
20 NRR 2-50, NRR 2-51 - Installation & Operating Manual - 819180-05
Bringing into service continued
Setting the measuring range
Note
nAdjusting the measuring range: The advantage of option 2 is that the tank only has to
be partially filled.
Level electrode NRG 2.-.. only : Setting the measuring range, option 1
Action Indication Function
Reduce water level to lower end of measuring range A.
Select parameter CAL.L. After a short time, a hexadecimal number
flashes.
Lower end of measuring
range is calibrated.
Press the pushbutton. The current hexadecimal number flashes.
Press and hold the pushbutton
again within 3 sec.
done is displayed. After this, the display toggles
between CAL.L and the hexadecimal number.
After a short time, 0000% appears.
Input is confirmed.
(actual value output = 4 mA)
Fill tank to upper end of measuring range B.
Select parameter CAL.H and
press pushbutton.
After a short time, a hexadecimal number
flashes.
Upper end of measuring
range is calibrated.
Press the pushbutton. The current hexadecimal number flashes.
Press and hold the
pushbutton again within 3 sec.
done is displayed. After this, the display
toggles between CAL.H and the hexadecimal
number. After a short time, 0100% appears.
Input is confirmed.
(actual value output = 20 mA)
Level electrode NRG 2.-.. only : Setting the measuring range, option 2
Action Indication Function
Reduce water level to lower end of measuring range A.
Select parameter CAL.L. After a short time, a hexadecimal number
flashes.
Lower end of measuring
range is calibrated.
Press the pushbutton. The current hexadecimal number flashes.
Press and hold the pushbutton
again within 3 sec.
done is displayed. After this, the display toggles
between CAL.L and the hexadecimal number.
After a short time, 0000% appears.
Input is confirmed.
(actual value output = 4 mA)
Fill tank up to at least 25% of the measuring range.
Select parameter CAL.H. After a short time, a hexadecimal number
flashes.
Calibrate at a percentage of the
measuring range, e.g. 25%.
Press the pushbutton. The current hexadecimal number flashes.
Press and hold the
pushbutton again within 3 sec.
done is displayed. After this, the display
toggles between CAL.H and the hexadecimal
number. After a short time, 0100% appears.
Input is confirmed.
(actual value output = 20 mA)
Select parameter CAL.P and set and save a percentage,
e.g. 25%.
CAL.P interpolates the value measured at CAL.H to find
100% of the measuring range. CAL.P can be adjusted
between 25 and 100%.