DIGITAL MULTIMETER
OPERATOR’S MANUAL
1. Overview
The multimeter is characterized at slim size, portable, stable performance and
anti-dropping capacity. Using 4000 counts digit LCD monitor with character
25mm high, they offer clear readings. With overall circuitry design centering on
large-scale IC A/D converters in conjunction and over-load protection circuit, the
meters give excellent performance and exquisite making as a handy utility
instrument.
The meters can be used to measure DC & AC voltage, DC & AC current,
resistance, temperature, battery test, positive diode voltage fall and audible
continuity.
2. Panel Layout
1. Test lead fixture: Fix the test lead.
2. Battery test green light: the green light show the battery being tested is good,
it can continue to be used.
3. Battery test yellow light: the yellow light show the battery being tested is close
to bad, it can be used but need to changed as quickly as possible.
4. Battery test red light: the red light show the battery being tested is bad, it can
not continue to be used.
5. LCD display: 3 1/2 digits LCD display, full function symbol display.
6. SELECT key: This key work on the " " range, press the key to choose
resistance, diode or continuity test, on the voltage or current range, change
to DC or AC, on the range, change to °C or °F test.
7. RANGE Key: Press the “RANGE” key, the meter enters manual range mode,
press it more than 2 seconds again, return to auto mode.
8. Key: Press the “ ” key over 2 seconds to light the back light, press it
again for more than 2 seconds to lights off it, it can lights off automatically after
15 seconds too.
9. M.HOLD key: Press the “M.HOLD” key to lock MAX value, and the “M.HOLD”
sign will appear on the display, press it again to exit. If press and hold the
“M.HOLD”key to power on, “Auto Power Off ” function will be disabled.
10.D.HOLD key: Press the “D.HOLD” key to lock display value, and the
“D.HOLD” sign will appear on the display, press it again to exit.
11. Rotary Switch: Use this switch to select functions and ranges.
12. COM: COM and Temperature “-” Input Jack
13. 10A: 10A Input Jack
14.VΩmA: V/μAmA/ BATT/ /Input Jack
15. Crust of meter
16. Protective casing
3. Safety Information
3-1 The meter is designed according to IEC-1010 concerning electronic
measuring instruments with an over-voltage category 600V (CAT Ⅲ) and
pollution 2.
3-2 Follow all safety and operating instructions to ensure that the meter is used
safely and is kept in good operating condition.
3-3 safety symbols:
Important safety information, refer to the operating manual.
Dangerous voltage may be presence.
Double insulation (protection Class II)
4. Special Cautions for Operation
4-1 The meters can be safe only according to standard procedures when used in
conjunctions with the supplied test leads. To replace damaged test leads with
only the same model or same electric specifications.
4-2 To avoid risk of electric shock, do not use the meters before the cover is
in place.
4-3 The range switch should be right position for the testing.
4-4 To avoid electric shock and damaging the instruments, the input signals are
forbidden to exceed the specified limits.
4-5 When measuring TV set or switched power, attention should be paid to the
possible pulses that may bring destruction to the circuit.
4-6 Range switch position is forbidden to be changed at random during
measurement.
4-7 Take caution against shock in the course of measuring voltage higher than
DC 60V & AC 30V.
4-8 Protection fuse should be replaced only with same type and same
specification.
4-9 After operation is finished, set function switch at OFF to save battery power.
4-10 If the meter is without usage for long time, take out battery to avoid damage
by battery leakage.
5. GENERAL SPECIFICATIONS
5-1 Max Voltage between input terminal and Earth Ground: CAT Ⅲ 600V
5-2 Over-range Indication: display “OL” for the significant digit.
5-3 Automatic display of negative polarity “-” .
5-4 Low Battery Indication: “ ” displayed.
5-5 Display: 3 1/2 digit LCD with a max. reading of 1999.
5-6 Auto range control
5-7 Auto Power Off: When measurement exceeds 15 minutes without switching
mode and pressing key, the meter will switch to standby mode. Press any
key to exit standby mode. When restart the system, press and hold M.HOLD
key to disable auto power off.
5-8 Fuse protection: 500mA/500V Fast Fuse, 10A/500V Fast Fuse
5-9 Power supply: 9V 6F22 battery
5-10 Operating Temp.: 0℃ to 40℃ (relative humidity <85%)
5-11 Storage Temp.: -10℃ to 50℃ (relative humidity <85%)
5-12 Guaranteed precision Temp.: 23±5℃ (relative humidity <70%)
5-13 Dimension: 150x100x36mm
5-14 Weight: approx. 250g (including battery)
6. Testing Specifications
Accuracy is specified for a period of year after calibration and at 18 ℃ to
28 ℃ (64℉ to 82℉) with relative humidity to 70%.
6-1 DC Voltage
Range
Resolution
Accuracy
200mV
0.1mV
±(0.5% of rdg + 2 digits)
2V
1mV
20V
10mV
200V
100mV
600V
1V
±(0.8% of rdg + 2 digits)
-- Impedance: 10MΩ, More than 100MΩ on 200mV range
-- Overload protection: 600V DC or AC rms
6-2 AC Voltage
Range
Resolution
Accuracy
2V
1mV
±(0.8% of rdg + 3 digits)
20V
10mV
200V
100mV
600V
1V
±(1.2% of rdg + 3 digits)
-- Impedance: 10MΩ
-- Overload protection: 600V DC or AC rms
-- Frequency Range: 40 to 400Hz
-- Response: average, calibrated in rms of sine wave
6-3 DC Current
Range
Resolution
Accuracy
200μA
0.1μA
±(1.2% of rdg + 2 digits)
2000μA
1μA
20mA
10μA
200mA
100μA
2A
1mA
±(2.0% of rdg + 3 digits)
10A
10mA
-- Overload protection: 500mA/500V Fast Fuse
10A/500V Fast Fuse, 10A up to 10 seconds
6-4 AC Current
Range
Resolution
Accuracy
200μA
0.1μA
±(1.5% of rdg + 3 digits)
2000μA
1μA
20mA
10μA
200mA
100μA
2A
1mA
±(2.5% of rdg + 5 digits)
10A
10mA
-- Overload protection: 500mA/500V Fast Fuse
10A/500V Fast Fuse, 10A up to 10 seconds
-- Frequency Range: 40 to 400Hz
-- Response: average, calibrated in rms of sine wave
6-5 Resistance
Range Resolution Accuracy
200Ω 0.1Ω ±(1.0% of rdg + 3 digits)
2kΩ 1Ω
±(1.0% of rdg + 2 digits)
20kΩ 10Ω
200kΩ 100Ω
2MΩ 1kΩ
20MΩ 10kΩ ±(1.5% of rdg + 3 digits)
-- Overload protection: 500V DC or AC rms
6-6 Temperature
Range Accuracy Resolution
℃-20~150℃± ( 3℃+ 1digit ) 1℃
150~1000℃± ( 3% of rdg + 2digits )
℉-4~302℉± ( 5℉+ 2digits ) 1℉
302~1832℉± ( 3% of rdg + 3digits )
-- NiCr-NiSi K-type sensor
-- Overload protection: 500mA/500V Fast Fuse
6-7 Battery test
Range Accuracy Load current Resolution
1.5V
±(5.0% of rdg + 5 digits)
Approx. 10mA 1mV
6V Approx. 100mA 10mV
9V Approx. 10mA 10mV
12V Approx. 200mA 10mV
-- Overload protection: 500mA/500V Fast Fuse
LED light
Range Green LED Yellow LED Red LED
1.5V 1.30V ±0.08V
and up.
0.94V ±0.08V to
1.29V ±0.08V
0.15V ±0.08V to
0.93V ±0.08V
6V 5.22V ±0.3V
and up.
3.76V ±0.3V to
5.21V ±0.3V
0.6V ±0.3V to
3.75V ±0.3V
9V 7.83V ±0.45V
and up.
5.64V ±0.45V to
7.82V ±0.45V
0.9V ±0.45V to
5.63V ±0.45V
12V 10.44V ±0.6V
and up.
7.52V ±0.6V to
10.43V ±0.6V
1.2V ±0.6V to
7.51V ±0.6V
-- When the voltage of the battery being tested is lower than 10% of its
rated voltage, the red LED will not light.
-- When the battery connect to the meter is reverse, the meter will appear
negative on the LCD display, but the LED will not light.
6-8 Diode and Audible continuity test
Range Description Test Condition
Display read approximately
forward voltage of diode
Forward DC current
approx. 0.4mA
Reversed DC voltage
approx. 1.5V
Built-in buzzer sounds if
resistance is less than 100Ω
Open circuit voltage
approx. 0.5V
Overload protection: 500V DC or AC rms
7. OPERATING INSTRUCTIONS
7-1 Attention before operation
7-1-1 Check battery. When the battery voltage drop below proper operation
range, the “ ” symbol will appear on the LCD display and the battery need to
changed.
7-1-2 Pay attention to the “ ” besides the input jack which shows that the input
voltage or current should be within the specified value.
7-1-3 The range switch should be positioned to desired range for measurement
before operation.
7-2 Measuring DC & AC Voltage
7-2-1 Connect the black test lead to COM jack and the red to VΩmA jack.
7-2-2 Set the rotary switch at the desired “V” range position, it shows symbol
for testing DC voltage, if you want to test AC voltage, push “SELECT” button
switch.
7-2-3 Connect test leads across the source or load under measurement.
7-2-4 You can get reading from LCD. The polarity of the red lead connection will
be indicated along with the DC voltage value.
NOTE:
1.“ ” means you can’t input the voltage more than 600V, it’s possible to show
higher voltage, but it may destroy the inner circuit or pose a shock.
2. Be cautious against shock when measuring high Voltage.
7-3 Measuring DC & AC Current
7-3-1 Connect the black test lead to COM jack and the red to the VΩmA jack for
a maximum 200mA current , for a maximum 2A or 10A current, move the red
lead to the 10A jack.
7-3-2 Set the rotary switch at the desired “uA ”& “mA ”& “A”range
position, it shows symbol for testing DC current, if you want to test AC current,
push “SELECT” button switch.
7-3-3 Connect test leads in series with the load under measurement.
7-3-4 You can get reading from LCD. The polarity of the red lead connection will
be indicated along with the DC current value.
NOTE:
1. When the value scale to be measured is unknown beforehand, set the range
selector at the highest position.
2. When only “OL” is displayed, it indicates over-range situation and the higher
range has to be selected.
3. “ ” means the socket mA’s maximum current is 500mA and 10A’s maximum
current is 10A, over 500mA or 10A current can be protected by the fast fuse.
4. On the 10A range, the measuring time should be less than 10 seconds to
prevent precision from affecting by circuit heating.
7-4 Measuring Resistance
7-4-1 Connect the black test lead to COM jack and the red to VΩmA jack.
7-4-2 Set the rotary switch at the desired “ ” range position.
7-4-3 Connect test leads across the resistance under measurement.
7-4-4 You can get reading from LCD.
NOTE: Max. input overload: 500V rms<10sec
1. For measuring resistance above 1MΩ, the mete may take a few seconds to
get stable reading.
2. When the input is not connected, i.e. at open circuit, the figure ‘OL’ will be
displayed for the over-range condition.
3. When checking in-circuit resistance, be sure the circuit under test has all
power removed and that all capacitors have been discharged fully.
7-5 Battery Testing
7-5-1 Connect the black test lead to COM jack and the red to VΩmA jack.
7-5-2 Set the rotary switch at the desired BATT “1.5V”, “6V”, “9V”or “12V” range
position to test 1.5V, 6V, 9V or 12V battery.
7-5-3 Connect the red test lead to the positive of the battery and the black test
lead to the negative of the battery.
7-5-4 You can get reading from LCD and the corresponding LED will light.
NOTE:
1. the green LED light show the battery being tested is good, it can continue to
be used; the yellow LED light show the battery being tested is close to bad, it
can be used but need to changed as quickly as possible; the red LED light
show the battery being tested is bad, it can not continue to be used.
2. When the battery connect to the meter is reverse, the meter will appear
negative on the LCD display, but the LED will not light.
7-6 Diode & Audible continuity Testing
7-6-1 Connect the black test lead to COMjack and the red to VΩmA jack.
7-6-2 Set the rotary switch at the “ ” range position, push “SELECT” to
choose Diode or Audible continuity measurement.
7-6-3 On diode range, connect the test leads across the diode under
measurement, display shows the approx. forward voltage of this diode.
7-6-4 On Audible continuity range, connect the test leads to two point of circuit,
if the resistance is lower than approx. 100Ω, the buzzer sounds.
NOTE: Make sure the power is cut off and all capacitors need to be discharged
under this measurement.
8. Battery replacement
8-1 When the battery voltage drop below proper operation range the " "
symbol will appear on the LCD display and the battery need to changed.
8-2 Before changing the battery, set the selector switch to “OFF” position and
remove the test leads from the terminals. Open the cover of the battery
cabinet by a screwdriver.
8-3 Replace the old battery with the same type battery (9V 6F22).
8-4 Close the cover of the battery cabinet and fasten the screw.
9. Fuse replacement
9-1 This meter is provided with a 500mA/500V fast fuse to protect the battery
test, temperature test and the current measuring circuits which measure up to
200mA, with a 10A/500V fuse to protect the 10A range.
9-2 Ensure the meter is not connected to any external circuit, set the selector
switch to “OFF” position and remove the test leads from the terminals. Open
the cover of the battery cabinet by a screwdriver.
9-3 Replace the old fuse with the same type and rating: 6×30mm 500mA/500V
fast fuse or 6×30mm 10A/500V fast fuse.
9-4 Close the cover of the battery cabinet and fasten the screw.
10. Maintenance
10-1 You must replace the test leads if the lead is exposed, and should adopt the
leads with the same specifications as origin.
10-2 Use only moist fabric or small amount of detergent but not chemical
solution for cleaning.
10-3 Do not use the meter before the back cover is properly closed and screw
secured. Upon any abnormality, stop operation immediately and send the
meter for maintenance.
10-4 Please take out the battery when not using for a long time.
11. Accessories
[1] Test Leads: electric rating 1000V 10A
[2] Operator’s Manual

Other manuals for 39B

1

Other HoldPeak Multimeter manuals

HoldPeak 770G User manual

HoldPeak

HoldPeak 770G User manual

HoldPeak HP-36C User manual

HoldPeak

HoldPeak HP-36C User manual

HoldPeak 90B User manual

HoldPeak

HoldPeak 90B User manual

HoldPeak HP-37A User manual

HoldPeak

HoldPeak HP-37A User manual

HoldPeak 6205 User manual

HoldPeak

HoldPeak 6205 User manual

HoldPeak 850E User manual

HoldPeak

HoldPeak 850E User manual

HoldPeak 6208 User manual

HoldPeak

HoldPeak 6208 User manual

HoldPeak HP-770H User manual

HoldPeak

HoldPeak HP-770H User manual

HoldPeak 39B User manual

HoldPeak

HoldPeak 39B User manual

HoldPeak 860D User manual

HoldPeak

HoldPeak 860D User manual

HoldPeak 870CR User manual

HoldPeak

HoldPeak 870CR User manual

HoldPeak 39K User manual

HoldPeak

HoldPeak 39K User manual

HoldPeak HP-90E User manual

HoldPeak

HoldPeak HP-90E User manual

HoldPeak 760J Installation manual

HoldPeak

HoldPeak 760J Installation manual

HoldPeak 870N User manual

HoldPeak

HoldPeak 870N User manual

HoldPeak HP-90F User manual

HoldPeak

HoldPeak HP-90F User manual

HoldPeak 870N User manual

HoldPeak

HoldPeak 870N User manual

HoldPeak 770C User manual

HoldPeak

HoldPeak 770C User manual

HoldPeak 770HD User manual

HoldPeak

HoldPeak 770HD User manual

HoldPeak 4201 User manual

HoldPeak

HoldPeak 4201 User manual

HoldPeak 6300 User manual

HoldPeak

HoldPeak 6300 User manual

HoldPeak HP-37B User manual

HoldPeak

HoldPeak HP-37B User manual

HoldPeak 860N User manual

HoldPeak

HoldPeak 860N User manual

HoldPeak HP-760E User manual

HoldPeak

HoldPeak HP-760E User manual

Popular Multimeter manuals by other brands

Steren MUL-020 manual

Steren

Steren MUL-020 manual

PCE Instruments PCE-DC2 user manual

PCE Instruments

PCE Instruments PCE-DC2 user manual

Tenma 72-10465 manual

Tenma

Tenma 72-10465 manual

Agilent Technologies U1401A User's guide and service guide

Agilent Technologies

Agilent Technologies U1401A User's guide and service guide

Simpson Electric 360-2 Operator's manual

Simpson Electric

Simpson Electric 360-2 Operator's manual

RND lab 355-00010 user manual

RND lab

RND lab 355-00010 user manual

ProTechTrader PDM-711 user guide

ProTechTrader

ProTechTrader PDM-711 user guide

RS IDM 61 instruction manual

RS

RS IDM 61 instruction manual

Wavetek Meterman DM2 Operator's manual

Wavetek

Wavetek Meterman DM2 Operator's manual

Circuit-test DCL-620 user manual

Circuit-test

Circuit-test DCL-620 user manual

UEi DM525 instruction manual

UEi

UEi DM525 instruction manual

Amprobe AM-1200 operating instructions

Amprobe

Amprobe AM-1200 operating instructions

Steinberg Systems SBS-CM-600 user manual

Steinberg Systems

Steinberg Systems SBS-CM-600 user manual

Cojali Jaltest JT2001 user manual

Cojali

Cojali Jaltest JT2001 user manual

Actron Digital Multimeter CP7676 operating instructions

Actron

Actron Digital Multimeter CP7676 operating instructions

Kaiweets KM312AB user manual

Kaiweets

Kaiweets KM312AB user manual

Atten ATM3500A user manual

Atten

Atten ATM3500A user manual

Steren MUL-016 instruction manual

Steren

Steren MUL-016 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.