Gima V8V8C User manual

LETTO OSPEDALIERO ELETTRICO
ELECTRIC HOSPITAL BED
LIT MÉDICALISÉ ÉLECTRIQUE
CAMA ELÉCTRICA DE HOSPITAL
M44746-M-Rev.0-01.24
MANUALE D’USO – USER MANUAL – MODE D’EMPLOI – MANUAL DE USO
V8V8C (Gima 44746)
Jiangsu Saikang Medical Equipment Co.,Ltd
No. 35 Lehong Road, Modern Agriculture Demonstration
Park, Zhangjiagang City, Jiangsu Province, China
Made in China
Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy
gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
SUNGO Europe B.V.
Fascinatio Boulevard 522, Unit 1.7,
2909VA Capelleaan den IJssel, The Netherlands
-10˚C
+40˚C
• È necessario segnalare qualsiasi incidente grave verificatosi in relazione al dispositivo medico da noi fornito al fabbricante e all’autorità
competente dello Stato membro in cui si ha sede.
• All serious accidents concerning the medical device supplied by us must be reported to the manufacturer and competent authority of the member
state where your registered office is located.
• Il est nécessaire de signaler tout accident grave survenu et lié au dispositif médical que nous avons livré au fabricant et à l’autorité compétente de
l’état membre où on a le siège social.
• Es necesario informar al fabricante y a la autoridad competente del Estado miembro en el que se encuentra la sede sobre cualquier incidente
grave que haya ocurrido en relación con el producto sanitario que le hemos suministrado.
500hPa
1060hPa
0%
95%
%

2
ENGLISH
USER MANUAL
GENERAL INFORMATION
The electric hospital bed is a bed specially designed for hospitalized paents or
others in need of some form of health care.
Essenal Performance: This bed is designed for posioning and transporng pa-
ents, as an aid to diagnosing, monitoring, treang and alleviang illnesses or com-
pensang for injuries or disabilies.
The bed itself is not life sustaining or life supporng. The bed has no medical indi-
caon.
User Environment: hospitals and medical clinics in applicaon environments 2 or 3.
The installaon room must be equipped with electrical installaon in accordance
with current standards.
Personnel for use of the product: specialized operators and medical personnel.
Supervision and responsibility: the bed should be used under the supervision of
a doctor.
Limits of use: the bed can be used only as described in this manual
Essenal Performance: This bed is designed for posioning and transporng.
Note for Users:
Please carefully read this manual and strictly follow the steps before installing and
using.
We require all users to have the knowledge of user’s manual and all the cauons.
Before carrying out any operaon on the device, qualied operators and technicians
should read carefully the instrucon contained in this publicaon.
This manual is applicable to V8V8C electric hospital bed. It contains theinstallaon
method, operang instrucons and maintenance check points.
The graphics in this manual are for users' reference only. In actual use,please refer
to the real object.
This user manual includes the descripon, instrucon, maintenance andtechnical
parameters of the product.
Any quesons, please contact the aer-sale service center or distributor.
Anyone using or operang this product should read and comply with it.
Except for some unexpected situaons, users are required to be familiar withthe
usage and precauons.
Explanaon of symbols:
WARNINGS
1) DUTY OF THE USERS
2) Before using a hospital bed, you, as the user, must check each me that the ho-
spital bed is fully funconal and in perfect working order and must observe the in-
strucons in the instrucon manual, parcularly in the safety
informaon, during operaon and maintenance.
This is the only way to prevent operang errors and ensure correct handling in order
to prevent from injuries and damage from occurring.
3) Pay special aenon here to the safe roung of all loose connector cables, tubing
etc… if any
4) Ensure that no obstacles such as bedside cabinets, supply rails or chairs could
impede adjustments to the bed
5) Consult the manufacturer of the equipment in queson, or the distributor, if there
are any uncertaines.
2) ELECTRICAL CABLING AND CONNECTIONS
Connect the bed only to a mains electricity supply with an earth wire
TECHNICAL SPECIFICATION
► Model:V8V8C
► Double auto-regression funcon for bed plaorm
► Integrated handle facilitates the paents to get up/o the bed.
Technical Parameters:
► External Size (LxWxH): 2085*1025*(455-745)mm
► Maress Plaorm: 1925x900mm
► Safe Working Load: 250kg
► Back-rest Adjustment: 0-75°(±5°)
► Knee-rest Adjustment: 0-32°(±5°)
► Reverse Trendelenburg: 0-12°
► Trendelenburg: 0-12° Technical Conguraon:
► PP Plaorm with Integrated Maress Retainer, removable for sucient sterilizaon
► Linear Motor: 4pcs
► 5" Double Side Secure®Castors, Central Locking: 4pcs
► PP Side Rail with Integrated Control Panel: 1set
► PP Bed Ends plug-in type, with safe lock: 1set
► Bumper wheels diameter 65mm: 4pcs
► IV Pole 4 hooks: 1set
► IV Pole Prevision: 4pcs
► Drainage Hook: 2pcs
► Central Locking Pedal: 2 pcs
► With hook type nurse control panel at foot end:1 pc
► Cardiac Chair Posion
SUMMARY
GENERAL INFORMATION................................................................... 2
WARNINGS......................................................................................... 2
TECHNICAL SPECIFICATION ............................................................... 2
1.
PRODUCT INSTALLATION GUIDE ................................................... 3
............................................................ 3
................................................... 3
............................................................. 3
......................................................................... 3
2.
BED OPERATION............................................................................. 4
................................................................................. 4
............................................................................... 4
................................................................................. 4
.......................................................... 4
.............................................. 5
............................................................. 5
............................................15
3.
SPARE PARTS REPLACEMENT AND MAINTENANCE ...................... 5
.......................... 6
........................................... 6
................................................... 6
4. TROUBLESHOOTING....................................................................... 6
...................................................... 6
...................... 6
.................. 7
5.
BED MAINTENANCE....................................................................... 8
6.
BED CLEANING PRECAUTIONS....................................................... 8
7.
MATTRESS MAINTENANCE ............................................................ 8
8.
CASTERS INSPECTION AND MAINTENANCE.................................. 8
9.
SIDE RAILS MAINTENANCE............................................................ 8
10.
LIST OFWEARING AND SPARE PARTS.......................................... 8
11.
PACKAGING, TRANSPORTATION AND STORAGE CONDITIONS ... 9
.................................................................. 9
SYMBOL EXPLANATION:.................................................................... 10

3ENGLISH
► Back Rest 30 Degree Posion Smart Stop
► Emergency Stop Buon
► Manual CPR
► Weighing system as oponal (extra price)
1. PRODUCT INSTALLATION GUIDE
1.1 LIST OF STANDARD ACCESSORIES
Aer opening the packing case, please check whether the following parts are com-
plete:
Table a:
№Name Picture Specication Qty Unit
1Head foot board
packing
box check table a.1 below 1 box
2IV pole Φ16-19/o-white 1 pc
3 Bed platform 2048*1047 1 pc
4 Screw pack / Check table a.2 below 1 bag
Table a.1
№Name Picture Specication Qty Unit
1Head board 990*460 1 pc
2Foot board 990*460 1 pc
Table a.2
№Name Picture Specication Qty Unit
1 30-50 drainage hook 30-50(plastic) 2 pc
2 Pole 154 φ21.7*154 2 pc
3Pole 184 φ21.7*184 2 pc
4Flat washer φ12 4 pc
5Jam nut M12 4 pc
1.2 HEAD AND FOOT BOARD INSTALLATION
(the picture does not exactly match the real object, for installaon guide reference
only).
Head board poles: φ21.7x154, foot board poles: φ21.7x184,the installaon method
is as shown in the gure below:
Pole
Shim φ12
Jam nut M12
Unlocked state Locked state
1.3 DRAINAGE HOOK INSTALLATION
Prepare drainage hook, turn the concave buckle of the drainage hook downwards,
then buckle it down on the beam of the bed frame, see picture:
dr. hook
1.4 IV POLE INSTALLATION
Primary part installaon: place the primary plasc hook part on top of the 16mm Ø
stainless steel pole and press it down.
Secondary part installaon: place the secondary plasc hook part on top of primary,
align the secondary hook part’s snap with the primary’s socket and press it down,
aer you hear a “click” sound, check if the clip is ght, if so, the installaon is com-
pleted, ifnot, connue to press down. Place IV pole vercally in one of the IV pole
previsions.
IV pole prevision 1 IV pole prevision 3
IV pole prevision 2 IV pole prevision 4

4
ENGLISH
Secondary
hook part
Primary
hook part
2
1
prevision
4
3
2. BED OPERATION
543
2
1
76 8
1- Head board 2- IV pole 3- Side rail 4- Bed board
5- Foot board 6- Caster 7- Bed frame 8- Drainage hook
Technical parameters:
1 Bed plaorm dimension length 1925±10
width 900±10
height 455±10
2 Overall dimension length 2085±10
width 1025±10
height (455-745) ±10
3 Tilt range back rest 0 ~ 75±5°
knee rest 0 ~ 32±5°
reverse trendelenburg 0 ~ 12°
trendelenburg 0 ~ 12°
4Weight Safe working load 250
This bed is not suitable for paents weighing less than 40 kg. The maximum recom-
mended paent weight is 218 kg. The safe working load of the bed is 250 kg.
The safe working load is calculated as follows (in accordance with EN 60601-2-52):
Maximum paent weight 218kg
Maress 7kg
Accessories (including aached loads) 25kg
TOTAL 250kg
2.1 CASTERS USAGE
In this product the central locking system is used, which can lock all four wheels by
controlling the brake pedal on one side.
Press the green pedal down to unlock casters, press the red pedal down to lock caters.
The gures below show two states of the wheels:
Locked state Unlocked state
2.2 SIDE RAIL USAGE
(The picture is for reference only, the actual product shall prevail.)
To raise side rail: hold the rail and pull it up, when you hear “click” sound means the
rail is locked, see the pic:
rail safe lock
rail To lower side rail: unlock the rail, slowly lower it down, see the pic:
2.3 IV POLE USAGE
IV pole height adjustment: rst unscrew the outer rotang sleeve, counterclockwise,
insert the inner tube of the IV pole into the rotang sleeve, and then adjust the IV
pole height as needed. Aer adjusng the required height, ghten the outer rotang
sleeve, clockwise. Safe working load of each individual hook is 2kg, total safe working
load is 8kg.
(height adjustment range: 890-1540mm, the picture is for reference only)
rotang sleeve
2.4 SIDE RAIL CONTROL PANEL USAGE
Side rail outer control panel
Side rail inner control panel
2.4.1 Side rail inner control panel

5ENGLISH
1. Paent mobilizaon 1
2
3
4
5
6
8
7
9
10
2. Raise back rest
3. Lower back rest
4. Raise back+knee rest
5. Lower back+knee rest
6. Raise knee rest
7. Lower knee rest
8. Raise bed
9. Lower bed
10. Power on/o
6
1
2
8
3
4
5
7
9
10
2.4.2 Side rail outer control panel
1. CPR 1
8
9
3
10
11
12
2
4
6
5
7
2. Raise back rest
3. Lower back rest
4. Power on/o
5. Raise back+knee rest
6. Lower back+knee rest
7. Trendelenburg
8. Raise knee rest
9. Lower knee rest
10. Raise bed
11. Reverse Trendelenburg
12. Lower bed
2
1
3
4
5
6
7
8
10
9
12
11
(Note: The hand controller is an oponal accessory, is supplied according to require-
ments) (the picture is for reference only)
1. Raise back rest
2. Lower back rest
3. Raise knee rest
4. Lower knee rest
5. Raise back+knee rest
6. Lower back+knee rest
7. Raise bed
8. Lower bed
9. Reverse Trendelenburg
10. Trendelenburg
11. CPR
12. Power on/o
13. Flash light
2.6 NURSE CONTROL PANEL USAGE
1. Raise back rest 14. Lock Trendelenburg
2. Lower back rest 15. Power on/o
3. Raise knee rest 16. Emergency stop
4. Lower knee rest 17. Trendelenburg
5. Raise back+knee rest 18. Cardiac posion
6. Lower back+knee rest 19. Nursing posion
7. Raise bed 20. Paent mobilizaon
8. Lower bed 21. CPR
9. Reverse Trendelenburg
10. Trendelenburg
11. Lock back rest
12. Lock knee rest
2.7 ELECTRONIC CONTROL SYSTEM COMPOSITION
2.7.1 Electronic control system diagram
Operator priority:
(1)Nurse control panel > Hand controller > Side rail control panel; (2)One priority but-
ton can interrupt and terminated the same priority buon operaon (except hand
controller), when one buon is released and another one is sll pressed, the opera-
on of the last one is connued;
(3)High-priority keys (except CPR, TR, and STOP) can interrupt and terminate low-pri-
ority keys. When one buon is released and another one is sll pressed, the ope-
raon of the last one is connued. When CPR, TR or STOP buon is pressed during
other buon operaon, CPR, TR or STOP operaon is executed.
2.7.2 Electronic control parts list
Name Qty
JC motor 200 vertical(manual CPR) 1
JC knee rest motor 100 horizontal 1
JC bed platform height motor 150 horizontal 2
JC 5-functions control panel 1
JC 3m power cord (environmental) 1
JC 5-functions nurse control panel 1
JC 5-functions hand controller 1
Emergency stop switch box 1
JC night light 1
Adapter box JCP35C5 1
JC battery box 1
JC Adapter cable 2
3. SPARE PARTS REPLACEMENT AND MAINTENANCE
3.1 Caster disassembly and replacement

6
ENGLISH
Use 10”inch wrench unscrew the bolt,pull out the screw, at spring wa-
shers, brake pedal,plasc retaining ring, hexagonal rod, etc, replace with spare ca-
ster, place the bolt back into the socket, ghten the nut.
3.2 SIDE RAIL GAS SPRING DISASSEMBLY AND REPLACEMENT
(the picture is for reference only)
Use external calipers to take out the head and tail pins’ snap springs, remove the gas
spring, replace it with the spare one, reinstall it, test the rail.
snap springs
3.3 SIDE RAIL DISASSEMBLY AND REPLACEMENT
First remove the gas spring, using two 14”inch wrenches loosen the nut,
remove the nut, remove the bearing, remove the rail. (The picture is for reference
only, the actual product shall prevail)
3.4 MOTOR REMOVAL AND REPLACEMENT
Aer removing the R-type lock pins at both ends of the motor, pull out the lock pins.
Aer motor replacement, install the bolts in the order shown below. (the pic. is for
reference only).
Perno di bloccaggio di po R Perno di bloccaggio di po R
Perno di bloccaggio Perno di bloccaggio
4. TROUBLESHOOTING
Issue Possible reason Troubleshooting
Bed cannot be
adjusted using
positioning buttons
Did not power on Press power on button
Motor is powered o, Motor
malfunction Check power cords Replace motor,
see 3.4
Power plug isn’t plugged correctly Plug in power plug correctly
Control system malfunction Contact technical support
Bed height
adjustment failure Motor is powered o Check power cords
Motor malfunction Replace motor, see 3.4
Power plug isn’t plugged correctly Plug in power plug correctly
Power box malfunction Contact technical support
Control system malfunction Contact technical support
Back plate cannot
be lowered from
upright position
Foreign objects under the
backplane or in the drive unit Remove foreign objects
The guardrail
cannot be adjusted Dirty guardrail lock Clean the locking mechanism
Side rail safe lock malfunction Replace side rail, see 3.3
Casters can't turn The rotating shaft is stuck by strips
of dirt Clean the mechanism rotating
Malfunction of the internal
mechanism of the caster Replace caster, see 3.1
Unable to install
head and foot board Wrong installation method Check the locking mechanism,
properly position the head and foot
board, see 1.2
Mechanical malfunction Contact technical support
4.1 PUSH ROD MALFUNCTION HANDLING
Push rod malfuncon handling:
(1) Causes of push rods malfuncon:
1. push rod’s abnormal Hall signal;
2. rst me programming;
3. nurse desk, hand controller, guardrails control panel 6S system malfuncon.
(2) Push rod malfuncon prompng method: if the M1, M3, M4 malfuncons are
detected, JCH35-JSSK Saikang custom hand controller blue indicator is ashing.
(3) Push rod malfuncon troubleshoong:
1. By pressing corresponding buons, raise the M1pus rod to the top posion or
lower it to the boom posion, so the push rod completes the “rebound”, and
the malfuncon is xed;
2. By pressing corresponding buons, lower the M3, M4 push rods to the boom
posion, make them complete “rebound”, and x the malfuncon;
3. When the M3, M4 rods are completely malfunconed, Reverse TR/TR buons
are not valid;
4. When any push rod is malfunconed, paent mobilizaon, cardiac posion,
TR, emergency buons are not valid.
4.2 TEST METHOD FOR UNRESPONSIVE SIDE RAIL CONTROL PANEL
Step 1, check if the emergency stop switch is pressed
A. Check: if the emergency stop switch is pressed;
B. Check method: turn the emergency stop switch clockwise to open(it bounces out
for a short distance), see pic 1-1;
C. Dene a problem: if the emergency stop switch pops up, it means that it has been
pressed; if the emergency stop switch doesn’t pop up, it means that it is already in

7ENGLISH
Pic. 1-4
C. Problem determinaon: if the circuit board connector is loose, plug it in ghtly,
see pic.1-5;
Connector
Pic. 1-5
D. Conclusion: aer the above three steps, check the side rail control panel, if it sll
unresponsive, means the control panel is damaged, and should be replaced.
4.3 TEST METHOD FOR UNRESPONSIVE FOOT BOARD CONTROL
PANEL
Step 1: check if the emergency stop switch is pressed
A. Check the content: if the emergency stop switch is pressed;
B. Check method: turn the emergency stop switch clockwise to open(it bounces out
for a short distance), see pic.1-1;
C. Problem determinaon:if the emergency stop switch pops up, it means that it
has been pressed; if the emergency stop switch doesn’t pop up, it means that it is
already in the open state, and there is no need to rotate it;
D. Conclusion: aer the above three steps, check the side rail control panel, if it sll
unresponsive, move to the Step 2.
Pressed state Open
Pic. 2-1
Step 2: check whether the guardrail buckle is loose or the pin cores are deformed
A. Check the content: foot board buckle is loose, pin cores are deformed;
B. Check method: follow the connecng line of the foot board, nd the foot board
adaptersee pic.2-2;
C Problem determinaon: buckle is loose, as on pic.2-2 lock the buckle; pin cores
are deformed, using a at-blade screwdriver recfy pin cores, make sure it’s con-
nected ghtly with the adapter;
D. Conclusion: aer the above three steps, check the side rail control panel, if it sll
unresponsive, move to the Step 3.
Adattatore
Fibbia
nucleo
the open state, and there is no need to rotate it;
D. Conclusion: aer the above three steps are nished, test the guardrail, if there is
sll no response, go back to step 2.
Pressed state Open
Pic. 1-1
Step 2: check whether the guardrail buckle is loose or the pin cores
are deformed.
A. Check the content: check whether side rail buckle is loose, or pin cores are de-
formed;
B. Check method: follow the connecng line under back board side rail, nd the side
rail adapter, see pic.1-2;
C. Problem determinaon:the buckle is loose, as on pic.1-2 lock the buckle; pin
cores are deformed, using a at-blade screwdriver recfy pin cores, make sure it’s
connected ghtly with the adapter;
D. Conclusion:aer the above three steps, check the side rail control panel, if it sll
unresponsive, move to the Step 3.
adapter
buckle
Pin cores
Pic. 1-2
Step 3: Check if the controller plug is loose
A. Check the content:check if the controller plug is loose;
B. Check method:
① Check if the controller clip on the bed frame under the back plate is loose or
dislodged, see pic.1-3;
② Check if the plug is loose;
Clip is loose
Pic. 1-3
C. Problem determinaon:loose clips can easily lead to loose plugs, please fasten
the clips aer the plugs are rmly inserted;
D. Conclusion: aer the above three steps, check the side rail control panel, if it sll
unresponsive, move to the Step 4.
Step 4: Check if the side rail control circuit board connector is loose
A. Check the content: side rail circuit board connector is loose;
B. Check method: peel o the cover scker from the side rail control panel, using
screwdriver unscrew 4 screws, see pic.1-4;

8
ENGLISH
Pic. 2-2
Step 3: Check if the controller plug is loose
A. Check the content: controller plug is loose;
B. Check method:
①Check if the controller clip on the bed frame under the back plate is loose or
dislodged, see pic.2-3;
②Check if the plug is loose;
La clip è allentata
Pic. 2-3
C. Problem determinaon: loose clips can easily lead to loose plugs, please fasten the
clips aer the plugs are rmly inserted;
D. Conclusion: aer the above three steps, check the side rail control panel, if it sll
unresponsive, means the control panel is damaged, and should be replaced.
5. BED MAINTENANCE
1. In order to use the bed safely, it is necessary to conduct regular safety inspecons
on the bed, it’s recommended to conduct a comprehensive inspecon every six mon-
ths to ensure the connecon parts are not loose and the liing funcon of the bed
is operang normally.
2. The connecon of the frequently moving parts of the bed needs to be lubricated
with an appropriate amount of oil during daily inspecon; is serious wear is found, it
should be replaced immediately to ensure safety. The moving part in gure below:
3. Avoid hing the bed and scratching the coang on the bed surface with sharp
objects.
4. Please prevent the bed and maress from absorbing water and moisture, which
may cause rust, abnormal noises and bacterial growth. If there was water spilled,
please wipe it o immediately.
5. When the bed ages and reaches its service life me, the metal part and the plasc
parts can be recycled.
6. BED CLEANING PRECAUTIONS
1. When cleaning, please wring the cloth soaked in neutral detergent diluted with
water, wipe it dry, then wring the cloth soaked in clean water, wipe o the residual
detergent ingredients, and nally wipe with a dry cloth.
2. Do not use volale items (thinners, propellants, gasoline, etc)
that may cause chemical reacons and damage the bed.
3. When cleaning with a disinfectant, be sure to use it aer dilung it at the specied
concentraon. Depending on the composion of the disinfectant, it may corrode me-
tal parts, resin parts, etc, causing discoloraon, deformaon and other undesirable
consequences. Therefore, the recommended content of the disinfectant is as
follows:
0.05~0.2% ammonium chloride
0.05~0.2% chlorinated phenyl
0.05% dichlorobenzene biguanide ethane
0.05~0.2% sodium hypochlorite
Do not use a smoking sterilizer or autoclave, and do not use methyl phenol to clean
the head foot board, bed feet, etc, which may cause corrosion, discoloraon, dete-
rioraon.
Note: 1. Unplug power cable before cleaning and maintenance
1. During the cleaning, DO NOT splash water to prevent short circuit and electric
shock.
2. When cleaning the bed or changing bedding, pay aenon to the corners, edges
and screw parts of the frame to prevent scratches.
7. MATTRESS MAINTENANCE
1. Avoid scratching the fabric with sharp-angled appliances or knives when using it,
cover it with sheets or cleaning pads, and clean it frequently to keep it clean and dry.
2. Regularly clean the maress cover with a vacuum cleaner, it can be washed direct-
ly with water or detergent.
3. Air the maress oen to keep your sleep fresh and comfortable, but be careful not
to expose the maress to the sun for too long.
4. If the fabric is accidentally stained, please refer to the corresponding recommen-
ded cleaning method in the instrucons. Do not use corrosive chemicals to clean
the fabric.
5. If you accidentally get chewing gum or other glue-like objects on the fabric, do
not wipe it forcibly. You can use am ice pack to allow it to harden and then peel it
o gently.
6. If the thread is found to be loose due to wear and tear, it cannot be broken by
hand, and it should be cut short with scissors.
8. CASTERS INSPECTION AND MAINTENANCE
1. Check the casters regularly, avoid hing heavy objects or overloading, which will
cause damage to the casters.
2. The casters should be applied with lubricang grease regularly, the frequency of
appliance should be determined according to the specic actual use situaon. Under
normal condions, the grease should be applied every six month.
3. Aer cleaning the caster equipment, pay aenon to adding lubricang grease.
The selecon of lubricang grease should be based on the actual use environment,
special high low temperature environment, etc.
4. It’s forbidden to push the bed with force when the casters are braked, which will
cause damage to the braking system.
5. The wear of the re tread of the caster can be detected by visual inspecon. Some
accumulaons such as thread and spun yarn may wrap around the caster. Remove
the bolts and nuts of thecaster, clean up the debris, and check whether the bearing
of the caster is damaged. If the parts are not damaged, the caster can be reassem-
bled and used.
6. Decision whether to replace enre caster is according to the wear condion. Aer
the caster is replaced or reinstalled, make sure that the axle bolt and nut are con-
nected reliably, and use lock washers or lock nuts as much as possible.
7. For casters equipped with brakes, it is necessary to regularly check whether the
brakes work normally, and check the brakes every day or before each use. For a ho-
spital bed equipped with mulple brake casters, you can lock only one brake caster
at a me and try to push the bed to check whether the braking performance of each
caster is good. If the caster brake fails to funcon due to wear or damage to the ca-
ster, replace the caster immediately and retest the brake system.
8. If the mechanism of the brake system is damaged, and the brake needs to be re-
paired or replaced, please contact our aer- sales or an authorized dealer. Every me
the brakes arereplaced, the brake performance of the casters needs to be retested.
9. SIDE RAILS MAINTENANCE
1. Please check whether the side rails are installed correctly to prevent the lying
paent from shiing and rolling o.
2. Please always check the screws of the side rail and the side rail ngs, to prevent
loosening and falling o.
3. Please use the red part of the side rail handle correctly, do not use brute force to
pull hard, which will cause damage to the side rail.
10. LIST OF WEARING AND SPARE PARTS
№Name Material code Picture Specication
1 V21 head board GPSL101TS648 990*460
2 V21 foot board GPSL101TS648 990*460
3V21 bed board 1 GPSL101TS610 733*903/PP/white
4 V21 bed board 2 GPSL101TS611 333*903/PP/white
5 V21 bed board 3 GPSL101TS613 373*903/PP/white

9ENGLISH
6 V21 bed board 4 GPSL101TS614 473*903/PP/white
7 V21 back board
guardrail,with
control panel, left,
blow molding
GPSL101TS604 924*350mm/PP/wh ite
8 V21 back board
guardrail,with
control panel, right,
blow molding
GPSL101TS605 924*350mm/PP/wh ite
9 V21 foot board
guardrail, left, blow
molding
GPSL101TS608 924*350mm/PP/wh ite
10 V21 foot board
guardrail, right,
blow
molding
GPSL101TS609 924*350mm/PP/wh ite
11 V21 guardrail
5-functions control
panel, left
LGPDJ101TS82 0 Back rest, knee rest,
back+knee rest, bed
height, lock, CPR,
patient mobilization
12 V21 guardrail
5-functions control
panel, right
LGPDJ101TS81 9 Back rest, knee rest,
back+knee rest, bed
height, lock, CPR,
patient mobilization
13 Nurse control panel
//
14 V21 back board
guardrail
sticker
LGPFU101TS4 16 780*35*0.4
15 V21 foot board
guardrail sticker LGPFU101TS4 17 750*35*0.4
16 Guardrail
damper(white) LGPQT101TS0 24 18-58-168-35N
17 30-50
drainage hook BYKBGKM0280 6 30-50(plastic)
18 Stainless Steel
IV pole with 4
plastic hooks
S0FJSY010003 Φ16-19/o-white
19 Foot cover (left) BYKBGKM0240 2 white
20 Foot cover (right) BYKBGKM0240 3 white
21 bed foot cover BYKBGKM0240 4 white
22 Brake pedal (left) A BYKBGKM0281 7 green
23 Brake pedal (left) B BYKBGKM0281 8 red
24 Brake pedal
(right) A BYKBGKM0281 9 green
25 Brake pedal
(right) B BYKBGKM0282 0 red
26 5 inch central
control double-
sided wheel
BYDYGX10513 4 /
27 JC motor 200
vertical (manual
CPR, Hall)
LGPDJ101TS82 1 JC35L10Y4424200
/385VGDZK2102D 0
28 JC knee rest motor
100 horizontal LGPDJ101TS80 5 JC35D106324100/
275HGDZD0102D 0
29 JC height motor
150 horizontal LGPDJ101TS80 6 JC35D1Y6324150/
325HGDZD0102D 0
30 JC 3m power cord LGPDJ101TS69 4 / GB 3 plug
31 JC 5-functions
controller LGPDJ101TS81 8 JCB35R-AM-4-3-G
-R0088-100-03-14
32 Emergency stop
switch box LGPDJ101TS59 0 JCP35Q-G-4D1-2
D0
33 JC 5-functions
nurse control panel LGPDJ101TS72 1 /
34 Adapter box
JCP35C5 LGPDJ101TS76 8 /JCP35C5-4-G
35 JC Adapter cable LGPDJ101TS72 2
/
1.5m o-white 10 core
crystal head waterproof
bend 9-pin waterproof
motor socket plug 10-
pin transfer cable
36 JC Adapter cable LGPDJ101TS83 9
/
1.5m o-white 10 core
crystal head waterproof
bend 8 core yellow
motor straight plug
6-pin patch cord
Note: Due to the renewal of product lineup, the parts of the original product may not
be available. (The company ensures availability of the parts for sold products within
6 years aer their producon ends.)
11. PACKAGING, TRANSPORTATION AND STORAGE CONDI
TIONS
1. The packaging of the manual bed is carried out according to the contract or pro-
duct standard.
2. During the transportaon of the product, avoid rushing out, violent vibraon, pre-
vent direct sun and rain access.
3. Store condions:
a) Ambient temperature: -10℃~ 40℃
b) Relave humidity: ≤ 95%
c) Atmospheric pressure: 500hPa ~ 1060hPa
ELECTRIC VARIABLE HEIGHT BED – V8v8c
Supply voltage V100-230V a.c.
Network frequency Hz 50/60Hz
Operating voltage VA 500 VA
Class of electrical protection -I
Applied part -B
Battery capacity Ah 12V d.c. 1.3Ah
IP Protection IPX4
Symbol explanaon:
Noce!Read aachment paper IPX4 splash proof
Protecve grounding
1. Please preserve the aachment papers and invoice of this product, you need to
present those to enjoy aer-sales maintenance service.
2. If you have any queson during use, please contact us any me, so that our com-
pany can provide you with accurate and ecient technical support and maintenance
services in a mely manner.
3. Starng from the purchase date and during one year, if any malfuncon or damage
occurs when the product is installed and used correctly according to the instrucons,
you can enjoy free warranty and lifeme maintenance service presenng a “cer-
cate” or invoice.
4. Do not disassemble the internal parts of this product by yourself to avoid unneces-
sary damage. I you nd that the use is aected by quality issues, please contact the
company’s aer-sales service department or the authorized distributor.
5. Starng from the date of purchase within 1 year, if the product cannot work nor-
mally due to quality problems, the company will repair it for clients for free.
6. The warranty is guaranteed according to the contract requirements. Outside the
warranty period, the company is responsible for maintenance, providing accessories,
and implemenng life-long service for clients.
7. Aer-sales service unit: Jiangsu Saikang Medical Equipment Co., Ltd.
Noce!any serious incident that has occurred in relaon to the device shall be
reported to the manufacturer and the competent authority of the Member State in
which the user and/or the paent is established.
EMC
1) This product needs special precauons regarding EMC and needs to be installed
and put into service according to the EMC informaon provided, and this unit can be
aected by portable and mobile RF communicaons equipment.
2) * Do not use a mobile phone or other devices that emit electromagnec elds,
near the unit. This may result in incorrect operaon of the unit.
3) Cauon: This unit has been thoroughly tested and inspected to assure proper per-
formance and operaon!

10
ENGLISH
4) * Cauon: This machine should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, this machine should be
observed to verify normal operaon in the conguraon in which it will be used.
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The V8v8c is intended for use in the following electromagnetic environment. The customer or user
of the V8v8c should make sure it is used in such an environment
Emission test Compliance Electromagnetic Environment - Guidance
RF Emission to
CISPR 11 Group 1 The bed uses RF energy for its internal function
RF Emission to
CISPR 11 Class B
The bed is intended for use in all types of establishment
including residential and similar user that are directly
connected to a public supply network that also serves
buildings used for residential purposes
Harmonics
according to
IEC 61000-3-2
Class A
Voltage
uctuations/
icker acc. to
IEC 61000-3-3
Complies
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The V8v8c is intended for use within healthcare facility environment. Users or customers must use
bed in this environment
Immunity test IEC 60601-1-2/ EN 60601-1-2
Test level Electromagnetic Environment
-Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
/EN 61000-4-2
Contact discharge: +8kV
Air discharge: +2kV, +4kV, +8kV,
+15kV
Contact discharge: +8kV
Air discharge: +2kV, +4kV, +8kV, +15kV
Floor should be made of wood and
concrete or be tiled with ceramic tiles.
If the oor is covered with synthetic
ooring material, the relative air
humidity must be at least 30%
Can be used when higher ESD levels
are present
Electrical fast
transient/burst
IEC 61000-4-4
/EN 61000-4-4
+2kV Power supply line
+1kV Input/Output lines +2kV Power supply line
The quality of the supply voltage
should be equivalent to that of a typical
business or hospital environment.
Surge
IEC 61000-4-5
/EN 61000-4-5
Line - line: +0.5kV, +1.0kV
Line - ground: +0.5kV, +1.0kV,
+2.0kV
Line - line: +0.5kV, +1.0kV
Line - ground: +0.5kV, +1.0kV, +2.0kV
The quality of the supply voltage
should be equivalent to that of a typical
business or hospital environment.
Power Frequency
Magnetic Fields
IEC 61000-4-8
/EN 61000-4-8
30 A/m
50Hz/60Hz 30 A/m
50Hz/60Hz
Magnetic elds with a network
frequency should be equivalent to
those to be found in a typical business
or hospital environment.
Voltage Dips IEC
61000-4-
11
/EN 61000-4-11
0% UT: 0.5 cycles
Phase: 00, 450, 900, 1350, 1800,
2250, 2700, 3150
0% UT: 0.5 cycles
Phase: 00, 450, 900, 1350, 1800, 2250,
2700, 3150
0% UT: 1 cycle
70% UT: 25/30
cycles Single phase: 00
0% UT ;
300 cycle
0% UT: 1 cycle
70% UT: 25/30
cycles Single phase: 00
0% UT ; 300 cycle
The quality of the supply voltage
should be equivalent to that of a typical
business or hospital environment.
If the person using the bed requires
that the bed functions must continue
despite any interruptions in the energy
supply, it is recommended that the
bed be connected to an uninterruptible
electricity supply or a battery
Short interuptions
IEC 61000-4-11
/EN 61000-4-11
0% UT: 250/300 cycles 0% UT: 250/300 cycles
The quality of the supply voltage
should be equivalent to that of a typical
business or hospital environment.
If the person using the bed requires
that the bed functions must continue
despite any interruptions in the energy
supply, it is recommended that the
bed be connected to an uninterruptible
electricity supply or a battery
Note: UT is the AC Power voltage before applying in the test level
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The V8v8c is intended for use in an electromagnetic environment that used wireless RF transceiver.
Electromagnetic interference can be prevented by maintaining the minimum distance between RF
transceiver and V8v8c depend on the maximum output and frequency of the communication device
as recommended below.
Immunity test IEC 60601-1-
2
/EN 60601-1-
2
Test level
Complianc
elevel
(V/m)
Band
(MHz) Service Modulation Maximu
mpower
(W)
Distance
(m)
Enclosure port
to RF wireless
communication
sequipment IEC
61000-4-
3
/EN 61000-4-
3
27V/m
385MHz 27 380-
390 TETRA 400 Pulse
modulatio
n18Hz 1.8 0.3
28V/m
450MHz 28 430-
470 GMRS 460,
FRS 460
FM+5
kHz
deviation 1
kHz sine
2 0.3
9V/m
710MH
9704-
787
LTE Band
13,
17
Pulse
modulatio
n217Hz
0.2 0.3
9V/m
745MHz
9V/m
780MHz
28V/m
810MHz
28 800-
960
GSM
800/900,
TETRA 800,
iDen 820,
CDMA 850
LTE Band 5
Pulse
modulatio
n18Hz 2 0.3
28V/m
870MHz
28V/m
930MHz
28V/m
1720MHz
28 1700-
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band
1,3,
4,25; UMT S
Pulse
modulatio
n217Hz 2 0.3
28V/m
1845MHz
28V/m
1970MHz
Enclosure port
to RF wireless
communication
sequipment IEC
61000-4-
3
/EN 61000-4-
3
28V/m
2450MHz 28 2400-
2570
Bluetooth,
WLAN,
802.11 b/g/n
RFID 2450,
LTE band 7
Pulse
modulatio
n18Hz 2 0.3
9V/m
5240MHz
95100-
5800
WLAN
802.11
a/n
Pulse
modulatio
n217Hz 0.2 0.3
9V/m
5500MHz
9V/m
5785MHz
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The V8v8c is intended for use in the following electromagnetic environment. The customer or user
of the V8v8c should make sure it is used in such an environment.
Immunity test IEC 60601-1-
2
/EN 60601-1-
2
Test level
Compliance level Electromagnetic Environment -
Guidance
RF Conductivity IEC
61000-4-6
/EN 61000-4-6
3V
0.15MHz-80MHz 3V
0.15MHz-80MHz Field intensity from the xed RF
transmitter determined by a eld
investigation of the magnetic eld
must be lower than the compliance
level of the respective frequency
band.
Near devices with the following
symbol,
interference could occur.
6V in ISM
bands between
0.15MHz-
80MHz
80%AM at 1kHz
6V in ISM bands
between 0.15MHz-
80MHz
80%AM at 1kHz
Radioactive RF
electromagnetic eld
IEC 61000-4-3
/EN 61000-4-3
3V/M
80MHZ-2.7GHz
80%AM at 1kHz
3V/M
80MHZ-2.7GHz
80%AM at 1kHz
a. The intensity of a magnec eld from xed transmiers such as a radio (mobile/
radio) telephone base staon, land mobile radio, amateur radio, AM and FM radio
broadcasng, and TV broadcasng cannot be theorecally predicted accurately. To
determine the electromagnec environment for a xed RF transmier, consider inve-
sgang the electromagnec eld in the locaon where the bed is used. If the inten-
sity of the magnec eld in the locaon where the V8v8c is used exceeds the above
RF compliance level, make sure to monitor if the V8v8c operates appropriately. If
abnormal movement is found, take addional measures as needed, such as changing
the direcon or posion of the V8v8c.
b. These guidelines may not apply in all situaons. Electromagnec propagaon is
aected by absorpon and reecon from structures, objects, and people.
SYMBOL EXPLANATION:
Keep away from sunlight Medical Device compliant with
Regulation (EU) 2017/745
Product code Follow instructions for use
Date of manufacture Lot number
Caution: read instructions (warnings)
carefully Temperature limit
Imported by Serial number
Atmospheric pressure limit Keep in a cool, dry place

11
Manufacturer Medical Device
Authorized representative in the
European community
%
Humidity limit
Type B applied part WEEE disposal
Press the red handle to lower the side rail(unlock), to secure press the red handle and
push forward until you hear click
Disposal: The product must not be disposed of along with other domesc waste. The
users must dispose of this equipment by bringing it to a specic recycling point for electric
and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
ENGLISH
This manual suits for next models
1
Table of contents
Other Gima Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Huvitz
Huvitz HDR-9000 Service manual

Morzine Medical
Morzine Medical Doak MK4 user guide

Bard
Bard Sherlock 3CG Tip Confirmation System Instructions for use

Reichert
Reichert TONO-PEN XL Quick reference guide

Nuvo
Nuvo B2AV800 Series instruction manual

Novametrix Medical Systems
Novametrix Medical Systems CO2SMO user manual

Integra
Integra MAYFIELD Pediatric Radiolucent Horseshoe... instruction manual

Hitachi
Hitachi UST-5710-7.5 manual

Akron
Akron 8601B operating manual

Kinetec
Kinetec Spectra Essential user manual

Joerns
Joerns Oxford Stature quick start guide

Veridian Healthcare
Veridian Healthcare VH SonicMist instruction manual