HP 421A Service manual

,
НР
427A
SM:
94)—
opp
YP
OPERATING
AND
SERVICE
MANUAL
PROPERTY
OF
ELECTRICAL
ENGINEERING
VANDERBILT
UNNER
SITY
VOLTMETER
421A
HEWLETT
Åp;
PACKARD
Vilt
dH

K4XL's
8.
BAMA
This
manual
is
provided
FREE
OF
CHARGE
from
the
“BoatAnchor
Manual
Archive’
as
a
service
to
the
Boatanchor
community.
lt
was
uploaded
by
someone
who
wanted
to
help
you
repair
and
maintain
your
equipment.
If
you
paid
anyone
other
than
BAMA
for
this
manual,
you
paid
someone
who
is
making
a
profit
from
the
free
labor
of
others
without
asking
their
permission.
You
may
pass
on
copies
of
this
manual
to
anyone
who
needs
it.
But
do
it
without
charge.
Thousands
of
files
are
available
without
charge
from
BAMA.
Visit
us
at
http://bama.sbc.edu

OPERATING
AND
SERVICE
MANUAL
-hp-
Part
No.
00427-90004
MODEL
427A
VOLTMETER
Serials
Prefixed:
947-
Appendix
C,
Manual
Backdating
Changes,
С
%
adapts
manual
to
serials
prefixed
550-,
621-,
731-
|
| |
|
4
3
|
|
1
ў
3
|
|
Copyright
Hewlett-Packard
Company
1969
P.O.
Box
301,
Loveland, Colorado,
80537
U.S.A.
Printed:
JAN
1970

a
a
e
e
Table
of
Contents
TABLE
OF
CONTENTS
Model
427A
Section
Page
I
GENERAL
INFORMATION
.
.
......
1-1
1-1.
Description.
.........
..
1-1
1-5.
Вайегу..............
1-1
1-7.
Instrument
and
Manual
Identification.
. .
.......
1-1
1-10.
Available
Accessories
......
1-1
1-12.
Cables
and
Adapters
.....
1-1
1-13.
Probes,
“22524
04
asa
a
Jal
Section
Page
п
INSTALLATION
.............
25-1
2-1.
Introduction.
...........
2-1
2-3
Initial
Inspection
. .
. .
.
557221
2-9.
Power
Requirements,
. . . . .
,
2-1
2-7.
Grounding
Requirements.
.
...
2-1
2-10.
Installation.
...........
2-1
2-12,
Bench
Mounting
........
2-1
2-14.
Rack
Mounting.
........
2-1
2-16.
Combination
Mounting
....
2-1
2-18.
Repackaging
for
Shipment
....
2-1
Section
Page
ІП
OPERATING
INSTRUCTIONS
......
3-1
3-1.
Introduction.
...........
3-1
3-3.
Front
and
Rear
Panel
Description
.........
Operating
Instructions
,....
Mechanical
Zero
Adjustment
Turn-On
Procedure.
.
.
.
.
1
DC
Measurements.
,....
2.
Resistance
Measurements
3.
AC
Measurements.
....
А
6
Section
IV
THEORY
OF
ОРЕБВАТІОМ,......
4-1.
General.
............
4-5.
DC
Орегаһоп..........
Ohms
Operation
........
.
AC
Operation.
.........
2.
Circuit
Descriptions,
.....
3.
DC
Amplifier
(A1).
. .
.,.
6.
DC
Offset
(А2)
...
.
9.
AC
Metering
Circuit
(A2)
ния
3.
Battery
Regulator
Al....
Section
У
MAINTENANCE
............
5-1,
Introduction
..........
9-3.
Required
Equipment
,......
5-5.
Performance
Checks
DB
Measurements.
,.....
.
У
ФА)
Q3 C2
Q3
023
2
C3
CO
и
|
09
se
(D
po
~
Eo
са
eR
ры
ы
PPP
PP
SP
hh
нь
н
!
1
3
|P
J
93
I
1
w
оо
tO
DD
to
ма
=
—
*
өл
өл
лс
р
ны
ка
ка
ка
(D
Section
У
MAINTENANCE
(Сопе'
4)
9-7.
АС
Accuracy
and
Frequency
Page
Response
Check.
....
5-1
5-14.
Resistance
2.
Check.
5-2
5-16,
DC
Accuracy.
4
4.50...»
5-2
5-18.
AC
Rejection
Check
.
sé
aaa
des
5-20.
Input
Impedance
Check
....
5-3
5-25.
Adjustment
and
Calibration
LIST
OF
TABLES
Number
Page
1-1.
Specifications
..,.............
1-0
3-1.
Effects
of
Harmonic
Distortion.
.
.
.
.
3-2
3-2.
DB
Range
Identification.
.
. .
.
..
..
3-2
5-1.
Required
Test
Equipment,
.......
5-0
5-2.
Resistance
Accuracy
Check.
. .
. .
.
.
5-2
01791-2
Procedures
.........
.
5-4
5-21.
Cover
Removal
........
5-4
5-30.
Amplifier
Balance
and
Zero
Offset
Adjustment.
.
. .
. .
5-5
5-31.
Regulator
Adjust
.......
5-5
5-32.
AC
Calibration.
........
5-5
5-36,
Resistance
Infinity
Adjustment
5-6
5-38.
DC
Calibration.
........
5-6
9-40.
Alternate
Performance
Checks
and
Calibration
Procedures.
.
5-6
5-42.
AC
Сїтсїв..........
5-6
5-45.
Resistance
Circuits.
. . .
..
5-6
5-47,
DC
Circuitfs
32,22
xxx
5-7
5-50.
АС
Rejection
.........
5-7
5-52.
Input
Impedance
Check
....
5-7
5-54.
Battery
Replacement.
.
.
....
5-7
5-56.
Troubleshooting.
.........
5-7
5-60.
Power
Supply
.........
5-8
5-63.
DC
Amplifier
.........
5-8
5-67.
AC
Metering
Circuit,
.....
5-8
5-71.
Adjustment
of
Factory
Selected
Components
......
....
5-8
5-13.
MRI
1.15114.
xw
ca
кез
5-8
5-75.
A2C2*
så xg
Xov
5-9
5-77.
A2R14*
T
.
5-9
5-80.
A2R16*
.....
4-8
ызык
va
10-9
5-82.
Etched
Circuit
Board
Repair.
5-9
Section
Page
VI
CIRCUIT
ПЛАСВАМ$.,,.......
2.
6-1
6-1,
Introduction
............
6-1
Section
Page
ҮП
REPLACEABLE
PARTS..........
7-1
1-1.
Introduction
............
7-1
7-4.
Ordering
Information.
.
.
. . .
. .
7-1
7-6.
Non-Listed
Parts.
.
.
...
века
41-1
Appendix
А
CODE
LIST
OF
MANUFACTURERS
B
SALES
AND
SERVICE
OFFICES
С
MANUAL
BACKDATING
CHANGES
Number
Page
5-3.
1%
Resistors............
Ux.
5-7
5-4.
Troubleshooting.
..........
..
5-7
5-5.
Power
Supply
Voltages
,........
5-8
5-6
AC
Metering
Circuit
Voltages
.....
5-8
7-1.
Replaceable
Parts.
..
.........
7-1
iii

Table
of
Contents
Model
427A
Voltmeter
, .
.
11096A
High
Frequency
Probe.
.
Location
of
Front
Panel
Controls
and
Indicators.
.
.
Impedance
Correction
Graph
Simplified
Block
Diagram
.
.
DC
ОрегаПоп.........
Ohms
Operation
.......
AC
Operation
DC
Offset
Circuit,
,.
» .
+
+=
AC
Accuracy
and
Frequency
Response
Check.
.
.
.
Моде!
427А
OF
ILLUSTRATIONS
Page
Number
Page
.
1-0
5-9,
Resistance
Accuracy
Сһеск.......
5-2
‚
1-1
5.3.
DC
Accuracy
and
Linearity
Check.
...
5-2
5-4,
AC
Rejection
Сһеск...........
5-3
.
3-0
5-5,
Input
Impedance
Check
.........
5-4
.
3-3
5-6.
Location
of
Internal
Adjustments
....
5-5
.
4-0
5-7,
Battery
Керісетепі...........
5-7
.
4-1
.
4-2
6-1.
Location
of
Switch
Components
....
-
6-2
.
4-2
6-2.
Component
Госапоп,..........
6-3
4-3
6-3.
Schematic
Diagram
...........
6-3
.
5-0
1-1.
Location
of
Mechanical
Рагів......
7-0
01791-2
а
~
а
da

Section
I
Model
427A
в
a
ын
| |
ҮМЕТЕН
HEWLETT
·
PACKARD
@
4274
VOLI
»
FUNCTION
3
EEE
Figure
1-1.
Model
427A
Voltmeter
Table
1-1.
Specifications
DC
VOLTMETER
Voltage
Ranges:
+100
mV
to
+1000
У
full
scale
па
1,
3,
10
sequence
(9
ranges).
AC
VOLTMETER
(Cont'd)
==
а
bl
си
ОПТ
а)
Input
Impedance;
10
megohms
shunted
by
40pF
on
10
mV
to
1
V
ranges;
20
pf
on
3
V
to
300
V
ranges.
Response:
Responds
to
the
average
value
of
the
in-
put;
calibrated
in
rms
voltsfor
а
sine
wave
input.
Overload:
300
V/rms
momentarily,
1
V
range
and
Accuracy:
+2%
of
full
scale
on
any
range
(09C
to
50°C),
Input
Resistance:
10
megohms
on
all
ranges.
Ре
у
below
AC
Rejection:
Superimposed
peak
ac
voltages
(60
M
:
Hz
and
above)
100
times
greater
than
full
scale
425
V/rms
maximum
above
1
V
range.
affects
reading
less
than
1%.
Maximum
450
OHMMETER
volts
peak.
Overload:
1200
Vdc
on
any
range.
AC
VOLTMETER
Voltage
Ranges:
10
mV
to
300
V
rms
full
scale
in
a
1,
3,
10
sequence
(10
ranges).
Frequency
Range:
10
Hz
to
1
MHz.
Accuracy:
(0°C
to
50°C),
RANGE
+2%
OF
FULL
SCALE
0.01
V
-
30
V
10
Hz
-
1
МН?
100
V
-
300
V
10
Hz
-
100
KHz
Frequency
Response:
Resistance
Ranges:
10
ohms
center
scale
to
10
meg-
ohms
center
scale
(7
ranges).
Accuracy:
+5%
of
reading
at
midscale
(00C
to
+50°C),
Polarity:
Common
terminal
negative,
source
Current:
OPEN
CIRCUIT
|
SHORT
CIRCUIT
RANGE
VOLTAGE
CURRENT
X10
3d8
edB
IdB
О
GENERAL
Floating
Input:
May
be
operated
up
to
500
Vdc
above
ground.
(Ohms
input
open
in
any
function
except
ohms--volts
input
open
when
instrument
is
in
off
position.)
Power:
22-1/2
volt
dry
cell
battery.
(Eveready
No.
763
ог
RCA
VS102.)
Option
01:
Battery
operation
and
ac
line
opera-
tion
(selectable
on
rear
panel).
115
or
230
V
+20%,
50
Hz
to
400Н2,
1/2
W.
IHz
SHz
lOHz
IMHz
eMHz 3MHz
4MHz
Frequency
response
10
mV
to 30
V
ranges,
1-0
01791-2
44

е
~
e
p
=
Model
427A
Section
I
SECTION
1
GENERAL
1-1.
DESCRIPTION.
1-2.
The
Hewlett-Packard
Model
427A
is
a
versatile,
compact,
self-contained
voltmeter.
It
is
capable
of
making
de
measurements
from
1
mV
to
1000
volts,
ac
measurements
from
0.
3
туѓо
300
volts
at
frequencies
from
10
Hz
to
1
MHz,
and
resistance
measurements
from
0.2
ohms
to
500
megohms.
With
the
01
option,
the
Model
427A
may
be
powered
either
by
a
115
or
230
volt
line
or
by
an
internal
22-1/2
volt
dry
cell
battery.
1-3,
The
use
of
solid
state
components
throughout
gives
the
Model
427A
both
ruggedness
and
reliability,
Battery
operation
makesthe
instrument
ideal
for
field
use
or
isolation
from
common
mode
ground
loops.
1-4,
Figure
1-1showsthe
Model427A,
and
Table
1-1
contains
а
list
of
the
Model
427A
Specifications,
1-5.
BATTERY.
1-6.
The
battery
used
in
the
Model
427A
is
a
22-1/2
volt
dry
cell,
Eveready
No.
763
or
an
RCA
VS102.
Typical
battery
life
is
morethan
300
hours
continuous
operation
or
700
hours
intermittent
operation.
1-7.
INSTRUMENT
AND
MANUAL
IDENTIFICATION.
1-8.
Hewlett-Packard
uses
a
two-section
eight-digit
serial
number
(000-00000).
If
the
first
three
digits
(prefix)
of
the
serial
number
on
your
instrument
do
not
agree
with
those
on
the
title
page
of
this
manual,
INFORMATION
change
sheets
supplied
with
the
manual
will
define
differences
between
your
instrument
and
the
Model
427A
described
in
this
manual.
1-9.
If
a
letter
prefixes
the
serial
number,
the
instrument
was
manufactured
outside
the
United
States.
1-10.
AVAILABLE
ACCESSORIES.
1-11.
The
following
accessories
are
available
to
in-
crease
the
test
capabilities
of
the
Model
427A.
1-12.
CABLES
AND
ADAPTERS.
-hp-
110014
45"
test
lead,
dual
banana
to
BNC
male
-hp-
11002А
5'
test
lead-dual
banana
to
alligator
clips
-hp-
110034
5"
test
lead-dual
banana
to
pencil
probe
and
alligator
clip
-hp-
10111A
shielded
BNC
female
to
banana
plug
adapter
1-13.
PROBES.
-hp-
Model
11074A
10:1
Voltage
Divider
Probe
extends
the
voltage
range
of
Model
427А
by
a
factor
of
10.
-hp-
Model
11039A
1000:1
Capacitive
Voltage
Divider
for
measuring
voltages
up
to
24
KV.
-hp-
Model
11096A
High
Frequency
Probe
ex-
tends
the
frequency
range
ofthe
427A
to
500
MHz.
The
voltage
range
is
0.25V
to
30V.
Figure
1-2.
11096А
High
Frequency
Probe
01791-2
1-1

e^
e
Mode1
427А
Section
П
SECTION
II
INSTALLATION
2-1.
INTRODUCTION.
2-2.
This
section
contains
information
and
instruc-
tions
necessary
for
installation
and
shipping
of
Model
427A
Voltmeter.
Included
are
initial
inspection
pro-
cedures,
power
and
grounding
requirements,
install-
ation,
and
instructions
for
repackaging
for
shipment,
2-3.
INITIAL
INSPECTION.
2-4.
This
instrument
was
carefully
inspected
both
mechanically
and
electrically
before
shipment.
It
should
be
physically
free
of
mars
or
scratches
and
in
perfect
electrical
order
upon
receipt.
To
confirm
this,
the
instrument
should
be
inspected
for
physical
damage
in
transit.
Also,
test
the
electrical
perform-
ance
of
the
instrument
using
the
procedure
outlined
in
Paragraph
5-5.
If
there
is
damage
or
deficiency,
see
the
warranty
on
the
inside
front
cover
of
this
manual,
2-5.
POWER
REQUIREMENTS.
2-6.
The
Model
427A
uses
а
22.
5
volt
dry
cell
battery
for
its
primary
power
source.
However,
if
Option01
is
included,
the
Model
427A
can
be
operated
from
any
source
of
115
or
230
volts
(420%)
at
50
to
400
Hz.
With
the
instrument
disconnected
from
the
ac
power
source,
move
the
slide
switch
(located
on
the
rear
panel)
until
the
desired
line
voltage
appears.
Power
dissipation
is
less
than
1/2
watt
maximum.
2-7.
GROUNDING
REQUIREMENTS.
2-8.
To
protect
operating
personnel,
the
National
Electrical
Manufacturers'
Association
(NEMA)
recom-
mends
that
the
instrument
panel
and
cabinet
be
ground-
ed.
The
Option
01
427A
is
equipped
with
a
three-
conductor
power
cable
which,
when
plugged
into
an
appropriate
receptacle,
grounds
the
instrument.
The
offset
pin
on
the
power
cable
three-prong
connector
is
the
ground
wire.
2-9.
To
preserve
Ше
protectionfeature
when
operating
the
instrument
from
atwo-contact
outlet,
use
а
three-
prong
to
two-prong
adapter
and
connect
the
pigtail
on
the
adapter
to
ground.
2-10.
INSTALLATION.
2-11.
The
Model
427A
is
fully
transistorized;
there-
fore,
no
special
cooling
is
required.
However,
the
instrument
should
not
be
operated
where
the
ambient
temperature
exceeds
+559С
(1310F).
For
additional
information,
address
inquiries
to
your
local
-hp-
Sales
and
Service
Office.
(See
Appendix
B
for
office
locations.
)
2-12.
BENCH
MOUNTING,
2-13,
Model
427A
is
shipped
with
plastic
feet
and
tilt
stand
in
place,
ready
for
use
as
а
bench
instrument.
01791-3
2-14.
RACK
MOUNTING.
2-15.
Model
427A
may
be
rack
mounted
by
using
an
Adapter
Frame
(-hp-
Part
No.
5060-0797).
The
adapt-
er
frame
is
a
rack
frame
that
accepts
any
combination
of
submodular
units.
It
can
be
rack
mounted
only.
2-16.
COMBINATION
MOUNTING,
2-17.
The
Model
427A
may
be
mounted
incombination
with
other
submodular
units
by
using
a
Combining
Case
(-һр-
Model
1051A).
If
the
427A
is
equipped
witha
carrying
handle,
it
will
be
necessary
to
remove
the.
top
cover
on
427A
before
inserting
it
into
the
Com-
bining
Case,
The
Combining
Case
is
a
full-module
unit
which
accepts
various
combinations
of
submodular
units.
Being
a
full-module
unit
itself,
itcan
be
bench
or
rack
mounted
and
is
analogous
to
any
full-module
instrument.
2-18.
REPACKAGING
FOR
SHIPMENT.
2-19.
The
following
paragraphs
contain
а
general
guide
for
repackaging
ofthe
instrumentfor
shipment,
Refer
to
Paragraph
2-20
if
the
original
container
is
to
be
used;
2-21
if
it
is
not,
NOTE
If
the
instrument
is
to
be
shipped
to
Hewlett-Packard
for
service
or
repair,
attach
a
tag
to
the
instrument
identifying
the
owner
and
indicating
the
service
or
repair
to
be
accomplished;
include
the
model
number
and
full
serial
number
of
the
instrument.
In
any
correspondence,
identify
the
instrument
by
model
number,
serial
number,
and
serial
number
prefix.
2-20.
If
original
container
is
to
be
used,
proceed
as
follows:
a.
Place
instrument
in
original
container.
If
original
container
is
not
available,
a
suitable
container
can
be
purchased
from
your
near-
est
-hp-
Sales
and
Service
Office.
b,
Ensure
that
container
is
well
sealed
with
strong
tape
or
metal
bands.
2-21,
И
original
container
is
notto
ре
used,
proceed
as
follows:
a.
Wrap
instrument
in
heavy
paper
or
plastic
before
placing
in
an
inner
container.
b.
Place
packing
material
around
all
sides
of
in-
strument
and
protect
panel
face
with
cardboard
strips.
c.
Place
instrument
and
inner
container
inaheavy
cartonor
wooden
box
and
seal
with
strong
tape
or
metal
bands.
d.
Mark
shipping
container
with
"DELICATE
INSTRUMENT,"
"FRAGILE"
etc,
2-1

Қ
шек
TET
сесия
таче
tn
oe
wama
=
Section
Ш
Model
427A
a
í
MODEL
427A
FRONT
PANEL
MODEL
427А
REAR
PANEL
(OPTION
Ol)
HEWLE1)
-
PACKARD
//7
427А
VOLTMETER
4)
FUNCTION
RANGE
АСУ
хок
IO
XIOOK
5
ре
30
хм
(
IOO
хюм
N
300
>
1000
—
ону
ос
ZERO/I
00
427А-8-0529
BATT/LINE
(option
01
only)
slide
switch:
Se-
COM
terminal:
Connects
to
instrument
com-
lects
either
battery
or
line
operation.
mon.
(Common
input
for
ACV,
OHMS
and
DCV.)
115/230
slide
switch
(option
01
only):
Selects
either
115
Vac
or
230
Vac
for
line
operation.
VOLTS
terminal:
Applies
unknown
ac
or
dc
voltage
to
instrument.
Line
input
(option
01
only):
Connects
ac
erati
er
to
instrument.
op
ing
pow
жігі:
2
FUNCTION
switch:
Selects
mode
of
operation.
Selections
are
OFF,
BATT,
+DC,
-DC,
OHMS
and
AC.
RANGE
switch:
Selects
appropriate
range
of
unknown
input.
e
©
DC
ZERO/2
«thumbwheel:
Usedto
electrical-
Mechanical
zero:
Mechanically
zeroes
the
ly
zerothe
instrument
in
DC
mode
and
to
infinity-
meter,
set
the
instrument
in
OHMS
mode.
©
Meter
face:
Displays
the
magnitude
of
unknown
OHMS
terminal:
Connects
unknown
resistance
resistance
or
voltage
in
ohms
or
volts
respec-
to
instrument.
tively.
о
00000
е
ғ
Figure
3-1.
Location
of
Front
Panel
Controls
and
Indicators
3-0
01791-2

жө
Model
427A
SECTION
INSTRUCTIONS
OPERATING
3-1.
INTRODUCTION.
3-2.
The
Model
427А
may
be
operated
ав
а
дс
volt-
meter,
ac
voltmeter,
ohmmeter
or
dB
meter.
This
section
contains
operating
instructions
for
each
mode
of
operation.
3-3.
FRONT
AND
REAR
PANEL
DESCRIPTION.
3-4.
Figure
3-1
shows
the
location
of
all
the
Model
427A
controls
and
indicators
and
explainsthe
function
of
each.
Тһе
Option
01
rear
panel
is
shown,
Тһе
standard
rear
panel
is
blank.
3-5.
OPERATING
INSTRUCTIONS.
NOTE
To
obtain
maximum
battery
life,
set
the
FUNCTION
switch
of
OFF
when
the
instrument
is
not
in
use.
3-6.
MECHANICAL
ZERO
ADJUSTMENT.
3-7.
Before
any
measurements
are
made,
complete
the
Mechanical
Zero
Adjustment
inthe
following
steps.
а.
Be
sure
instrument
has
been
off
for
at
least
one
minute.
b.
Rotate
Mechanical
Zero
Adjustment
screw
CLOCKWISE
until
meter
pointer
istothe
left
of
zero
and
moving
upscale
toward
zero.
c.
Continue
to
rotate
adjustment
screw
clockwise,
STOP
when
needle
is
exactly
on
zero,
If
needle
overshoots,
repeat
step
b.
а.
When
pointer
is
exactly
over
zero,
rotate
adjustment
screw
slightly
COUNTERCLOCK-
WISE
to
relieve
tension
on
suspension.
И
the
pointer
іпоуев
to
the
left,
repeat
whole
procedure,
but
make
the
counterclockwise
rotation
less.
3-8.
TURN-ON
PROCEDURE.
3-9,
STANDARD
INSTRUMENT,
a,
Rotate
the
FUNCTION
Switch
to
BATT/1.
5
MIN
position,
b.
Тһе
meter
should
read
1.
5
or
higher
on
the
0-3
scale
on
any
range,
indicating
that
the
battery
voltage
is
15
volts
or
higher.
И
the
reading
is
below
1.5,
replace
the
battery
according
to
the
steps
in
Paragraph
5-54.
01791-2
Section
ІП
3-10.
OPTION
01
INSTRUMENT.
а.
3-11.
а.
b.
с.
е,
3-12.
а.
Select
either
battery
ог
line
operation
withthe
rear
panel
BATT/LINE
slide
switch.
H
bat-
tery
operation
is
selected,
check
the
battery
according
to
Paragraph
3-9.
If
line
operation
is
selected,
set
the
115/230
slide
switch
to
indicate
the
proper
line
volt-
age,
Rotate
the
FUNCTION
switch
to
the
desired
function.
During
line
operation,
the
BATT/
1.5
MIN
check
position
displays
the
output
of
the
Option
01
power
supply.
The
reading
should
be
1,5
or
higher
on
the
0-3
scale
оп
any
range,
indicating
а
power
supply
output
of
15
volts
or
more.
This
serves
as
а
соп-
venient
check
of
the
Option
01
power
supply.
DC
MEASUREMENTS.
Rotate
the
FUNCTION
switch
to
+DCV
or
-DCV
depending
on
the
polarity
of
the
input.
Short
the
VOLTS
input
tothe
COM
input,
го-
tate
RANGE
to
0.
1,
and
adjustthe
DC
ZERO/
Өс
thumbwheel
for
zero
meter
deflection.
Remove
shorting
connection.
Н
there
is
a
zero
offset
with
COM
and
VOLTS
open,
refer
to
Paragraph
4-16
and
Paragraph
5-30.
Select
approximate
range
of
input
with
RANGE
switch,
DO
NOT
APPLY
MORE
THAN
1200
VDC
TO
ANY
DC
RANGE.
Connect
input
across
VOLTS
and
COM
ter-
minals
and
read
magnitude
of
input
on
meter.
RESISTANCE
MEASUREMENTS.
DAMAGE
ТО
427A
INPUT
CIR-
CUIT
МАҮ
RESULT
IF
DC
OR
AC
VOLTAGE
IS
APPLIED
TO
OHMS
TERMINAL.
Rotate
the
FUNCTION
switch
to
OHMS.
Selectthe
approximate
range
withthe
RANGE
switch;
and
withthe
input
terminals
open,
ай-
just
the
DC
ZERO/Qe
thumbwheel
for
an
=
indication
onthe
ohms
scale.
(Pointer
should
rest
on
the
mark
just
to
the
left
of
х).
3-1

Section
Ш
с.
Connect
the
unknown
resistance
across
the
OHMS
and
COM
terminals.
Read
the
resis-
tance
value
on
the
ohms
scale.
NOTE
For
best
accuracy,
select
an
ohms
range
‘that
will
place
the
meter
pointer
near
the
center
of
the
scale,
3-13.
AC
MEASUREMENTS.
3-14,
The
Model
427А
responds
to
the
average
value
о
the
ac
input
and
is
calibrated
in
rms
volts
for
a
sine
wave
input.
value
of
а
non-sinusoidal
signal
are
different,
any
distortion
on
the
input
will
affect
the
accuracy
of
the
reading.
Table3-1
shows
the
effect
of
harmonic
dis-
tortion
on
å
reading.
NOTE
The
following
table
is
uni-
versal
in
application
since
these
errors
are
inherent
in
all
average-responding
volt-
meters.
The
error
shown
above
may
vary
with
the
phase
relation-
ship
between
the
harmonic
and
fundamental.
Table
3-1.
Effects
of
Harmonic
Distortion
INPUT
VOLTAGE
TRUE
RMS
METER
CHARACTERISTICS
VALUE
INDICATION
umaman
=
10
|
юш
_|
ко
|
Fundamental
+
10%
second
harmonic
Fundamental
+
20%
102
second
harmonic
Fundamental
+
50%
112 100
-
110
second
harmonic
Fundamental
+
10%
third
һагтпопіс
Fundamental
+
20%
third
harmonic
3-15.
Use
the
following
stepsto
make
ап
ас
measure-
Fundamental
+
50%
ment.
third
harmonic
а,
Rotate
FUNCTION
switch
to
АСУ,
NOTE
With
the
input
shorted,
there
may
be
а
zero
offset
of
about
two
minor
divisions.
This
is
caused
by
the
bias
current
through
the
meter
bridge
and
does
not
affect
the
accuracy
of
ac
measurements
as
the
meter
moves
upscale,
Since
the
average
value
and
the
rms
b.
с.
Model
427А
a
м
Rotate
RANGE
switch
to
approximate
range
of
input
voltage.
DO
NOT
APPLY
MORE
THAN
425V
RMS
WHEN
THE
INSTRUMENT
IS
ON
RANGES
ABOVE
3,
OR
MORE
THAN
300V
RMS
ON
RANGES
BE-
LOW
3.
Connectthe
signal
to
be
measured
to
the
VOLTS
and
COM
terminals
and
read
the
magnitude
on
the
voltage
scale.
3-16.
DB
MEASUREMENTS.
а,
Making
а
dB
ог
dBm
measurement
is
essen-
tially
the
same
as
making
an
ac
voltage
mea-
surement,
Follow
Ше
steps
іп
Paragraph
3-13,
$
but
read
the
magnitude
on
the
dB
scale.
The
1
volt
position
of
the
RANGE
switch
is
the
0
dBm
range.
Each
position
above
1
volt
?
is
а
10
dB
increase,
and
each
position
below
1
volt
is
а
10
АВ
decrease.
Table
3-2
lists
the
dB
value
of
each
range.
Table
3-2.
DB
Range
Identification
Å
given
dB
reading
is
equal
to
the
algebraic
sum
of
the
range
andthe
meter
reading.
For
example,
if
the
meter
reading
were
-6
and
the
instrument
were
on
the
10
volt
(+20
dB)
range,
the
final
reading
would
be
20
dB
-
6
dB
-
14
dB.
The
427A
meter
is
calibrated
in
dBm.
0dBm
is
equivalent
to
0.775
volt
dropped
acrossa
6002
load.
Consequently,
any
dBm
measure-
ments
must
be
made
acrossatotal
impedance
of
6002.
Measurements
across
other
imped-
ances
will
be
in
dB,
not
dBm.
To
convert
a
dB
reading
to
dBm,
usethe
Im-
pedance
Correction
Graph
(Figure
3-2).
For
example,
to
convert
a
+30
dB
reading
made
across
а
502
load
to
dBm,
locate
the
500
load
impedance
on
the
bottom
of
the
graph.
Follow
the
impedance
line
to
the
heavy
black
line
and
read
the
meter
correction
at
that
point.
Тһе
correction
for
500
is
+10.5
dBm,
and
the
corrected
reading
is
+40.
5
dBm.
01791-2

С
Ма
eee
ee
JE
7,”
4
`
ез
©
«те:
Model
4274
беспоп
Ш
pm
а
--
L
у
л
И
ЕН
ШЕНІНЕ
1
1
I
а
шш
ІНЕН
БП
Ча]
Esi
L
РЕНИ
Г
БЕ
БН
НЕНІ
ИЕ
НЕД
ШШ
БЕН
БЕН
ЕН
ЕНЕ
НЕ
СЦ
0
Г;
HHHH
=H
="
LHP
SO
шен
ші
ші
1915!
ЕЕЕ
иш
k.
БЕН
Б
4011]
__|_|
111
ЕГРИП
TT
I
EN
e
—
з
н
ee
a
А:
‚шш
нт
[7
__|__|_|
|
шши
ГЕТЕ
ш
pref
ft
+
©
ы
j
`
[1
т
ERE
ПА
шан
ESE
ЕЙ
Fi
S
I.
ЖЕН
ЕЕ
ЕЕ
|
Ll
ЕНСЕ
а
+
on
L1
|
ЕЛ
БЕ
БЕН
ЕНІН
ТОТИ
ӘБЕН
БЕН
ШЕВ.
КН
[
|
|
(1106
eee
|
1]
[
1
[]
ПЕР
|211
ПЕ
|__
||
METER
CORRECTION
-
DBM
5Тр-8-0477
IMPEDANCE
-
ОНМ5
„~
+.
Figure
3-2.
Impedance
Correction
Graph
01791-1
3-3
FEE
TEEN
ЕЕ
АЙ

=
а
er
а
ee
ны
EE
cS
EM
S
UNS.
EE
Pv
Omm
"P.
"rue
ааа
элит,
ғ
9
ела
I
Г
№
а
"
№
Æ
da
<
weraq
yora
periduns
ст-р
ә10319
=
S
=ч
<
Ë
ла
TÀ
Ф
>
*
=
vYOH-
912%
S110A
г9-
S39V110A
NONINO
кӨн
наш
AHILLVE
ATddNS
Анз.1уй
SITOA
¿`9
+
(2v)
LINDYID
HOLVNNILLV
HOLVNNILLV
Алдап5
эм
|
ONINILIN
27
1508
27
JONVY
DV
измоа
10
NOILdO
vo
SWHO
(9)
|172
SAHO
|
ov)
MOLYNNILIV
90193735
|
IndNI
SL
OA
5752
на
0997
зомуч
SWHO
NOLLAN
дпам вино
(гу)|
192
50
ТЕГЕ"!
(2%)
Lsnrav
HOLVNNILLV
135440
20
39474
20
>
m
E
o
=
o
т
ul
4-0

ør
A
ph
=
po
Model
427A
Section
IV
SECTION
IV
THEORY
OF
4-1.
GENERAL.
4-2.
The
Model
427A
measuresac
voltage,
dc
voltage,
and
resistance.
It
is
battery
operated,
but
with
Op-
tion
01,
can
be
powered
by
line
voltage.
Figure
4-1
shows
a
simplified
block
diagram
of
the
427A.
4-3.
The
Battery
Regulator
(А1)
regulates
the
bat-
tery
output
and
provides
+6.
7
and
-6.
7
volt
bias
sup-
ply
to
the
amplifiers.
The
427A
uses
two
amplifiers,
the
DC
Amplifier
(А1)
and
the
AC
Metering
Circuit
(A2).
The
former
is
a
high
input
impedance
unity
gainamplifier
used
toamplify
de
and
resistance
inputs.
It
also
serves
as
å
preamplifier
for
ac
signals.
The
AC
Metering
Circuit
amplifies
ac
signals
from
the
preamplifier,
converts
them
to
dc
signals
proportional
to
the
average
ac,
and
connects
them
to
the
meter.
The
meter
is
calibrated
to
display
the
rms
value
of
the
ac
input.
4-4,
The
DC
Offset
Adjust
(A2)
compensates
for
leakage
current
from
the
de
amplifier,
and
DC
CAL
and
OHMS
CALareresistive
circuits
used
for
calibration.
4-5.
DC
OPERATION.
4-6.
Figure
4-2
shows
Ше
Model
427A
in
the
DC
Mode
of
Operation.
The
ас
input
is
first
applied
to
the
DC
Range
Attenuator
where
it
is
attenuated
by
10dB
for
each
stepoftheattenuator.
Thedc
signal
from
the
at-
tenuator
goes
to
the
DC
Filter,
and
the
filter
rejects
any
ac
superimposed
noise
that
may
be
present
on
the
input.
The
dc
output
of
the
filter
goes
to
the
DC
Am-
рііНег
(А1)
and
then
to
the
meter.
The
DC
Amplifier
matches
the
high
impedance
of
the
attenuator
to
the
low
impedance
of
the
meter.
The
DC
CAL
circuits
ATTENUATOR
Figure
4-2.
01791-2
OPERATION
are
resistive
circuits
in
series
with
the
meter
used
to
adjust
the
meter
current
to
calibrate
the
lower
ranges.
4-7.
OHMS
OPERATION.
4-8.
Figure
4-3
15
a
block
diagram
of
the
Model
427A
in
the
OHMS
Mode
of
Operation.
With
the
input
open,
Ra
and
Rp
forma
voltage
divider.
The
voltage
across
R,,
causes
full
scale
current
toflow
through
the
meter.
The
OHMS
CAL
circuit
adjusts
the
meter
current
for
an
indication
of
»
with
the
input
open.
When
R,
is
equal
tothe
parallel
combination
of
Ва
and
Rp,
the
total
re-
Bistance
from
the
OHMS
terminal
to
ground
will
be
half
the
parallel
combination
of
R4
and
Ву,
the
voltage
into
the
amplifier
will
be
halved,
and
the
meter
indi-
cation
will
be
half
scale.
The
Model
427A
is
designed
во
that
thefull
RANGE
setting
will
be
displayed
in
the
center
of
the
scale.
For
example,
10
Поп
the
X10
range
is
a
center
scale
reading.
4-9.
AC
OPERATION.
4-10.
Figure
4-4
shows
a
block
diagram
of
the
427A
in
the
AC
Mode
of
Operation,
The
input
signal
goes
to
the
AC
Range
Attenuator.
On
the
1
volt
range
and
be-
low,
the
signal
is
not
affected
by
the
AC
Range
Attenuator;
but
on
all
the
higher
ranges,
the
signal
is
attenuated
by
504В,
Capacitor
C3adjusts
the
frequency
response
of
the
attenuator
with
а
3
volt
100
КН?
input.
Тһе
signal
from
the
AC
Range
Attenuator
goes
through
the
DC
Amplifier
to
the
AC
Post
Attenuator
where
it
is
attenuated
by
10
dB
for
each
step
of
the
RANGE
selector.
The
DC
Amplifier
matches
the
low
impe-
dance
of
the
Post
Attenuator
to
the
high
impedance
of
the
Range
Attenuator,
acting
ав
а
preamplifier.
ос
ос
FILTER
AMPLIFIER
FEEDBACK
DC
Operation

Section
IV
OHMS
ATTENUATOR
427А-НО
ty
Model
427A
DC
AMPLIFIER
FEEDBACK
Figure
4-3.
Ohms
Operation
AC
RANGE
AT
TENUATOR
DC
AMPLIFIER
FEEDBACK
427A—ROA
AC
POST
ATTENUATOR
AC
METERING
CIRCUIT
AC
FEEDBACK
Figure
4-4.
AC
Operation
4-11.
The
AC
Metering
Circuit
contains
both
a
feedback
Stabilized
ac
amplifier
andan
averaging
meter
circuit,
The
meter
circuit
converts
the
ac
signal
to
a
dc
volt-
age
proportional
to
the
average
of
the
ac
amplifier
output.
A2R17
adjusts
the
current
through
the
meter
so
that
the
scale
reading
is
in
rms
volts.
4-12.
CIRCUIT
DESCRIPTIONS.
4-13.
DC
AMPLIFIER
(А1).
4-14.
Figure
6-3
is
the
schematic
diagram
of
the
Model
427A.
The
input
to
the
DC
Amplifier
(А1)
is
applied
through
Pin
2
to
the
impedance
converter
A1Q6.
A
field
effect
transistor
is
used
as
the
impe-
dance
converter
because
of
its
characteristicaily
high
input
impedance.
Transistors
А107
and
A1Q9
make
пра
two-stage
amplifier,
with
4109
as
ап
emitter
follower
output
stage.
The
signal
from
the
emitter
of
А109
is
fed
back
to
the
base
of
А107
for
gain
sta-
bilization.
4-15.
A1Q8actsasa
constant
current
source
for
A1Q6.
The
constant
current
from
A1Q8
assures
linearity
in
A1Q6
and
helps
minimize
drift.
4-2
4-16.
DC
OFFSET
(A2).
4-17.
The
gate
leakage
from
А196
(Iggg)
amounts
to
а
fraction
of
a
nanoamp.
This
current
is
insignificant
with
the
input
shorted;
but
with
the
input
open,
it
flows
through
the
DC
Range
Attenuator resistance
of
10
M2
resulting
in
a
few
millivolts
deflection
on
the
lower
ranges.
Consequently,
the
meter
indications
with
the
input
open
and
with
the
input
shorted
can
be
different.
When
measuring
across
a
low
impedance,
Iggg
flows
mostly
through
the
low
impedance;
and
this
difference
does
not
effect
the
measurement.
Unless
1065
is
compensated
for,
it
will
effect
measurements
made
across
high
impedances
(>1
М
9).
4-18.
The
DC
Offset
Circuit
(A2) (see
Figure
4-5)
compensates
for
the
leakage
from
A1Q6.
A
positive
voltage
tapped
from
A2R7
reverse
biases
A2CR1.
The
reverse
leakage
current
from
A2CR1
(Ir)
meets
ап
extremely
high
resistance
at
the
gate
of
A1Q6
and
passes
through
the
much
lower
resistance
of
the
at-
tenuator,
іп
a
direction
opposite
to
that
of
Iqgg.
Ву
adjusting
A2R7,
Ip
can
be
set
so
that
it
and
Теса
cancel
eachother.
The
net
voltage
dropacross
the
attenuator
01791-2
ғ
4.

ж
+
ж
~
Model
427A
РУ
А2
RG
OF
4127
<,
RESISTANCE
<,
OF
ATTENUATOR
<,
(IO
MEGOHMS)
Y
4274-8-0560
Figure
4-5.
DC
Offset
Circuit
would
then
be
zero.
Ап
alternative
method
of
mini-
mizing
the
effects
of
Іс
would
beto
zero
the
instru-
ment
with
the
impedance
of
the
measurement
source
connected
to
Ще
input
terminals
rather
than with
the
input
shorted,
A2CR1
has
а
temperature
charac-
teristic
quite
similar
to
that
of
A1Q6.
"Therefore,
01791-2
Е
Iu
|
|
|
Section
IV
both
Теса
and
Ig
change
by
approximately
the
same
amount
with
temperature
and
the
amount
of
additional
error
at
higher
temperatures
is
lessened.
4-19.
AC
METERING
CIRCUIT
(A2).
4-20.
Figure
6-3
contains
the
schematic
of
the
AC
Metering
Circuit.
The
ac
input
from
the
ac
post
at-
tenuator
is
applied
through
A2C1
tothe
base
of
A2Q1.
A2Q2
is
an
emitter
follower
that
provides
impedance
matching
to
common
emitter
output
stage
A2Q3.
Capacitor
A2C4
provides
an
ac
feedback
path
for
gain
stabilization.
4-21.
The
output
from
the
collector
of
A2Q3
is
recti-
fied
by
A2CR3
and
A2CR4
and
applied
to
the
meter
movement.
4-22.
Resistor
A2R17
is
used
to
calibrate
the
ampli-
Нег
at
low
frequency
ага
is
adjusted
for
full
scale
meter
deflection
witha
10
ту,
400
Hz
input.
Capacitor
А2С3
is
used
to
calibrate
the
amplifier
at
high
fre-
quency.
With
а
10
mV,
1
MHz
input,
A2C3
is
adjusted
for
full
scale,
4-23.
BATTERY
REGULATOR
Ai
(See
Figure
6-3).
4-24.
The
Battery
Regulator
regulates
the
dc
voltage
fromthe
22,
5
volt
battery
or
from
the
optional
power
supply.
The
voltage
drop
across
A1CR2
and
the
emitter-base
junction
of
A1Q3
provides
the
-6,
7
volt
referencefor
the
power
supply
andthe
-6.
7
volt
output.
A1Q5
senses
changes
in
the
+6.
7
volt
output
and
соп-
trols
A1Q2,
the
positive
series
regulator.
А101
isa
current
source
for
A1Q5,
and
diodes
A1CR3,
CR4,
and
CR5
provide
temperature
stabilization.
Тгап-
sistor
A1Q4
provides
a
current
shunt
for
A1CR2
to
keep
its
voltage
constant
under
heavy
current
loads,

Section
У
Model
427A
Table
5-1,
Required
Test
Equipment
INSTRUMENT
REQUIRED
RECOMMENDED
TYPE
CHARACTERISTICS
USE
MODEL
DC
Voltage*
Range:
0-300
V
Performance
Checks
~hp-
Model
738BR
уой-
Standard
Accuracy:
+0.2%
meter
Calibrator
Test
Oscillator*
|
Frequency
Range:
10
Hz
to
1
MHz
Performance
Checks
|
-hp-
Model
652A
Oscil-
Output:
0
to
3
Vac
lator
Variable
Line
Frequency
Range:
50
to
60
Hz
Performance
Checks
Powerstat
Type
116B
Transformer
Voltage
Range:
0
to
120
V
Decade
Resistor
|
Range:
102
to 10
MQ
Performance
Checks
|
General
Radio
Model
Accuracy:
+0.5%
1432Z
Decade
Resistor
DC
Voltmeter
Range:
0-50
Vdc
Troubleshooting
-hp-
Model
427A
Volt-
Accuracy:
+2%
meter
AC
Voltmeter
Range:
0-50
Vac
Troubleshooting
-hp-
Model
427A
Volt-
Accuracy:
+2%
meter
Resistors
100
ко
+,
1%
Performance
Checks
-hp-
Part
Ко.
0811-0460
1
МО
+
.1%
-hp-
Part
No,
0811-0473
*
Contained
іп
-hp-
E02-738BR
VTVM
Calibration
System
VOLTMETER
CALIBRATION
SYSTEM
hpE02-7386R
hp
427A
VOLTMETER
VOLTMETER
CALIBRATOR
Артзавв
TEST
OSCILLATOR
Apes2a
4274-B-1180
Figure
5-1.
AC
Accuracy
and
Frequency
Response
Check
5-0
01791-2
A^
а

ж
o
Model
427A
EE
==
Section
V
SECTION
V
MAINTENANCE
5-1.
INTRODUCTION.
5-2.
This
section
contains
information
necessary
to
maintain
the
Model
427A.
The
following
paragraphs
describe
the
Performance
Checks,
the
Calibration
Procedures,
and
the
Troubleshooting
Procedures.
5-3.
REQUIRED
EQUIPMENT.
5-4.
Table
5-1
15а
list
of
the
equipment
needed
to
properly
maintain
the
Model
427A.
If
the
recommended
model
is
not
available,
use
any
substitute
that
meets
the
required
characteristics.
5-5.
PERFORMANCE
CHECKS.
5-6,
The
Performance
Checks
are
"in
cabinet"
tests
that
compare
the
Model
427A
with
its
specifications.
These
procedures
can
be
used
both
for
incoming
in-
spection
and
periodic
inspection.
A
Performance
Check
Test
Cardis
providedat
the
endof
this
section
for
recording
the
performance
of
the
Model
427A
during
the
performance
checks.
The
card
canbe
removed
from
the
manual
and
used
as
a
permanent
record
of
the
incoming
inspection
or
of
a
routine
performance
check.
if
the
Model
427A
does
not
meet
its
specifi-
cations,
refer tothe
Adjustment
and
Calibration
Pro-
cedures,
Paragraph
5-25,
5-7.
AC
ACCURACY
AND
FREQUENCY
RESPONSE
CHECK.
5-8.
The
AC
Accuracy
and
Frequency
Response
Check
requires
a
test
oscillator
that
is
flat
within
+
0.5%
from
10
Hz
to
1
MHz.
The
absolute
value
of
the
ap-
plied
voltage
must
be
accurate
within
+0.2%.
The
-hp-
Model
738BR
Voltmeter
Calibrator
produces
a
400
Hz
signal
that
is
within
less
than
0.2%
of
the in-
dicated
output.
The
-hp-
Model
652A
Test
Oscillator
can
be
adjusted
to
within
less
than0.
25%
of
a
desired
reference
from
10
Hz
to
10
MHz.
NOTE
Before
beginning
the
Performance
Checks,
be
sure
to
adjust
the
me-
chanical
meter
zero
according
to
the
steps
in
Paragraph
3-6.
5-9.
AC
ACCURACY
CHECK.
5-10.
Connect
the
test
oscillator,
and
the
voltmeter
calibrator
as
shown
in
Figure
5-1.
a.
Set
the
voltmeter
calibrator
for
an
output
of
0.01
volts
at
400
Hz.
b,
Set
51
in
Figure
5-1
to
position
A.
c.
Set
the
Model
427A
FUNCTION
to
ACV
and
the
range
to
0.01,
d.
The
427A
should
read
0.01
volts
rms
+
2%
(2
minor
scale
division).
01791-3
е.
Repeat
steps
а
through
d
for
each
ас
RANGE
selection
through
300
V
by
setting
the
volt-
meter
calibrator
output
tothe
full
scale
value
for
each
range.
Тһе
Model
427A
indication
should
not
vary
from
the
known
input
by
more
than
2%
оп
any
range.
2%
corresponds
to
2
minor
divisions
on the
0-1
scale
and
1-1/2
minor
divisions
on
the
0-3
scale,
f.
Set
the
Model
427A
to
the
1
volt
range
and
apply
0.9,
0.8,
0.7, 0,6,
0,5, 0.3,
0,2
and
0.1
volt
signals.
In
each
case
the
reading
should
be
within
two
minor
divisions
of
the
known
input
signal.
е.
Set
the
Model
427A
to
the
3
volt
range
and
apply
2.5,
2,
1.5,
1,
and
0.5
volt
signals.
In
each
case
the
reading
should
be
within
1-1/2
minor
divisions
of
the
knowninput
sig-
nal.
9-11.
FREQUENCY
RESPONSE
CHECK.
9-12.
The
frequency
range
of
the
-hp-
Model
652A
із
10
Hz
to
10
MHz.
The
set
level
indication
men-
tioned
in
the
following
steps
is
used
to
re-establish
the
proper
amplitude
each
time
the
frequency
is
changed.
a.
Set
S1
in
Figure
5-1
to
position
B.
Set
the
Model
427A
RANGE
to
0.01.
b.
Setthe
test
oscillator
to
400
Hz
and
adjust
the
output
for
a
0.95
indication
on
the
427A
0-1
scale.
с.
Set
test
oscillator
to
expand,
and
adjust
the
set
level
control
(REF
SET)
for
convenient
meter
indication.
d.
Record
set
level
indication
for
use
as
a
re-
ference
in
steps
e
and
f.
e.
Change
the
oscillator
frequency
to
10
Hz
and
reset
the
oscillator
am
plitude
to
the
set
level
indication
establishedinstepc.
DO
NOT
re-
adjust
the
set
level
control
on
the
oscillator.
The
427A
indication
should
not
vary
by
more
than
+2
minor
scale
divisions
(x
296).
f.
Repeat
step
e
for
100
Hz,
1
kHz,
10
kHz,
100
kHz,
500
kHz
and
1
MHz.
in
each
case
the
427A
indication
should
not
vary
by
more
than
+2
minor
scale
divisions
(+
29),
5-13.
Кереаі
Paragraph
5-11
with
the
427А
RANGE
set
to
3
volts,
and
test
oscillator
output
set
to
give
3
V
reading
оп
427A
іп
step
b.
Тһе
427A
reading
should
not
vary
by
more
than
+2%
(+1-1/2
small
scale
divisions).
5-1

Section
У
5-14.
RESISTANCE
ACCURACY
CHECK.
5-15,
To
check
the
resistance
accuracy,
precision
resistances
are
needed.
Figure
5-2
shows
the
resis-
tance
accuracy
check
using
а
General
Radio
Model
14322
Decade
Resistor.
The
resistance
used
should
be
accurate
to
within
+0.
5%
and
should
have
а
range
of
10
ohms
to
10
М
ohms,
a.
With
the
input
open,
set
the
Model
427A
FUNCTION
to
OHMS
and
the
RANGE
to
X10.
Adjust
the
DC
ZERO/Sx
adjustment
for
an
indication
of
«
on
the
meter
face.
(The
needle
should
rest
on
the
mark
just
to
the
left
of
о).
b.
Connect
the
decade
resistor
and
the
Model
421A
as
shown
in
Figure
5-2,
and
set
the
decade
for
10
0.
c.
Тһе
427A
meter
should
read
within
+5%
of
the
known
resistance
(-
one
small
scale
divi-
sion,
+
one-half
small
division).
d.
Repeat
steps
a
through
c
using
the
RANGE
and
decade
resistor
settings
listed
in
Table
5-2.
DECADE
RESISTOR
GENERAL.
RADIO
14322
990090090869
—
—
Model
427А
т
=
МОТЕ
The
DC
ZERO/
adjustment
need
only
be
made
on
the
X10
and
X1K
ranges.
Table
5-2,
Resistance
Accuracy
Check
RANGE
5-16.
DC
ACCURACY.
5-11.
The
DC
Accuracy
Check
requires
а
дс
voltage
standard
that
is
accurate
to
within
+0.
2%
of
its
indi-
cated
output
from
0.1
volt
to
300
V.
The
-hp-
Model
738BR
Voltmeter
Calibrator
is
accurate
to
within
+0,
1%
of
the
indicated
output,
Figure
5-3
shows
the
test
setup
for
the
DC
Accuracy
Check.
~
а.
Set
the
Model
427A
FUNCTION
to
«ОСУ
and
the
RANGE
to
0.1.
hp
427A
VOLTMETER
@@@
=
427А4-8-0533
Figure
5-2.
Resistance
Accuracy
Check
VOLTMETER
CALIBRATOR
hp
7388R
hp
427A
VOLTMETER
°
427A-B-0330
ғ
«ғ.
Figure
5-3.
DC
Accuracy
and
Linearity
Check
01791-3

p
o.
=
~
Model
427A
b.
Adjust
the
DC
ZERO/Q«
for
zero,
This
ad-
justment
need
only
be
made
on
the
0.1
volt
range.
c.
Connect
the
voltmeter
calibrator
and
the
Model
427A
as
shown
in
Figure
5-3.
а.
Set
ће
voltmeter
calibrator
output
to
0.1
volts
dc.
The
Model
427A
should
read
0.1
volt
«275,
e.
Repeat
steps
a
through
e
for
each
RANGE
selection
to
300
V
by
setting
the
voltmeter
calibrator
output
to
the
full
scale value
for
each
range.
The
Model
427A
indication
should
not
vary
from
the
known
input
by
more
than
+2%
on
any
range.
2%
corresponds
to
2
min-
or
divisions
on
the
0-1
scale
and
1-1/2
minor
division
on
the
0-3
scale.
1.
Set
the
dc
RANGE
to
1000
V
and
the
voltmeter
calibrator
to
300
V.
The
Model
427A
should
read
300
V
+2%
of
full
scale
on
the
1000
volt
range.
г.
Set
the
Model
427A
to
the
1
volt
range
and
apply
0.9,
0.8,
0.7,
0,6,
0.5, 0.4,
0.3, 0.2,
and
0,1
volt
dc
signals
from
the
voltmeter
calibrator,
1n
each
case
the
reading
should
be
within
two
minor
divisions
of
the
known
input
voltage.
В.
Set
the
Model
427A
to
the
3
volt
range
and
apply
2.5,
2,
1,5,
1,
and
0.5
volt
dc
signals.
In
each
case
the
reading
should
be
within
1/2
minor
division
on
the
0-3
scale.
i.
Repeat
setps
a
through
f
with
the
FUNCTION
set
to
-DCV
and
the
voltmeter
calibrator
set
for
a
negative
output.
Тһе
test
results
should
be
the
same.
5-18.
AC
REJECTION
CHECK.
3-19.
A
peak
ac
superimposed
noise
signal
100
times
the
full
scale
input
should
affect
the
Model
427A
read-
ing
less
than
1%.
Figure
5-4
shows
the
test
setup
using
a
variable
line
transformer
as
a
noise
generator.
А
7.07
volt
rms
output
from
the
variable
line
trans-
former
corresponds
to
a
10
volt
peak
noise
signal.
The
10
volt
noise
signal
will
be
applied
to
the
0.
1
volt
dc
range.
VARIABLE
LINE
TRANSFORMER
42TA-B-O53t
Section
V
а.
Set
the
427A
FUNCTION
to
-DCV
and
short
the
input,
Set
the
RANGE
to
0.
1
and
adjust
the
DC
?ЕНО/Фо
for
0
meter
indication.
b.
Switch
the
FUNCTION
to
ACV
and
the
RANGE
to
10.
Connect
the
variable
line
transform-
er
and
the
Model
427A
as
shown
in
Figure
5-4.
c.
Adjust
the
transformer
output
for
a
reading
of
7,
07
volts
rms
on
the
427A.
d.
Switch
the
427A
FUNCTION
to
+DCV
andthe
RANGE
to
0.1.
e.
Тһе
meter
reading
shouldbe
0
+1%
(1
minor
scale
division).
NOTE
The
meter
may
move
upscale
momentarily
and
then
return
to
zero.
This
indicatesthe
charging
of
the
DC
Filter
capacitors
and
is
normal.
1.
Repeat
steps
a
through
e
using
the
1
volt
dc
range
and
70,
7
volt
rms
signal
from
thetrans-
foriner.
5-20,
INPUT
IMPEDANCE
CHECK.
5-21.
INPUT
RESISTANCE
CHECK.
5-22.
Figure
5-5
shows
the
setup
for
the
input
resis-
tance
check,
А
1
МО
resistor
is
connected
in
series
with
the
input,
and
the
voltage
drop
across
the
input
resistance
will
be:
Rin
Ев
=
Fapplied
|
Reeries
+
Rin
With
1
volt
applied,
Ев
will
be
0.91
volt
if
the
input
resistance
is
10
МО.
Ед
varies
directly
with
changes
in
Rin.
а,
Connect
the
Model
427A
and
the
voltmeter
calibrator
to
position
Å
illustrated
in
Figure
5-5.
Set
the
Model
427A
FUNCTION
to+DCV
and
RANGE
to
1
V.
b.
Set
the
voltmeter
calibrator
for
+1
V
output.
с.
Connect
the
voltmeter
calibrator
to
position
B
shown
in
Figure
5-5,
hp
427A
VOLTMETER
Figure
5-4.
AC
Rejection
Check
01791-3
Table of contents
Other HP Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Victron energy
Victron energy BMV-702 Quick installation guide

Dytran Instruments
Dytran Instruments 3256C operating guide

Oxford PRECISION
Oxford PRECISION OXD-331-7420K Operator instructions

Siemens
Siemens SITRANS RD300 operating instructions

Kathrein
Kathrein MFK 16 operating manual

ALFAMATIC
ALFAMATIC Press-Right user manual

KHAE
KHAE GW6 Series instruction manual

Endress+Hauser
Endress+Hauser Deltapilot S FMB70 operating instructions

janitza
janitza UMG 103-CBM User manual and technical data

Tyredog
Tyredog TD2000A-X owner's manual

WIKA
WIKA A2G-30 operating instructions

Hanna Instruments
Hanna Instruments HI6321 Quick reference guide