GE druck dpi 104 User manual

Druck DPI 104
Digital pressure indicator
User manual - K394
GE Industrial
Sensing
99 Washington Street
Melrose, MA 02176
Fax 781-665-0780
TestEquipmentDepot.com

A1.1 A1.2
A1.3
A2
1
3
2
4
6
A2
5
7
8
9
10
14
12
16
13 11
15
B1
B2
Customer service
Visit our web site: www.gesensing.com
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com

[EN] English - 1
K394 Issue 1
Table of Contents
Introduction ......................................................................... 1
Safety .................................................................................... 1
Marks and symbols on the DPI 104 ....................................... 1
To Start ................................................................................. 2
Location of items ........................................................................... 2
Items on the display ..................................................................... 2
Prepare the instrument ............................................................... 2
Power on or off ................................................................................ 2
Menu operation ............................................................................... 3
Installation ........................................................................... 4
Battery .................................................................................................. 4
Position ................................................................................................ 4
Pressure connections ................................................................... 4
Electrical connections .................................................................. 4
Operation ............................................................................. 5
Menu: Set units ................................................................................ 5
Menu: Set tare .................................................................................. 5
Menu: Monitor maximum/minimum ..................................... 5
Menu: Monitor a pressure switch ............................................ 6
Menu: Calibration ........................................................................... 6
Menu: Set low/high alarm .......................................................... 6
Menu: Supply voltage output (Vout) ...................................... 6
Menu: Set Vout scale factor ...................................................... 7
Menu: Set automatic power OFF ............................................. 7
Menu: Set lock code ....................................................................... 7
Menu: Set scan rate ....................................................................... 7
Menu: Monitor external IDOS .................................................... 7
Menu: Set FSO low/high registers ........................................... 8
Set up a DPI 104 network ........................................................... 8
Error indications .............................................................................. 8
Maintenance ........................................................................ 9
Replace the batteries .................................................................... 9
Restore the original configuration .......................................... 9
Calibration ............................................................................ 9
Equipment and conditions ......................................................... 9
Procedures ......................................................................................... 9
Specification data ............................................................. 11
General ............................................................................................. 11
Pressure measurement ............................................................ 11
Electrical .......................................................................................... 11
Customer service ............................................... Back cover
© 2005 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.
Introduction
The Druck DPI 104 is a digital pressure indicator that
measures the pressure of liquid, gas or vapour and shows
the pressure value on a liquid crystal display (LCD). It also
has the Intelligent Digital Output Sensor (IDOS) technology
to use data from a Universal Pressure Module (UPM).
The DPI 104 includes these functions:
Safety
Before you use the DPI 104, make sure that you read and
understand all the related data. This includes: all local
safety procedures, and this publication.
WARNING
• Some liquid and gas mixtures are dangerous. This
includes mixtures that occur because of
contamination. Make sure that the DPI 104 is safe to
use with the necessary media.
• It is dangerous to ignore the specified limits for the
DPI 104 or to use the DPI 104 when it is not in its
normal condition. Use the applicable protection and
obey all safety precautions.
• To prevent a dangerous release of pressure, isolate
and bleed the system before you disconnect a
pressure connection.
• Do not use the DPI 104 in locations with explosive
gas, vapour or dust. There is a risk of an explosion.
Before you start an operation or procedure, make sure
that you have the necessary skills (if necessary, with
qualifications from an approved training establishment).
Follow good engineering practice at all times.
Safety - Marks and symbols on the DPI 104
Function
* Measure pressure - Accuracy: 0.05% full scale (FS)
* Refer to “Specification data”.
** Optional item
Large 5-digit main display with 11 pressure units
Adjustable Full Scale Output (FSO)
20 segment analogue dial in increments of 5% FSO
(large division marks = 10% increments).
2.5 digit percentage indicator (0-100% FSO)
8-pin connector port: For RS232, **IDOS UPM, external power
supply
Alarm output for high/low pressure conditions
Switch input to monitor an external pressure switch
Analogue voltage output (Vout): 0 - 5 Vdc
Other functions: Maximum/minimum, tare, Vout scale factor,
automatic power off
Complies with European Union directives

2- [EN] English K394 Issue 1
To Start
To start - Location of items
To start - Items on the display
Continued
To start - Prepare the instrument
Before you use the instrument for the first time:
• Make sure that there is no damage to the instrument,
and that there are no missing items.
• Install the battery (refer to “Installation”). Then
re-attach the display bezel [A1: item 6].
To start - Power on or off
Press the buttons in the sequence shown below.
•Poweronsequence:
• Power off sequence:
When the power is off, the last set of configuration options
stays in memory.
Note: The DPI 104 uses a small quantity of power while it is
OFF. If you put it into storage for a long period, disconnect
the battery (refer to “Installation”).
Item Description
1. 8-pin connector for external power supplies,
RS232/UPM connections and signal input/output.
2. •Power on button
•Menu mode: Press for 2 seconds to show the first
menu option. To move down the menu structure,
press again and again, or press and hold.
•Reject or stop the change to a value.
•In Maximum/minimum mode. Press to show the
maximum and minimum values since the last reset.
= maximum = minimum
3. In menu mode:
- On/OFF selection
- Increase/decrease a value
- Move the decimal left/right
4. Pressure sensor and connector with 320° of turn:
gauge (g), absolute (a) or Sealed gauge (sg). Refer to
“Specification data”.
5. •In menu mode:
- Accepts a menu selection
- Shows the next menu level
- Accepts a value
•In Tare mode: Tare the pressure value on the display.
•In Maximum/minimum mode. Reset the
maximum/minimum values.
6. Display bezel with 348° of turn.
7. O-ring.
8. Battery connector
9. Battery: 9 V Alkaline (supplied but not installed). Refer
to “Installation”.
10. Battery clamp with two screws.
Item Description
11. 5-digit main display.
12. 2.5-digit percentage indicator (0-100% FSO).
%FSO = [ Applied Pressure /(FSOHigh—FSOLow)] * 100
13. 20 segment analogue dial in increments of 5% FSO
(large division marks = 10% increments).
%FSO = [ Applied Pressure /(FSOHigh—FSOLow)] * 100
14. The units of measurement: kPa, MPa, kg/cm2, psi,
mbar, bar, mmHg, mmH2O, mH2O, inH2O, inHg
A1
A2
Item Description
15. Mode indication.
Voltage output (Vout) mode - On.
RS232 connection. The data transmit/receive function
is active.
Switch mode - On. To monitor an external pressure
switch.
= switch closed = switch open
Menu Lock mode - On. To restrict access to the menu
functions.
Alarm mode - On. The symbol flashes when the
measured value satisfies one of the alarm conditions.
= High alarm = Low alarm
Maximum/minimum mode - On.
IDOS UPM mode - On. To monitor pressure from a UPM
16. Low battery power indication: Battery life ≤15%.
1First display =
FS limit
Then:
Normal output
Normal output 1Menu: 2
OFF
(For 2 seconds)

To start - Menu operation
Menu Description Steps Result/Subsequent steps
1 2
OFF =Power supply: OFF only - Power goes off
unitS =Set units: (A2: item 14). Pressure value changes to the applicable units:
psi, mbar, bar ...
t OFF =Set tare: Set to On or OFF. On ➤tA 00.000 : Set a tare value (Refer to Table 4)
OFF =Monitor maximum/minimum:
Set to On or OFF
Monitor function is set on or off
OFF =Monitor a pressure switch:
Set to On or OFF.
Monitor function is set on or off
C _ _ _ _ =Calibration: To continue, set the
correct calibration access code =
Serial number: Last four digits.
C0 (Correct the zero offset value) ➤
C2 (Do a two-point pressure calibration) ➤
V2 (Do a two-point voltage calibration). Refer to “Calibration”.
A OFF =Set low/high alarm: Set to On or OFF. On ➤000.0 ↓ ➤100.0 ↑
Set a value for the low and/or high alarm (0 to 105% FSO).
OFF =Supply voltage output (Vout):
Set to OFF, P-V, or US.
P-V: Vout is proportional to the pressure value on the display. Make
sure the Vout scale factor is correct.
US ➤000.0:Set a Vout value (0 to 100%) to control an external
pressure regulator. Make sure the Vout scale factor is correct.
S 1.00 =Set Vout scale factor: A Vout
adjustment.
If applicable, set a new Vout scale factor (0.01 to 9.99).
Factory value = 1.00
Au On =Set automatic power OFF:
Set to On or OFF.
On ➤Au 15 : Set the period for automatic power OFF (1 to 99
minutes). Factory value = 15 minutes.
L OFF
=Set lock code: A menu protection
facility. Set to On or OFF.
On ➤L 000 : Set a new lock code (If necessary).
Factory code = 000.
Sc 02 =Set scan rate: A rate to update the
maximum/minimum values
Set an applicable rate (02 to 10 Hz).
Factory value = 02 Hz.
OFF
=Monitor external IDOS: Set to On or
OFF.
Monitor function is set on or off
FS ↓=Set FSO low register: To set a
different range for these functions:
analogue display, %, alarm.
Set a value for the low end of the range (Refer to Table 5).
Factory value = Factory calibration value.
FS ↑=Set FSO high register: To set a
different range for these functions:
analogue display, %, alarm.
Set a value for the high end of the range (Refer to Table 5).
Factory value = Factory calibration value.
Normal
display
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com

4- [EN] English K394 Issue 1
Installation
This section shows how to install and connect the DPI 104.
Before you start:
• Read and understand the “Safety” section.
• Do not use a damaged DPI 104.
Installation - Battery ...
Use the procedures in Table 1 to install or replace the
battery.
Table 1: Installation procedures - Battery
Installation - Position
Make sure that you attach the DPI 104 in a safe
configuration that prevents unwanted stress (for example
vibration, physical impact, shock, mechanical and thermal
stresses).
To get the best installation position, you can turn the
pressure connector (A1: item 4) and the display bezel
(A1: item 6) to give the best view of the display (Figure B1).
End stops set the limits in each axis.
CAUTION: To prevent damage when you are setting the
best view of the display, do not use force to turn the
pressure connector or the bezel farther than the end
stops.
Installation - Pressure connections
CAUTION: To prevent damage, do not use the body of the
DPI 104 to tighten the pressure connection. Use the flat
faces on the pressure connector.
Use an applicable method to seal the pressure
connections, and then tighten to the applicable torque
(Figure 1 and Table 2).
Figure 1: Connection methods
Table 2: Key to figure 1
Installation - Electrical connections
The DPI 104 includes an 8-pin electrical connector
(A1: item 1). Table 3 shows the pin connections.
Table 3: Connections for the 8-Pin connector
These optional accessories use the connector:
• RS232 PC connection
• external IDOS UPM connection + power supply adaptor
Note: Use only original parts supplied by the manufacturer.
You can also use the RS232 interface to make a serial
network of units (maximum: 99). Refer to “Operation - Set
up a DPI 104 network”.
Electrical connections - External power
We recommend you use an external power supply for
these functions and operations:
• Functions: Maximum/minimum, Switch,
Low/high alarm, Vout, IDOS
• Operations that use the DPI 104 for long periods
Step Procedure
1If applicable, set the power to off and isolate the
external power supply.
2Remove the display bezel (Figure A1.2).
3Make sure that the o-ring [A1: item 7] and the related
surfaces are serviceable. Use only original parts
supplied by the manufacturer.
4Remove the battery clamp [A1: item 10].
5If applicable, disconnect the battery connector
[A1: item 8] and *discard the used battery.
* Use an applicable recycling facility.
6Attach the battery connector [A1: item 8] to the new
battery.
7Install the new battery (Figure A1.3) and re-attach
the battery clamp [A1: item 10].
8Push the display bezel [A1: item 6] back into position
until it is fully engaged.
A1.2 A1.3
Item Description
1. Applicable DPI 104 pressure connector. Maximum
torque:
1/4 NPT: 68 Nm (50 lbf.ft)
G1/4: 20 Nm (15 lbf.ft)
9/16 x 18 UNF cone: 34 Nm (25 lbf.ft)
2. (1/4 NPT only) Thread with an applicable sealant
3. (G1/4 only) Applicable bonded seal
Connector Pin Input/
Output
Description
1. Input 12- 24V dc power supply
2. Input Signal ground
3. Output RS232 transmit
4. Input RS232 receive
5. Output Voltage output
6. Output Alarm output
7. Input Pressure switch input
8. Output No connection
1
3
1
2
1c) 9/16 x 18 UNF cone:
Pressure
≥
1000 bar (15000 psi)
a) 1/4 NPT:
Pressure
<
1000 bar (15000 psi)
b) G1/4:
Pressure
<
1000 bar (15000 psi)

[EN] English - 5
K394 Issue 1
Operation
This section shows how to use the DPI 104. Before you
start:
• Read and understand the “Safety” section.
• Make sure that the installation is complete (Refer to the
“Installation” section).
• Do not use a damaged DPI 104.
Operation - Menu: Set units
There are 11 different units to measure pressure. Refer to
the “Specification data” section.
Units - Set up
Refer to “To start - Menu operation”.
Operation - Menu: Set tare
Use the tare function to adjust the pressure value on the
display. For example: To make an adjustment for
atmospheric pressure. Refer to Table 4.
Table 4: Permitted tare values
If you set a value that is not in the permitted range, the
value is reset to the nearest applicable limit.
Tare - Set up and use
Menu: Set this function to On (Refer to “To start - Menu
operation”).
When this function is On, there are two options to set a tare
value (tA):
•Menu option: Set the menu “t On”, then set a tA value:
Repeat steps 1 + 2for each digit and for the decimal
point.
•Zero option: Step 1lets you set a value for tA. Press and
hold.
When tA is not zero, the last segment on the analogue dial
flashes.
To make sure that there is an indication of the correct
pressure while tare is On, the analogue dial and %
indication show values calculated from the calibrated
range without the tare adjustment.
Tare - With Lock
If the menu lock is On with a lock code set < 500, the zero
option is rejected - Error code (E0002).
Tare - With Alarm and/or Vout
If you set a tare value (tA) while the alarm and/or Vout
functions are On, the display counts down from:
tArE9 to tArE0.
If the tA value is set, the alarm and Vout functions start to
use values calculated from the calibrated range.
Tare - With FSO values
To make sure that there is an indication of the correct
pressure while tare is On, the FSO Low and/or FSO High
values are not used.
Operation - Menu: Monitor maximum/minimum
When this function is On, it updates its data at five times
the specified scan rate (02 to 10 Hz). Refer to “To start -
Menu operation”.
To save battery power, we recommend that you use an
external power supply with this function.
Maximum/minimum - Set up and use
Menu: Set this function to On (Refer to “To start - Menu
operation”).
When this function is On, use steps 1+ 2to show the
maximum/minimum since the last reset.
Step 3lets you reset the values for maximum/ minimum.
Press and hold.
Range Permitted tare values
All ranges: a, sg 0 to 105% FS
g: 0.7 bar (10 psi) -0.7 bar (-10 psi) to 105% FS
g: ≥2 bar (30 psi) -1 bar (-15 psi) to 105% FS
tA display 12 tA display
0 to 9, or - 0 to 9
Normal
output 1Normal
output
donE
tA = 0 tA = 70 mbar
To cancel the specified tA value, press this button
OR let the count complete.
To continue with the specified tA value, press this
button.
1maximum 2minimum 3
donE
Reset

6- [EN] English K394 Issue 1
Operation - Menu: Monitor a pressure switch
Use this function to measure the performance of a
pressure switch (mechanical operation and hysteresis).
To save battery power, we recommend that you use an
external power supply with this function.
Pressure switch input - Set up and use
1. Connect the instrument (Figure 2/Table 3).
2. Menu: Set this function to On (Refer to “To start - Menu
operation”).
Figure 2: Example configuration - Switch input
Figure 2 shows the display when the switch condition
changes (open or closed). The analogue dial and the %
indication continue to monitor the normal pressure.
The switch symbol and the value on the main display flash
to give the switch condition and the switch pressure.
Operation - Menu: Calibration
Refer to the “Calibration” section.
Operation - Menu: Set low/high alarm
Use the alarm function to show when the pressure is not in
the specified limits for the system.
Set applicable values in the range 0 to 105% FSO:
Note: If you set a tare value, the alarm function uses the
calibrated range (Refer to “Operation - Menu: Set tare”).
The alarm indication is available on the display and as a
signal output (Table 3).
While there is an alarm condition, the applicable alarm
symbol (high or low) flashes on the display (A1: Item 15).
To save battery power, we recommend that you use an
external power supply with this function.
Low/high alarm - Set up and use
Menu: Set this function to On (Refer to “To start - Menu
operation”). Then use these steps to set the low and/or
high alarm.
5. To finish, repeat steps 3 + 4for each digit.
If the value you enter is not correct, the value resets to the
nearest permitted value. That is:
• a value in the range 0 to 105% FSO
• a low alarm value < high alarm value
To accept or change the new value, repeat steps 1 to 5.
Operation - Menu: Supply voltage output (Vout)
Use the Vout function to supply a voltage output (0 to 5V)
to an external system. There are two options:
To save battery power, we recommend that you use an
external power supply with this function.
P-V mode voltage calculation
Example DPI 104: FSO = 20 bar (or 300 psi),
Vout scale factor = 1.00.
If you apply 10 bar (or 150 psi) to this DPI 104:
US mode voltage calculation
This calculation uses the values set up for the Vout register
and the Vout scale factor.
If the pressure ranges for the DPI 104 and the regulator are
different, set a new Vout scale factor (refer to “Operation -
Menu: Set Vout scale factor”).
Example if the Vout register is set to 25%, and the Vout
scale factor is set to 0.5:
Voltage output (Vout) - Set up and use
1. Connect the instrument (Figure 3).
2. Menu: Set this function to OFF, P-V, US (Refer to “To
start - Menu operation”).
To reset the monitor function, press this button.
%FSO = [ Applied Pressure / (FSO High — FSO Low) ] * 100
Pin 1 (+), 2 (GND)
12- 24V dc
Pin 7 (Switch)
Pin 2 (GND)
Alarm
(% FSO) 12 Alarm
(% FSO) 34
Digit = 0 or 1 Digit = 0 to 9
To cancel the new value, press this button.
P-V: Vout is proportional to the pressure value on the
display.
US: User mode. Set a value in the Vout register (0 to
100%) to control an external pressure regulator.
Vout = ( 10/20 * 5V ) / 1.0 OR ( 150/300 * 5V) / 1.0 = 2.5V
Vout = ( Vout register / 100 ) * 5V
Vout scale factor
Vout = ( 25/100 * 5V ) / 0.5 = 2.5 V

[EN] English - 7
K394 Issue 1
Figure 3: Example configuration: Vout
Operation - Menu: Set Vout scale factor
When the Vout function is set to P-V or US mode, the Vout
scale factor becomes part of the Vout calculation (refer to
“Operation - Menu: Supply voltage output (Vout)”).
If the pressure ranges for the DPI 104 and the external
pressure regulator are different, you must set an
applicable scale factor (0.01 to 9.99).
Example - To get a 25 bar (or a 375 psi) line pressure with:
• an External pressure regulator:
FSO = 100 bar (or 1500 psi)
• a DPI 104: FSO = 200 bar (or 3000 psi)
In this example:
To get a 25 bar (or a 375 psi) line pressure, the DPI 104 uses
these values to supply the Vout value shown below:
Operation - Menu: Set automatic power OFF
Use this function to save battery power. The power goes
off a specified period after the last button or external
software operation. To get the maximum battery life, we
recommend you use this function.
Note: The DPI 104 uses a small quantity of power while it is
OFF. If you put it into storage for a long period, disconnect
the battery (refer to “Installation”).
Automatic power OFF - Set up and use
Menu: Set this function to On. Then set an applicable value
in the range 1 to 99 minutes. Refer to “To start - Menu
operation”.
Note: If continuous operation is important, set this function
to OFF and use an external power supply.
Operation - Menu: Set lock code
Use the lock function to prevent accidental changes to the
configuration. There are two options:
•Lock code < 500: This locks the menu and the tare
function. Factory code = 000
•Lock code > 499: This locks the menu but you can still
use the zero option to set a tare value.
Refer to “Operation - Menu: Set tare”.
Lock code - Set up and use
Menu: Set this function to On (Refer to “To start - Menu
operation”). Then use these steps to set a new code.
3. To finish the lock code, repeat steps 1 + 2for each digit.
The next time you want to change the menu options, the
display shows: L - - -
Enter the applicable code. To reset the code to the factory
code, you must do a restore operation. Refer to
“Maintenance”.
Operation - Menu: Set scan rate
Use the scan rate function to set the rate used to update
the maximum/minimum values.
Note: When you increase the scan rate, you increase the
power consumption. The maximum/minimum function
updates its data at five times the specified scan rate.
The rate for the normal pressure display is always 2 Hz.
Scan rate - Set up and use
Menu: Set an applicable value in the range 2 to 10 Hz. Refer
to “To start - Menu operation”.
Operation - Menu: Monitor external IDOS
Use this function to read the pressure from an external
IDOS UPM. The calibration function is not available but you
can use all the other DPI 104 pressure functions. Example:
Set tare, Monitor maximum/minimum.
This function does not supply power to the IDOS UPM. To
use it, you must have the optional external IDOS UPM
connection + power supply. This optional accessory uses
the RS232 connections in the 8-pin connector (Table 3).
Monitor external IDOS - Set up and use
1. Connect the instrument (Figure B2).
2. Menu: Set this function to On (Refer to “To start - Menu
operation”).
3. If necessary, set the applicable additional functions
specified in this manual.
Scale factor = 100/200 OR 1500/3000 = 0.5
Vout register (%)
(DPI 104) = (25/200) * 100 OR (375/3000) * 100
= 12.5%
Vout = ( 12.5/100 * 5V ) / 0.5 = 1.25 V
Pin 5 (Vo)
Pin 2 (GND)
Pin 1 (+), 2 (GND)
12- 24V dc
Vout: 0 to 5V
Lock 12 Lock
Digit = 0 or 9 Digit = 0 to 9

Operation - Menu: Set FSO low/high registers
Use the FSO low/high registers to set a different range for
these functions: analogue display, % indication, low/high
alarm.
Initially, these register values are set to the factory
calibration values. Example:
Calibrated range: 0.7 bar (10 psi) gauge.
Selected units: mbar
Table 5 gives the permitted alternative values you can use.
Table 5: Permitted FSO values
FSO low/high registers - Set up and use
Menu: Setthe menu option to the FSO low register (Refer to
“To start - Menu operation”). Then use these steps to set an
applicable value in the permitted range (Table 5):
3. Repeat steps 1 + 2for each digit and for the decimal
point.
If the value you enter is not correct, the value resets to the
nearest permitted value (Table 5).
To accept or change the new value, repeat steps 1 to 3.
4. If necessary, repeat the procedure for the FSO high
register.
Operation - Set up a DPI 104 network
You can set up a network of up to 99 units in series (‘daisy
chain’). Figure 4 showsthe electrical connections to do this
(Refer to Table 3).
Figure 4: Connections for a DPI 104 network
Operation - Error indications
Table 6: Error codes/indications
FSO low FSO high
0 bar (0 psi) 700 mbar (10 psi)
Range Permitted FSO values
All ranges: a, sg 0 to 105% FS
g: 0.7 bar (10 psi) -0.7 bar (-10 psi) to 105% FS
g: ≥2 bar (30 psi) -1 bar (-15 psi) to 105% FS
All ranges FSO low < FSO high
FSO low 12 FSO low
0 to 9, or - 0 to 9
To cancel the new value, press this button.
Code Description
E0001 Incorrect unlock code. Use the correct code.
E0002 The tare facility is not available because the menu
lock is On and the lock code < 500. Change the menu
configuration.
E0004 Start up error. Do a restore operation (Refer to
“Maintenance”).
E0005 External IDOS UPM not found. Make sure that all the
related equipment and connections are serviceable.
E0006 Incorrect calibration access code. Use the correct
code.
E0007 The power supply is too low to do a calibration. Use an
external power supply or replace the battery.
E0009 Unable to supply the specified Vout. Example:
• Low battery. Use an external power supply or
replace the battery.
• Bad connection. Make sure that all the related
equipment and connections are serviceable.
OLoAd Applied pressure ≥110% FS. Reduce the pressure.
99999/
-9999
There are not enough digits in the main display to give
the correct pressure value. Change the measurement
units.
DPI 104 (1)
RxGND+ve
PC: 9-way, D-type
connector
12- 24V dc
DPI 104 (2)
GND Tx
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com

[EN] English - 9
K394 Issue 1
Maintenance
Clean the case with a moist, lint-free cloth and a weak
detergent. Do not use solvents or abrasive materials.
Make sure that there is no damage to the threads and
O-rings, and that they are free of grit and other
obstructions.
You must return the unit to the supplier for all repairs.
Maintenance – Replace the batteries
To replace the batteries, refer to “Installation”. All the
configuration options stay in memory.
Maintenance – Restore the original configuration
If it is necessary to restore the unit to the original factory
configuration, press and hold all four buttons until the
display goes off ( ≈five seconds). The unit then restarts.
“To start - Menu operation” shows the factory settings. The
lock code is reset to the factory code (000).
Calibration
Note: GE can provide a calibration service that is traceable
to international standards.
We recommend that you return the DPI 104 to the
manufacturer or an approved service agent for
calibration.
If you use an alternative calibration facility, make sure that
it uses these standards.
Calibration - Equipment and conditions
To do an accurate calibration, you must have:
• the calibration equipment specified in Table 7.
• a stable temperature environment: 20 ± 1°C (68 ± 2°F)
Table 7: Calibration equipment
Calibration - Procedures
1. Connect the applicable calibration equipment (Table 7).
2. Menu: Set the menu option to C _ _ _ _ . Then set the
calibration access code = last four digits of the serial
number (Refer to “To start - Menu operation”).
There are three calibration options (Table 8):
Table 8: Calibration options
Calibration - C0 (Zero offset)
The DPI 104 shows these displays:
1. The calibration point to be used for C0. This value is
only adjustable for an absolute type DPI 104 (Table 8).
C0 - Gauge = 0000.0
5. Repeat steps 3 + 4for each digit and for the decimal
point. The value is ignored if it is not in the permitted
limits (Table 8).
This value is then used as the Set Point (SP) on the
subsequent displays.
6. This sequence of displays will follow:
The SP value is followed by the measured pressure -
Current Pressure (CP). This sequence continues until you
accept or reject the offset value.
7. When the pressure is stable:
The value is ignored if it is not in the permitted limits (5% FS)
or if the CP value is not stable.
Function Calibration equipment
Pressure An applicable pressure standard (primary or
secondary) with a total uncertainty of 0.01% reading
or better.
Make the pressure connection to A1 - item 4. Refer to
“Installation”.
Volts (V) Volts calibrator.
Accuracy: 0.025% reading or better.
Make the Vout connection to A1 - item 1. Refer to
“Installation”
Option Description
C0: Set the necessary offset value for the instrument to give
the correct pressure related to zero:
All ranges g or sg: Zero (bar/psi);
Ranges a: Ambient pressure*
* adjustable by 5% FS; ** adjustable by 50 mV
C2: Do a two-point pressure calibration.
All ranges g or sg: P1 = Zero (bar/psi); P2* = FS
Ranges a: P1* = Ambient pressure; P2* = FS
V2: Do a two-point voltage calibration.
All ranges: P1** = 0.1000 V; P2** = 5.0000 V
To move to the next option without a change to the
values, press this button.
To stop and make changes to a value, press this
button.
To return to the normal display, wait eight seconds.
C0 - Absolute 2C0 - Absolute 34
8 seconds 0 to 9
C0
Example sequence: Absolute type
To accept the new offset value, press this button.
The display shows “donE”, and then the next
calibration option (C2).
To reject the new offset value and move to the next
calibration option (C2), press this button.

10 - [EN] English K394 Issue 1
Calibration - C2 (two-point pressure calibration)
Point 1 (P1) - The DPI 104 shows these displays:
1. The calibration point to be used for C2 - Point 1. This
value is only adjustable for an absolute type DPI 104
(Table 8). C2 - Point 1 (Gauge) = 0000.0
5. Repeat steps 3 + 4for each digit and for the decimal
point. The value is ignored if it is not in the permitted
limits (Table 8).
This value is then used as the Set Point (SP) for point 1
on the subsequent displays.
6. This sequence of displays will follow:
The SP value is followed by the measured pressure - CP.
This sequence continues until you accept or reject the
point 1 value.
7. When the pressure is stable:
The value is ignored if it is not in the permitted limits (5% FS)
or if the CP value is not stable.
Point 2 (P2) - Use the same steps (1to 5above) to set
C2 - Point 2. This is the FS value and it is adjustable for the
absolute and gauge type DPI 104 (Table 8).
6. This sequence of displays will follow:
The SP value is followed by the measured pressure - CP.
This sequence continues until you accept or reject the
point 2 value.
7. When the pressure is stable:
The value is ignored if it is not in the permitted limits (5% FS)
or if the CP value is not stable.
Calibration - V2 (two-point voltage calibration)
Point 1 (P1) - The DPI 104 shows these displays:
1. The calibration point to be used for V2 - Point 1.
After step 2, the DPI 104 sets Vout to 0.1 V. Correct the
value (P1) to the value shown on the voltage calibrator.
5. Repeat steps 3 + 4for each digit. The value is ignored if
it is not in the permitted limits (Table 8).
Point 2 (P2) - If P1 has a permitted value, the DPI 104 shows
these displays:
1. The calibration point to be used for V2 - Point 2.
After step 2, the DPI 104 sets Vout to 5.0 V. Correct the
value (P2) to the value shown on the voltage calibrator.
5. Repeat steps 3 + 4for each digit. The value is ignored if
it is not in the permitted limits (Table 8).
6. If P2 has a permitted value, the DPI 104 uses the new
P1/P2 values to adjust the output (Vt):
C2 - Absolute 2C2 - Absolute 34
8 seconds 0 to 9
P1
Example sequence: Absolute type
To accept the new P1 value, press this button. The
display shows the calibration point C2 - point 2 (C2).
To reject the new P1 value and move to the next
calibration option (V2), press this button.
P2
Example sequence: Absolute type
To accept the new P2 value, press this button. The
display shows “donE”, and does a two-point
calibration. The instrument then restarts.
To reject the new P2 value and move to the next
calibration option (V2), press this button.
234
8 seconds 0 to 9
234
8 seconds 0 to 9
To accept the V2 calibration, press this button. The
display shows “donE”. The instrument then restarts.
To reject the V2 calibration and move to the next
menu option, press this button.

Specification data
Specification - General
Specification - Pressure measurement
Specification - Electrical
Operating temperature -10 to 50°C (14 to 122°F)
Storage temperature -20 to 70°C (-4 to 158°F)
Ingress Protection IP65 (Dust-tight, jets of water)
Materials Case: Acrylonitrile Butadiene Styrene (ABS)
Refer also to Media notes.
Humidity 0 to 95% without condensation
(Def Stan 66-31, 8.6 cat III)
Shock/Vibration BS EN 61010:2001; Def Stan 66-31, 8.4 cat III
EMC BS EN 61326-1:1998 + A2:2001
Safety Electrical - BS EN 61010:2001;
Pressure Equipment Directive - Class: Sound Engineering Practice (SEP);
Approved CE Marked
Size Diameter = 95 mm (3.74 in); Depth = 55 mm (2.2 in)
Typical length (with connector) ≈120 mm (4.7 in)
Weight 350 g (12.5 oz)
Power supply 9V, Alkaline (MN1604) - Supplied; OR Use an external 12-24 V dc supply
Note: For maximum battery life, use a 9V, Li (MN1604).
Battery life Up to one year for pressure measurements: Au (power save facility) - On;
maximum/minimum, alarm, Vout, switch - All set to OFF
Range: gauge (g), absolute (a),
sealed gauge (sg)
Resolution Maximum Working
Pressure (MWP)
Media
notes
bar* psi* Type mbar psi bar psi
(-0.7) 0 to 0.7 (-10.0) 0 to 10 g* 0.01 0.001 4 60 1
(-1.0) 0 to 2.0 (-15.0) 0 to 30 g* or a 0.1 0.001 8 120 1
(-1.0) 0 to 7.0 (-15.0) 0 to 100 g* or a 0.1 0.01 28 400 2
(-1.0) 0 to 20 (-15.0) 0 to 300 g* or a 1 0.01 80 1200 2
(-1.0) 0 to 70 (-15.0) 0 to 1000 g* or a 1 0.1 140 2000 2
0 to 200 0 to 3000 sg 10 0.1 400 6000 2
0 to 350 0 to 5000 sg 10 0.1 700 10000 2
0 to 700 0 to 10000 sg 10 1 900 13000 2
0 to 1000 0 to 15000 sg 100 1 2000 29000 3
0 to 1400 0 to 20000 sg 100 1 2000 29000 3
* Negative ranges shown in ( ... ) - gauge units only.
Media notes:
1. Non-corrosive, non-conductive liquid or Non-corrosive, dry gas
2. Media applicable to stainless steel (316)
3. Media applicable to Inconel 625
Accuracy (0 to FS) 0.7 bar (10 psi): 0.15% FS
All ranges ≥2 bar (30 psi): 0.05% FS
Units kPa, MPa, kg/cm2, psi, mbar, bar, mmHg, mmH2O, mH2O, inH2O, inHg
Pressure
connections
Ranges ≤700 bar (10000 psi): 1/4 NPT male OR G1/4 male
Ranges >700 bar (10000 psi): 9/16 x 18 male cone
Switch input Maximum impedance: 200Ω(mechanical contact only)
Alarm output Type: Open drain Field Effect Transistor (FET)
Maximum (mA): 250 mA; Maximum (V): 24 V dc
Analogue output 0 to 5 V dc; Minimum load: 500 Ω; Maximum zero offset error: 50 mV
Accuracy: 0.1% FS at 20°C (68°F) - User mode only.
Temperature coefficient: 0.007% FS / °C (0.0035% FS / °F)
RS232/Network capacity For: IDOS UPM, external software, or up to 99 units in series (‘daisy chain’).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Other manuals for druck dpi 104
1
Table of contents
Other GE Touch Panel manuals

GE
GE IC754VSL06CTD User manual

GE
GE SIMON XT User manual

GE
GE K0470 User manual

GE
GE ATP2100 User manual

GE
GE IC754VSI06STD User manual

GE
GE QuickPanel+ IC754VGI06STD User manual

GE
GE Druck DPI 104-IS Instruction Manual

GE
GE IC755CxW07CD SERIES User manual

GE
GE IC754CSL06MTD User manual

GE
GE Druck PACE1000 Instruction Manual

GE
GE IC754VBI12MTD User manual

GE
GE 17453- - Motion Sensing Utility Light 1 LED User manual

GE
GE IC754VGB06MTD Installation manual

GE
GE 60-924-RF-TS User guide

GE
GE SIMON XT Assembly instructions

GE
GE HiTech User manual

GE
GE druck dpi 104 User manual

GE
GE IC754VSB06MTD Installation manual

GE
GE IC754CSL12CTD User manual

GE
GE Druck DPI 150 Use and care manual