②Start the measurement
MEASCAL DATA
LIGHT
SETMODE
ENT
①Enter the measurement mode
Measurement
④Start the third point calibration
Immerse the electrode in the pH 7 standard solution
until the liquid junction is fully submerged.
Immerse the electrode in the pH 4 standard solution
until the liquid junction is fully submerged.
To measurement or second point calibration
To measurement or third point calibration
②Start the rst point calibration
③Start the second point calibration
①Enter the calibration mode
Immerse 3 cm
or more.
Immerse 3 cm
or more.
To measurement
When the measurement sample is known to be acidic, perform two-point calibration with pH
7 and pH
4.
When it is known to be alkaline, perform two-point calibration with pH
7 and pH
9.
When the sample pH is not known, perform three-point calibration with pH
4, pH
7 and pH
9.
Calibration
Press
Conrm that
the instantaneous value
is stable, then
press .
When changes
from the blinking state
to the lit state and
appears, calibration is
complete.
When performing the
second point calibration,
press .
Conrm that
the instantaneous value
is stable, then
press .
When changes
from the blinking state
to the lit state and
appears, calibration is
complete.
When performing the
third point calibration,
press .
Immerse the electrode in the pH 9 standard solution
until the liquid junction is fully submerged.
※The calibration method is the same as first
and second point calibration.
When changes from the blinking state to
the lit state and appears, calibration is complete.
Press
MEASCAL DATA
SETMODE
MEASCAL DATA
SETMODE
Immerse 3 cm
or more.
Press
When changes from the blinking state
to the lit state, measurement is complete.
When the batteries run out or the instrument
will not be used for a long time, remove the batteries.