Insufficient Unit Weight Indication () :
Cursor appears on the bottom edge of UNIT
WEIGHT LCD (above of S.W printed on the
display-overlay), and indicates the averaging unit
weight is too light.
The internal resolution of this scale is 60000. If the
averaging weight per unit is less than “max capacity
/ 60000”, cursor will flash on UNIT WEIGHT LCD.
For example:
With 30kg max capacity and an internal resolution of
60000, 30kg / 60000 = 0.5g. If the average unit
weight of measured subjects is lighter than 0.5g,
cursor will flash on UNIT WEIGHT LCD (above of S.W
printed on the display-overlay).
Indication of Weighing Range () :
Cursor appears on the bottom edge of UNIT
PRICE LCD (above of W1 or W2 printed on the
display-overlay), and indicates the present weighing
range to be Weighing Range 1 (W1) or Weighing
Range 2 (W2).
!!
Notice:The indication of weighing range is
only available for scales with multiple range (2
ranges).