3 1/2 DIGITAL DUAL DISPLAY
AC CLAMPMULTIMETER
OPERATION MANUAL
1. SAFETY INFORMATION
The following safety information must be observed to insure
maximum personal safety during the operation at this meter.
1.1 Do not operate the meter if the body of meter or the test lead
look broken.
1.2 Check the main function dial and make sure it is at the correct
position before each measurement.
1.3 Do not perform resistance, temperature, diode and continuity
test on a live power system.
1.4 Do not apply voltage between the test terminals and test
terminal to ground that exceed the maximum limit record in
this manual.
1.5 Exercise extreme caution when measuring live system with
voltage greater than 60V DC or 30VAC.
1.6 Keep the fingers after the protection ring when measuring
through the test lead.
1.7 Change the battery when the symbol appears to avoid incorrect
data.
2. Panel Layout
1) Rotary Switch: use this switch to select functions and
ranges on the display.
2) Back light: Push the LIGHT button to light the backlight then it
will auto light off after approx. 5 seconds.
3)Data hold: Push the HOLD button and “DH” sign will appear
4) Phase lamp
5) lack Phase LED light
6) LCD Display
7) a b c Input Jack
3. SPECIFICATIONS
3.1 GENERAL SPECIFICATIONS
Display: 3 1/2 digit LCD with a max. reading of 1999.
Polarity: Automatic negative polarity indication.
Zero adjustment: Automatic.
Over range indication: Only the "1" or "-1" display.
Sample rate : 2 times/sec
Low battery indication: “ ”
Safety Standards: The meter is up to the standards of IEC1010
Double Insulation, Pollution Degree 2, Over voltage Category
Ⅲ.
Clamp opening size: 45mm.
Operating Environment: Temperature 32~104°F(0~40℃),
humidity<75%RH.
Storage Environment: Temperature -4~140°F(-20~60℃),
humidity<90%RH.
Power supply: 9V Zinc-carbon battery.
Dimension: 225(H)×77(W)×45(D)mm.
Weight: Approx. 330g (including battery).
3.2 ELECTRICAL SPECIFICATIONS
Accuracies are ±(% of reading + number in last digit) at 23±5℃,
≤70%RH.
3.2.1 DC Voltage
Range
Accuracy
Resolution
1000V
±(1.5% of rdg+5digits)
1V
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ
3.2.2 AC Voltage
Range
Accuracy
Resolution
Frequency
750V
±(1.8% of rdg+10digits)
1V
50~100Hz
Average sensing, calibrated to rms of sine wave
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ
3.2.3 AC Current
Range
Accuracy
Resolution
Frequency
20A
±(3.0% of rdg+10digits)
10mA
50~60Hz
1000A
0~800
±(3.5% of rdg+10digits)
1A
800~1000
±(6.5% of rdg+10digits)
Average sensing, calibrated to rms of sine wave
Overload protection: 1000A rms within 60 seconds
3.2.4 Resistance
Range
Accuracy
Resolution
200Ω
±(1.0% of rdg+15digits)
0.1 Ω
2000Ω
±(1.5% of rdg+15digits)
1 Ω
Overload protection: 250V DC/250Vrms AC
3.2.5 Diode test
Range
Description
Test condition
Display read
approximately forward
Voltage of diode.
Forward DC current approx
1.5mA
.Reversed DC voltage
approximately 3V
Overload protection: 250V DC/250Vrms AC
3.2.6 Audible continuity Test
Range
Description
Test condition
Built-in buzzer sounds if
resistance is less than 30Ω.
Open circuit voltage
approx 3V
Overload protection: 250V DC/250Vrms AC.
4. OPERATION
4.1 DC Voltage Measurement
1) Connect the black test lead to "COM" socket and red test
lead to the "V" socket.
2) Set the selector switch to "1000V " position.
3) Measure the voltage by touch the test lead tips to the test
circuit where the value of voltage is needed.
4) Read the result from the LCD panel.
5) Push the HOLD button to lock display value, push it again
to exit.
6) Push the LIGHT button to light the back light.
4.2 AC Voltage Measurement
1) Connect the black test lead to "COM" socket and red test
lead to the "V" socket.
2) Set the selector switch to "750V~" position.
3) Measure the voltage by touch the test lead tip to the test
circuit where the value of voltage is needed.
4) Read the result from the LCD panel.
5) Push the HOLD button to lock display value, push it again
to exit.
6) Push the LIGHT button to light the back light.
4.3 AC Current Measurement
1) Set the selector switch to desired "20A~" or "1000A~"
position.
2) Open the clamp by pressing the jaw-opening handle and
insert the cable to be measured into the jaw.
3) Close the clamp and get the reading from the LCD panel.
4) Push the HOLD button to lock display value, push it again
to exit.
5) Push the LIGHT button to light the back light.
Note:
Before this measurement, disconnect the test lead with the meter
for safety.
4.4 Resistance Measurement
1) Connect the black test lead to "COM" socket and red test
lead to the "Ω" socket.
2) Set the selector switch to "200Ω" or “2000Ω” position.
3) Connect tip of the test leads to the points where the value of
the resistance is needed.
4) Read the result from the LCD panel.
5) Push the HOLD button to lock display value, push it again
to exit.
6) Push the LIGHT button to light the back light.
Note:
When take resistance value from a circuit system, make sure
the power is cut off and all capacitors need to be discharged.
4.5 Diode Measurement
1) Connect the black test lead to "COM" socket and red test
lead to the "Ω" socket.
2) Set the selector switch to " " position.
3) Connect the test leads across the diode under measurement,
display shows the approx. forward voltage of this diode.
4) Push the HOLD button to lock display value, push it again
to exit.
5) Push the LIGHT button to light the back light.
Note:
Make sure the power is cut off and all capacitors need to be
discharged under this measurement.
4.6 Audible Continuity Test
1) Connect the black test lead to "COM" socket and red
test lead to the "Ω" socket.
2) Set the selector switch to " " position.
3) Connect the test leads to two point of circuit, if the
resistance is lower than approx. 30Ω,the buzzer sounds .
4) Push the HOLD button to lock display value, push it again
to exit.
5) Push the LIGHT button to light the back light.
Note:
Make sure the power is cut off and all capacitors need to be
discharged under this measurement.
4.7 Three Phase Live Wire Identification
Set the rotary switch to “
”,And connect the yellow
measuring cable to the “VΩa”jack ,connect the black measuring
cable to the “COM b”jack ,connect red the measuring cable to
the. “c”jack .Afterward connect three measuring cables to
Three Phase Live Wire By testing .
a. Phase lamp light and three lack Phase LED light ,show
phase order of three phase live wire is :a , b , c
b. Phase lamp not light and three lack phase LED light ,
show phase order of three phase live wire is : c, b , a
c. Phase lamp not light and one of three lack phase LED
not light, show lack phase.
4.8 Live Wire Differentiate
Set the rotary switch to “
”,And connect the red measuring
cable the “c”jack ,connect the black measuring cable the
“COM b ”jack .Then you hold to the black shank by hand
(Note : not brush up against test metal head of the black shank ),
connect the red test head to the wire by test , if “ ” sign wink ,
show the wire is live wire . Otherwise show the wire is not live
wire .Here LCD display is faradism.
5. Battery replacement
1) When the battery voltage drop below proper operation range
the “ ”symbol will appear on the LCD display and the
battery need to changed.
2) Before changing the battery, set the selector switch to "OFF".
open the cover of the battery cabinet by a screwdriver.
3) Replace the old battery with the same type battery.
4) Close the battery cabinet cover and fasten the screw.
6. MAINTENANCE
1) Before open the battery door, disconnect both test lead and
never uses the meter before the battery door is closed.
2) To avoid contamination or static damage, do not touch the
circuit board without proper static protection.
3) If the meter is not going to be used for a long time, take out
the battery and do not store the meter in high temperature or
high humidity environment.
4) When take current measurement, keep the cable at the center
of the clamp will get more accurate test result.
5) Repairs or servicing not covered in this manual should
only by qualified personal.-
6) Periodically wipe the case with a dry cloth and detergent.
Do not use abrasives or solvents on this instruments.
7) Please take out the battery when not using for a long time.
Above picture and content just for your reference.
Please be subject to the actual products if anything
different or updated. Please pardon for not
informing in advance.

This manual suits for next models

1

Other HoldPeak Multimeter manuals

HoldPeak 870CR User manual

HoldPeak

HoldPeak 870CR User manual

HoldPeak HP-90E User manual

HoldPeak

HoldPeak HP-90E User manual

HoldPeak HP-760E User manual

HoldPeak

HoldPeak HP-760E User manual

HoldPeak 770HD User manual

HoldPeak

HoldPeak 770HD User manual

HoldPeak 90B User manual

HoldPeak

HoldPeak 90B User manual

HoldPeak 90F User manual

HoldPeak

HoldPeak 90F User manual

HoldPeak 770A User manual

HoldPeak

HoldPeak 770A User manual

HoldPeak 39K User manual

HoldPeak

HoldPeak 39K User manual

HoldPeak 870N User manual

HoldPeak

HoldPeak 870N User manual

HoldPeak 860N User manual

HoldPeak

HoldPeak 860N User manual

HoldPeak 770A User manual

HoldPeak

HoldPeak 770A User manual

HoldPeak 760A Installation manual

HoldPeak

HoldPeak 760A Installation manual

HoldPeak HP-90F User manual

HoldPeak

HoldPeak HP-90F User manual

HoldPeak 39B User manual

HoldPeak

HoldPeak 39B User manual

HoldPeak 770C User manual

HoldPeak

HoldPeak 770C User manual

HoldPeak 760J Installation manual

HoldPeak

HoldPeak 760J Installation manual

HoldPeak 39B User manual

HoldPeak

HoldPeak 39B User manual

HoldPeak 770G User manual

HoldPeak

HoldPeak 770G User manual

HoldPeak HP-770H User manual

HoldPeak

HoldPeak HP-770H User manual

HoldPeak 850E User manual

HoldPeak

HoldPeak 850E User manual

HoldPeak HP-37B User manual

HoldPeak

HoldPeak HP-37B User manual

HoldPeak 6208 User manual

HoldPeak

HoldPeak 6208 User manual

HoldPeak HP-36C User manual

HoldPeak

HoldPeak HP-36C User manual

HoldPeak 6688D User manual

HoldPeak

HoldPeak 6688D User manual

Popular Multimeter manuals by other brands

Keysight U3606A Programmer's reference

Keysight

Keysight U3606A Programmer's reference

CARPOINT MAS838 instruction manual

CARPOINT

CARPOINT MAS838 instruction manual

Gossen MetraWatt METRA MAX 12 operating instructions

Gossen MetraWatt

Gossen MetraWatt METRA MAX 12 operating instructions

C-LOGIC 3200 instruction manual

C-LOGIC

C-LOGIC 3200 instruction manual

Vorel 81801 Original instructions

Vorel

Vorel 81801 Original instructions

Chauvin Arnoux F407 Start guide

Chauvin Arnoux

Chauvin Arnoux F407 Start guide

Ziegler MULTI 12S user manual

Ziegler

Ziegler MULTI 12S user manual

Actron SunPro CP7678 Operating instrucktions

Actron

Actron SunPro CP7678 Operating instrucktions

Sealey TA330 manual

Sealey

Sealey TA330 manual

Topex 94W102 manual

Topex

Topex 94W102 manual

Perel EVM890N user manual

Perel

Perel EVM890N user manual

Amprobe AD105A user manual

Amprobe

Amprobe AD105A user manual

Steinberg SBS-DMB-1000 user manual

Steinberg

Steinberg SBS-DMB-1000 user manual

Extech Instruments SDL350 user manual

Extech Instruments

Extech Instruments SDL350 user manual

Maxwell Digital Multimeters 25302 user manual

Maxwell Digital Multimeters

Maxwell Digital Multimeters 25302 user manual

TTI 1604 instruction manual

TTI

TTI 1604 instruction manual

Elenco Electronics M-1008K Assembly and instruction manual

Elenco Electronics

Elenco Electronics M-1008K Assembly and instruction manual

Agilent Technologies U3402A User's and service guide

Agilent Technologies

Agilent Technologies U3402A User's and service guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.