HP 3457A User manual

HP
3457A
Multimeter
HP
3457A
Service
Manual
|
@
HEWLETT
PACKARD
Copyright
©
Hewlett-Packard
Company,
1988
Manual
Part
Number:
03457-90012
Edition
3
Microfiche
Part
Number:
03457-99012
E0288

Printing
History
The
Printing
History
shown
below
lists
the
printing
dates
of
all
Editions
and
Updates
created
for
this
manual,
The
Edition
number
changes
as
the
manual
undergoes
subsequent
revisions,
Editions
are
num-
bered
sequentially
starting
with
Edition
1.
Updates,
which
are
issued
between
Editions,
contain
in-
dividual
replacement
pages
which
the
customer
uses
to
update
the
current
Edition
of
the
manual
Updates
are
numbered
sequentially
starting
with
Update
|.
When
a
new
Edition
is
created,
all
Updates
associated
with
the
previous
Edition
are
merged
into
the
manual.
Each
new
Edition
or
Update
also
in
cludes
a
revised
copy
of
this
printing
history
page.
Many
product
updates
and
revisions
do
not
require
manual
changes
and,
conversely,
manual
corrections
may
be
done
without
accompanying
product
changes.
Therefore,
do
not
expect
a
one-to-one
cor-
respondence
between
product
updates
and
manual
updates.
Edition
|
(Part
Number
03457-90010)
+--+
+>
EER
VME
SF
earl
Ria
APRIL,
1985
Update
|
(Part
Number
03457-90201)
++
eet
+
+
OCTOBER,
1985
Edition
2
(Part
Number
03457-90011)
>
+--+
+--+
ips
i
aol
-aibte’
erm
4
MAY,
1986
Update
|
(Part
Number
03457-90090)
-
-
-
ee
tees
JUNE,
1987
ition
3
(Part
Number
03457-90012)-
++
> -
*
tress
PEBRUARY,
1988
RESTRICTED
RIGHTS
LEGEND
Use.
duplication.
or
disclosure
by
the
Government
is
subject
to
restrictions
as
set
forth
in
subdivision
(bX
3
Mii)
of
the
Rights
in
Technical
Data
and
Computer
Software
clause
at
532.227-7013.
Hey
lett-Packard
Company
3000
Hanover
Street,
Palo
Alto,
California
94304

|
YD
HEWLETT
PACKARD
SAFETY
SUMMARY
The
following
general
safety
precautions
must
be
observed
during
all
phases
of
operation,
service,
and
repair
of
this
instrument.
Failure
to
comply
with
these
precautions
or
with
specific
warnings
elsewhere
in
this
manual
violates
safety
standards
of
design,
manufacture,
and
intended
use
of
the
instrument.
Hewlett-Packard
Company
assumes
no
liability
for
the
customer's
failure
to
comply
with
these
requirements.
GROUND
THE
INSTRUMENT
To
minimize
shock
hazard,
the
instrument
chassis
and
cabinet
must
be
connected
to
an
electrical
ground.
DO
NOT
OPERATE
IN
AN
EXPLOSIVE
ATMOSPHERE
Do
not
operate
the
instrument
in
the
presence
of
flammable
gases
or
fumes.
Operation
of
any
electrical
instrument
in
such
an
environment
constitutes
a
definite
safety
hazard.
KEEP
AWAY
FROM
LIVE
CIRCUITS
Operating
personnel
must
not
remove
instrument
covers.
Component
replacement
and
internal
adjustments
must
be
made
by
qualified
maintenance
personnel.
Under
certain
conditions,
dangerous
voltages
may
ex-
ist
even
with
the
instrument
switched
off.
To
avoid
injuries,
always
disconnect
input
voltages
and
discharge
circuits
before
touching
them.
DO
NOT
SERVICE
OR
ADJUST
ALONE
Do
not
attempt
internal
service
or
adjustment
unless
another
person,
capable
of
rendering
first
aid
and
resuscitation,
is
present.
DO
NOT
SUBSTITUTE
PARTS
OR
MODIFY
INSTRUMENT
Because
of
the
danger
of
introducing
additional
hazards,
do
not
install
substitute
parts
or
perform
any
unauthorized
modification
to
the
instrument.
Return
the
instrument
to
a
Hewlett-Packard
Sales
and
Ser-
vice
Office
for
service
and
repair
to
ensure
that
safety
features
are
maintained.
DO
NOT
OPERATE
A
DAMAGED
INSTRUMENT
Whenever
it
is
possible
that
the
safety
protection
features
built
into
this
instrument
have been
impaired,
either
through
physical
damage,
excessive
moisture,
or
any
other
reason,
REMOVE
POWER
and
do
not
use
the
instrument
until
safe
operation
can
be
verified
by
service-trained
personnel.
If
necessary,
return
the
instrument
to
a
Hewlett-Packard
Sales
and
Service
Office
for
service
and
repair
to
ensure
that
safety
features
are
maintained.

|
@
HEWLETT
PACKARD
Operating
and
Safety
Symbols
Symbols
Used
On
Products
And
In
Manuals
AC
line
voltage
input
receptacle.
Instruction
manual
symbol
affixed
to
product.
Cautions
the
user
to
refer
to
respective
instruction
manual
procedures
to
avoid
possible
damage
to
the
product.
Indicates
dangerous
voltage
—
terminals
connected
to
interior
voltage
exceeding
1000
volts.
Protective
conductor
terminal.
Indicates
the
field
wiring
terminal
that
must
be
connected
to
earth
ground
before
operating
equipment
—
protects
against
electrical
shock
in
case
of
fault.
Clean
ground
(low-noise).
Indicates
terminal
that
must
be
connected
to
earth
ground
before
operating
equipment
—
for
single
common
connections
and
protection
against
electrical
shock
in
case
of
fault.
Frame
or
chassis
ground.
Indicates
equipment
chassis
ground
rH
oon
LE
;
; ;
:
terminal
—
normally
connects
to
equipment
frame
and
all
metal
parts.
Affixed
to
product
containing
static
sensitive
devices
—
use
anti-static
Aas
handling
procedures
to
prevent
electrostatic
discharge
damage
to
ATTENTION
ne
components.
NOTE
Calls
attention
to
a
procedure,
practice,
or
condition
that
requires
special
attention
by
the
reader.
CAUTION
CAUTION
Calls
attention
to
a
procedure,
practice,
or
condition
that
could
possibly
cause
damage
to
equipment
or
permanent
loss
of
data.
WARNING
WARNING
Calls
attention
to
a
procedure,
practice,
or
condition
that
could
possibly
cause
bodily
injury
or
death.
vi

Section
|
Section
Il
Section
Ill
TABLE
OF
CONTENTS
GENERAL
INFORMATION
Berea
|
I-1.
Introduction.
.....
1-3.
Manual
Description
1-5.
Instrument
Description...
..........
1-7.
Safety
Considerations
1-9,
Instruments
Covered
by
this
Ms
anual
.
113:
Specifications
ovo
s
p.oels
4%
FF
1-15.
Options
and
Accessories
..
.
be
28
117,
Tesh
Eetpments
5.6
6%
wed
4
4:5
Sas
es
INSTALLATION
2-1.
Introduction.
nas
2-3.
Initial
Inspection.
. .
.
2-5.
Preparation
for
Use.
2-8.
Line
Voltage
Selection
2-10.
Power
Cords.
xz
2-12,
Grounding
Requirement.
gy
fake
2-14)
BenghvOse
i
5.3.0
Hea
a
ace
2-16.
Rack
Mounting...
.........-45
2-18.
Interface
Connections.
on
We
2-19.
Control
Interface...
.........--
2-21.
HP-IB
Address
Selection...
........
2-23.
Changing
the
Address
from
the
Front
Panel...
.
.
Sri
bate
es
Hae
2-24,
Changing
the
Address
from
the
CONCOMET
i
eset
is
a
inderas.
wed
baey
How
HZ
2-25.
Installation
of
Options
...........
.2
2-28.
Rear
Terminal
Assembly
Removal.
. .
2-7
2-30.
Plug-In
Assembly
Installation...
.
. .
2-8
2-32.
Plug-In
Assembly
Configuration
..
.
.
2-9
2-38.
Wiring
Block
Preparation
.......
.
2-11
OPERATION
...
esse
ese
reerencces
3-1.
Introduction...
......
3-3.
General
Information
.
3-4,
AC
Power.
. .
3-6.
Power
On.
. .
é%
aapk
‘%
29:
ONCPAUON,
505.34
id
3
Synr8'
eee
ag
Eade
3-11.
Voltage
Measurements
.........-
3-18.
Resistance
Measurements.
.
. .
3-25.
Current
Measurements
379
14
3-32.
Frequency/Period
Measurements.
.
.
.
3-6
3-39,
HP
44491A
General
Purpose
IMAI
DIOR
CS
cece
ris)
syste
et
ei
ieid
oa
3-6
3-45.
HP
44492A
Reed
Relay
Multiplexers
cncc
2
6
nines
a
cate
at
are
e
Page
Section
IV
PERFORMANCE
TESTS
..
4-1.
Introduction
4-3.
Equipment
Required
4S
Test
Resoid«.
dinccd
9
Gtk
ee
»
Hy
4-7.
Calibration
Cycle
.
4-9.
Test
Considerations
ALOMGSETAL
sii
oo
Praia
s
wy
ered
Be
4-12.
Performance
Tests...
.
rarer
a
DC
Voltage
Performance
Tests........
4-
4-13.
Preliminary
Steps.
...
.
ag
4
4-14.
DC
Voltage
Function
-
Offset
Test
|
4-16.
DC
voltage
Function
-
Gain
Test...
.
4-
4-18,
DC
Voltage
Function
-
Lineari
RUGS:
hoot
sisogduri
do
enGhi
aad
wade
AC
Voltage
Performance
Tests:
4-20.
AC
Voltage
Function
-
Gain
Test
.
4-22.
AC
Voltage
Function
-
Frequency
RESPONSE
WERE.
&
shins)
CD
0
a
DC
Current
Performance
Tests.
4-24.
DC
Current
Function
-
Offset
Test
.
.
4-8
4-26.
DC
Current
Function
-
Gain
Test...
4-8
AC
Current
Performance
Tests........
4-9
4-28.
AC
Current
Function
-
Gain
Test...
4-9
2-Wire
Ohms
Performance
Tests
.....
4-10
4-30,
2-Wire
Ohms
Function
-
Offset
Te
4-32.
2-Wire
Ohms
Function
-
Gain
Test
.
4-11
4-Wire
Ohms
Performance
Tests
.....
4-11
4-34.
4-Wire
Ohms
Function
-
Offset
bh
oh
eee
Coed
ere
4-12
4-36.
4-Wire
Ohms
Function
-
Gain
Test
.
4-13
Frequency
Counter
Performance
Tests
.
4-13
4-38.
Frequency
Counter
-
Accuracy
LOST,
.
3
Gnsste
dive
abked
&,
Bulag
nie
al
ak
4-14
Rear
Input
Performance
Tests
(Standard
Instrument)...6
2
6
ecceccawcanne
4-40.
Preliminary
Steps.
.
4-4].
DC
Voltage
Function
-
Rear
Terminal
OFfRCG
Teh
6
2.3
att
AR
ets
Oe
4-14
4-43,
DC
Current
Function
-
Rear
Terminal
Offset,
Test.
20
ec0
ce
ss
a
craters
HA
Hes
4-15
4-45,
2-Wire
Ohms
Function
-
Offset
BORG
ras
stieis
a
a4
aarainl
ed
eet
He
$e
4-15
4-47,
4-Wire
Ohms
Function
-
or
fset
TEs
5.4
Wi
as
KAKA
ARR
dD
4-16
Rear
Input
Performance
Tests
(44491A
General
Purpose
Relay
Assy)
....-
4-17
4-49.
Preliminary
Steps.............
-

TABLE
OF
CONTENTS
(Cont'd)
Section
IV
Page
PERFORMANCE
TESTS
(Cont'd)
4-50.
DC
Voltage
Function
-
44491A
Offset
Test
.
~
417
4-52.
2-Wire
Ohms
Function
-
444914
OP
ESOG
ESE
osc
eo
bebe
Gasney
woard
4-18
4-54,
4-Wire
Ohms
Function
-
44491
A
OPGSBt
TSE,
odstus
5
4.
o
cans
ea
3
tw
mem
4-18
4-56.
DC
Current
Function
-
44491.
OFESEE
THOSE.
stonise
ora
dade
avs!
8
akane
a
ad
4-19
Rear
Input
Performance
Tests
(44492A
10
Channel
Multiplexer
Assy).......
4-20
4-58.
Preliminary
Steps...............
4-20
4-59.
DC
Voltage
Function
-
444924
OFtRel
TOSh
3:25.36
been
SS
eee
os
>
4-20
4-61.
2-Wire
Ohms
Function
-
44492A
ORR
est
iia sie
ns
aariiius
Daaasie
4
9m
4-21
4-63.
Operational
Verification
Tests
.....
.
4-22
viii
DC
Voltage
Operational
Verification
TE
BLS
Ss,
ocd
iso
apegiraa
0,
Gcakepes
On
epee
ee
ee
4-65.
Preliminary
Steps.
4-66,
DC
Voltage
Function
-
Of
4-68.
DC
Voltage
Function
-
Gain
Test...
.4-23
AC
Voltage
Operational
Verification
TOSS.
cece
cress
enna
veneene
4-24
4-70.
AC
Voltage
Function
-
‘Gain
Test...
.4-24
4-72.
AC
Voltage
Function
-
Frequency
Response
Fest
oi:
'a:
o
sigsiss
6
6d,
Sraine
ig
%
avn
4-24
DC
Current
Operational
Verification
PRESS
a
5,
as
G
iheh
5.
SO
eat
&
Riki
4:20
Go
4-25
4-74,
DC
Current
Function
-
Of
.
4-26
4-76.
DC
Current
Function
-
4-26
AC
Current
Operational
Verificati
o
.
-
4-27
4-78.
AC
Current
Function
-
Gain
wee
4-27
2-Wire
Ohms
Operational
Verification
Tests
4-27
4-80.
2-Wire
Ohms
Function:
TEs
a
Aas
ed
aya
A
4
Sigurt
4:
Ks
4-27
4-82.
2-Wire
Ohms
Function
-
Gain
Test
.
.
.4-28
4-Wire
Ohms
Operational
Verification
TOSS.
eciee
cv
ew
ses
ecene
scene
4-29
4-84.
4-Wire
Ohms
Function
-
Offset
OBES
gf
macnn
unioh
mately
Vial
mass
Jus
athe
4-29
4-86.
4-Wire
Ohms
Function
-
Gain
Test
.
.
.4-30
Frequency
Counter
Operational
Verification
TestS.
2.0.
66
cs
aes ees
4-31
4-88.
Frequency
Counter
-
Accuracy
OSES
es
Rut
love
trace
&
ego
tuate.¥
dob
ies
4-31
Rear
Input
Operational
Verification
Test
(Standard
Instrument).........
4-32
Section
IV
PERFORMANCE
TESTS
(Cont'd)
4-90.
Preliminary
Steps.............
+
4-32
4-91,
DC
Voltage
Function
-
Rear
Terminal
Offset
Test...
...
4-93.
DC
Current
Function
-
Rear
Terminal
Offset
Test...
0...
see
os
4-95,
2-Wire
Ohms
Function
-
Rear
Terminal
Offset
Test
4-97.
4-Wire
Ohms
Function
-
Rear
Terminal
Offset
Tests
ccc60
6
gadis
«
Rear
Input
Operational
Verification
Tests
(44491A
General
Purpose
Relay
Assy)
4-99,
Preliminar:
4-100.
DC
Voltage
Fu
Offset
Test
4-102.
2-Wire
Ohms
Function
-
44491A
Offset
Test
<9
4-104.
4-Wire
Ohms
Function
-
44491A
Offset
Test
4-106.
DC
Current
Function
-
444914
Offset
Test
Rear
Input
Operational
Verification
Test
(44492A
10
Channel
Multiplexer
Assy)...
........005
4-108.
Preliminary
Steps
4-109.
DC
Voltage
Function
-
44492A
Offset
Test
4-111.
2-Wire
Ohms
Function
-
44492A
Offset
Test:
Limits:
054.45
49
sae
sa
ction
-
44491A
Section
V
CALIBRATION
5-1.
Introduction.
.
5-4.
Calibration
Secu
5-6.
Security
Code.
5-8.
Changing
the
Security
Code.
5-10.
Calibration
Number
5-13.
Preliminary
Calibration
ProcedurGsia.
so
s.Qy
ss
eames
ee
eSB
5-15.
Input
Amplifier
Offset
Adjustments
..........02.02+.
34
5-17.
AC
Convertor
Frequency
Response
Adjustment
5-19.
Calibration
Procedures
-
Front
PaneliIiputs:
«
3...0sies
v2
6
dates
wD
5-21.
DC
Volts
Offset
Calibration.
.
5-5
5-23.
DC
Volts
Gain
Calibration.
..
2.2...
5-6
5-25.
DC
Volts
Linearity
Calibration...
. .
5-6

TABLE
OF
CONTENTS
(Cont'd)
Section
V
Page
Section
VI
Page
CALIBRATION
(Cont'd)
REPLACABLE
PARTS
(Cont'd)
5-27.
2-Wire
Ohms
Offset
Calibration...
...
5-8
Digital
(Al)
Assembly
Removal.
. . .
.
6-4
5-29,
2-Wire
Ohms
Gain
Calibration
.
Plug-In
Support
Assembly
5-31.
4-Wire
Ohms
Offset
Calibration
.
.
.
Removal
..<.
esac
¢ 5
aaa
Edy
4
0-5
5-33,
4-Wire
Ohms
Gain
Calibration
2
6-29.
AC
Convertor
(A3)
Assembly
5-35.
DC
Current
Offset
Calibration...
...
5-10
Removal
0-5
5-37.
DC
Current
Gain
Calibration.
.....
.
5-10
6-31.
Analog
(A2)
Assembly
Removal
.
0-5
5-39.
AC
Volts
Offset
&
Gain
6-33.
Front
Panel
Removal...
2.2.0...
6-0
Calibration...
.
.
5-41,
AC
Current
Section
VII
Page
5-43.
Frequency
Calibration
......
.
2
512
MANUAL
CHANGES.........--555%
7A
5-45.
Calibration
Procedures
Rear
Panel
FAs
INtrOdCHOH:.
si i<
sm
acpe
ae
Ha
ged
7-1
Inputs
5-47,
DC
Volts
Offset
Ca!
bration.
..
ee.
5-12
Section
VIII
Page
5-49,
2-Wire
Ohms
Offset
Calibration...
.
.
5-13
SERVICE
accede
nesters
eens
8-1
5-51.
4-Wire
Ohms
Offset
Calibration...
.
.5-14
8-1.
Introduction.
.
. .
2
ted
er
cays.
Bal
:
DC
Current
Offset
Calibration...
..
.5
14
8-2.
Safety
Considerations.
........-
.
8-1
5-55.
Calibration
Procedures
-
44491
A
8-3.
Static
Handling
.
8-2
General
Purpose
Relay
Assy
8-4.
Recommended
Test
Equipment
....
.
8-2
5-56.
Wiring
Block
Preparation
. .
=e
8-5.
General
Block
Diagram
Theory
5-58.
DC
Volts
Offset
Calibration,
...
2...
5-16
of
Operation
8-2
5-59,
2-Wire
Ohms
Offset
Calibration...
..
5-17
8-6.
Chassis
Common
C
8-3
5-60.
4-Wire
Ohms
Offset
Calibration
.
SeL7
8-10.
Floating
Common
Circuitr
8-3
5-61,
DC
Current
Offset
Calibration.
.....
5-17
8-18.
Detailed
Theory
of
Operation
.....
8-5
5-62.
Calibration
Procedures
-
444924
8-19.
Input
Circuitry
. . .
8-5
10
Channel
Multiplexer
Assy.
.
8-26.
Ohms
Current
.
8-11
5-63.
Wiring
Block
Preparation
.
8-31.
A/D
Converter...
0...
6.005
8-13
5-65.
DC
Volts
Offset
Calibration.
.
5-19
8-35:.A/D
‘Controllers.
«isis
es as
8-16
5-66.
2-Wire
Ohms
Offset
Calibration...
.
.
5-19
8-40.
Main
Controller
(03457-66501)...
8-18
8-49.
Main
Controller
(03457-66511)...
8-23
Section
VI
8-55.
AC
to
DC
Converter...
...
REPLACABLE
PARTS........-+++++%
8-63.
44491
A
General
Purpose
6-1.
Introduction.
...
2...
Relay’
Gatdsas
asi
sf
ares.
34-2
stirs
.
8-32
6-3.
Ordering
Information
8-69.
44492A
10
Channel
Reed
GA)
[lated
RARE:
acca,
weeded
wm
SaewS
ee
Multiplexer...
.....
8-35
6-6.
Non-Listed
Parts
.
&
8-71.
Troubleshooting
.
eu
8-35
6-8.
Direct
Mail
Order
System
ee
ee
6-1
8-72.
General
Information
....
soa,
+
(8-36
6-11.
Part
Changes.......
~
8-73.
Al
Assy.
Failure
Indications.
.
..
.
8-36
6-13.
Proprietary
Parts...
. -
6-2
8-74.
A2/A3
Assy.
Failure
Indications
. .
8-36
6-15.
Disassembly/Assembly
Procedures.
.
. .
.6-3
B75.
SE
TESS
6
asiss
coe
eatin
wa
eR
8-36
6-17.
Rear
Terminal
Assembly
Removal.
. .
6-3
8-78.
Self
Test
Changes
for
New
Main
6-19.
Optional
Plug-In
Assembly
Controller
(03457-66511)...
0.6.4
8-44
TRENIOVEL
¢
we
acunsgses
Jr
ee
Ow
Dea
aks
6-3
8-81.
Autocal
(ACAL)
. .
2.
8-46
6-21.
Top
Cover
Removal...
......--5
6-4
8-82.
Plug-in
Cards
.........
...
8-60
8-85.
Replacement
Procedures.
......
8-02

&S
SECTION
|
GENERAL
INFORMATION
The
information
contained
in
this
manual
is
intended
for
the
use
of
service
trained
personnel
who
understand
electronic
circuitry
and
are
aware
of
the
hazards
involved.
Do
not
attempt
to
perform
any
of
the
procedures
outlined
in
this
manual
unless
you
are
qualified
to
do
so.
1-1,
INTRODUCTION
1-2.
This
manual
contains
information
relating
to
the
installation,
operation,
performance
testing,
calibration,
and
service
of
the
HP
3457A
Multimeter.
The
information
is
designed
for
the
use
of
service trained
personnel.
Other
users
should
refer
to
the
HP
3457A
Operating
Manual.
1-3,
MANUAL
DESCRIPTION
1-4,
The
following
paragraphs
provide
a
general
description
of
the
manual
layout
and
content:
&
Section
|.
General
Information
Section
I
contains
a
brief
description
of
the
manual,
and
a
general
description
of
the
instrument.
This
section
also
contains
the
instrument
specifications
and
information
concerning
options
and
accessories.
Section
Il,
Installation
Procedures
Section
IL
contains
information
to
prepare
the
HP
3457A
for
use,
Included
in
this
section
is
power
requirements,
line
voltage
selection,
and
interfacing
information.
Section
Ill.
Operating
Instructions
Section
III
contains
operating
instructions
necessary
to
service
the
HP
3457A.
These
operating
instruc-
tions
are
a
condensed
version
of
those
contained
in
the
Operating,
Programming
and
Configuration
Manual.
Section
IV.
Performance
Tests
Section
IV
contains
information
and
procedures
required
to
test
the
HP
3457A
to
its
rated
specifica-
tions,
Also
included
in
this
section
is
a
condensed
version
of
the
performance
test
(operational
verifica-
tion
test)
which
will
provide
a
more
rapid
test
and
give
a
90%
confidence
that
the
instrument
will
meet
its
specifications.
Section
V.
Calibration
&
Section
V
contains
procedures
required
to
calibrate
the
HP
3457A
to
its
rated
specifications.

General
Information
3457A
Section
VI.
Replaceable
Parts
Section
VI
lists
part
numbers
for
cabinet
parts
and
printed
circuit
assemblies.
Section
VII.
Manual
Backdating
Section
VII
contains
information
required
to
adapt
this
manual
to
instruments
whose
serial
numbers
are
lower
than
those
listed
on
the
title
page.
Section
VIIl.
Service
Section
VII
contains
a
block
diagram
theory
of
operation,
schematic
diagrams,
and
procedures
to
aid
in
troubleshooting
the
HP
3457A.
1-5.
INSTRUMENT
DESCRIPTION
1-6.
The
HP
3457A
is
a
versatile
high
precision
digital
multimeter
which
measures
ac
or
de
volts,
ac
or
de
current,
actde
volts
or
current,
resistance,
frequency
and
period.
It
measures
voltages
up
to
300
Vac
or
300
Vde
with
high
de
resolution
and
accuracy.
The
instrument
is
designed
for
both
bench
and
system
use.
Program
and
reading
storage
are
included
to
allow
high
speed
measurement
applications
with
reading
rates
of
greater-than
1000
read-
ings
per
second.
Provision
has
been
made
for
optional
assemblies
such
as
a
general
purpose
scanner
or
multiplexer
assembly.
1-7.
SAFETY
CONSIDERATIONS
1-8.
The
HP
3457A
is
a
safety
class
|
instrument
(provided
with
a
protective
earth
terminal).
The
instrument
and
manual
should
be
reviewed
for
safety
markings
and
instructions
before
operation.
Refer
to
the
Safety
Summary
preceding
this
section
for
appropriate
safety
instructions
and
markings
covering
this
instrument.
1-9.
INSTRUMENTS
COVERED
BY
THIS
MANUAL
1-10,
Instruments
covered
by
this
manual
are
identified
by
a
serial
number
prefix
listed
on
the
title
page.
Hewlett-Packard
uses
a
two
part
serial
number
in
the
form
xxxxayyyyy,
where
xxxx
is
the
serial
prefix,
a
is
the
country
of
origin
(A=USA)
and
yryyy
is
the
serial
suffix.
The
serial
number
prefix
identifies
a
series
of
identical
instruments.
The
serial
number
suffix
is
assigned
sequentially
and
is
unique
to
each
instrument,
1-11.
If
the
serial
number
prefix
of
your
instrument
is
greater
than
the
one
listed
on
the
title
page,
a
yellow
Manual
Changes
supplement
will
explain
how
to
adapt
this
manual
to
your
instrument.
1-12.
If
the
serial
number
prefix
of
your
instrument
is
lower
than
the
one
listed
on
the
title
page,
information
contained
in
Section
VII
(Manual
Backdating)
will
explain
how
to
adapt
this
manual
to
your
instrument.
1-13.
SPECIFICATIONS
1-14.
Specifications
for
the
HP
3457A
Multimeter
are
listed
in
Table
|-1.
These
specifications
are
the
perfor-
mance
standards
which
the
instrument
is
guaranteed
to
meet.

3457A
General
Information
Table
1-1.
Specifications
DC
VOLTS
INPUT
CHARACTERISTICS:
MAXIMUM
RESOLUTION
READING
RANGE
|6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits|3
1/2
digits
30
mv
|
30.30000
10
nv
100
nv
1
uv
10
uv
300
mv}
303.0000
100
nv
1
uv
10
uv
100
uv
3
Vv
3.030000
1
uv
10
uv
100
uv
1
mv
30
Vv
30.30000
10
uv
100
uv
1
mv
10
mv
300
Vv
303.0000
100
uv
1
mv
10
mv
100
mv
INPUT
RESISTANCE
(OHMS):
30
millivolt
through
3
volt
range
-
10
gigaohms
30
volt
and 300
volt
range
-
10
megohms
(+/-
1%)
MAXIMUM
INPUT
VOLTAGE:
(non-destructive)
HI
or
LO
to
Earth
Ground:
+
450V
peak
MEASUREMENT
ACCURACY:
+(%
of
reading
+
Number
of
Counts)
Auto-Zero
ON
24
HOUR:
Tcal
+
1°C
2
hour
warm-up
Accuracy
relative
to
calibration
standard
24
hour
specifications
apply
if
the
instrument
is
calibrated
between
90
and
100%
of
full-scale.
For
calibration
between
33
and
90%
of
full-scale,
add
.00015%
to
%
Reading.
%
READING
NUMBER
OF
COUNTS
100
PLC*
10
PLC*
1
PLC*
+1
PLe*
+005
PLC*
+0005
PLC*
6
1/2
digits|6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits|3
1/2
digits
215
235
350
24 25
35
5
6 6
10
20
6 6
90
DAY:
Tcal
+
5°C
After
1
hour
warm-up
RANGE
|
%
READING
NUMBER
OF
COUNTS
100
PLC*
10
PLC*
1
PLC*
+1
PLO*
+005
PLC*
+0005
PLC*
6
1/2
digits|6
1/2
digits|6
1/2
digits]5
1/2
digits|4
1/2
digits|3
1/2
digits
365 385
500
39 40 50
6
Ka
7
41%
20 30
6
c4
Kg

General
Information
3457A
Table
1-1.
Specifications
DC
VOLTS
(Cont’d)
1
YEAR:
Teal
+
5°C
After
1
hour
warm-up
RANGE
|
%
READING
NUMBER
OF
COUNTS
100
PLC*
10
PLC*
1
PLC*®
-1
PLC*
-005
PLC*
-0005
PLC*
6
1/2
digits|]6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits|3
1/2
digits
30
mv
0045
365
500
70
300
mv
0035
39
50
3
0025
6 *
30
Vv
+0040
19 30
300
v
0055
6
7
*
PLC
=
Integration
Time
in
Power
Line
Cycles.
When
using
integration
times
of
100
PLC,
10
PLC,
or
1PLC;
multiply
Number
of
Counts
in
the
Measurement
Accuracy
Tables
by
0.1
for
5
1/2
digit
readings,
0.01
for
4
1/2
digit
readings
and
0.001
for
3
1/2
digit
readings.
Teal
=
the
temperature
of
the
environment
where
the
3457A
was
calibrated.
Calibration
should
be
performed
with
the
environment
temperature
between
18
and
28
degrees
centigrade.
TEMPERATURE
COEFFICIENT:
*(%
of
Reading
+
Number
of
Counts)/°C
6
1/2
Digits
Displayed
RANGE
|
Auto-Zero
ON
Auto-Zero
OFF*
30
mv
+0005
-0005
+
40
300
mv
~0005
0005
+
13
3
0005
.
-0005
+
10.3
30
Vv
~0005
.
+0005
+
11.0
300
Vv
0005
‘
-0005
+
10.3
*
Specifications
given
are
for
a
stable
environment
(+
1°C)
and over
a
ten
minute
period.
Multiply
Number
of
Counts
by
0.1
for
5
1/2
digit
readings,
0.01
for
4
1/2
digit
readings
and
0.001
for
3
1/2
digit
readings.
NOISE
REJECTION:
With
1
Kohm
imbalance
in
the
LO
lead
and
line
frequency
(50
or
60
Hz)
INTEGRATION
TIME
100
PLC
|
10
PLC
|
1
PLC
|
-1
PLC
|
-005
tc
|
.0005
PLC
AC
NMR
90 db 80 db
60
db
0
db
0
db
0
db
AC
ECMR
160
db
156
db
136
db
76
db
76
db 76
db
DC
CMR
140
db
140
db
140
db
140
db
140
db
140
db

3457A
General
Information
Table
1-1.
Specifications
DC
VOLTS
(Cont’d)
MAXIMUM
READING
RATES:
(Readings/Second)
INTEGRATION
|
DIGITS
READING
RATE
TIME
DISPLAYED
|60
HZ
|50
Hz
0.0005
PLC
31/2
1350
1350
0.005
PLC
4
1/2
1250
1250
0.1
PLC
5
1/2
360
312
1
PLC
6
1/2
53
45
10
PLC
6
1/2
4.8 4.0
100
PLC
6
1/2
48
-40
Reading
Rates
apply
with
the
keyboard
lock
ON,
Auto-Zero
OFF,
Display
OFF,
Math
Func-
tion
OFF,
Delay
Time
set
to
0.0,
Manual
ranging,
readings
stored
in
internal
memory
using
Timer
Trigger
and
single
precision
integer
format.
DC
CURRENT
INPUT
CHARACTERISTICS:
MAXIMUM
RESOLUTION
SHUNT
READING
RESISTANCE
RANGE
|6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits|3
1/2
digits
300
uA|
303.0000
uA
100
pA
1
mA
10
nA
100
nA
Ohm
3
mA}
3.030000
ma
1
nA
10
nA
100
nA
1
uA
Ohm
30
mA}
30.30000
mA
10
nA
100
nA
1
uA
10
uA
10
Ohm
mA]
303.0000
mA
nA
1
uA
10
uA
100
uA
1
Ohm
A
1.000000
A uA
10
uA
100
uA
1
mA
«1
Ohm
MAXIMUM
BURDEN
VOLTAGE:
0.35V
-
300UA,
3mA and
30mA
ranges.
0.6V
-
300mA
range.
1
V
-
1A
range.
MEASUREMENT
ACCURACY:
+(%
of
reading
+
Number
of
Counts)
After
1
hour
warmup,
with
Auto-zero
ON.
24
HouR:
Tcal
+
1°C
After
two
hour
warm-up.
Accuracy
relative
to
calibration
standard.
RANGE
|
%
READING
NUMBER
OF
COUNTS
100
PLC*
10
PLC*
1
PLC*
«1
PLC*
-005
PLC*
+0005
PLC*
6
1/2
digits|6
1/2
digits]6
1/2
digits|5
1/2
digits]4
1/2
digits|3
1/2
digits
54
54
65
9
54
54
65
9
54
54
65
154
504
155

General
Information
3457A
Table
1-1.
Specifications
@
DC
CURRENT
(Cont’d)
90
DAY:
Tcal
+
5°C
After
one
hour
warm-up.
RANGE
|
%
READING
NUMBER
OF
COUNTS
100
PLC*
10
PLC*
Vv
PLe®
«1
PLC*
005
PLC*
~0005
PLC*
6
1/2
digits]6
1/2
digits|]6
1/2
digits|5
1/2
digits|4
1/2
digits|3
1/2
digits
104
104 11S
14
104
104 115
14
104 104
115
14
204 204 215
24
604 604
615
64
1
YEAR:
Teal
+
5°C
After
one
hour
warm-up.
RANGE
|%
READING
NUMBER
OF
COUNTS
100
PLC
10
PLC*
1
PLC
+1
PLc*
+005
PLC*
+0005
PLC*
6
1/2
digits|6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits|3
1/2
digits
0.04
104 104 115
14
0.04
104
104
115
14
0.04
104
104
115
14
0.08
204 204 215
24
0.08
604 604 615
64
PLC
=
Integration
Time
in
Power
Line
Cycles.
When
using
integration
times
of
100
PLC,
10
PLC,
or
1PLC;
multiply
Number
of
Counts
in
the
Measurement
Accuracy
Tables
by
0.1 for
5
1/2
digit
readings,
0.01
for
4
1/2
digit
readings
and
0.001
for
3
1/2
digit
readings.
Tcal
=
the
temperature
of
the
environment
where
the
3457A
was
calibrated.
Calibration
should
be
performed
with
the
environment
temperature
between
18
and
28
degrees
centigrade.
TEMPERATURE
COEFFICIENT:
+(%
of
Reading
+
Number
of
Counts)/°C
6
1/2
Digits
Displayed
RANGE
|
Auto-Zero
ON
Auto-Zero
OFF*
300
uA
002
+
20
-002
+
30
3
mA
-002
+
20
-002
+
30
30
mA
002
+
20
-002
+
30
300
mA
007
+
20
-007
+
30
1A
007
+
20
007
+
30
*
Specifications
given
are
for
a
stable
environment
(+
1°C)
and
over
a
ten
minute
period.
Multiply
Number
of
Counts
by
0.1
for
5
1/2
digit
readings,
0.01
for
4
1/2
digit
readings
and
0.001
for
3
1/2
digit
readings.

3457A
General
Information
Table
1-1.
Specifications
DC
CURRENT
(Cont’d)
MAXIMUM
READING
RATES:
(Readings/Second)
INTEGRATION
|
DIGITS
READING
RATE
TIME
DISPLAYED
|
60
HZ
[50
Hz
0.0005
PLC
3
1/2
1350
|
1350
0.005
PLC
41/2
1250
|
1250
0.1
PLC
5
1/2
360
312
1
PLC
6
1/2
53
45
10
PLC
6
1/2
4.8
4.0
6
100
PLC
we
-48
+40
Reading
Rates
apply
with
the
keyboard
lock
ON,
Auto-Zero
OFF,
Display
OFF,
Math
Func-
tion
OFF,
Delay
Time
set
to
0.0,
Manual
ranging,
readings
stored
in
internal
memory
using
Timer
Trigger
and
single
precision
integer
format.
RESISTANCE
(2-Wire
Ohms,
4-Wire
Ohms,
Offset
Compensated
Ohms)
INPUT
CHARACTERISTICS:
MAXIMUM
RESOLUTION
CURRENT
RANGE
READING
THROUGH
(OHMS)
16
1/2
digits
digits|5
1/2
digi
4
1/2
digits|3
1/2
digits
UNKNOWN
30
30.30000
Ohm
uOhm
100
uOhm
1
mohm
10
mohm
1
mA
300
303.0000
Ohm
udhm
1
mOhm
10
mOhm
100
mOhm
1
3
3.030000KOhm
mohm
10
mOohm
mOhm
1
Ohm
1
mA
30
30.30000kKOhm
mOhm
100
mOhm
1
Ohm
10
Ohm
100
uA
303.
0000KOhm
mohm
1
Ohm
10
Ohm Ohm
10
uA
3.030000MOhm
Ohm
10
Ohm
100
Ohm
KOhm
1
uA
30.30000M0hm
Ohm
100
Ohm
1
KOhm
kKOhm
100
nA
303.
0000MOhm
Ohm
1
KOhm
10
KOhm
KOhm
100
na
3.030000GOhm
kOhm
10
KOhm
100
KOhm MOhm
100
na
On
the
Extended
Ohms
ranges
(300
MOhm
&
3
GOhm),
the
current
source
is
100
nA
in
parallel
with
10
MOhms.
INPUT
PROTECTION:
(non-destructive)
HI
sense
to
LO
sense:
+350V
peak.
HI
or
LO
to
Earth
ground:
+450V
peak.
MAXIMUM
OPEN
CIRCUIT
VOLTAGE:
30
Ohm
to
3
Megohm
ranges:
12
volts
30
Megohm
to
3
Gigaohm
ranges:
8.5
volts

General
Information
3457A
Table
1-1.
Specifications
RESISTANCE
(Cont’d)
MEASUREMENT
ACCURACY
-
FOUR-WIRE
OHMS:
+(%
of
reading
+
Number
of
Counts)
Auto-zero
ON.
24
HOUR:
Tcal
+
1°C
After
two
hour
warm-up.
Accuracy
relative
to
calibration
standard.
RANGE
|
%
READING
NUMBER
OF
COUNTS
(OHMS)
100
PLC*
10
PLC*
1
PLC
+1
PLO*
-005
pLc*
|
.0005
PLC*
6
1/2
digits|6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits|3
1/2
digits
30
-003
215
235 350
+0015
24 35
3
+001
5
6
30
-001
5
6
300
+001
6
7
8
3
+003
12
16
30
02
80
93
90
DAY:
Teal
+
5°c
After
one
hour
warm-up.
RANGE
|
%
READING
NUMBER
OF
COUNTS
(OHMS)
100
PLC*
10
PLC*
1
PLC*
-1
PLC*
+005
PLC*
-0005
PLC*
6
1/2
digits|6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits]3
1/2
digits
0.0065
315
335
450
65
0.0045
35 45
0035
7
rs
.
0035
7
7
004
8
9
.0055
14
16
025
83 93
YEAR:
Teal
+
5°C
After
one
hour
warm-up.
RANGE
|
%
READING
NUMBER
OF
COUNTS
(OHMS)
100
PLC*
10
PLC*
1
PLC*
«1
PLC*
+005
PLC*
-0005
PLC*
6
1/2
digits]6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits]3
1/2
digits
315
335
450
34
35
45
6 7
3
6
if
7
f 8
9
12 14
16
80
83
93

3457A
General
Information
Table
1-1.
Specifications
RESISTANCE
(Cont’d)
MEASUREMENT
ACCURACY
-
TWO-WIRE
OHMS:
+(%
of
reading
+
number
of
counts)
Auto-zero
ON.
24
HOUR:
Teal
+
1°C
After
two
hour
warm-up.
RANGE
|
%
READING
NUMBER
OF
COUNTS
OHMS)
100
PLC*
10
PLC*
1
PLC*
»
3:
PLOS
005
PLC*
-0005
PLC*
6
1/2
digits|6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits
|3
1/2
digits
it
.003
20215 20235
20350
20055
20017
20006
0015
2024
2025
2035
2006
2004
2004
001
205
206
206
204
204 204
.001
25
26 26
24
24
24
-001
8
9
10
6
6
003
12
14 16
7 5
0.02
80 83 93
14
6
0.6
1000 1000 1000
100
10
6.0
1000
1000
1000
100
10
*
specifications
apply
for
a
stable
environment
(+1°C)
and
after
Ohms
Auto-Cal
(Auto-Cal
3).
90
DAY:
Tcal
+
5°C
After
one
hour
warm-up.
RANGE
|
%
READING
NUMBER
OF
COUNTS
(OHMS)
100
PLC*
10
PLC*
1
PLC*
a1
PLG*
+005
PLC*
+0005
PLC*
6
1/2
digits|6
1/2
digits|6
1/2
digits|5
1/2
digits|4
1/2
digits]3
1/2
digits
20315
20335
20450
20065 20018
20006
2034
2035 2045
2008 2004 2004
206
207
207
204
204
204
26 27
27
24 24 24
9
10
+
6 6
12
14
16
, 5
80
83
93
14
6
1000
10
1000
10

General
Information
3457A
Table
1-1.
Specifications
RESISTANCE
(Cont’d)
1
YEAR:
Teal
+
5°C
After
one
hour
warm-up.
RANGE
|
%
READING
NUMBER
OF
COUNTS
(OHMS)
100
PLC*
10
PLC*
1
PLc*
=
Puc¥
-005
PLC
-0005
PLC*
6
1/2
digits|6
1/2
digits|6
1/2
digits|5
1/2
digits]4
1/2
digits|3
1/2
digits
20315
20335
20450
20065
20018
20006
2034
2035
2045 2008
2004
2004
206 207
204 204 204
26 27
24 24 24
9
1
6 6 6
12
16
.
5 5
80
93
14
6
5
1000 1000
100
10
1
1000 1000
1000
100
10
1
Integration
Time
in
Power
Line
Cycles.
When
using
integration
times
of
100
PLC,
10
PLC,
or
1PLC;
multiply
Number
of
Counts
in
the
Measurement
Accuracy
Tables
by
0.1
for
5
1/2
digit
readings,
0.01
for
4
1/2
digit
readings
and
0.001
for
3 1/2
digit
readings.
Tcal
=
the
temperature
of
the
environment
where
the
3457A
was
calibrated.
Calibration
should
be
performed
with
the
environment
temperature
between
18
and
28
degrees
centigrade.
TEMPERATURE
COEFFICIENT:
*(%
of
Reading
+
Number
of
Counts)/°C
6
1/2
Digits
Displayed
RANGE
|
Auto-Zero
ON
Auto-Zero
OFF*
30
0005
50
-0005
+
510
300
.0005
5.0
+0005
+
60
3
0005
0.5
-0005
+
15
30
0005
0.5
0005
+
15
-0008
+
0.5
0008
+
15
0010
0.5
+0010
+
15
~0025
0.5
-0025
+
15
-3500
0.0
+3500
+
10
3.5000
+
0.0
3.5000
+
10
*
Specifications
given
are
for
a
stable
environment
(+
1°C)
and over
a
ten
minute
period.
For
integration
times
of
>=
1
PLC,
multiply
Number
of
Counts
by
0.1
for
5
1/2
digit
readings,
0.01
for
4
1/2
digit
readings
and
0.001
for
3
1/2
digit
readings.

3457A
General
Information
Table
1-1.
Specifications
RESISTANCE
(Cont’d)
RANGE
|
MAXIMUM
LEAD
MAXIMUM
OFFSET
|
PRE-PROGRAMMED
(OHMS)
|
RESISTANCE
VOLTAGE
FOR
OFFSET]
SETTLING
TIME
FOUR-WIRE
OHMS
|
COMPENSATED
OHMS
(FULL
SCALE)
1
ohm
1
mv
10
Ohm
10
mv
100
ohm
100
mv
1
KOhm
applicable
10
KOhm
applicable
100
KOhm
applicable
1
Mohm
applicable
not
applicable applicable
not
applicable applicable
RESPONSE
TIME:
First
reading
meets
accuracy
specification
with
pre-programmed
settling
times
and
<200
pF
circuit
capacitance.
An
additional
delay
of
0.1
seconds
is
necessary
after
a
range
or
function
change
to
meet
rated
accuracy.
MAXIMUM
READING
RATES:
(Readings/Second)
30
Ohm
to
30
KOhm
ranges
INTEGRATION
|
DIGITS
READING
RATE
TIME
DISPLAYED
|60
HZ
|50
Hz
0.0005
PLC
3
1/2
1350 1350
0.005
PLC
41/2
1250 1250
0.1
PLC
5
1/2
360
312
1
PLC
6
1/2
53 45
10
PLC
6
1/2
4.8
4.0
100
PLC
6
1/2
48
-40
Reading
Rates
apply
with
the
keyboard
lock
ON,
Auto-Zero
OFF,
Display
OFF,
Math
Func-
tion
OFF,
Delay
Time
set
to
0.0,
Manual
ranging,
readings
stored
in
internal
memory
using
Timer
Trigger
and
single
precision
integer
format.

General
Information
3457A
1-12
Table
1-1.
Specifications
TRUE
RMS
AC
VOLTAGE
(AC
and
DC
Coupled)
INPUT
CHARACTERISTICS:
MAXIMUM
RESOLUTION
READING
RANGE
|6
1/2
digits|6
1/2
diyits|5
1/2
digits]4
1/2
digits|3
1/2
digits
30
mv
|
32.50000
mv
10
nv
100
nv
1
uv
10
uv
300
mv}
325.0000
mv
100
nv
1
uv
10
uv
100
uv
3
3.250000
v
Tw
10
uv
100
uv
1m
30
V
32.50000
v 10
uv
100
uv
1
mv
10
mv
300
Vv
303.0000
v
100
uv
1
mv
10
mv
100
mv
INPUT
IMPEDANCE
:
1
megohm
1%
shunted
by
<
90
pF
(¢
<
115
pF
rear
input).
MAXIMUM
INPUT
VOLTAGE:
(non-destructive)
Input
Terminals:
+
450V
peak.
HI
or
LO
to
Earth
Ground:
+
450V
peak.
CREST
FACTOR:
3.5
to
1
at
full-scale.
MEASUREMENT
ACCURACY
-
AC
COUPLED:
+(%
of
reading
+
number
of
counts)
Specifications
apply
for
sine-wave
inputs
>
10%
of
range
with
DC
component
<
10%
of
the
AC
component,
with
Auto-zero
ON,
AC
slow
filter
ON
(ACBAND
20),
and
after
2
hour
warm-up.
For
AC
inputs
between
3%
and
10%
of
full-scale
and
<
20KHz,
add
0.7%
to
the
percent
of
read-
ing
figures.
For
pre-programmed
settling
times,
add
.1%
of
input
voltage
step
to
accuracy
specifications
for the
first
reading.
HOUR:
Tcal
+
1°
(ACV
Function)
After
2
hour
warm
up
Accuracy
relative
to
calibration
standard
30
mV
to
30
V
ranges.
FREQUENCY
PERCENT
NUMBER
OF
COUNTS
OF
>=
1
PLC
0.1
PLC
+005
PLC
-0005
PLC
READING
|6
1/2
digits]5
1/2
digits|4
1/2
digits]3
1/2
digits
20Hz2-45Hz
-
720
76
12
45Hz-100Hz
720
76 12
100Hz2-20KHz
720
76
ce
**
400Hz-20KHz
a
720
55
20kiz-100KHz
bs
1700
23
100KHz-300KHz
=
9300
98
300KHz-
IMHz
66000
664

3457A
General
Information
Table
1-1.
Specifications
TRUE
RMS
AC
VOLTAGE
(Cont’d)
24
HOUR:
Teal
+
1°C
(ACV
Function)
After
2
hour
warm-up
Accuracy
relative
to
calibration
standard
300
V
range.
FREQUENCY
PERCENT
NUMBER
OF
COUNTS
oF
>=
1
PLC
0.1
PLC
-005
PLC
-0005
PLC
READING
|6
1/2
digits|5
1/2
digits|4
1/2
digits|3
1/2
digits
20Hz-45Hz
0.56
720
76
12
45Hz-
100Hz
0.21
720
76
12
1OOHz-20KiIz
0.13
720
76
12
**
GOOHZ-20KHz
0.14
720
510
55
20KNz-
100KHz
1.0
3300
334
38
90
DAY:
Teal
+
5°C
(ACV
Function)
After
1
hour
warm-up
30
mv
to
30
V
ranges.
FREQUENCY
*
PERCENT
NUMBER
OF
COUNTS
OF
>=
1
PLO
0.1
PLC
-005
PLC
-0005
PLC
READING
|6
1/2
digits]5
1/2
digits|4
1/2
digits]3
1/2
digits
20Hz2-45Hz
0.56
1120
116
16
45Hz-100Hz
0.21 1120
116
16
100H2-20KHz
0.13
1120
116
16
**
4OOHZ-20KHz
0.14
1120
550
59
20KHz-100KHz]
0.66 2100
224
27
100KHz-300KHz|
3.16
9700
974
102
30OKH2-
IMHz
10.16
66400
6640
668
90
DAY:
Teal
+
5°c
(ACV
Function)
After
1
hour
warm-up
300
V
range.
FREQUENCY
*
PERCENT
NUMBER
OF
COUNTS
OF
>=
1
PLC
0.1
PLC
-005
PLC
-0005
PLC
READING
|6
1/2
digits]5
1/2
digits|4
1/2
digits]3
1/2
digits
20Hz-45Hz
0.62
1120
116
16
45Hz-100Hz
0.27
1120
116
16
100Hz-20KHz
0.19
1120
116
16
**
400HZ-20KHz
0.2
1120
550
59
20KHz-100Kiiz}
1.06
3700
374
42
*
Specifications
apply
within
one
week
of AC
auto-cal
(ACAL
2)
for
stable
temperature
conditions.
(Teal
£5°C).
If
AC
auto-cal
is
not
used,
add
0.6
to
the
percent
of
reading
figures.
**
Using
AC
fast
filter
(ACBAND
400)
for
frequencies
above
400Hz.
For
1
year
specifications,
add
.1%
to
24
hour
percent
of
reading
figures,
add 600
to
6
1/2
digit
counts,
add
60
to
5
1/2
digit
counts,
add
6
to
4
1/2
digit
counts,
add
.6
to
3
1/2
digit
counts
(Teat
+5°c).
1-13
Other manuals for 3457A
3
Table of contents
Other HP Multimeter manuals