Hioki L2003 User manual

PIN TYPE PROBE
Instruction Manual
L2003
EN
November 2014 Edition 1
Printed in Japan
L2003A981-00 14-11H
Inspection/Maintenance
When you receive the productinspect it carefully to ensure
that no damage occurred during shipping. If damage is
evidentor if it fails to operate according to the specifications
contact your authorized Hioki distributor or reseller.
Preliminary checks
WARNING
Before using the productensure that the insulation on
the probes are not damaged and conductors are not
exposed. Using the product in such conditions could
cause an electric shockso contact your authorized
Hioki distributor or reseller for repair.
Maintenance and service
IMPORTANT
Never use cleaning agents containing benzenealcohol
acetoneetherketonesthinners or gasoline. These can
deform and discolor the case.
•To clean the productwipe it gently with a soft cloth moistened with
water or mild detergent.
•If the product seems to be malfunctioningcontact your authorized
Hioki distributor or reseller. Pack the product so that it will not get
damaged during shippingand include a description of existing
damages. Hioki cannot be responsible for damage that occurs
during shipment.
Disposal
Handle and dispose of the product in accordance with local
regulations.
Operating Precautions
Installing the product in inappropriate locations may cause
the probe to malfunction or lead to accidents. Please do not
install in the following locations. For details on the operating
temperature and humiditysee the specifications.
WARNING
• Exposed to direct sunlight or high temperatures
• Exposed to corrosive or combustible gases
• Exposed to strong electromagnetic fields or
electrostatic charges
• Near induction heating systems (such as high-
frequency induction heating systems and IH cooking
equipment)
• Susceptible to vibration
• Exposed to wateroilchemicalsor solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
CAUTION
•The ends of the probes are sharp. Be careful to avoid
injury.
•To prevent probe damagedo not step on probes or
pinch them between other objects. Do not bend or pull
on probes at their base.
Safety Information
Before using the productbe certain to carefully read the
following safety notes.
WARNING
•Customers are not allowed to modifydisassemble
or repair the instrument. Failure to observe these
precautions may result in fireelectric shockor injury.
•Ensure that the input does not exceed the maximum
input voltage or current to avoid product damage
resulting from heat building. Excessive voltage and
current can cause the product to malfunction.
CAUTION
• Mishandling could result in malfunction of the probe.
Be certain that you understand the instructions and
precautions in the manual before use.
•If persons unfamiliar with electricity measuring product
are to use the productanother person familiar with
such instruments must supervise operations.
Safety symbols
Indicates cautions and hazards. When the symbol is
printed on the productrefer to a corresponding topic
in the Instruction Manual.
Indicates DC (Direct Current).
Notation
In this manualthe risk seriousness and the hazard levels are
classified as follows.
WARNING
Indicates a potentially hazardous situation that may
result in death or serious injury to the operator.
CAUTION
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the
operator or damage to the product or malfunction.
IMPORTANT
Indicates information related to the operation of
the product or maintenance tasks with which the
operators must be fully familiar.
Indicates a prohibited action.
Indicates actions which must be performed.
Overview
L2003 Pin Type Probe is a four-terminal probe. The probe can be
used to contact various types of batteries since the measuring unit is
pin shaped.
Specifications
Measurable frequency
range
DC to 1050 Hz
Maximum rated voltage 30 V peak
Maximum rated current 2.5 A peak
Structure Four-terminal pair connection structure
Length Approx. 1500 mm (59.06″)
Mass Approx. 200 g (7.1 oz.)
Cable used 50 ΩCoaxial cable
Terminal structure Gold plating process
Tip shape: Needle
Used tip pin can be replaced.
Distance between pins 2.5 mm (0.10″)
Operating temperature
and humidity
Temperature; 0°C to 40°C (32°F to 104°F)
Humidity; 80% RH or less (no condensation)
Storage temperature
and humidity
Temperature; -10°C to 50°C (14°F to 122°F)
Humidity; 80% RH or less (no condensation)
Operating environment IndoorsPollution degree 2altitude up to
2000 m (6562 ft.)
Product warranty
period
1 year
Accessories Return Cable (400 mm)
Return Cable (550 mm)
Return Cable (800 mm)
Instruction Manual
Option 9772-90 Tip Pin
Part Names
Tip pin
Red (HIGH)
Black (LOW)
Return
Cable
(550 mm)
Mounting holes
d (HIGH)
Probe’s maximum rating is 30 Vbut be careful to avoid
exceeding the maximum rating of the measuring instrument.
IMPORTANT
Return Cable (550 mm) is shipped in a state connected to
the instrument.
Accessories
Plug
400 mm
800 mm
Connection Method
IMPORTANT
Clean the tip of the tip pin and the surface of the measuring
object before zero adjustment and measurement.
Connecting to the instrument
1
Adjust
23
Turn to the rightInsert
13-09
Introduction
Thank you for purchasing the HIOKI L2003 Pin Type Probe.
To obtain maximum performance from the product please
read this manual firstand keep it handy for future reference.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

PIN TYPE PROBE
Instruction Manual
L2003
EN
November 2014 Edition 1
Printed in Japan
L2003A981-00 14-11H
Inspection/Maintenance
When you receive the productinspect it carefully to ensure
that no damage occurred during shipping. If damage is
evidentor if it fails to operate according to the specifications
contact your authorized Hioki distributor or reseller.
Preliminary checks
WARNING
Before using the productensure that the insulation on
the probes are not damaged and conductors are not
exposed. Using the product in such conditions could
cause an electric shockso contact your authorized
Hioki distributor or reseller for repair.
Maintenance and service
IMPORTANT
Never use cleaning agents containing benzenealcohol
acetoneetherketonesthinners or gasoline. These can
deform and discolor the case.
•To clean the productwipe it gently with a soft cloth moistened with
water or mild detergent.
•If the product seems to be malfunctioningcontact your authorized
Hioki distributor or reseller. Pack the product so that it will not get
damaged during shippingand include a description of existing
damages. Hioki cannot be responsible for damage that occurs
during shipment.
Disposal
Handle and dispose of the product in accordance with local
regulations.
Operating Precautions
Installing the product in inappropriate locations may cause
the probe to malfunction or lead to accidents. Please do not
install in the following locations. For details on the operating
temperature and humiditysee the specifications.
WARNING
• Exposed to direct sunlight or high temperatures
• Exposed to corrosive or combustible gases
• Exposed to strong electromagnetic fields or
electrostatic charges
• Near induction heating systems (such as high-
frequency induction heating systems and IH cooking
equipment)
• Susceptible to vibration
• Exposed to wateroilchemicalsor solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
CAUTION
•The ends of the probes are sharp. Be careful to avoid
injury.
•To prevent probe damagedo not step on probes or
pinch them between other objects. Do not bend or pull
on probes at their base.
Safety Information
Before using the productbe certain to carefully read the
following safety notes.
WARNING
•Customers are not allowed to modifydisassemble
or repair the instrument. Failure to observe these
precautions may result in fireelectric shockor injury.
•Ensure that the input does not exceed the maximum
input voltage or current to avoid product damage
resulting from heat building. Excessive voltage and
current can cause the product to malfunction.
CAUTION
• Mishandling could result in malfunction of the probe.
Be certain that you understand the instructions and
precautions in the manual before use.
•If persons unfamiliar with electricity measuring product
are to use the productanother person familiar with
such instruments must supervise operations.
Safety symbols
Indicates cautions and hazards. When the symbol is
printed on the productrefer to a corresponding topic
in the Instruction Manual.
Indicates DC (Direct Current).
Notation
In this manualthe risk seriousness and the hazard levels are
classified as follows.
WARNING
Indicates a potentially hazardous situation that may
result in death or serious injury to the operator.
CAUTION
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the
operator or damage to the product or malfunction.
IMPORTANT
Indicates information related to the operation of
the product or maintenance tasks with which the
operators must be fully familiar.
Indicates a prohibited action.
Indicates actions which must be performed.
Overview
L2003 Pin Type Probe is a four-terminal probe. The probe can be
used to contact various types of batteries since the measuring unit is
pin shaped.
Specifications
Measurable frequency
range
DC to 1050 Hz
Maximum rated voltage 30 V peak
Maximum rated current 2.5 A peak
Structure Four-terminal pair connection structure
Length Approx. 1500 mm (59.06″)
Mass Approx. 200 g (7.1 oz.)
Cable used 50 ΩCoaxial cable
Terminal structure Gold plating process
Tip shape: Needle
Used tip pin can be replaced.
Distance between pins 2.5 mm (0.10″)
Operating temperature
and humidity
Temperature; 0°C to 40°C (32°F to 104°F)
Humidity; 80% RH or less (no condensation)
Storage temperature
and humidity
Temperature; -10°C to 50°C (14°F to 122°F)
Humidity; 80% RH or less (no condensation)
Operating environment IndoorsPollution degree 2altitude up to
2000 m (6562 ft.)
Product warranty
period
1 year
Accessories Return Cable (400 mm)
Return Cable (550 mm)
Return Cable (800 mm)
Instruction Manual
Option 9772-90 Tip Pin
Part Names
Tip pin
Red (HIGH)
Black (LOW)
Return
Cable
(550 mm)
Mounting holes
d (HIGH)
Probe’s maximum rating is 30 Vbut be careful to avoid
exceeding the maximum rating of the measuring instrument.
IMPORTANT
Return Cable (550 mm) is shipped in a state connected to
the instrument.
Accessories
Plug
400 mm
800 mm
Connection Method
IMPORTANT
Clean the tip of the tip pin and the surface of the measuring
object before zero adjustment and measurement.
Connecting to the instrument
1
Adjust
23
Turn to the rightInsert
13-09
Introduction
Thank you for purchasing the HIOKI L2003 Pin Type Probe.
To obtain maximum performance from the product please
read this manual firstand keep it handy for future reference.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

PIN TYPE PROBE
Instruction Manual
L2003
EN
November 2014 Edition 1
Printed in Japan
L2003A981-00 14-11H
Inspection/Maintenance
When you receive the productinspect it carefully to ensure
that no damage occurred during shipping. If damage is
evidentor if it fails to operate according to the specifications
contact your authorized Hioki distributor or reseller.
Preliminary checks
WARNING
Before using the productensure that the insulation on
the probes are not damaged and conductors are not
exposed. Using the product in such conditions could
cause an electric shockso contact your authorized
Hioki distributor or reseller for repair.
Maintenance and service
IMPORTANT
Never use cleaning agents containing benzenealcohol
acetoneetherketonesthinners or gasoline. These can
deform and discolor the case.
•To clean the productwipe it gently with a soft cloth moistened with
water or mild detergent.
•If the product seems to be malfunctioningcontact your authorized
Hioki distributor or reseller. Pack the product so that it will not get
damaged during shippingand include a description of existing
damages. Hioki cannot be responsible for damage that occurs
during shipment.
Disposal
Handle and dispose of the product in accordance with local
regulations.
Operating Precautions
Installing the product in inappropriate locations may cause
the probe to malfunction or lead to accidents. Please do not
install in the following locations. For details on the operating
temperature and humiditysee the specifications.
WARNING
• Exposed to direct sunlight or high temperatures
• Exposed to corrosive or combustible gases
• Exposed to strong electromagnetic fields or
electrostatic charges
• Near induction heating systems (such as high-
frequency induction heating systems and IH cooking
equipment)
• Susceptible to vibration
• Exposed to wateroilchemicalsor solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
CAUTION
•The ends of the probes are sharp. Be careful to avoid
injury.
•To prevent probe damagedo not step on probes or
pinch them between other objects. Do not bend or pull
on probes at their base.
Safety Information
Before using the productbe certain to carefully read the
following safety notes.
WARNING
•Customers are not allowed to modifydisassemble
or repair the instrument. Failure to observe these
precautions may result in fireelectric shockor injury.
•Ensure that the input does not exceed the maximum
input voltage or current to avoid product damage
resulting from heat building. Excessive voltage and
current can cause the product to malfunction.
CAUTION
• Mishandling could result in malfunction of the probe.
Be certain that you understand the instructions and
precautions in the manual before use.
•If persons unfamiliar with electricity measuring product
are to use the productanother person familiar with
such instruments must supervise operations.
Safety symbols
Indicates cautions and hazards. When the symbol is
printed on the productrefer to a corresponding topic
in the Instruction Manual.
Indicates DC (Direct Current).
Notation
In this manualthe risk seriousness and the hazard levels are
classified as follows.
WARNING
Indicates a potentially hazardous situation that may
result in death or serious injury to the operator.
CAUTION
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the
operator or damage to the product or malfunction.
IMPORTANT
Indicates information related to the operation of
the product or maintenance tasks with which the
operators must be fully familiar.
Indicates a prohibited action.
Indicates actions which must be performed.
Overview
L2003 Pin Type Probe is a four-terminal probe. The probe can be
used to contact various types of batteries since the measuring unit is
pin shaped.
Specifications
Measurable frequency
range
DC to 1050 Hz
Maximum rated voltage 30 V peak
Maximum rated current 2.5 A peak
Structure Four-terminal pair connection structure
Length Approx. 1500 mm (59.06″)
Mass Approx. 200 g (7.1 oz.)
Cable used 50 ΩCoaxial cable
Terminal structure Gold plating process
Tip shape: Needle
Used tip pin can be replaced.
Distance between pins 2.5 mm (0.10″)
Operating temperature
and humidity
Temperature; 0°C to 40°C (32°F to 104°F)
Humidity; 80% RH or less (no condensation)
Storage temperature
and humidity
Temperature; -10°C to 50°C (14°F to 122°F)
Humidity; 80% RH or less (no condensation)
Operating environment IndoorsPollution degree 2altitude up to
2000 m (6562 ft.)
Product warranty
period
1 year
Accessories Return Cable (400 mm)
Return Cable (550 mm)
Return Cable (800 mm)
Instruction Manual
Option 9772-90 Tip Pin
Part Names
Tip pin
Red (HIGH)
Black (LOW)
Return
Cable
(550 mm)
Mounting holes
d (HIGH)
Probe’s maximum rating is 30 Vbut be careful to avoid
exceeding the maximum rating of the measuring instrument.
IMPORTANT
Return Cable (550 mm) is shipped in a state connected to
the instrument.
Accessories
Plug
400 mm
800 mm
Connection Method
IMPORTANT
Clean the tip of the tip pin and the surface of the measuring
object before zero adjustment and measurement.
Connecting to the instrument
1
Adjust
23
Turn to the rightInsert
13-09
Introduction
Thank you for purchasing the HIOKI L2003 Pin Type Probe.
To obtain maximum performance from the product please
read this manual firstand keep it handy for future reference.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

PIN TYPE PROBE
Instruction Manual
L2003
EN
November 2014 Edition 1
Printed in Japan
L2003A981-00 14-11H
Inspection/Maintenance
When you receive the productinspect it carefully to ensure
that no damage occurred during shipping. If damage is
evidentor if it fails to operate according to the specifications
contact your authorized Hioki distributor or reseller.
Preliminary checks
WARNING
Before using the productensure that the insulation on
the probes are not damaged and conductors are not
exposed. Using the product in such conditions could
cause an electric shockso contact your authorized
Hioki distributor or reseller for repair.
Maintenance and service
IMPORTANT
Never use cleaning agents containing benzenealcohol
acetoneetherketonesthinners or gasoline. These can
deform and discolor the case.
•To clean the productwipe it gently with a soft cloth moistened with
water or mild detergent.
•If the product seems to be malfunctioningcontact your authorized
Hioki distributor or reseller. Pack the product so that it will not get
damaged during shippingand include a description of existing
damages. Hioki cannot be responsible for damage that occurs
during shipment.
Disposal
Handle and dispose of the product in accordance with local
regulations.
Operating Precautions
Installing the product in inappropriate locations may cause
the probe to malfunction or lead to accidents. Please do not
install in the following locations. For details on the operating
temperature and humiditysee the specifications.
WARNING
• Exposed to direct sunlight or high temperatures
• Exposed to corrosive or combustible gases
• Exposed to strong electromagnetic fields or
electrostatic charges
• Near induction heating systems (such as high-
frequency induction heating systems and IH cooking
equipment)
• Susceptible to vibration
• Exposed to wateroilchemicalsor solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
CAUTION
•The ends of the probes are sharp. Be careful to avoid
injury.
•To prevent probe damagedo not step on probes or
pinch them between other objects. Do not bend or pull
on probes at their base.
Safety Information
Before using the productbe certain to carefully read the
following safety notes.
WARNING
•Customers are not allowed to modifydisassemble
or repair the instrument. Failure to observe these
precautions may result in fireelectric shockor injury.
•Ensure that the input does not exceed the maximum
input voltage or current to avoid product damage
resulting from heat building. Excessive voltage and
current can cause the product to malfunction.
CAUTION
• Mishandling could result in malfunction of the probe.
Be certain that you understand the instructions and
precautions in the manual before use.
•If persons unfamiliar with electricity measuring product
are to use the productanother person familiar with
such instruments must supervise operations.
Safety symbols
Indicates cautions and hazards. When the symbol is
printed on the productrefer to a corresponding topic
in the Instruction Manual.
Indicates DC (Direct Current).
Notation
In this manualthe risk seriousness and the hazard levels are
classified as follows.
WARNING
Indicates a potentially hazardous situation that may
result in death or serious injury to the operator.
CAUTION
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the
operator or damage to the product or malfunction.
IMPORTANT
Indicates information related to the operation of
the product or maintenance tasks with which the
operators must be fully familiar.
Indicates a prohibited action.
Indicates actions which must be performed.
Overview
L2003 Pin Type Probe is a four-terminal probe. The probe can be
used to contact various types of batteries since the measuring unit is
pin shaped.
Specifications
Measurable frequency
range
DC to 1050 Hz
Maximum rated voltage 30 V peak
Maximum rated current 2.5 A peak
Structure Four-terminal pair connection structure
Length Approx. 1500 mm (59.06″)
Mass Approx. 200 g (7.1 oz.)
Cable used 50 ΩCoaxial cable
Terminal structure Gold plating process
Tip shape: Needle
Used tip pin can be replaced.
Distance between pins 2.5 mm (0.10″)
Operating temperature
and humidity
Temperature; 0°C to 40°C (32°F to 104°F)
Humidity; 80% RH or less (no condensation)
Storage temperature
and humidity
Temperature; -10°C to 50°C (14°F to 122°F)
Humidity; 80% RH or less (no condensation)
Operating environment IndoorsPollution degree 2altitude up to
2000 m (6562 ft.)
Product warranty
period
1 year
Accessories Return Cable (400 mm)
Return Cable (550 mm)
Return Cable (800 mm)
Instruction Manual
Option 9772-90 Tip Pin
Part Names
Tip pin
Red (HIGH)
Black (LOW)
Return
Cable
(550 mm)
Mounting holes
d (HIGH)
Probe’s maximum rating is 30 Vbut be careful to avoid
exceeding the maximum rating of the measuring instrument.
IMPORTANT
Return Cable (550 mm) is shipped in a state connected to
the instrument.
Accessories
Plug
400 mm
800 mm
Connection Method
IMPORTANT
Clean the tip of the tip pin and the surface of the measuring
object before zero adjustment and measurement.
Connecting to the instrument
1
Adjust
23
Turn to the rightInsert
13-09
Introduction
Thank you for purchasing the HIOKI L2003 Pin Type Probe.
To obtain maximum performance from the product please
read this manual firstand keep it handy for future reference.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

PIN TYPE PROBE
Instruction Manual
L2003
EN
November 2014 Edition 1
Printed in Japan
L2003A981-00 14-11H
Inspection/Maintenance
When you receive the productinspect it carefully to ensure
that no damage occurred during shipping. If damage is
evidentor if it fails to operate according to the specifications
contact your authorized Hioki distributor or reseller.
Preliminary checks
WARNING
Before using the productensure that the insulation on
the probes are not damaged and conductors are not
exposed. Using the product in such conditions could
cause an electric shockso contact your authorized
Hioki distributor or reseller for repair.
Maintenance and service
IMPORTANT
Never use cleaning agents containing benzenealcohol
acetoneetherketonesthinners or gasoline. These can
deform and discolor the case.
•To clean the productwipe it gently with a soft cloth moistened with
water or mild detergent.
•If the product seems to be malfunctioningcontact your authorized
Hioki distributor or reseller. Pack the product so that it will not get
damaged during shippingand include a description of existing
damages. Hioki cannot be responsible for damage that occurs
during shipment.
Disposal
Handle and dispose of the product in accordance with local
regulations.
Operating Precautions
Installing the product in inappropriate locations may cause
the probe to malfunction or lead to accidents. Please do not
install in the following locations. For details on the operating
temperature and humiditysee the specifications.
WARNING
• Exposed to direct sunlight or high temperatures
• Exposed to corrosive or combustible gases
• Exposed to strong electromagnetic fields or
electrostatic charges
• Near induction heating systems (such as high-
frequency induction heating systems and IH cooking
equipment)
• Susceptible to vibration
• Exposed to wateroilchemicalsor solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
CAUTION
•The ends of the probes are sharp. Be careful to avoid
injury.
•To prevent probe damagedo not step on probes or
pinch them between other objects. Do not bend or pull
on probes at their base.
Safety Information
Before using the productbe certain to carefully read the
following safety notes.
WARNING
•Customers are not allowed to modifydisassemble
or repair the instrument. Failure to observe these
precautions may result in fireelectric shockor injury.
•Ensure that the input does not exceed the maximum
input voltage or current to avoid product damage
resulting from heat building. Excessive voltage and
current can cause the product to malfunction.
CAUTION
• Mishandling could result in malfunction of the probe.
Be certain that you understand the instructions and
precautions in the manual before use.
•If persons unfamiliar with electricity measuring product
are to use the productanother person familiar with
such instruments must supervise operations.
Safety symbols
Indicates cautions and hazards. When the symbol is
printed on the productrefer to a corresponding topic
in the Instruction Manual.
Indicates DC (Direct Current).
Notation
In this manualthe risk seriousness and the hazard levels are
classified as follows.
WARNING
Indicates a potentially hazardous situation that may
result in death or serious injury to the operator.
CAUTION
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the
operator or damage to the product or malfunction.
IMPORTANT
Indicates information related to the operation of
the product or maintenance tasks with which the
operators must be fully familiar.
Indicates a prohibited action.
Indicates actions which must be performed.
Overview
L2003 Pin Type Probe is a four-terminal probe. The probe can be
used to contact various types of batteries since the measuring unit is
pin shaped.
Specifications
Measurable frequency
range
DC to 1050 Hz
Maximum rated voltage 30 V peak
Maximum rated current 2.5 A peak
Structure Four-terminal pair connection structure
Length Approx. 1500 mm (59.06″)
Mass Approx. 200 g (7.1 oz.)
Cable used 50 ΩCoaxial cable
Terminal structure Gold plating process
Tip shape: Needle
Used tip pin can be replaced.
Distance between pins 2.5 mm (0.10″)
Operating temperature
and humidity
Temperature; 0°C to 40°C (32°F to 104°F)
Humidity; 80% RH or less (no condensation)
Storage temperature
and humidity
Temperature; -10°C to 50°C (14°F to 122°F)
Humidity; 80% RH or less (no condensation)
Operating environment IndoorsPollution degree 2altitude up to
2000 m (6562 ft.)
Product warranty
period
1 year
Accessories Return Cable (400 mm)
Return Cable (550 mm)
Return Cable (800 mm)
Instruction Manual
Option 9772-90 Tip Pin
Part Names
Tip pin
Red (HIGH)
Black (LOW)
Return
Cable
(550 mm)
Mounting holes
d (HIGH)
Probe’s maximum rating is 30 Vbut be careful to avoid
exceeding the maximum rating of the measuring instrument.
IMPORTANT
Return Cable (550 mm) is shipped in a state connected to
the instrument.
Accessories
Plug
400 mm
800 mm
Connection Method
IMPORTANT
Clean the tip of the tip pin and the surface of the measuring
object before zero adjustment and measurement.
Connecting to the instrument
1
Adjust
23
Turn to the rightInsert
13-09
Introduction
Thank you for purchasing the HIOKI L2003 Pin Type Probe.
To obtain maximum performance from the product please
read this manual firstand keep it handy for future reference.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

Use a cable of suitable length for the distance between the
terminals of the measuring object from the three types of
accessory cables.
Return cable length Distance between the terminals
of the measuring object
400 mm 100 mm or less
550 mm
(mounted before shipment) 100 mm to 250 mm
800 mm 250 mm to 500 mm
Fixing Using the Mounting
Holes
The probe can be fixed using the mounting holes. (Please
prepare your own jig or equipment to fix.)
Hole diameter: 3.2 mmuse a M3 screw.
Jig or
equipment
to fix
M3 screw
12 mm
12 mm + thickness of jig or equipment
2.5 mm
9.15 mm
1.8 mm
First contact
pressure:
Approx. 60 g
Full stroke: 3 mm±0.2 mm
(condensing pressure
approx. 230 g)
Recommended
stroke:
2 mm
(condensing pressure
approx. 170 g)
Zero Adjustment
Execute the zero adjustment using the zero adjustment board
provided with the connected instrument.
Insert the tip pins of the probes into through holes having the
same width as the terminals of the measuring objectand carry
out the zero adjustment with the connected instrument.
Through hole Scale
Measuring object
Zero
adjustment
board
L2003
Pin Type
Probe
Select through holes having the
same number scale for both
HIGH and LOW terminal.
IMPORTANT
• Adjust the return cable between the terminals so that it
does not sag.
• If there is no through hole with the same width as the
terminals of the measuring objectperform zero adjustment
with the most nearest width scale of the through hole as
the terminals.
• Perform zero adjustment by facing the probe’s (both red
and black) SENSE side inwards.
Replacing the Used Tip Pin
Replace the used tip pin with a new one when the tip pin
breaks or is worn out. To purchase Model 9772-90 Tip Pin
for replacementcontact your authorized Hioki distributor or
reseller.
<Tools to be prepared> Model 9772-90 Tip Pinpliersetc.
1
Turn off the power to the connected instrument and
remove the return cables of the probes.
2
Grip the tip pin to be replaced with pliers and pull it out.
3
Insert a new tip pin in the socketand securely
push to the back by pressing with a hard board so
that the tip pin does not fly out.
Socket
Tip pin
4
Measure a known measuring object to check if the
measured value is correct.
Four-terminal Pair Method
L2003 uses the four-terminal pair method as the measurement
method.
Four-terminal pair method
In the four-terminal pair methodthe current flows backward (current
returns) with the same magnitude as the measuring current in the
shields of the SOURCE cablesand then cancels the magnetic field
of the measuring current. This method suppresses the induced
electromotive force induced at the SENSE terminalsand detects the
voltage actually generated in the object being measured.
Four-terminal pair method when using the optional probe
When the L2003 is usedthe four-terminal pair method is structured
as described below. Vicinity of the measurement object will not
entirely be four-terminal pairand will be affected by an inductive
magnetic field. The shape of the return cable should not be changed
and kept away from metals when use. (When there are metals
inductive magnetic field occurs due to eddy currentand leads to
measurement errors.)
SOURCE-H
SOURCE-L
SENSE-L
SENSE-H
V
A
Shield Tip end: SOURCE-H
Tip end: SENSE-H
Return cable
Tip end:
SENSE-L
Tip end:
SOURCE-L
Measuring
object
• Keep the loop area between the return cable and measurement
object as small as possible.
• Keep the loop shape and wiring position always the same.
• Keep away from metals.
When the measurement value fluctuates
Keep the return cable in shapeor the affects of the magnetic field
changes and the measurement value may flactuate.
Twist the return cable to keep a fixed shape.
(The return cable is indicated black in the illustration.)
Connecting the return cable
IMPORTANT
• Connect the return cable before measurement.
• Adjust the fixed position of the probe so that the return
cable between the probes does not sag.
1
Securely insert the plug of the return cable up to
the back of the probe (both red and black).
2
Arrange the probes so that the distance between
the pin tips of the probe is the same as that
between the terminals of the actual measuring
objectwith the SENSE sides of the probes (both
red and black) facing inwards.
Measuring object
SENSE
3
Adjust the position of the probes such that the
return cable between the probes does not sagand
fix by pushing the return cable into the grooves of
the probes.
Measuring
object Groove
When viewed from the front
When viewed from the rear
(The return cable is indicated black in the illustration above.)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Use a cable of suitable length for the distance between the
terminals of the measuring object from the three types of
accessory cables.
Return cable length Distance between the terminals
of the measuring object
400 mm 100 mm or less
550 mm
(mounted before shipment) 100 mm to 250 mm
800 mm 250 mm to 500 mm
Fixing Using the Mounting
Holes
The probe can be fixed using the mounting holes. (Please
prepare your own jig or equipment to fix.)
Hole diameter: 3.2 mmuse a M3 screw.
Jig or
equipment
to fix
M3 screw
12 mm
12 mm + thickness of jig or equipment
2.5 mm
9.15 mm
1.8 mm
First contact
pressure:
Approx. 60 g
Full stroke: 3 mm±0.2 mm
(condensing pressure
approx. 230 g)
Recommended
stroke:
2 mm
(condensing pressure
approx. 170 g)
Zero Adjustment
Execute the zero adjustment using the zero adjustment board
provided with the connected instrument.
Insert the tip pins of the probes into through holes having the
same width as the terminals of the measuring objectand carry
out the zero adjustment with the connected instrument.
Through hole Scale
Measuring object
Zero
adjustment
board
L2003
Pin Type
Probe
Select through holes having the
same number scale for both
HIGH and LOW terminal.
IMPORTANT
• Adjust the return cable between the terminals so that it
does not sag.
• If there is no through hole with the same width as the
terminals of the measuring objectperform zero adjustment
with the most nearest width scale of the through hole as
the terminals.
• Perform zero adjustment by facing the probe’s (both red
and black) SENSE side inwards.
Replacing the Used Tip Pin
Replace the used tip pin with a new one when the tip pin
breaks or is worn out. To purchase Model 9772-90 Tip Pin
for replacementcontact your authorized Hioki distributor or
reseller.
<Tools to be prepared> Model 9772-90 Tip Pinpliersetc.
1
Turn off the power to the connected instrument and
remove the return cables of the probes.
2
Grip the tip pin to be replaced with pliers and pull it out.
3
Insert a new tip pin in the socketand securely
push to the back by pressing with a hard board so
that the tip pin does not fly out.
Socket
Tip pin
4
Measure a known measuring object to check if the
measured value is correct.
Four-terminal Pair Method
L2003 uses the four-terminal pair method as the measurement
method.
Four-terminal pair method
In the four-terminal pair methodthe current flows backward (current
returns) with the same magnitude as the measuring current in the
shields of the SOURCE cablesand then cancels the magnetic field
of the measuring current. This method suppresses the induced
electromotive force induced at the SENSE terminalsand detects the
voltage actually generated in the object being measured.
Four-terminal pair method when using the optional probe
When the L2003 is usedthe four-terminal pair method is structured
as described below. Vicinity of the measurement object will not
entirely be four-terminal pairand will be affected by an inductive
magnetic field. The shape of the return cable should not be changed
and kept away from metals when use. (When there are metals
inductive magnetic field occurs due to eddy currentand leads to
measurement errors.)
SOURCE-H
SOURCE-L
SENSE-L
SENSE-H
V
A
Shield Tip end: SOURCE-H
Tip end: SENSE-H
Return cable
Tip end:
SENSE-L
Tip end:
SOURCE-L
Measuring
object
• Keep the loop area between the return cable and measurement
object as small as possible.
• Keep the loop shape and wiring position always the same.
• Keep away from metals.
When the measurement value fluctuates
Keep the return cable in shapeor the affects of the magnetic field
changes and the measurement value may flactuate.
Twist the return cable to keep a fixed shape.
(The return cable is indicated black in the illustration.)
Connecting the return cable
IMPORTANT
• Connect the return cable before measurement.
• Adjust the fixed position of the probe so that the return
cable between the probes does not sag.
1
Securely insert the plug of the return cable up to
the back of the probe (both red and black).
2
Arrange the probes so that the distance between
the pin tips of the probe is the same as that
between the terminals of the actual measuring
objectwith the SENSE sides of the probes (both
red and black) facing inwards.
Measuring object
SENSE
3
Adjust the position of the probes such that the
return cable between the probes does not sagand
fix by pushing the return cable into the grooves of
the probes.
Measuring
object Groove
When viewed from the front
When viewed from the rear
(The return cable is indicated black in the illustration above.)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Use a cable of suitable length for the distance between the
terminals of the measuring object from the three types of
accessory cables.
Return cable length Distance between the terminals
of the measuring object
400 mm 100 mm or less
550 mm
(mounted before shipment) 100 mm to 250 mm
800 mm 250 mm to 500 mm
Fixing Using the Mounting
Holes
The probe can be fixed using the mounting holes. (Please
prepare your own jig or equipment to fix.)
Hole diameter: 3.2 mmuse a M3 screw.
Jig or
equipment
to fix
M3 screw
12 mm
12 mm + thickness of jig or equipment
2.5 mm
9.15 mm
1.8 mm
First contact
pressure:
Approx. 60 g
Full stroke: 3 mm±0.2 mm
(condensing pressure
approx. 230 g)
Recommended
stroke:
2 mm
(condensing pressure
approx. 170 g)
Zero Adjustment
Execute the zero adjustment using the zero adjustment board
provided with the connected instrument.
Insert the tip pins of the probes into through holes having the
same width as the terminals of the measuring objectand carry
out the zero adjustment with the connected instrument.
Through hole Scale
Measuring object
Zero
adjustment
board
L2003
Pin Type
Probe
Select through holes having the
same number scale for both
HIGH and LOW terminal.
IMPORTANT
• Adjust the return cable between the terminals so that it
does not sag.
• If there is no through hole with the same width as the
terminals of the measuring objectperform zero adjustment
with the most nearest width scale of the through hole as
the terminals.
• Perform zero adjustment by facing the probe’s (both red
and black) SENSE side inwards.
Replacing the Used Tip Pin
Replace the used tip pin with a new one when the tip pin
breaks or is worn out. To purchase Model 9772-90 Tip Pin
for replacementcontact your authorized Hioki distributor or
reseller.
<Tools to be prepared> Model 9772-90 Tip Pinpliersetc.
1
Turn off the power to the connected instrument and
remove the return cables of the probes.
2
Grip the tip pin to be replaced with pliers and pull it out.
3
Insert a new tip pin in the socketand securely
push to the back by pressing with a hard board so
that the tip pin does not fly out.
Socket
Tip pin
4
Measure a known measuring object to check if the
measured value is correct.
Four-terminal Pair Method
L2003 uses the four-terminal pair method as the measurement
method.
Four-terminal pair method
In the four-terminal pair methodthe current flows backward (current
returns) with the same magnitude as the measuring current in the
shields of the SOURCE cablesand then cancels the magnetic field
of the measuring current. This method suppresses the induced
electromotive force induced at the SENSE terminalsand detects the
voltage actually generated in the object being measured.
Four-terminal pair method when using the optional probe
When the L2003 is usedthe four-terminal pair method is structured
as described below. Vicinity of the measurement object will not
entirely be four-terminal pairand will be affected by an inductive
magnetic field. The shape of the return cable should not be changed
and kept away from metals when use. (When there are metals
inductive magnetic field occurs due to eddy currentand leads to
measurement errors.)
SOURCE-H
SOURCE-L
SENSE-L
SENSE-H
V
A
Shield Tip end: SOURCE-H
Tip end: SENSE-H
Return cable
Tip end:
SENSE-L
Tip end:
SOURCE-L
Measuring
object
• Keep the loop area between the return cable and measurement
object as small as possible.
• Keep the loop shape and wiring position always the same.
• Keep away from metals.
When the measurement value fluctuates
Keep the return cable in shapeor the affects of the magnetic field
changes and the measurement value may flactuate.
Twist the return cable to keep a fixed shape.
(The return cable is indicated black in the illustration.)
Connecting the return cable
IMPORTANT
• Connect the return cable before measurement.
• Adjust the fixed position of the probe so that the return
cable between the probes does not sag.
1
Securely insert the plug of the return cable up to
the back of the probe (both red and black).
2
Arrange the probes so that the distance between
the pin tips of the probe is the same as that
between the terminals of the actual measuring
objectwith the SENSE sides of the probes (both
red and black) facing inwards.
Measuring object
SENSE
3
Adjust the position of the probes such that the
return cable between the probes does not sagand
fix by pushing the return cable into the grooves of
the probes.
Measuring
object Groove
When viewed from the front
When viewed from the rear
(The return cable is indicated black in the illustration above.)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Use a cable of suitable length for the distance between the
terminals of the measuring object from the three types of
accessory cables.
Return cable length Distance between the terminals
of the measuring object
400 mm 100 mm or less
550 mm
(mounted before shipment) 100 mm to 250 mm
800 mm 250 mm to 500 mm
Fixing Using the Mounting
Holes
The probe can be fixed using the mounting holes. (Please
prepare your own jig or equipment to fix.)
Hole diameter: 3.2 mmuse a M3 screw.
Jig or
equipment
to fix
M3 screw
12 mm
12 mm + thickness of jig or equipment
2.5 mm
9.15 mm
1.8 mm
First contact
pressure:
Approx. 60 g
Full stroke: 3 mm±0.2 mm
(condensing pressure
approx. 230 g)
Recommended
stroke:
2 mm
(condensing pressure
approx. 170 g)
Zero Adjustment
Execute the zero adjustment using the zero adjustment board
provided with the connected instrument.
Insert the tip pins of the probes into through holes having the
same width as the terminals of the measuring objectand carry
out the zero adjustment with the connected instrument.
Through hole Scale
Measuring object
Zero
adjustment
board
L2003
Pin Type
Probe
Select through holes having the
same number scale for both
HIGH and LOW terminal.
IMPORTANT
• Adjust the return cable between the terminals so that it
does not sag.
• If there is no through hole with the same width as the
terminals of the measuring objectperform zero adjustment
with the most nearest width scale of the through hole as
the terminals.
• Perform zero adjustment by facing the probe’s (both red
and black) SENSE side inwards.
Replacing the Used Tip Pin
Replace the used tip pin with a new one when the tip pin
breaks or is worn out. To purchase Model 9772-90 Tip Pin
for replacementcontact your authorized Hioki distributor or
reseller.
<Tools to be prepared> Model 9772-90 Tip Pinpliersetc.
1
Turn off the power to the connected instrument and
remove the return cables of the probes.
2
Grip the tip pin to be replaced with pliers and pull it out.
3
Insert a new tip pin in the socketand securely
push to the back by pressing with a hard board so
that the tip pin does not fly out.
Socket
Tip pin
4
Measure a known measuring object to check if the
measured value is correct.
Four-terminal Pair Method
L2003 uses the four-terminal pair method as the measurement
method.
Four-terminal pair method
In the four-terminal pair methodthe current flows backward (current
returns) with the same magnitude as the measuring current in the
shields of the SOURCE cablesand then cancels the magnetic field
of the measuring current. This method suppresses the induced
electromotive force induced at the SENSE terminalsand detects the
voltage actually generated in the object being measured.
Four-terminal pair method when using the optional probe
When the L2003 is usedthe four-terminal pair method is structured
as described below. Vicinity of the measurement object will not
entirely be four-terminal pairand will be affected by an inductive
magnetic field. The shape of the return cable should not be changed
and kept away from metals when use. (When there are metals
inductive magnetic field occurs due to eddy currentand leads to
measurement errors.)
SOURCE-H
SOURCE-L
SENSE-L
SENSE-H
V
A
Shield Tip end: SOURCE-H
Tip end: SENSE-H
Return cable
Tip end:
SENSE-L
Tip end:
SOURCE-L
Measuring
object
• Keep the loop area between the return cable and measurement
object as small as possible.
• Keep the loop shape and wiring position always the same.
• Keep away from metals.
When the measurement value fluctuates
Keep the return cable in shapeor the affects of the magnetic field
changes and the measurement value may flactuate.
Twist the return cable to keep a fixed shape.
(The return cable is indicated black in the illustration.)
Connecting the return cable
IMPORTANT
• Connect the return cable before measurement.
• Adjust the fixed position of the probe so that the return
cable between the probes does not sag.
1
Securely insert the plug of the return cable up to
the back of the probe (both red and black).
2
Arrange the probes so that the distance between
the pin tips of the probe is the same as that
between the terminals of the actual measuring
objectwith the SENSE sides of the probes (both
red and black) facing inwards.
Measuring object
SENSE
3
Adjust the position of the probes such that the
return cable between the probes does not sagand
fix by pushing the return cable into the grooves of
the probes.
Measuring
object Groove
When viewed from the front
When viewed from the rear
(The return cable is indicated black in the illustration above.)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

PIN TYPE PROBE
Instruction Manual
L2003
EN
November 2014 Edition 1
Printed in Japan
L2003A981-00 14-11H
Inspection/Maintenance
When you receive the productinspect it carefully to ensure
that no damage occurred during shipping. If damage is
evidentor if it fails to operate according to the specifications
contact your authorized Hioki distributor or reseller.
Preliminary checks
WARNING
Before using the productensure that the insulation on
the probes are not damaged and conductors are not
exposed. Using the product in such conditions could
cause an electric shockso contact your authorized
Hioki distributor or reseller for repair.
Maintenance and service
IMPORTANT
Never use cleaning agents containing benzenealcohol
acetoneetherketonesthinners or gasoline. These can
deform and discolor the case.
•To clean the productwipe it gently with a soft cloth moistened with
water or mild detergent.
•If the product seems to be malfunctioningcontact your authorized
Hioki distributor or reseller. Pack the product so that it will not get
damaged during shippingand include a description of existing
damages. Hioki cannot be responsible for damage that occurs
during shipment.
Disposal
Handle and dispose of the product in accordance with local
regulations.
Operating Precautions
Installing the product in inappropriate locations may cause
the probe to malfunction or lead to accidents. Please do not
install in the following locations. For details on the operating
temperature and humiditysee the specifications.
WARNING
• Exposed to direct sunlight or high temperatures
• Exposed to corrosive or combustible gases
• Exposed to strong electromagnetic fields or
electrostatic charges
• Near induction heating systems (such as high-
frequency induction heating systems and IH cooking
equipment)
• Susceptible to vibration
• Exposed to wateroilchemicalsor solvents
• Exposed to high humidity or condensation
• Exposed to high quantities of dust particles
CAUTION
•The ends of the probes are sharp. Be careful to avoid
injury.
•To prevent probe damagedo not step on probes or
pinch them between other objects. Do not bend or pull
on probes at their base.
Safety Information
Before using the productbe certain to carefully read the
following safety notes.
WARNING
•Customers are not allowed to modifydisassemble
or repair the instrument. Failure to observe these
precautions may result in fireelectric shockor injury.
•Ensure that the input does not exceed the maximum
input voltage or current to avoid product damage
resulting from heat building. Excessive voltage and
current can cause the product to malfunction.
CAUTION
• Mishandling could result in malfunction of the probe.
Be certain that you understand the instructions and
precautions in the manual before use.
•If persons unfamiliar with electricity measuring product
are to use the productanother person familiar with
such instruments must supervise operations.
Safety symbols
Indicates cautions and hazards. When the symbol is
printed on the productrefer to a corresponding topic
in the Instruction Manual.
Indicates DC (Direct Current).
Notation
In this manualthe risk seriousness and the hazard levels are
classified as follows.
WARNING
Indicates a potentially hazardous situation that may
result in death or serious injury to the operator.
CAUTION
Indicates a potentially hazardous situation that
may result in minor or moderate injury to the
operator or damage to the product or malfunction.
IMPORTANT
Indicates information related to the operation of
the product or maintenance tasks with which the
operators must be fully familiar.
Indicates a prohibited action.
Indicates actions which must be performed.
Overview
L2003 Pin Type Probe is a four-terminal probe. The probe can be
used to contact various types of batteries since the measuring unit is
pin shaped.
Specifications
Measurable frequency
range
DC to 1050 Hz
Maximum rated voltage 30 V peak
Maximum rated current 2.5 A peak
Structure Four-terminal pair connection structure
Length Approx. 1500 mm (59.06″)
Mass Approx. 200 g (7.1 oz.)
Cable used 50 ΩCoaxial cable
Terminal structure Gold plating process
Tip shape: Needle
Used tip pin can be replaced.
Distance between pins 2.5 mm (0.10″)
Operating temperature
and humidity
Temperature; 0°C to 40°C (32°F to 104°F)
Humidity; 80% RH or less (no condensation)
Storage temperature
and humidity
Temperature; -10°C to 50°C (14°F to 122°F)
Humidity; 80% RH or less (no condensation)
Operating environment IndoorsPollution degree 2altitude up to
2000 m (6562 ft.)
Product warranty
period
1 year
Accessories Return Cable (400 mm)
Return Cable (550 mm)
Return Cable (800 mm)
Instruction Manual
Option 9772-90 Tip Pin
Part Names
Tip pin
Red (HIGH)
Black (LOW)
Return
Cable
(550 mm)
Mounting holes
d (HIGH)
Probe’s maximum rating is 30 Vbut be careful to avoid
exceeding the maximum rating of the measuring instrument.
IMPORTANT
Return Cable (550 mm) is shipped in a state connected to
the instrument.
Accessories
Plug
400 mm
800 mm
Connection Method
IMPORTANT
Clean the tip of the tip pin and the surface of the measuring
object before zero adjustment and measurement.
Connecting to the instrument
1
Adjust
23
Turn to the rightInsert
13-09
Introduction
Thank you for purchasing the HIOKI L2003 Pin Type Probe.
To obtain maximum performance from the product please
read this manual firstand keep it handy for future reference.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Table of contents
Other Hioki Measuring Instrument manuals

Hioki
Hioki MEMORY HiCORDER MR8870-20 User manual

Hioki
Hioki 3441 User manual

Hioki
Hioki CM4141-90 User manual

Hioki
Hioki CHT3548 User manual

Hioki
Hioki MR8880 User manual

Hioki
Hioki BT4560 User manual

Hioki
Hioki FT3432-20 User manual

Hioki
Hioki PW3390 User manual

Hioki
Hioki CM4003 User manual

Hioki
Hioki 9465-10 User manual
Popular Measuring Instrument manuals by other brands

Extech Instruments
Extech Instruments MA410T user guide

AST
AST P-GM+ user manual

Task Force Tips
Task Force Tips BLITZFIRE Series Instruction for operation and maintenance

Fluke Calibration
Fluke Calibration 9190A Getting started

Itm
Itm SCALE MASTER CLASSIC 6020 user guide

Trotec
Trotec BA30WP operating manual