Hach APA 6000 User manual

DOC023.97.80365
APA 6000
05/2013, Edition 1
Installation
Installation
Instalación

English...................................................................................................................................................................................................3
Français..............................................................................................................................................................................................21
Español...............................................................................................................................................................................................40
2

Table of contents
Specifications on page 3
General information on page 5
Installation on page 10
Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification Details
Dimensions (W x D x
H)
626.11 x 527.05 x 522.73 mm (24.65 x 20.75 x
20.58 in.)
Enclosure Rating: UL Type 4X (indoor), IEC 529 (IP 66) when air
purge is used and the door is closed and latched
Material: PC/ABS case, PC door, PC hinges and
latches, 316 SST hardware
Weight 22.4 kg (49.3 lb) dry, 28.9 kg (63.7 lb) wet with full
analyzer bottles1 (approximately)
Mounting Wall or bench
Protection class I
Pollution degree 2
Installation category II
Power requirements 100–240 VAC, 50/60 Hz, 100 VA
Connection: 18 to 14 AWG wire, 18 AWG stranded
recommended
Operating temperature 5 to 50 °C (41 to 122 °F); best performance: 5 to 35 °C
(41 to 95 °F)
Operating humidity 5 to 95% relative humidity, non-condensing
Maximum altitude 2000 m (6560 ft)
Table 1 General specifications (continued)
Specification Details
Recorder outputs Two 4–20 mA outputs applicable for recorders; load
impedance: 675 Ω maximum; 130 VAC isolation from
earth ground
Connection: 22 to 14 AWG wire, 22 to 20 AWG
recommended, twisted pair shielded wire
Alarm relay outputs Two dry contact SPDT relays, not powered, each rated
at 5 A resistive, 240 VAC maximum
Connection: 18 to 14 AWG wire, 18 AWG stranded
recommended
Analyzer bottles 1 liter, high-density polyethylene with polypropylene
caps
Fuses Two, 1.6 A, 250 V, 5 mm x 20 mm
External plumbing
connections
Sample lines: 6.35 mm (0.25-in.) OD push-in fitting for
poly tubing
Chemical drain (drain block): 12.7 mm (0.5-in.) ID barb
fitting for hose
Case drain: 25.4 mm (1-in.) ID for hose
Bottle enclosure drains: 9.53 mm (0.375-in.) ID hose
Air purge line: 6.35 mm (0.25-in.) OD push-in fitting for
poly tubing
Sample pressure No minimum or maximum requirement
Sample flow rate 100–2000 mL/minute; 500–1000 mL/minute
recommended
Sample temperature 5 to 50 °C (41 to 122 °F)
Sample particle size 0.5 µm maximum; sample goes through a 0.5 µm filter
in the sample conditioning block before analysis
Number of sample
streams
1; optional sample sequencing kit adds a second
sample stream, sample sequence is programmable
Node-to-node distance 400 m (1320 ft) maximum between analyzers
English 3

Table 1 General specifications (continued)
Specification Details
Air purge 15 scfh (7 L/min), dry instrument air, inert gas
Certifications Certified to UL and CSA by ETL, CE
1Six 1-liter bottles of water = 6 kg. The small amount of fluids in the analyzer =
0.5 kg. Add another 0.5 kg for chemicals that weigh more than water.
Table 2 Measurement specifications (part 1 of 2)
Specification Alkalinity Ammonia LR Ammonia/
Monochloramine
Light source Class 1M LED,
peak wavelength
600 nm
Class 1M LED,
peak wavelength
650 nm
Class 1M LED,
peak wavelength
650 nm
Parameters
measured
Total (T) alkalinity
Phenolphthalein
(P) alkalinity
Hydroxide
alkalinity
Carbonate
alkalinity
Bicarbonate
alkalinity
Free ammonia1Total ammonia,
monochloramine
Measurement
range
T alkalinity:
1–1000 mg/L as
CaCO3
P alkalinity:
5–1000 mg/L as
CaCO3
0.02–2 mg/L as N 0.1–10 mg/L as
Cl2
Accuracy2±5% or ±1.0 mg/L
(the larger value)
±5% or
±0.02 mg/L (the
larger value)
±5% or
±0.10 mg/L (the
larger value)
Precision/
Repeatability
±3% or ±0.6 mg/L
(the larger value)
±3% or
±0.01 mg/L (the
larger value)
±3% or
±0.05 mg/L (the
larger value)
Table 2 Measurement specifications (part 1 of 2) (continued)
Specification Alkalinity Ammonia LR Ammonia/
Monochloramine
Response time < 10 minutes3< 5 minutes3< 5 minutes3
Cycle time
(average)
8 minutes 6.2 minutes 6.2 minutes
Minimum
detection limit
≤ 0.10 mg/L ≤ 0.01 mg/L ≤ 0.05 mg/L
1The ammonia analyzer measures background monochloramine during
calibration only.
2Based on a calibration frequency of once a day
3For a 90% response to a step change at the sample inlet and one sample
stream
Table 3 Measurement specifications (part 2 of 2)
Specification Hardness LR Hardness HR
Light source Class 1M LED, peak
wavelength 520 nm
Class 1M LED, peak
wavelength 600 nm
Parameters measured Total hardness Total hardness
Measurement range 0.05–10 mg/L as CaCO310–1000 mg/L as CaCO3
Accuracy1±5% or ±0.05 mg/L (the
larger value)
±5% or ±2 mg/L (the larger
value)
Precision/ Repeatability ±3% or ±0.03 mg/L (the
larger value)
±5% or ±2 mg/L (the larger
value)
Response time < 5 minutes2< 17 minutes2
Cycle time (average) 4 minutes 8.2 minutes
Minimum detection limit 0.05 mg/L 10 mg/L
1Based on a calibration frequency of once a day
2For a 90% response to a step change at the sample inlet and one sample
stream
4 English

General information
In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without
notice or obligation. Revised editions are found on the manufacturer’s
website.
Safety information
NOTICE
The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent permitted
under applicable law. The user is solely responsible to identify critical application
risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.
Please read this entire manual before unpacking, setting up or operating
this equipment. Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage
to the equipment.
Make sure that the protection provided by this equipment is not impaired.
Do not use or install this equipment in any manner other than that
specified in this manual.
Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided, will
result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate
injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the instrument.
Information that requires special emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol on the
instrument is referenced in the manual with a precautionary statement.
This symbol, if noted on the instrument, references the instruction
manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of
in European public disposal systems after 12 August of 2005. In
conformity with European local and national regulations (EU Directive
2002/96/EC), European electrical equipment users must now return
old or end-of-life equipment to the Producer for disposal at no charge
to the user.
Note: For return for recycling, please contact the equipment producer or supplier
for instructions on how to return end-of-life equipment, producer-supplied
electrical accessories, and all auxiliary items for proper disposal.
This symbol indicates that a risk of electrical shock and/or
electrocution exists.
This symbol indicates the need for protective eye wear.
This symbol indicates that the marked item requires a protective earth
connection. If the instrument is not supplied with a ground plug on a
cord, make the protective earth connection to the protective
conductor terminal.
English 5

This symbol indicates that the marked item is the functional ground
connection for cable shield wires. Do not connect the ground wire of a
power cord or conduit to this item.
This symbol, when noted on the product, identifies the location of a
fuse or current limiting device.
Certification
Canadian Radio Interference-Causing Equipment Regulation,
IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la
réglementation canadienne sur les équipements provoquant des
interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device
complies with Part 15 of the FCC Rules. Operation is subject to the
following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by
the party responsible for compliance could void the user's authority to
operate the equipment. This equipment has been tested and found to
comply with the limits for a Class A digital device, pursuant to Part 15 of
the FCC rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference
to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference, in which case the user will be
required to correct the interference at their expense. The following
techniques can be used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or
is not the source of the interference.
2. If the equipment is connected to the same outlet as the device
experiencing interference, connect the equipment to a different
outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the
interference.
5. Try combinations of the above.
Product overview
DANGER
Chemical or biological hazards. If this instrument is used to monitor a
treatment process and/or chemical feed system for which there are
regulatory limits and monitoring requirements related to public health,
public safety, food or beverage manufacture or processing, it is the
responsibility of the user of this instrument to know and abide by any
applicable regulation and to have sufficient and appropriate
mechanisms in place for compliance with applicable regulations in the
event of malfunction of the instrument.
This analyzer continuously measures the hardness or the concentration
of a specific substance (alkalinity, ammonia or ammonia and
monochloramine) in drinking water, power water and industrial water.
Refer to Figure 1–Figure 3 for an overview of the analyzer.
6 English

Figure 1 Instrument overview – front
1 Door14 Drain block (sample drain)
2 Bottle enclosure for reagents25 Pressure gauge for drain block
3 Bottle enclosure for standards and
cleaning solution
6 Keypad and display
1Keep the door and door latch closed during operation so that the temperature
control system operates correctly.
2Reagents are light sensitive. Keep the bottle enclosure closed when bottles
are not being replaced.
Figure 2 Instrument overview – back
1 Air purge fitting 4 Case drain
2 Tubing strain relief holes 5 Sample conditioning block (sample
inlet)
3 Electrical access holes for external
cabling
English 7

Figure 3 Instrument overview – interior
1 Mixing module 4 Colorimeter module
2 Valve module 5 Pump (autoburette) module
3 Power switch
Product components
Make sure that all components have been received. Refer to
Figure 4–Figure 7. If any items are missing or damaged, contact the
manufacturer or a sales representative immediately.
Figure 4 Instrument components
1 APA 6000 analyzer 5 Reagent bottles1, 1 L
2 Tool kit (Figure 5)6 Standard solution bottles, 1 L (2x)
3 Installation kit (Figure 6)7 Cleaning solution bottle, 1 L
4 Maintenance kit (Figure 7)8 Powder funnel2
1The number of reagent bottles varies by analyzer model.
2Supplied only with the high range hardness analyzer
8 English

Figure 5 Tool kit
1 Tubing cutter, for Tefzel 0.030 ID x
0.062 OD tubing
7 Hex nut, 10-32, for 0.062-in. OD
tubing (4x)
2 Flangeless nut extender 8 Ferrule with locking ring, for 0.062-
in. OD tubing (10x)
3 Wrench, ¼-in., for the rotary valve 9 Flangeless nut, ¼-28, for 0.062-in.
OD tubing (10x)
4 Wrench, allen, 7/64-in. 10 Flangeless nut, plug, ¼-28
5 Alignment tool for the pump module 11 Fitting, union, Tefzel, ¼-in. (4x)
6 Ferrule, for 0.062-in. OD tubing (4x) 12 Tubing, Tefzel, 0.030 ID x
0.062 OD, 4.6 m (15 ft)
Figure 6 Installation kit
1 Grab sample kit 3 Ceramic filters, 0.5 µm (4x)
2 Tubing, polyethylene, 0.25-in. OD
x .040-in. ID, 6m (20 ft)
Figure 7 Maintenance kit
1 Valve rotors (4x) 3 Pump O-rings, fluorocarbon,
0.676 ID x 0.070 wide (4x)
2 Pump piston seals, flanged,
0.722 OD x 0.104 long (8x)
4 High temperature, non-flammable
grease
English 9

Installation
W A R N I N G
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
Installation guidelines
• Install the analyzer as near to the sampling point as possible.
• Make sure to leave sufficient space above the analyzer to install
analyzer bottles in the bottle enclosures. Refer to Figure 8.
• Make sure to leave sufficient space in front of the analyzer to open the
analyzer door. Refer to Figure 8.
• Install the analyzer indoors, in a non-hazardous environment.
• Do not install the analyzer in direct sunlight or near a heat source.
• Make sure that the ambient conditions are within operating
specifications. Refer to Specifications on page 3.
Figure 8 Analyzer dimensions
10 English

Mechanical installation
W A R N I N G
Personal injury hazard. Instruments or components are heavy. Use
assistance to install or move.
W A R N I N G
Personal injury hazard. The object is heavy. Make sure that the
instrument is securely attached to a wall, table or floor for a safe
operation.
W A R N I N G
Personal injury hazard. Make sure that the wall mounting is able to
hold 4 times the weight of the equipment.
Analyzer mounting
Install the analyzer on a bench or wall. Refer to Figure 9 and Figure 10.
Figure 9 Bench mounting
English 11

Figure 10 Wall mounting Electrical installation
DANGER
Electrocution hazard. Always remove power to the instrument before
making electrical connections.
DANGER
Electrocution hazard. If this equipment is used outdoors or in
potentially wet locations, a Ground Fault Circuit Interrupt (GFCI/GFI)
device must be used for connecting the equipment to its main power
source.
DANGER
Electrocution hazard. Protective Earth Ground (PE) connection is
required.
Use shielded twisted-pair cable for all electrical connections except input
power. Use of non-shielded cable may result in radio frequency emission
or susceptibility levels higher than the allowed levels.
To prevent shock hazards from ground currents in inadequate ground
systems, connect the shield at only the analyzer end. Do not connect the
shield wire at both ends.
Open the module panel
Refer to the illustrated steps in Figure 11.
12 English

Figure 13 Wiring connections
1 Network connection 6 Ground stud for ground wire
2 JP11 setting (remote analyzer)17 Alarm relay connections
3 JP11 setting (master or standalone
analyzer), default
8 Recorder output (4–20 mA)
connections
4 Power connection 9 Ground stud for cable shields
5 Fuses (2x)
1Do not change the default setting of JP11 unless the network connection is
used to connect another analyzer.
General wiring procedure
DANGER
Electrocution hazard. Always remove power to the instrument before
making electrical connections.
Connect wiring to the controller as shown in the illustrated steps in
Figure 14. Make all electrical cable connections through strain relief
fittings installed in the electrical access holes as shown in Figure 14.
Tighten the strain relief fittings to hold the cables securely.
To keep the enclosure rating, make sure that all unused electrical
access holes are sealed with an access hole cover.
Figure 14 General wiring procedure
14 English

Wiring for power
DANGER
Electrocution hazard. Protective Earth Ground (PE) connection is
required.
DANGER
Electrical shock and fire hazards. Make sure to identify the local
disconnect clearly for the conduit installation.
W A R N I N G
Potential Electrocution Hazard. If this equipment is used outdoors or in
potentially wet locations, a Ground Fault Interrupt device must be
used for connecting the equipment to its mains power source.
NOTICE
Install the device in a location and position that gives easy access to the
disconnect device and its operation.
Supply power to the analyzer with conduit or a power cable. Make sure
that a circuit breaker with sufficient current capacity is installed in the
power line. The circuit breaker size is based on the wire gauge used for
installation.
For installation with conduit:
• Install a local disconnect for the analyzer within 3 m (10 ft) of the
analyzer. Put a label on the disconnect that identifies it as the main
disconnect device for the analyzer.
• Make sure that the power and safety ground service drops for the
analyzer are 18–12 AWG and the wire insulation is rated for 300 VAC
or higher.
• Connect equipment in accordance with local, state or national
electrical codes.
• Connect the conduit through a conduit hub that holds the conduit
securely and seals the enclosure when tightened.
• If metal conduit is used, make sure that the conduit hub is tightened
so that the conduit hub connects the metal conduit to safely ground.
For installation with a power cable, make sure that the power cable is:
• Less than 3 m (10 ft) in length
• Rated for at least 60 °C (140 °F) and applicable to the installation
environment
• Not less than 18 AWG with applicable insulation colors for local code
requirements
• A power cable with a three-prong plug (with ground connection) that is
applicable to the supply connection
• Connected through a cable gland (strain relief) that holds the power
cable securely and seals the enclosure when tightened
• Does not have a locking type device on the plug
Connect to power
NOTICE
The instrument is used for a single phase connection only.
NOTICE
Make sure the power switch is set to off. Dry operation may damage the
analyzer.
Connect power with conduit or a power cable. Refer to Figure 15 and
Table 4.
Make sure that the safety ground wire is securely connected to the
protective earth ground terminal. Make sure that the metal conduit is
securely bonded to the safety ground wire through the conduit backing
plate.
An AC power cable is not supplied with the analyzer. To order an AC
power cable, refer to Replacement parts and accessories in the
maintenance manual.
English 15

Figure 15 Wiring for power Table 4 AC wiring information
Terminal Description Color—North
America
Color—EU
Ground
stud
Protective Earth (PE)
Ground
Green Green with
yellow stripe
J10-1 Hot (Φ1) Black Brown
J10-2 Neutral (Φ2) White Blue
Connect the alarm relays (optional)
W A R N I N G
Potential Electrocution Hazard. Power and relay terminals are
designed for only single wire termination. Do not use more than one
wire in each terminal.
W A R N I N G
Potential fire hazard. Do not daisy-chain the common relay
connections or jumper wire from the mains power connection inside
the instrument.
CAUTION
Fire hazard. Relay loads must be resistive. Always limit current to the
relays with an external fuse or breaker. Obey the relay ratings in the
Specifications section.
The analyzer has two alarm relays. The relay outputs are not powered.
Each alarm relay is associated with one alarm (e.g., Alarm A or Alarm B)
and energizes when that alarm occurs. The alarm relays are enabled or
disabled in the alarm settings. Refer to the operations manual to
configure the alarm settings.
Note: More alarm relays (maximum of twelve) can be added. A maximum of six 2-
channel signal output modules or one 8-channel signal output module can be
connected to the analyzer.
Refer to Table 5 and Figure 13 on page 14 to connect the alarm relays
to a control or alarm device.
16 English

The relay terminals accept 22–12 AWG wire (as determined by load
application). Wire gauge less than 18 AWG is not recommended. Use
wire with an insulation rating of 300 VAC or higher.
When switching large inductive loads (e.g., motors and pumps) or
currents higher than 5 A, use an auxiliary relay to extend the analyzer
relay life.
Use the relays at either all high voltage (greater than 30 V-RMS and
42.2 V-PEAK or 60 VDC) or all low voltage (less than 30 V-RMS and
42.2 V-PEAK, or less than 60 VDC). Do not configure a combination of
both high and low voltage.
Table 5 Relay wiring information
NC COM NO
Normally closed Common Normally open
Connect the recorder outputs (optional)
DANGER
Electrocution hazard. Always remove power to the instrument before
making electrical connections.
The analyzer has two recorder outputs (Recorder A and Recorder B).
The recorder outputs are isolated, self-powered, 4-20 mA current source
outputs. The current value at the recorder output represents the analyzer
measurement. Refer to the operations manual to configure the recorder
outputs.
Note: More recorder outputs (maximum of twelve) can be added. A maximum of
six 2-channel signal output modules or one 8-channel signal output module can be
connected to the analyzer.
To connect a recorder output, refer to Table 6 and Figure 13 on page 14.
The recommended output impedance is 3.6 to 500 ohms.
Table 6 Recorder output wiring
Recorder wires Circuit board markings
Recorder + +
Recorder – –
Shield GND
Connect a network device (optional)
Use the Network connection to connect an external Aquatrend network
device to the analyzer or to connect two or more analyzers to make an
Aquatrend network. The external Aquatrend network devices that can be
connected to the analyzer include:
•Signal output module—supplies additional recorder outputs and
alarm relays.
•Serial I/O module—lets the user send RS232 data to a printer or
send/receive RS232 data to/from a computer.
•MOD I/O module— lets the user send/receive Modbus data to/from a
PLC or SCADA.
Connect the cable termination kit supplied with the optional external
Aquatrend network device to the analyzer. Refer to Table 7 and
Figure 13 on page 14.
When the network connection wiring is done, configure the network
connection to configure the device or analyzer to the network. Refer to
the operations manual.
When the network connection is used to connect another APA analyzer,
use jumper JP11 to set one analyzer as the master and the other
analyzer(s) as a remote. Refer to Figure 13 on page 14. The master has
network configuration capability. The remote(s) is used as a display
device.
English 17

Table 7 Network wiring information
Terminal Color
A White
B Green
Ground stud Silver1
1Connect the shield (silver) wire at only one end of the cable. Connect the
shield wire to only one analyzer in an Aquatrend network. Cover the non-
connected end of the cable with insulation (e.g., tape or heat-shrink tubing).
Plumbing
DANGER
Fire hazard. This product is not designed for use with flammable
liquids.
Make sure to use the specified tubing size.
Plumb the sample and drain lines
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in
accordance with local, regional and national regulations.
NOTICE
Do not connect the drain lines to other lines or damage to the analyzer will occur
due to backpressure. Make sure that the drain lines are open to air.
NOTICE
Do not connect a hose to the case drain unless necessary because the case
drain hose can supply a path for corrosive fumes to enter the analyzer from the
facility drain.
Plumb the drain hoses and the sample lines to the analyzer. Refer to
Figure 16 and Figure 17. The drain hoses are supplied by the user. The
sample tubing is supplied with the analyzer.
Make sure that all the drains are at least 305 mm (12 in.) below the drain
block of the analyzer and drain to a facility chemical drain. Remove the
knock-out plugs in the bottom of each bottle enclosure to connect the
drain hoses.
If a case drain hose is necessary, make sure that there is an air break
between the case drain hose and the facility drain and that the facility
drain has a fume trap.
18 English

Figure 16 Drain lines
1 Bottle enclosure drains (9.53 mm
(0.375-in.) ID hose)
4 Air break
2 Chemical drain (12.7 mm (0.5-in.)
ID hose)
5 Case drain (25.4 mm (1-in.) ID
hose)
3 Fume trap
Figure 17 Sample lines
1 To drain or process stream 4 Sample bypass drain (0.25-in. OD x
0.040-in. ID poly tubing)
2 Vent 5 Sample in (0.25-in. OD x 0.040-in.
ID poly tubing)
3 Drain tee 6 Shut-off valve
Sample line guidelines
Select a good, representative sampling point for the best instrument
performance. The sample must be representative of the entire system.
To prevent erratic readings:
• Collect samples from locations that are sufficiently distant from points
of chemical additions to the process stream.
• Make sure that the samples are sufficiently mixed.
• Make sure that all chemical reactions are complete.
English 19

Connect the sample stream
Install the sample line into a larger process pipe to minimize interference
from air bubbles or pipeline bottom sediment. A sample line that goes
into the center of a process pipe is best.
Figure 18 shows examples of good and bad methods of sample line
installation into a process pipe.
Keep the sample line as short as possible to decrease analysis delay.
Sediment can collect in long sample lines.
Figure 18 Sampling methods
1 Air 2 Sample flow
Sample considerations
For wastewater applications or other samples that contain a large
amount of solids, use a 40-mesh Y-strainer to remove large solids alone
or with the optional micro filter system. Refer to Replacement parts and
accessories in the maintenance manual.
For highly aerated samples, a sample heater is recommended.
Increasing the temperature of the samples removes some of the excess
air and stabilizes the analyzer readings.
Sample line and valve cleanup
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and
wear all of the personal protective equipment appropriate to the
chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
New tubing, valves and other sample conditioning equipment may be
contaminated with silicate-based substances (oils, dust). These may
contribute to slightly high readings until they are cleaned.
1. Flush the sample line with sample for one to two hours.
2. For an expedient procedure, inject one to four liters of a dilute caustic
solution such as 1N (5%) sodium hydroxide solution into the front
end of the sample line. Force the solution through the analyzer to
clean sample system components.
Attach an air purge line (optional)
If the analyzer is installed in an environment with high humidity or caustic
vapors, attach an air purge line to the analyzer. Refer to Specifications
on page 3 for air purge specifications.
Note: The case drain is the vent for the air purge.
1. Remove the plug from the air purge fitting. Refer to Figure 2
on page 7.
2. Push 0.25-in. OD x 0.040-in. ID poly tubing into the air purge quick-
connect fitting until two distinct stops are felt.
Note: If not fully installed, the tubing will disconnect from the air purge fitting
when air pressure is applied.
Preparation for use
The physical installation is now complete. Refer to the operations
manual to set up the analyzer for the first use.
20 English
Table of contents
Languages:
Other Hach Measuring Instrument manuals

Hach
Hach POLYMETRON 9125 User manual

Hach
Hach 2200 PCX Quick manual

Hach
Hach TL2360 Quick manual

Hach
Hach TL2360 Quick manual

Hach
Hach sensION+ DL User manual

Hach
Hach CLF10sc User manual

Hach
Hach Sension4 Reference guide

Hach
Hach TitraLab User manual

Hach
Hach HQ1110 Quick manual

Hach
Hach sensION+ DO6 User manual