GPI TM Series User manual

TABLE OF CONTENTS
English...........................................................1
Español..........................................................9
Deutsch .......................................................16
Italiano.........................................................23
Français.......................................................30
ENGLISH
IMPORTANT NOTICE
Use TM Series meters with water and other
chemicals compatible with wetted components
(seeSpecificationsSection).Donotusetometer
fuelor incompatiblechemicals.TMSeries meters
are available with either a computer for local
electronicdisplay, or a conditioned signal output
module to provide a digital signal to customer
interfacing equipment. TM Series meters with
computer display measure in gallons or litres.
Refer to the Calibration Section for details.
Thesemetersarenotlegalfortradeapplications.
TM Series meters are very sensitive to electric
noise if operated within 1 to 2 inches of some
electric motors or other sources of electronic
noise.
TM Series Electronic Water Meters
User Manual
Rev. A 920861-01
04/13
INSTALLATION
Connections
Install your meter in-line either horizontally or
vertically or at the end of the hose adjacent to
the nozzle. Installation to metal connections is
not recommended. Install as follows:
1. Plantoinstallturbinewith a minimumstraight
pipe length as follows:
• Upstreamfromtheturbine,allowamini-
mum straight pipe length of 10 times the
internal diameter of the turbine.
• Downstream from the turbine, allow a
minimum straight pipe length of 5 times
the internal diameter of the turbine.
2. For Spigot (Pipe) End use only primer and
solvents approved for PVC gluing.
Cut to Length The meter housing can be
shortened by the customer. Each meter has
a “dotted” line feature molded on the top
surface of the housing tube. The housing
can be cut up to this line without harming
any internals.
Mostglueonfittings will fit without interfering
with the computer display area. However,
the customer should check all parts before
attempting cut.
TM Meter with Computer Display TM Digital Pulse Meter
Save theSe inStructionS
www.calcert.com sales@calcert.com1.800.544.2843 0
5
10
15
20
25
30

2
For NPT Fittings wrap all connections with
3 to 4 wraps of thread tape. Make sure the
tape does not intrude into the flow path.
3. Attach meter with arrow pointed in the direc-
tion of fluid flow.
4. For NPT Fittings - Hand tighten the meter at
the housing ends. Do not use a wrench or
similar tool to tighten. This can damage the
housing.
5. For ANSI or DIN Flange Fittings - Customer
to provide:
• Ring Gaskets or Full-Face Gaskets ap-
proved for use with type flange installed
(ANSIorDIN)andthefluidbeingmonitored
(2 required).
• 5/8”boltsandnutsforANSIanges.Four
per side for 3-inch meters; eight per side
for 4-inch meters.
• 16 mm bolts and nuts for DIN anges.
Eightpersidefor3-inch and 4-inch meters.
• Torqueboltsusingastarpatternto25ft-lbs.
Supplied flanges are two-piece Van Stone
style and allow the meter to be oriented
regardless of the mating flanges position.
For best results, always verify accuracy before
use.
OPEN COLLECTOR SIGNAL OUTPUT
Wiring Diagram 1
+9 to 35 volt DC
Open Collector
Signal
3
2
1
4
Red
White
Black Common
(Ground)
Terminal
Block
Customer
Interfacing
Equipment
The terminal block is identified as follows:
Pin #1 6 volt square wave (not used)
Pin #2 +9 to 35 volt DC Input
Pin #3 Common Ground
Pin #4 Open Collector signal Output
Resistor
WARNING
Make sure flow and pressure have been
eliminated from process pipe prior to
installing or removing product.
Compatibility of this product’s material and
theprocessfluidand/or environmentshould
be considered prior to putting into service.
WARNING
Productshouldneverbe operated outsideits
published specifications for temperature or
pressure.Seespecificationsforyourmodel.
WARNING
Always use appropriate thread sealant or
flange gaskets when connecting product
to process piping.
WARNING
www.calcert.com sales@calcert.com1.800.544.2843 0
5
10
15
20
25
30

3
Conditioned Signal Output
Module Wiring
This conditioned signal output module can be
wired to provide an open collector signal output
or 6-volt square wave output.
Open Collector Signal Output
To achieve an open collector signal output,
reference Wiring Diagram 1. The terminal block
is located on the back side of the module. The
module is factory assembled for open collec-
tor signal output. Please provide the (820 ohm
minimum) resistor.
Ten feet (3 m) of cable is provided with the
module. Trim it to desired length or extend it as
necessary. Distances up to 5,000 feet (1,524 m)
can be achieved for open collector signal output.
Square Wave Output
To achieve square wave output, reference Wiring
Diagram 2 and use an Electronic Digital Meter
BatteryKit(sold separately) for battery power.The
terminalblockandbatterylocationarelocated on
the back side of the module. Access as follows:
SQUARE WAVE OUTPUT
Wiring Diagram 2
3
2
1
4
White
Black Common
(Ground)
Terminal
Block
Customer
Interfacing
Equipment
6 volt Square
Wave Signal
The terminal block is identified as follows:
Pin #1 6 volt square wave
Pin #2 +9 to 35 volt DC Input (not used)
Pin #3 Common Ground
Pin #4 Open Collector signal Output (not used)
CAUTION
To protect against leakage, seal all pipe
threads with an appropriate sealing com-
pound. Make sure the sealing compound
does not intrude into the flow path.
CAUTION
Installationnearhighelectromagnetic fields
and high current fields is not recommended
and may result in inaccurate readings.
NOTE: If connecting to new male pipe threads,
burrs and curls can adversely affect accu-
racy. Correct the problem prior to turbine
installation.
NOTE: Do not over tighten the flange bolts. This
may cause the gasket to be compressed
into the flow stream and may decrease the
accuracy of the meter.
www.calcert.com sales@calcert.com1.800.544.2843 0
5
10
15
20
25
30

4
1. Remove the four Phillips-head screws from
the front of the module and lift the module
from the turbine.
2. Tochangeterminalblock connections, loosen
the appropriate screws. Reconnect the wires
intheproperpositionsandtighten the screws.
3. Install the batteries. Make sure the positive
post is in the correct position.
4. Position the module on the turbine housing.
To avoid moisture damage, make sure the
seal is fully seated. Tighten the four screws
on the front of the module.
Ten feet (3 m) of cable is provided with the mod-
ule. Trim the cable to desired length or extend
it as necessary.
Verify Meter Accuracy
Before using, check the meter’s accuracy and
verify calibration.
1. Make sure there is no air in the system by
starting the flow until it runs steadily. Then,
stop or divert the flow using a valve or nozzle.
2. Meter an exact known volume into an ac-
curate container. For best results, meter with
one continuous full stream.
3. Check the volume against the display or
recording equipment. If the amount metered
is accurate, further calibration is not neces-
sary. If not, refer to the Calibration Section
for further instructions.
OPERATION
Computer Display –
Batch and Cumulative Totals
Thecomputermaintainstwototals.TheCumula-
tive Total provides continuous measurement and
cannot be manually reset. The Batch Total can
be reset to measure flow during a single use. The
Cumulative Total is labeled with TOTAL 1, Batch
Total is labeled TOTAL 2 BATCH.
When the Cumulative Total reaches a display
reading of 999,999 the computer will highlight
an X10 icon. This indicates to the operator that
a zero must be added to the 6 digits shown.
When the next rollover occurs, the computer
will highlight an X100 icon. This indicates to the
operator that two zeroes must be added to the
6 digits shown.
Press the DISPLAY button briefly to switch
between the TOTAL 1, TOTAL 2 BATCH and
FLOWRATE. Press DISPLAY briefly to display
the TOTAL 2 BATCH. Hold the DISPLAY button
for 3 seconds to reset the Batch Total to zero.
When fluid is flowing through the meter, a small
propeller icon is highlighted.
NOTE: Totalization counts total units without
differentiating between gallons, litres or field
calibrated units.
Flowrate Feature
To use this feature, press and release DISPLAY
until FLOWRATE icon appears. The factory
set time base will be highlighted to the right of
FLOWRATE (M = minutes, H = hours, D = days).
When FLOWRATE is invoked, the display will be
indicating rate of flow.
Activate the Meter
Computer is on continuously and always ready
to perform. The computer is powered by field
replaceable batteries. When display becomes
dim, faded or the low battery message appears
(see below), the batteries need to be replaced.
Reference the Maintenance Section for details.
Factory and Field Calibration
All calibration information is visible to the user
as icons on the top line of the display, above the
numeric digits.
All units are configured with a “factory” calibra-
tion.Bothgallonsandlitresareavailable(“GL”
or “LT” will be displayed). While holding the
CALIBRATE button, briefly press DISPLAY to
toggle between gallons and litres. This factory
calibration (indicated with FAC) is permanently
programmed into the computer and is not user
adjustable.
NOTE: Your computer may have other units of
measureprogrammedinto it. If so, holdingthe
CALIBRATE buttonandmomentarilypressing
the DISPLAY button will toggle through all
factory set units. Other possible units are:
IGL(imperialgallon),QT(quart),CF(cubic
feet), CM (cubic meter), BL (42 gal. barrel),
CC (cubic centimeter) or OZ (ounce).
www.calcert.com sales@calcert.com1.800.544.2843 0
5
10
15
20
25
30

5
Switching between different units will notcorrupt
theTotal’scontents.Forexample,inGLmode,
the computer totalizes 10.00 gallons, if the user
switches to LT mode, the display will read 37.85
litres (the same volume, different unit).
The “field” calibration may be setby the user, and
can be changed or modified at any time using
the calibration procedure described below in the
Calibration Section. Totals or flowrate derived
from the field calibration are invoked when the
FAC icon is no longer visible on the top line of
the display.
CALIBRATION
Verify Accuracy Before
Beginning Field Calibration
For the most accurate results, dispense at a
flowrate which best simulates your actual oper-
ating conditions. Avoid “dribbling” more fluid or
repeatedly starting and stopping the flow. This
can result in less accurate calibrations.
Makesureyoumeet themeter’sminimumflowrate
requirements:
TM Series Meters
3inchmeter 30GPM(113.6LPM)
4inchmeter 40GPM(151.4LPM)
The use of a uniformly dependable, accurate
calibration container is recommended for the
most accurate results. Due to high flowrate, it
is strongly recommended that calibration be
completed with a combination of volume and
weight using fine resolution scales.
For best results, the meter should be installed
and purged of air before field calibration.
Field Calibration with
Computer Display
Field Calibration and Factory Calibration are de-
finedintheOperation Section. Factory calibration
settings are programmed into each computer
during manufacturing, using water at 70° F (21°
C). Readings using the Factory Calibration (FAC)
may not be accurate in some situations, for ex-
ample, under extreme temperature conditions,
non-standard plumbing configurations or with
fluids other than water.
Field Calibration Procedures
(Correction Factor Method)
1. To calibrate, press and hold the CALIBRATE
and DISPLAY buttons for about 3 seconds
until you see FLdCAL. Release both buttons
and you will see CF - 00.0. You are now in
the field calibration mode and values from
-99.9% to +99.9% can be entered.
2. The
+/–positionappearseitherasan“under-
score” character for plus, or as a “hyphen”
character for minus. The DISPLAY button
selects the position and the CALIBRATE
button toggles this character.
3. The DISPLAY button can then be pushed to
select the numeric positions. Press the CALI-
BRATE button to scroll from 0 to 9. Enter the
percentage of change you want the display
to correct. When satisfied with the value,
pressbothCALIBRATE and DISPLAY buttons
simultaneously. CALEnd will be displayed
and unit will go back to normal operation,
less the FAC (factory calibration) icon.
4. All enabled units-of-measure remain visible
and selectable – the entered correction will
be applied to all enabled units.
5. To return to factory calibration (FAC), press
and hold both CALIBRATE and DISPLAY
buttons for about 3 seconds until FAcCAL is
displayed. Then release buttons. Unit should
return to normal operation and FAC icon is
visible.
Calibration with Conditioned
Signal Output Module
The K-factor of your meter appears on the
calibration report as the number of pulses per
gallon. The factor is determined during produc-
tion using water at 70° F (21° C). This K-factor
may be used for “single point” calibration and
provideacceptable accuracy. However,readings
may not be accurate when using this calibration
methodinsomesituations.For example, extreme
temperatureconditions, non-standard plumbing
configurations or with fluids other than water.
www.calcert.com sales@calcert.com1.800.544.2843 0
5
10
15
20
25
30

6
MAINTENANCE
Proper handling and care will extend the life and
service of the meter.
Turbine Rotor
The meter is virtually maintenance-free. How-
ever, it is important the rotor moves freely. Keep
the meter clean and free of contaminants.
Therotorcanberemovedforcleaning and inspec-
tion. Begin by unscrewing the nose cone from
the outlet end of the meter. A 1/4" square socket
extension can be used. Remove the lock and
flat thrust washers. Rotor can then be removed
from shaft. Note orientation because rotor is
not bi-directional. Remove debris or deposits
using soft brush or small probe. Reassemble in
reverse order.
If the display becomes dim, blank or the low
battery message appears (see below), replace
the batteries as follows:
1. Remove the four Phillips-head screws from
the face of the meter and lift the faceplate
from the turbine.
2. Remove the old batteries and clean any cor-
rosion from the terminals.
3. Install new batteries. Make sure the positive
post is in the correct position.
4. Whenthebatteriesarereplaced,thefaceplate
will power ON. Check the display to ensure
normal functions have resumed before as-
sembling again.
5. Reseat batteries, if necessary, and position
the faceplate on the turbine housing. To
avoid moisture damage, make sure the seal
is fully seated. Tighten the four screws on
the faceplate.
CAUTION
Blowingcompressedair throughthe turbine
assembly could damage the rotor.
CAUTION
Donot allow liquids to dry inside the turbine.
CAUTION
Handle the rotor carefully. Small scratches
or nicks can affect accuracy.
CAUTION
Batteries should ONLY be replaced with
P/N 113520-1 Kit (Includes two each P/N
902004-2Batteries).Do not mix oldwithnew.
Do not use other brands or technologies.
Open battery cells should be disposed
of in accordance with local regulations.
Lithium batteries are best disposed of as a
non-hazardous waste when fully or mostly
discharged. EPA does not list or exempt
Lithium as a hazardous waste. If waste
lithium batteries are still fully charged
or only partially discharged, they can be
considered a reactive hazardous waste
because of unconsumed lithium remaining
in the battery. Such batteries may qualify
as “Universal Waste” in many jurisdictions
within the U.S. and thus can be shipped for
disposal or recycling in accordance with
Universal Waste requirements.
NOTE: Make sure the arrow on the turbine outlet
is pointed in the direction of fluid flow.
Battery Replacement
The computer display is powered by two 3-volt
lithiumbatterieswhichmay be replacedwhilethe
meter is installed. When batteries are removed or
lose power, the batch and cumulative totals and
the field and factory calibrations are retained.
(Battery) – Avoid mechanical or electrical
abuse. Batteries may explode or cause
burns, if disassembled, crushed or exposed
to fire or temperatures in excess of 212°F
(100°C). Do not short circuit or install with
incorrect polarity. DO NOT INCINERATE.
WARNING
www.calcert.com sales@calcert.com1.800.544.2843 0
5
10
15
20
25
30

7
SPECIFICATIONS
Inlet and Outlet:
Spigot (Pipe) End Models:
TM300/TM300-P 3 inch Schd. 80,
Spigot (Pipe)
TM400/TM400-P 4 inch Schd. 80,
Spigot (Pipe)
NPT Models:
TM300-N/TM300-N-P 3 inch NPT
TM400-N/TM400-N-P 4 inch NPT
ANSI Flange Models:
TM300-F/TM300-F-P 3 inch 150#
ANSI Flange
TM400-F/TM400-F-P 4 inch 150#
ANSI Flange
DIN Flange Models:
TM300-D/TM300-D-P 3 inch PN10/PN16
DIN Flange
TM400-D/TM400-D-P 4 inch PN10/PN16
DIN Flange
Design Type: Turbine
Wetted Components:
Housing: PVC
Journal Bearings: PEEK
Shaft: Stainless Steel
Rotor and Nose Cone: Acetal
Washers: Stainless Steel
Max. Working Pressure:
TM300(AllModels):225PSIG@73°F
TM400(AllModels):225PSIG@73°F
(NonCE),135PSIG@73°F
(CE applications)
U.S. Measurement
Unit of Measure: Gallon
Flow Range:
3inch: 40-400GPM
(Linearity: ±3.0% of Reading)
30-600GPM
(Accuracy: ±2.0% Full Scale)
4inch: 60-600GPM
(Linearity: ±3.0 of Reading)
40-800GPM
(Accuracy ±2.0% Full Scale)
Accuracy with Computer:
± 3.0% of reading (Accuracy can be
improved with field calibration)
Operating Temperature: +32° to +140° F
(Do not allow fluid to freeze inside meter.)
Battery Life: 5 years
Storage Temperature: –40° to +158° F
Product Weight - lbs.:*
Spigot ANSI DIN
(Pipe) NPT Flange Flange
3 inch: 2.4 3.9 5.8 6.5
4 inch: 3.7 6.1 9.2 8.8
Dimensions - Inches (L x H x W):**
Spigot (Pipe): 3 inch: 11.5 x 5.34 x 3.5
4 inch: 13.5 x 6.34 x 4.5
NPT: 3 inch: 14.7 x 5.78 x 4.37
4 inch: 17.0 x 6.7 x 5.87
ANSI Flange:
3 inch: 12.0 x 7.5 x 7.5
4 inch: 14.0 x 9.0 x 9.0
DIN Flange:
3 inch: 12.0 x 8.0 x 8.0
4 inch: 14.0 x 8.5 x 8.5
*Add .3 lbs. for Conditioned Signal Module.
** Add 1.1 inches to height for Conditioned Signal
module.
Metric Measurement
Unit of Measure: Litre
Flow Range:
3 inch: 151-1514 LPM
(Linearity: ±3.0% of Reading)
113-2271 LPM
(Accuracy: ±2.0% Full Scale)
4 inch: 227-2271 LPM
(Linearity: ±3.0 of Reading)
151-3028 LPM
(Accuracy ±2.0% Full Scale)
Accuracy with Computer:
± 3.0% of reading (Accuracy can be
improved with field calibration)
Operating Temperature: 0° to +60° C
(Do not allow fluid to freeze inside meter.)
Battery Life: 5 years
Storage Temperature: –40° to +70° C
Product Weight - kgs:*
Spigot ANSI DIN
(Pipe) NPT Flange Flange
3 inch: 1.09 1.77 2.63 2.95
4 inch: 1.68 2.77 4.17 3.99
www.calcert.com sales@calcert.com1.800.544.2843 0
5
10
15
20
25
30

8
The Waste Electrical and Electronic
Equipment (WEEE) directive
(2002/96/EC) was approved by the
European Parliament and the Council
of the European Union in 2003. This
symbol indicates that this product
contains electrical and electronic
equipment that may include bat-
teries, printed circuit boards, liquid
crystal displays or other components that may be subject
to local disposal regulations at your location. Please
understand those regulations and dispose of this product
in a responsible manner.
CAUTION
Do not return the meter without specific
authority from the GPI Customer Service
Department. Due to strict regulations
governing transportation, handling, and
disposal of hazardous or flammable liquids,
GPIwillnotacceptmetersfor reworkunless
they are completely free of liquid residue.
Dimensions - cm (L x H x W):**
Spigot (Pipe): 3 inch: 29.2 x 13.6 x 8.9
4 inch: 34.3 x 16.1 x 11.4
NPT: 3 inch: 37.3 x 14.7 x 11.1
4 inch: 43.2 x 17.0 x 14.9
ANSI Flange:
3 inch: 30.5 x 20.3 x 20.3
4 inch: 35.6 x 22.9 x 22.9
DIN Flange:
3 inch: 30.5 x 20.3 x 20.3
4 inch: 35.6 x 21.6 x 21.6
*Add .14 kg to weight for Conditioned Signal
Module.
** Add 2.8 cm to height for Conditioned Signal
module.
PARTS
Thefollowingreplacementpartsand accessories
are available for the TM Series meters:
Part No. Description
113435-1 Conditioned Signal Output Module
113520-1 Battery Replacement Kit
12517601 Gasket
12519001 Computer Mount Assembly
901002-52 Seal
125518-01 3 in. (Spigot) Turbine Assy Kit
125518-02 3 in. (NPT) Turbine Assy Kit
125518-03 3 in. (Flange) Turbine Assy Kit
125520-01 4 in. (Spigot) Turbine Assy Kit
125520-02 4 in. (NPT) Turbine Assy Kit
125520-03 4 in. (Flange) Turbine Assy Kit
125519-01 3 in. Computer Assy Kit
125521-01 4 in. Computer Assy Kit
125518-04 3 in. (DIN Flange, PVC) Turbine Assy Kit
125520-05 4 in. (DIN Flange, PVC) Turbine Assy Kit
SERVICE
For warranty consideration, contact your local
distributor. If you need furtherassistance,contact
theGPICustomerServiceDepartmentat:
1-888-996-3837
You will need to:
• Provide information from the decal on your
meter.
• ReceiveaReturnAuthorizationnumber.
• Flushanyuidfromthemeterbeforeshipping
to the factory.
If possible leave customer installed fittings or
ample length of bare pipe for reinstallation.
WEEE DIRECTIVE
RoHS Compliant (2011/65/EU)
This product is in compliance with the RoHS
Directive of the European Parliament and of the
Council on the Restriction of the Use of Certain
Hazardous Substances in Electrical and Elec-
tronic Equipment.
Environmental Rating: IP65
www.calcert.com sales@calcert.com1.800.544.2843 0
5
10
15
20
25
30
Other manuals for TM Series
2
This manual suits for next models
16
Table of contents
Other GPI Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

ZAHNER
ZAHNER PAD4 Operation manual

Halma
Halma Ocean Optics LASER-785-IP-LAB Installation and operation manual

Agilent Technologies
Agilent Technologies VS Series Quick reference card

Vortran
Vortran Stradus VersaLase user manual

Thyracont
Thyracont VD82EX operating instructions

C-LOGIC
C-LOGIC 810 instruction manual