrow
as a guide. Transducer placement is at the inferior edge of the nipple. Bring the
Scanner Assembly straight down on the breast flattening the tissue equally on
all sides.
Lateral Lateral tissue: includes
nipple and superior tissue
to include axillary tail. Nipple
is in the inferior-medial
corner. Lotion is applied
from nipple to lateral
breast surface, to include
axilla
Scanner assembly placement is shifted towards the axilla and laterally. Bring the
scanner assembly straight down on the breast flattening the tissue equally on
all sides. Tilt superior, then laterally making contact with the lateral edge of the
breast. Apply pressure snugging the transducer into the lateral edge. Transducer
should be angled to follow the contour of the body.
If breast shape and body habitus do not allow good superior compression, the
scanner assembly can be placed at the lateral edge and rolled down onto the
breast without displacing the tissue medially. For larger breasts, if the nipple
is outside the field of view, a second lateral view should be done to include the
nipple.
Medial Medial tissue; includes
nipple and inferior tissue
including inframmary fold.
Nipple is superior and
medial to the nipple-
positioning arrow. Lotion
is applied from nipple to
sternum extending below
the inframammary fold.
Move to the opposite side of the patient turning the scanner assembly around.
Position scanner assembly so it extends 4-5 cm (1½-2 in.) below the inframammary
fold. Bring scanner assembly straight down then tilt medially aligning the bottom
edge of assembly with the sternum. Nipple should be superior to the transducer
and slightly above the nipple-positioning arrow. Apply even pressure flattening
the tissue equally then apply pressure snugging the transducer into the medial
edge. Transducer should be angled to follow the contour of the body.
Additional Views
View Type Description Transducer Positioning
Superior Superior tissue: used when
the AP view does not cover
superior tissue. Nipple will
appear in the center at
the inferior edge. Lotion is
applied from the clavicle
to the nipple.
Shift the scanner assembly towards the patient’s head so that the superior edge
of the assembly is just below the clavicle. Center the scanner assembly on the
breast using the nipple-positioning arrow as a guide. Bring the scanner assembly
straight down flattening the tissue equally on all sides. The nipple will appear in
the center at the inferior edge of the active scan area.
Inferior Inferior tissue: nipple will
appear in the center at
the superior edge of the
active scan area. Lotion
is applied from the nipple
to 5 cm (2 in.) below
inframammary fold.
Center the scanner assembly on the breast using the nipple-positioning arrow
as a guide. Shift the scanner assembly inferior so it extends 4-5 cm (1½-2 in.)
below the inframammary fold. Bring the scanner assembly straight down flattening
the tissue equally on all sides, then apply superior pressure flattening out the
inframammary fold. The nipple will appear in the center at the superior edge
of the active scan area.
Upper Outer
Quadrant
Superior and lateral tissue;
includes axillary tail. The
nipple may not appear in
the active scan area or will
be at the lower inner edge.
Lotion is applied to the
upper outer quadrant of
the breast extending into
the axilla.
A rolled towel can be placed behind the scapula to flatten out the armpit. Place
the scanner assembly superior and lateral to the breast extending up into the
axilla. Bring the scanner assembly straight down into the axilla following the
contour of the body. The nipple may not appear in the active scan area or will
be at the lower inner edge.
Depending on the indication, an Invenia ABUS
exam may be Bilateral or Unilateral. A standard
screening exam is Bilateral and is typically
composed of three standard views: AP (Anterior-
Posterior), LAT (Lateral), and MED (Medial).
Depending on breast size, additional views such
as SUP (Superior), INF (Inferior) and UOQ (Upper
Outer Quadrant) may be required to accommodate
all breast tissue.
77