GPI 01N31GM-U User manual

Owner’s Manual for
01N31GM-U Electronic Flowmeter
Rev. - 920862-0109/09
5252 East 36th Street North
Wichita, KS USA 67220-3205
TEL: 316-686-7361
FAX: 316-686-6746
IMPORTANT NOTICE
Use this meter with water, DEF or similar fluids.
Do not use with flammable liquids, fuels or chemicals
that are not compatible with the wetted materials
found in the Specifications section.
This meter is not legal for trade applications.
The meter is sensitive to electrical “noise” and
may not operate correctly if located near some
electrical equipment.
CALIBRATION
This meter has a permanent factory calibration
for light viscosity fluids. If installed and used cor-
rectly, inaccuracies of no greater than ±5% will be
obtained.
INSTALLATION
This meter may be installed in-line either horizon-
tally or vertically or at the end of the hose adjacent to
the nozzle,if present. Installationto metalconnections
is not recommended.
The use of pipe sealant or PTFE tape is accept-
able. Review your application for suitability.
Hand tighten the meter at the housing ends. Do
not use a wrench or similar tools to tighten the meter
housing. The plastic housing could be damaged.
Batteries included.
OPERATION
Batch and Cumulative Totals
The meter maintains two totals. The batch total
may be reset to measure flow during a single use.
The cumulative total provides continuous measure-
ment and may not be manually reset. The batch total
is labeled as TTL1. The cumulative total is labeled
TTL2.
When the cumulative total reaches a maximum
reading of 9999, it will automatically reset to zero.
Press the DISPLAY button briefly to switch
between the batch and cumulative total.
Activate the Meter
Turn the meter ON by starting water flow or briefly
pressing the DISPLAY button. The meter will display
the batch or cumulative total from last use.
Press DISPLAY briefly to display the batch total.
Hold the DISPLAY button down for three seconds
to reset the batch total to zero.
The meter is programmed to turn off automatically
if not used for about one minute.
MAINTENANCE
Proper handling and care will extend the life and
service of the meter.
Turbine Rotor
The meter is virtually maintenance-free. However,
it is important the rotor moves freely. Keep the meter
clean and free of contaminants.
Remove any debris or deposits from the rotor
using a soft brush or small probe. Be careful not to
damage the turbine rotor or supports.
Blowing compressed air through the turbine
assembly could damage the rotor.
CAUTION
Owner’s Manual for
01N31GM-U Electronic Flowmeter
Rev. - 920862-0109/09
5252 East 36th Street North
Wichita, KS USA 67220-3205
TEL: 316-686-7361
FAX: 316-686-6746
IMPORTANT NOTICE
Use this meter with water, DEF or similar fluids.
Do not use with flammable liquids, fuels or chemicals
that are not compatible with the wetted materials
found in the Specifications section.
This meter is not legal for trade applications.
The meter is sensitive to electrical “noise” and
may not operate correctly if located near some
electrical equipment.
CALIBRATION
This meter has a permanent factory calibration
for light viscosity fluids. If installed and used cor-
rectly, inaccuracies of no greater than ±5% will be
obtained.
INSTALLATION
This meter may be installed in-line either horizon-
tally or vertically or at the end of the hose adjacent to
the nozzle,if present. Installationto metalconnections
is not recommended.
The use of pipe sealant or PTFE tape is accept-
able. Review your application for suitability.
Hand tighten the meter at the housing ends. Do
not use a wrench or similar tools to tighten the meter
housing. The plastic housing could be damaged.
Batteries included.
OPERATION
Batch and Cumulative Totals
The meter maintains two totals. The batch total
may be reset to measure flow during a single use.
The cumulative total provides continuous measure-
ment and may not be manually reset. The batch total
is labeled as TTL1. The cumulative total is labeled
TTL2.
When the cumulative total reaches a maximum
reading of 9999, it will automatically reset to zero.
Press the DISPLAY button briefly to switch
between the batch and cumulative total.
Activate the Meter
Turn the meter ON by starting water flow or briefly
pressing the DISPLAY button. The meter will display
the batch or cumulative total from last use.
Press DISPLAY briefly to display the batch total.
Hold the DISPLAY button down for three seconds
to reset the batch total to zero.
The meter is programmed to turn off automatically
if not used for about one minute.
MAINTENANCE
Proper handling and care will extend the life and
service of the meter.
Turbine Rotor
The meter is virtually maintenance-free. However,
it is important the rotor moves freely. Keep the meter
clean and free of contaminants.
Remove any debris or deposits from the rotor
using a soft brush or small probe. Be careful not to
damage the turbine rotor or supports.
Blowing compressed air through the turbine
assembly could damage the rotor.
CAUTION

Rev. - 920862-01
GPI is a registered trademark of Great Plains Industries, Inc.
© 2009 GREAT PLAINS INDUSTRIES, INC., Wichita, KS
Printed in U.S.A. 09/09
5252 East 36th Street North
Wichita, KS USA 67220-3205
TEL: 316-686-7361
FAX: 316-686-6746
Battery Replacement
The meter is powered by two AAA alkaline
batteries which may be replaced while the meter is
installed. When batteries are removed or lose power,
the batch and cumulative totals reset to zero but the
factory calibration is retained.
If themeter displaybecomes dimor blank,replace
the batteries as follows:
1. Remove the four Phillips-head screws from the
face of the meter and lift the faceplate from the
turbine.
2. Remove the old batteries and clean any corrosion
from the terminals.
3. Install new batteries. Make sure the positive post
is in the correct position.
4. When the batteries are replaced, the faceplate will
power ON. Check the display to ensure normal
functions have resumed before assembling
again.
5. Reseat batteries, if necessary, and position the
faceplate on the turbine housing. To avoid mois-
ture damage, make sure the seal is fully seated.
Tighten the four screws on the faceplate.
SPECIFICATIONS
Model: 01N31GM-U
Unit of Measure: Gallons
Flowrate: 3 to 30 GPM
Working Pressure: 150 PSIG
Operating Temperature: +14°F to +130°F
Storage Temperature: -40°F to +158°F
Threads: 1 inch NPT
Housing: Nylon
Bearings: Ceramic
Shaft: Tungsten Carbide
Rotor: Nylon
Signal Generators: Ferrite
Rings: 316 Stainless Steel
Limited Warranty Policy
Great Plains Industries, Inc. 5252 E. 36th Street North, Wichita, KS USA 67220-3205, hereby provides a limited warranty
against defects in material and workmanship on all products manufactured by Great Plains Industries, Inc. This product
includes a 1 year warranty from date of purchase as evidenced by the original sales receipt. A 30 month warranty from
product date of manufacture will apply in cases where the original sales receipt is not available. Reference product
labeling for the warranty expiration date based on 30 months from date of manufacture. Manufacturer’s sole obliga-
tion under the foregoing warranties will be limited to either, at Manufacturer’s option, replacing or repairing defective
Goods (subject to limitations hereinafter provided) or refunding the purchase price for such Goods theretofore paid by
the Buyer, and Buyer’s exclusive remedy for breach of any such warranties will be enforcement of such obligations of
Manufacturer. The warranty shall extend to the purchaser of this product and to any person to whom such product is
transferred during the warranty period.
This warranty shall not apply if:
A. the product has been altered or modified outside the warrantor’s duly appointed representative;
B. the product has been subjected to neglect, misuse, abuse or damage or has been installed or operated
other than in accordance with the manufacturer’s operating instructions.
To make a claim against this warranty, contact the GPI Customer Service Department at 316-686-7361 or 800-835-0113.
Or by mail at:
Great Plains Industries, Inc.
5252 E. 36th St. North
Wichita, KS, USA 67220-3205
GPI will step you through a product troubleshooting process to determine appropriate corrective actions.
GREAT PLAINS INDUSTRIES, INC., EXCLUDES LIABILITY UNDER THIS WARRANTY FOR DIRECT, INDIRECT,
INCIDENTAL AND CONSEQUENTIAL DAMAGES INCURRED IN THE USE OR LOSS OF USE OF THE PRODUCT
WARRANTED HEREUNDER.
The company herewith expressly disclaims any warranty of merchantability or fitness for any particular purpose other
than for which it was designed.
This warranty gives you specific rights and you may also have other rights which vary from U.S. state to U.S. state.
Note: In compliance with MAGNUSON MOSS CONSUMER WARRANTY ACT – Part 702 (governs the resale availability
of the warranty terms).
Rev. - 920862-01
GPI is a registered trademark of Great Plains Industries, Inc.
© 2009 GREAT PLAINS INDUSTRIES, INC., Wichita, KS
Printed in U.S.A. 09/09
5252 East 36th Street North
Wichita, KS USA 67220-3205
TEL: 316-686-7361
FAX: 316-686-6746
Battery Replacement
The meter is powered by two AAA alkaline
batteries which may be replaced while the meter is
installed. When batteries are removed or lose power,
the batch and cumulative totals reset to zero but the
factory calibration is retained.
If themeter displaybecomes dimor blank,replace
the batteries as follows:
1. Remove the four Phillips-head screws from the
face of the meter and lift the faceplate from the
turbine.
2. Remove the old batteries and clean any corrosion
from the terminals.
3. Install new batteries. Make sure the positive post
is in the correct position.
4. When the batteries are replaced, the faceplate will
power ON. Check the display to ensure normal
functions have resumed before assembling
again.
5. Reseat batteries, if necessary, and position the
faceplate on the turbine housing. To avoid mois-
ture damage, make sure the seal is fully seated.
Tighten the four screws on the faceplate.
SPECIFICATIONS
Model: 01N31GM-U
Unit of Measure: Gallons
Flowrate: 3 to 30 GPM
Working Pressure: 150 PSIG
Operating Temperature: +14°F to +130°F
Storage Temperature: -40°F to +158°F
Threads: 1 inch NPT
Housing: Nylon
Bearings: Ceramic
Shaft: Tungsten Carbide
Rotor: Nylon
Signal Generators: Ferrite
Rings: 316 Stainless Steel
Limited Warranty Policy
Great Plains Industries, Inc. 5252 E. 36th Street North, Wichita, KS USA 67220-3205, hereby provides a limited warranty
against defects in material and workmanship on all products manufactured by Great Plains Industries, Inc. This product
includes a 1 year warranty from date of purchase as evidenced by the original sales receipt. A 30 month warranty from
product date of manufacture will apply in cases where the original sales receipt is not available. Reference product
labeling for the warranty expiration date based on 30 months from date of manufacture. Manufacturer’s sole obliga-
tion under the foregoing warranties will be limited to either, at Manufacturer’s option, replacing or repairing defective
Goods (subject to limitations hereinafter provided) or refunding the purchase price for such Goods theretofore paid by
the Buyer, and Buyer’s exclusive remedy for breach of any such warranties will be enforcement of such obligations of
Manufacturer. The warranty shall extend to the purchaser of this product and to any person to whom such product is
transferred during the warranty period.
This warranty shall not apply if:
A. the product has been altered or modified outside the warrantor’s duly appointed representative;
B. the product has been subjected to neglect, misuse, abuse or damage or has been installed or operated
other than in accordance with the manufacturer’s operating instructions.
To make a claim against this warranty, contact the GPI Customer Service Department at 316-686-7361 or 800-835-0113.
Or by mail at:
Great Plains Industries, Inc.
5252 E. 36th St. North
Wichita, KS, USA 67220-3205
GPI will step you through a product troubleshooting process to determine appropriate corrective actions.
GREAT PLAINS INDUSTRIES, INC., EXCLUDES LIABILITY UNDER THIS WARRANTY FOR DIRECT, INDIRECT,
INCIDENTAL AND CONSEQUENTIAL DAMAGES INCURRED IN THE USE OR LOSS OF USE OF THE PRODUCT
WARRANTED HEREUNDER.
The company herewith expressly disclaims any warranty of merchantability or fitness for any particular purpose other
than for which it was designed.
This warranty gives you specific rights and you may also have other rights which vary from U.S. state to U.S. state.
Note: In compliance with MAGNUSON MOSS CONSUMER WARRANTY ACT – Part 702 (governs the resale availability
of the warranty terms).
Other GPI Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Toshiba
Toshiba RemotEye 4 Procedure for Installing

ELT Sensor
ELT Sensor MB-350U-V user guide

IFM Electronic
IFM Electronic efector 300 SD6101 operating instructions

HK Instruments
HK Instruments DPG Series installation instructions

Vega
Vega VEGADIS 371 Ex operating instructions

Electro Industries/GaugeTech
Electro Industries/GaugeTech Shark 100T quick start guide