Hioki LR8450 User manual

i
1.12 Using the A/B Cursors .........................75
Reading values from the waveforms..........75
Specifying a waveform range ....................77
1.13 CongurationNavigator
(Quick Set).............................................78
Strain gage connection diagram ................78
External control terminal pin names...........79
2 Trigger Function 81
2.1 Trigger Meanings..................................83
2.2 Enabling the Trigger Function.............84
Shared settings.........................................84
2.3 Analog Triggers, Pulse Triggers,
Waveform Calculation Triggers
(Level, Window)....................................87
Level triggers............................................89
Window triggers........................................91
2.4 Logic Triggers (Patterns) .....................92
2.5 Applying Triggers Based on
External Sources ..................................94
2.6 Activating a Trigger at a Set Interval..95
Interval triggers.........................................95
2.7 Example Trigger Settings ....................97
3 Saving and Loading
Data 99
3.1 Data That Can Be Saved and
Loaded .................................................100
3.2 Formatting Media................................103
3.3 Saving Data.........................................105
Auto save (real-time save) .....................106
Manual saving (selective saving,
immediate saving)...................................112
Selective save operation .........................114
3.4 Loading Data.......................................117
3.5 Managing Data ...................................119
Switching media (drives) .........................119
Moving between levels (folders) ..............120
Deleting data ..........................................121
Renaminglesandfolders......................122
Copying data ..........................................123
Sortingles ............................................124
3.6 Acquiring Data with a Computer
(PC) ......................................................125
Connecting the USB cable ......................125
Activating USB drive mode......................126
Canceling USB drive mode .....................127
Contents
Introduction........................................................1
AbouttheNotationsUsedinThisManual.....2
How to Use This Manual .................................4
1 Settings and Operation 5
1.1 Performing Basic Operations................6
Instructions.................................................6
Value entry method .....................................7
Text entry method .......................................8
1.2 Setting Measurement Conditions.......10
Measurement module data refresh
intervals....................................................13
1.3 ConguringInputChannels ................16
Measuring voltage ....................................19
Measuring temperature
(with thermocouples).................................21
Measuring temperature
(with resistance temperature detectors) .....24
Measuring humidity...................................25
Measuring resistance................................26
Measuring strain .......................................27
Integrating pulses .....................................29
Measuring rotational speed .......................31
Measuring logic signals.............................34
Treatment of data that exceeds the
measurable range.....................................35
1.4 ConguringtheWaveformDisplay ....36
Conguringthedisplayofthevertical
axis ..........................................................36
Other display settings ...............................40
1.5 Using the Scaling Function .................42
1.6 Entering Comments .............................47
Title comments .........................................47
Channel comments...................................48
Moduleidentiers......................................49
1.7 ConguringChannelsinaList............50
Copying channel settings ..........................55
Conguringchannelsettingsatonce .........56
1.8 Performing Zero Adjustment ...............57
1.9 Checking Input Signals
(Monitor Function) ................................58
1.10 Starting and Stopping
Measurement ........................................59
1.11 Observing Waveforms .........................60
Waveform display .....................................62
Gage (scale) display .................................65
Numericalvaluedisplay ............................66
Moving waveforms (scrolling) ....................69
Scroll bar (waveform display position)........71
Enlarging and shrinking the waveform
horizontally ...............................................71
Waveform search......................................72
Jump function (changing the display
position) ...................................................74
10
9
8
7
6
5
4
3
2
1
Index
LR8450A964-00

Contents
ii
8 External Control
(EXT. I/O) 173
8.1 ConguringVoltageOutput
(VOUTPUT).........................................174
8.2 ConguringAlarmOutput
(ALARM) ..............................................175
8.3 ConguringExternalInput/Output
(I/O) Terminals ....................................177
External trigger input...............................179
Trigger output .........................................180
Simultaneously starting measurement
using external triggers.............................182
9 Communicating with a
Computer (PC) 183
9.1 Using the Logger Utility .....................185
9.2 ConguringandEstablishing
a USB Connection..............................186
Installing the USB driver..........................186
Connecting the instrument to the
computer with a USB cable .....................188
9.3 ConguringandEstablishing
aLANConnection ..............................189
Conguringthecomputer’snetwork
settings...................................................191
Conguringtheinstrument’sLAN
settings...................................................192
Connecting the instrument to a
computerwithaLANcable .....................196
9.4 Exporting Acquired Data to a
ComputerOverWirelessLAN ..........199
ConguringandEstablishingaWireless
LANConnection(LR8450-01only) ..........199
9.5 Performing Remote Measurement
with the HTTP Server.........................201
Connecting to the HTTP server ...............201
Remote control using a browser ..............203
Starting and stopping measurement ........204
Entering comments.................................205
9.6 Acquiring Data Using the FTP
Server...................................................206
9.7 Sending Data Using the FTP Client.210
Example computer FTP server settings ...211
Conguringautomaticsendingofdata.....225
Sendingatestle ...................................227
Checking FTP communications status .....228
9.8 Controlling the Instrument with
Communication Commands .............229
4 Alarm (Alarm Output) 129
4.1 ConguringAlarms.............................130
Setting shared alarm conditions for all
channels.................................................130
Conguringchannel-specicalarm
settings...................................................133
4.2 Checking Alarms.................................137
5 Marking Functionality 139
5.1 Assigning Event Marks during
Measurement ......................................140
5.2 Assigning Event Marks with an
External Signal....................................141
5.3 Assigning Event Marks When
Alarms Occur ......................................142
5.4 Searching for Event Marks................143
5.5 Reviewing Events in CSV Data........144
6 Numerical and
Waveform Calculations 145
6.1 PerformingNumericalCalculations .146
Conguringnumericalcalculations ..........147
Real-time numerical calculations
(automatic calculations) ..........................150
Numericalcalculationsafter
measurement (manual calculations) ........151
Partial numerical calculations ..................152
Numericalcalculationformulas................153
6.2 Performing Waveform Calculations .155
Conguringcalculationsonthe
calculation list screen..............................160
Copying calculation formulas...................161
Conguringwaveformcalculation
settings at once ......................................162
7 Conguring System
Settings 163
7.1 ConguringSettings...........................164
7.2 Controlling the System.......................167
Setting the time.......................................167
Synchronizing the time............................168
Initializing (resetting) the system .............169
Systemconguration ..............................170
Performing a self-check ..........................171

Contents
iii
10 Specications 231
10.1 BasicSpecications ...........................231
LR8450/LR8450-01 Memory HiLogger ....231
10.2 Plug-inModuleSpecications ..........243
U8550 Voltage/Temp Unit........................243
U8551 Universal Unit ..............................249
U8552 Voltage/Temp Unit........................255
U8553 High Speed Voltage Unit ..............262
U8554 Strain Unit ...................................265
11 Knowledge and
Information 269
11.1 Measuring Temperature.....................269
11.2 Measuring Strain ................................270
Tension and compression on a single
axis ........................................................270
Bending stress........................................271
Torsional stress.......................................272
Converting values to stress .....................273
Auto-balancing........................................273
Correcting for wiring resistance ...............274
Correcting for gage factor........................274
11.3 Digitalltercharacteristics ................275
11.4 NoiseCountermeasures....................276
Noisecontaminationmechanisms ...........276
Example noise countermeasures.............278
11.5 Scan Timing.........................................282
U8550, U8551, LR8530, and LR8531......283
U8552 and LR8532.................................284
U8553 and LR8533.................................285
11.6 Filenames ............................................286
11.7 Text Format .........................................287
11.8 File Size ...............................................289
11.9 Settings after Initialization
(System Reset) ...................................290
11.10 Maximum Recording Times ..............292
11.11 Application Measurement..................293
Recording instrumentation signals
(4-20 mA) ...............................................293
Measuring power consumption using
pulse output from a watt-hour meter ........296
11.12 Input Circuit Schematics....................299
11.13 Data Handling .....................................302
11.14DisplayingtheCertication
Number ................................................304
Index 305
9
8
7
6
5
4
Index
10
11

1
Introduction
Introduction
Thank you for purchasing the Hioki LR8450/LR8450-01 Memory HiLogger. To ensure your ability to
get the most out of the instrument over the long term, please read this manual carefully and keep it
available for future reference.
The LR8450-01 Memory HiLogger adds wireless LAN functionality to the LR8450.
The instrument comes with the following documentation. Please refer to these resources as
necessary in light of your specic application. Please review the separate “Operating Precautions”
before using the instrument.
Type Manual contents Printed
edition CD edition
Operating Precautions Information to ensure safe use of the instrument –
Precautions Concerning Use
of Equipment That Emits
Radio Waves
Precautions relating to use of equipment that
emits radio waves, countries in which the
instrument has been certied, etc.
–
Quick Start Manual Basic connection and operating instructions for
the instrument
Instruction Manual (this
manual)
Detailed information about the instrument and
specications –
Logger Utility*1
User Manual
Information about how to install and use the
computer application –
Communications
Commands*2User Manual
Explanation of communications commands for
controlling the instrument –
*1: For information about how to install and use the Logger Utility computer application, see “Logger Utility
User Manual” on the included CD (application disc).
*2: The instrument can be controlled by a LAN- or USB-connected computer (PC).
For more information about the communications commands used to control the instrument, see
“Communications Commands User Manual” on the included CD (application disc).
Intended audience
This manual has been written for use by individuals who use the product or provide information
about how to use the product.
In explaining how to use the product, it assumes electrical knowledge (equivalent of the knowledge
possessed by a graduate of an electrical program at a technical high school).
Trademarks
• Microsoft, Windows, Excel, Internet Explorer, and Visual Basic are the registered trademarks or
trademarks of Microsoft Corporation in the U.S., Japan, and other countries.
•Other products and company names are the trade names, registered trademarks or trademarks
of their respective owners.
Screen font
• The typefaces included herein are solely developed by DynaComware Taiwan Inc.

2
About the Notations Used in This Manual
About the Notations Used in This Manual
Safety notations
This manual classies seriousness of risks and hazard levels as described below.
DANGER Indicates an imminently hazardous situation that, if not avoided, will result in
death of or serious injury to the operator.
WARNING Indicates a potentially hazardous situation that, if not avoided, could result in
death of or serious injury to the operator.
CAUTION Indicates a potentially hazardous situation that, if not avoided, could result in
minor or moderate injury to the operator.
NOTICE Indicates the possibility of equipment damage.
IMPORTANT Indicates information or content that is particularly important from the standpoint
of operating or maintaining the instrument.
Indicates a high-voltage hazard. Failure to verify safety or improper handling of
the instrument could lead to electric shock, burns, or death.
Indicates an action that must not be performed.
Indicates action that must be performed.
Symbols on equipment
Indicates the need for caution or the presence of a hazard. For more information
about locations where this symbol appears on instrument components, see “Operating
Precautions” in the Quick Start Manual, warning messages listed at the beginning of
operating instructions, and the document entitled “Operating Precautions” that comes
with the instrument.
Indicates an instrument that has been protected throughout by double insulation or
reinforced insulation.
Indicates whether the power is on or o.
Indicates the ground terminal.
Indicates direct current (DC).
Indicates alternating current (AC).

3
About the Notations Used in This Manual
Notations related to standards compliance
Indicates compliance with the Waste Electrical and Electronic Equipment (WEEE)
Directive in EU member nations.
Li-ion
Indicates that the instrument is targeted for recycling under the Act on the Promotion of
Eective Utilization of Resources.
Indicates that the instrument complies with standards imposed by EU directives.
Other notations
Indicates useful advice concerning instrument performance and operation.
*Instructs the reader to see below for additional information.
Indicates the default setting. When initialized, the instrument will revert to this value.
(p. ) Indicates the page number to reference.
Bold The names of control keys are printed in bold.
[ ] The names of user interface elements on the screen are enclosed in brackets ([ ]).
Windows Unless otherwise noted, the term “Windows” is used generically to refer to Windows 7,
Windows 8, and Windows 10.
S/s
For the instrument, the number of times the analog input signal is digitized is indicated
in samples per second (S/s).
Example: 20 MS/s (20 megasamples per second) signies 20×106samples per
second.
Accuracy
Hioki denes tolerances for measured values in terms of f.s. (full scale), as indicated below.
f.s.
Maximum display value, scale magnitude
Indicates the maximum display value or scale magnitude. Generally speaking, the f.s.
gure indicates the range in current use.
Example: f.s. for the 1 V range = 1 V

4
How to Use This Manual
How to Use This Manual
How to open screens
: SET key
: Main tab
: Sub tab
Step number
Numbers are the same as those
used in the instructions.
Selections and explanations
Indicates the selections that
can be made by pressing the
ENTER key and explains them.
Indicates the default setting.

5
Unit Measure Channel Trigger Alarm Calculation System
1Settings and Operation
This chapter introduces basic settings and instrument
operation.
Before starting measurement, you must set measurement
conditions such as recording interval and range. You must
also congure input channel settings such as input signal
types and ranges.
1.1 Performing Basic Operations..............................................p. 6
1.2 Setting Measurement Conditions .......................................p. 10
1.3 Conguring Input Channels ................................................p. 16
1.4 Conguring the Waveform Display.....................................p. 36
1.5 Using the Scaling Function .................................................p. 42
1.6 Entering Comments .............................................................p. 47
1.7 Conguring Channels in a List ...........................................p. 50
1.8 Performing Zero Adjustment...............................................p. 57
1.9 Checking Input Signals (Monitor Function).......................p. 58
1.10 Starting and Stopping Measurement................................p. 59
1.11 Observing Waveforms........................................................p. 60
1.12 Using the A/B Cursors .......................................................p. 75
1.13 Conguration Navigator (Quick Set) ................................p. 78
1 Settings and Operation
1
Settings and Operation

6
Performing Basic Operations
1.1 Performing Basic Operations
Instructions
> > ): main tab; : sub tab)
33
22
In this manual, the active
item is said to have
“focus.” The background
of the selected item will
turn yellow.
1
Press the SET key to display the settings screen.
2
Select the main tab you wish to congure with the Left Arrow and Right Arrow keys.
Unit Measure Channel Trigger Alarm Calculation System
You can also switch among the main tabs using the SET key.
You can move the focus among sub tabs by pressing the ENTER key.
You can return the focus to the main tab by pressing the ESC key.
3
Select the sub tab whose settings you wish to congure with the Up Arrow and Down Arrow
keys
You can move the focus to the settings area by pressing the ENTER key.
You can return the focus to the sub tab by pressing the ESC key.
4
Select the setting you wish to congure with the Left Arrow, Right Arrow, Up Arrow, and
Down Arrow keys and then press the ENTER key.
The available settings will be displayed.
5
Select an option with the Up Arrow and Down Arrow keys and then press the ENTER key.
The setting will be accepted.
IMPORTANT
Unless otherwise directed, do not press multiple keys simultaneously. The instrument could
exhibit unintended behavior.
You can prevent accidental or unintended operation by enabling the key lock feature to disable
key operation.
See “Key lock (disabling keys)” in the Quick Start Manual.

7
Performing Basic Operations
Unit Measure Channel Trigger Alarm Calculation System
Value entry method
This section describes how to enter values.
Numerical value entry window
1122
33
Clear Clears the value.
BS Deletes one digit (backspace).
←Moves left one digit.
→Moves right one digit.
OK Accepts the value.
Cancel Closes the window without entering a value.
1
Select the desired value with the Left Arrow, Right Arrow, Up Arrow, and Down Arrow keys
and then press the ENTER key.
The selected value will be entered.
2
Select the SI prex.
P, T, G, M, k,
␣
, m, µ, n, p, f
The open box “
␣
” indicates a blank space.
3
Press the ENTER key while [OK] is selected. Alternatively, press the START key.
The window will close, and the value will be entered.
1
Settings and Operation

8
Performing Basic Operations
Text entry method
This section describes how to enter comments and lenames.
You can enter single-byte alphanumeric and characters only.
Text entry window
22
11
22
1
Select a character with the Left Arrow, Right Arrow, Up Arrow, and Down Arrow keys and
press the ENTER key.
2
Press the START key (OK). Alternatively, or press the ENTER key while [OK] is selected.
The text entry window will close.
When saved in text format, some characters are saved using alternative characters as follows:
Character
used on the
instrument
Saved
character
2^2
3^3
μ~u
Ω~o

Performing Basic Operations
Unit Measure Channel Trigger Alarm Calculation System 9
Text entry
You can perform the operations listed below by choosing the operation on the screen and pressing
the ENTER key.
You can also use the corresponding key to perform the same operation.
Operation on screen Corresponding key Description
OK START Accepts the character.
Cancel ESC Closes the window without entering any text.
Clear FILE Deletes all entered text.
BS Deletes the previous character (backspace).
DEL Deletes the next character (delete).
<< Moves the cursor to the left.
>> Moves the cursor to the right.
1
Settings and Operation

10
Setting Measurement Conditions
1.2 Setting Measurement Conditions
This section describes how to congure settings such as the recording interval and recording time.
Settings cannot be changed while measurement is in progress. Stop measurement and then
change the settings.
You can select the recording method.
Continuous recording
Sets the recording time setting to [Continuous].
Recording will continue until you press the STOP key. Recording can
be stopped with a trigger.
See “Stop triggers” (p. 84).
Time-specied recording
Sets the recording time setting to [Specied time].
The setting indicates the amount of time for which to record (the
recording length). Recording will stop once the set time has elapsed.
Recording can be stopped before the set time elapses with the STOP
key or a trigger.
Repeat recording
Sets the repeat recording setting to [ON].
Recording will resume after stopping (due to stop trigger conditions or
the completion of recording for the set recording time). Recording will
repeat until you press the STOP key. If the repeat recording setting is
[OFF], recording will stop after one iteration.

11
Setting Measurement Conditions
Unit Measure Channel Trigger Alarm Calculation System
>>
22
33
44
55
66
11
1
Enter a title comment in the [Title comment] eld (optional).
See “Title comments” (p. 47).
2
Under [Recording mode], select type of recording.
Normal Data is recorded in synchronization with the internal clock.
The setting cannot be changed from [Normal]. Hioki plans to add new functionality in the future.
3
Under [Recording interval], select the data capture interval.
Example: Choosing [10 ms] will cause data to be captured at a 10 ms interval (100 times per
second).
1 ms*1, 2 ms*1, 5 ms*1, 10 ms, 20 ms, 50 ms, 100 ms, 200 ms, 500 ms,
1 s, 2 s, 5 s, 10 s, 20 s, 30 s,
1 min, 2 min, 5 min, 10 min, 20 min, 30 min,
1 h
*1:
Setting available only when using a module with data refresh intervals including 1 ms. (when there is at
least one U8553, U8554, LR8533, or LR8534 module for which measurement is enabled).
4
Under [Repetitive recording], choose whether to repeat recording operation.
OFF Stops recording after one iteration.
ON Repeats recording operation.
Measurement will stop when you press the STOP key.
1
Settings and Operation

12
Setting Measurement Conditions
5
Under [Recording time], set the amount (length) of time for which you wish to record data.
Specied time Records for the set amount of time. (Max. 500 days)
Days, hours, minutes, seconds
Continuous Recording will continue until you press the STOP key.
Measurement can also be stopped with a trigger.
See “Stop triggers” (p. 84).
If you select [Specied time], the maximum time and data count will be displayed. The maximum amount of
time that you can record varies with the number of channels in use and the recording interval.
If you select [Continuous], measurement will continue while deleting past data from the internal buer
memory if the maximum capacity of the internal buer memory is exceeded. It is recommended to use auto
saving since deleted data cannot be restored.
See “Auto save (real-time save)” (p. 106).
6
Under [Alarm source], choose whether to record the alarm source channel during alarms.
Set to [ON] when you’re using an alarm and you wish to save data for the alarm source channel.
Set as necessary as the amount of data will increase.
OFF , ON
When the recording time is set to [Specied time] and repeat recording is set to
[ON]
Internal processing will take some time (downtime) between the time that recording stops after the
specied amount of time has elapsed and the start of the next recording operation. Recording is not
performed during this time.
Recording
Downtime
You can record without any downtime by setting the recording time to [Continuous] and enabling
auto saving and le segmentation. Saved data les can be segmented using the desired time
segment.

13
Setting Measurement Conditions
Unit Measure Channel Trigger Alarm Calculation System
Measurement module data refresh intervals
This section describes how to set the data refresh interval for each measurement module
separately from the instrument’s recording interval.
Data refresh interval Interval at which measurement modules refresh measurement data
Recording interval Interval at which the instrument captures data from measurement
modules
>
A list of connected modules will be displayed.
•[Unit 1] to [Unit 4]: Plug-in modules
•[Remote 1] to [Remote 7]: Wireless modules
1122
1
Under [Update], select the data refresh interval.
Auto , 1 ms, 2 ms, 5 ms, 10 ms, 20 ms, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s
Ordinarily, this setting should be set to [Auto]. When you choose [Auto], the shortest data refresh interval will
be set for each module based on the recording interval.
The available data refresh interval settings vary with the modules in use and the wire break detection setting.
Wire break detection U8550, U8551,
LR8530, LR8531 U8552, LR8532 U8553, U8554,
LR8533, LR8534
OFF From 10 ms*1From 20 ms*2
From 1 ms*4
ON From 20 ms*1From 50 ms*3
*1: Settings start at 100 ms when using a Pt1000 with the U8551 or LR8531.
*2: When using 15 or fewer channels, the 10 ms setting is available.
*3: When using 15 or fewer channels, the 20 ms setting is available.
*4: No wire break detection function is available.
• When the data refresh interval is set to a value other than [Auto], it is recommended to use
longer times.
This will allow you to reduce the digital lter’s cuto frequency to eliminate low-frequency noise.
•You can eliminate power supply frequency noise by setting the data refresh interval so that the
[Filter] setting is 50 Hz or 60 Hz.
1
Settings and Operation

14
Setting Measurement Conditions
2
Under [Filter], check the lter’s cuto frequency.
The lter’s cuto frequency will change depending on the data refresh interval setting.
Check the cuto frequency that is displayed for each module.
Relationship between the data refresh interval and the recording interval
• Measurement modules send data to the instrument once each data refresh interval.
• The instrument receives data from measurement modules once each recording interval.
• Even if a measurement module’s data refresh interval is short, it will not be possible to record
waveform peaks if the instrument’s recording interval is long.
Data refresh interval Recording interval
Short Long Short Long
Strength of the power
supply frequency lter Weak Strong – –
Data volume – – More data Less data
Waveform peaks Easier to capture* More dicult to
capture
Easier to capture* More dicult to
capture
*: If the data refresh interval and recording interval are short.
•For the U8550 to U8553 and LR8530 to LR8533 modules, the longer the data refresh interval,
the lower the digital lter cuto frequency, yielding more eective noise rejection. For more
information about cuto frequencies, see the section about each module’s digital lter in “10.2
Plug-in Module Specications” (p. 243).
•To maximize the eectiveness of the digital lter, congure the [Power frequency lter] setting
according to the power supply frequency in the region where the instrument is being used.
See “7.1 Conguring Settings” (p. 164).
•For modules whose data refresh interval is longer than the recording interval, the rst two data
points will be continuous, and there will be a delay.
Dierences in measurement data caused by data refresh intervals and recording intervals
: Point in measurement data
Peak
1 s
Module
Data refresh interval: 1 s
Peak
1 s
Module
Data refresh interval: 1 s
Unable to capture peak
2 s
LR8450/LR8450-1
Recording interval: 2 s
Able to capture peak
1 s
LR8450/LR8450-1
Recording interval: 1 s

15
Setting Measurement Conditions
Unit Measure Channel Trigger Alarm Calculation System
Example setting
What you want to do Data refresh interval Recording interval
Record a signal that’s changing quickly
(electrical signal, etc.)
Shorter Shorter
Record a signal that’s changing slowly
(temperature, etc.)
Longer Longer
Record fast and slow signals at the same time Shorter for modules used to
measure fast signals
Longer for modules used to
measure slow signals
Shorter
Since you can set the data refresh interval separately for each module, the instrument can be used
as follows:
• For module 1, set the data refresh interval to 2 s so that you can eliminate power supply noise
and reduce the eects of noise when performing temperature measurement using thermocouples.
• For module 2, set the data refresh interval to 10 ms so that you can record battery voltage
uctuations.
• For module 3, set the data refresh interval to 1 ms so that you can record changes in control
signals at the maximum speed.
•Set the instrument’s recording interval to 1 ms, reecting the shortest data refresh interval.
The instrument will record data from modules 1 through 3 every 1 ms.
If the instrument’s recording interval is shorter than a module’s data refresh interval, the same value
will be recorded for that module’s data.
Example: If the recording interval is 1 ms and the data refresh interval is 1 s, 1000 pieces of data all
of which have the same value will be recorded.
For more information about the module identier, see “Module identiers” (p. 49); about the lter
of U8554 and LR8534 Strain Unit, “Measuring strain” (p. 27).
Data refresh interval of pulses
Pulse data will be refreshed at the data refresh intervals.
The data refresh interval of pulses will automatically be set depending on input types.
Input type Data refresh interval
Integration 1 ms
Rotational speed r/s or r/min (Smoothing: 1 s) 10 ms
r/min (Smoothing: 2 s to 60 s) 50 ms
•The data refresh interval setting does not aect pulse counting processes.
•Even if the data refresh interval of pulses is equal to that for the measurement module, these data
will not simultaneously be refreshed when the recording interval is shorter than the data refresh
interval.
1
Settings and Operation

16
Conguring Input Channels
1.3 Conguring Input Channels
Congure input channels for voltage measurement, temperature measurement, etc.
Channel
Un-m
The letters nand mrepresent a module number and a channel
number, respectively.
Input Selects the type of measurement target.
Voltage, thermocouple, humidity, etc.
Range Sets the magnitude of the input signal.
Set the waveform color, scaling, and comments as necessary.
These settings can be congured either on each channel’s individual settings screen or on the
settings list screen for multiple channels.
Setting method
1
Set the main tab to [Channel].
2
Select the settings screen on the sub tab.
•[Individual]
Congure the settings on the individual settings screen for each channel.
•[Unit n] (n = 1, 2, . . .)
Congure the settings on the settings list screen for each module.
•[Pulse]
Congure the settings on the settings list screen.
3
Press the ENTER key.
The focus will move to the settings area.
You can return the focus to the sub tab by pressing the ESC key.
4
Select the setting you wish to congure with the Left Arrow, Right Arrow, Up Arrow, and
Down Arrow keys.
5
Press the ENTER key.
The available settings will be displayed.
6
Select the desired setting with the Up Arrow and Down Arrow keys and then press the
ENTER key.
The setting will be accepted.
Other manuals for LR8450
6
This manual suits for next models
1
Table of contents
Other Hioki Data Logger manuals

Hioki
Hioki LR8400-20 User manual

Hioki
Hioki LR5092-20 User manual

Hioki
Hioki 3910-20 User manual

Hioki
Hioki Hioki Instruction manual

Hioki
Hioki 8423 User manual

Hioki
Hioki 3638-20 User manual

Hioki
Hioki LR5092-20 User manual

Hioki
Hioki HiLogger 3145-20 User manual

Hioki
Hioki LR8512 User manual

Hioki
Hioki 3645-20 User manual

Hioki
Hioki lr5001 User manual

Hioki
Hioki 3640-20 User manual

Hioki
Hioki 3639-20 User manual

Hioki
Hioki LR8400-20 User manual

Hioki
Hioki lr5001 User manual

Hioki
Hioki LR8432-20 User manual

Hioki
Hioki LR5011 User manual

Hioki
Hioki LR8101 User guide

Hioki
Hioki LR8401-20 User manual

Hioki
Hioki LR8431-20 User manual