HT HT9020 User manual

Copyright HT ITALIA 2016 Release 1.02 - 01/03/2016
HT9020
User manual


HT9020
EN - 1
Table of contents:
1.PRECAUTIONS AND SAFETY MEASURES ........................................................................... 2
1.1.Preliminary instructions .................................................................................................................... 2
1.2.During use ........................................................................................................................................ 3
1.3.After use ........................................................................................................................................... 3
1.4.Definition of measurement (overvoltage) category .......................................................................... 3
2.GENERAL DESCRIPTION ....................................................................................................... 4
2.1.Measuring average values and TRMS values ................................................................................. 4
2.2.Definition of true root mean square value and crest factor .............................................................. 4
3.PREPARATION FOR USE ....................................................................................................... 5
3.1.Initial checks..................................................................................................................................... 5
3.2.Instrument power supply .................................................................................................................. 5
3.3.Calibration ........................................................................................................................................ 5
3.4.Storage............................................................................................................................................. 5
4.OPERATING INSTRUCTIONS ................................................................................................. 6
4.1.Instrument description...................................................................................................................... 6
4.1.1.Description of the controls ................................................................................................................. 6
4.1.2.Alignment marks ................................................................................................................................ 6
4.1.3.Hand protection ................................................................................................................................. 7
4.1.4.Indication of the conventional direction of Current............................................................................. 7
4.2.Function keys description................................................................................................................. 8
4.2.1.F1, F2, F3, F4/OK keys ..................................................................................................................... 8
4.2.2.H/ESC/ key .................................................................................................................................... 8
4.3.Initial screen ..................................................................................................................................... 8
5.OPERATING INSTRUCTIONS ................................................................................................. 9
5.1.Instrument settings ........................................................................................................................... 9
5.2.AC Voltage detection........................................................................................................................ 9
5.3.DC Voltage measurement.............................................................................................................. 10
5.4.AC/AC + DC Voltage measurement............................................................................................... 12
5.4.1.Voltage Harmonics measurement ................................................................................................... 14
5.4.2.Phase Sequence and Phase Conformity ......................................................................................... 15
5.5.DC Current measurement .............................................................................................................. 19
5.6.AC/AC + DC Current measurement ............................................................................................... 21
5.6.1.Current Harmonics measurement.................................................................................................... 23
5.6.2.Dynamic Inrush current measurement............................................................................................. 24
5.7.DC Power and Energy measurement ............................................................................................ 26
5.8.AC/AC+DC Power and Energy measurement ............................................................................... 30
5.9.Resistance and Continuity test measurement ............................................................................... 36
6.MAINTENANCE ...................................................................................................................... 39
6.1.General information........................................................................................................................ 39
6.2.Battery replacement ....................................................................................................................... 39
6.3.Cleaning the instrument ................................................................................................................. 39
6.4.End of life ....................................................................................................................................... 39
7.TECHNICAL SPECIFICATIONS............................................................................................. 40
7.1.Technical characteristics................................................................................................................ 40
7.1.1.Reference guidelines ....................................................................................................................... 42
7.1.2.General characteristics .................................................................................................................... 42
7.2.ENVIRONMENT............................................................................................................................. 42
7.2.1.Environmental conditions for use..................................................................................................... 42
7.3.Accessories provided ..................................................................................................................... 42
8.SERVICE ................................................................................................................................ 43
8.1.Warranty conditions........................................................................................................................ 43
8.2.Service ........................................................................................................................................... 43
9.APPENDIX – THEORETICAL OUTLINE ................................................................................ 44
9.1.Calculation of powers in “AC 1P” mode ......................................................................................... 44
9.2.Calculation of powers in “AC 3P” mode ......................................................................................... 44
9.3.Calculation of powers in “DC” mode .............................................................................................. 44
9.4.voltage and current harmonics....................................................................................................... 45
9.5.Limit values for harmonics ............................................................................................................. 46
9.6.Causes FOR the presence of harmonics ....................................................................................... 46

HT9020
EN - 2
1. PRECAUTIONS AND SAFETY MEASURES
The instrument has been designed in compliance with directive IEC/EN61010-1 relative to
electronic measuring instruments. For your safety and in order to avoid damaging the
instrument, please carefully follow the procedures described in this manual and read all
notes preceded by the symbol paying the utmost attention.
Before and after carrying out measurements, carefully observe the following instructions:
Do not carry out any voltage or current measurement in humid environments
Do not carry out any measurement in case of gas, explosive and inflammable materials
or dusty environments
Avoid contact with the circuit under test if no measurement is carried out
Avoid contact with exposed metal parts, with unused measuring probes, circuits, etc.
Do not carry out any measurement in case of instrument’s anomalies such as
deformation, breaks, substance leaks, absence of displayed screen, etc.
Pay special attention when measuring voltages higher than 20V, since a risk of
electrical shock exists
The following symbols are used in this manual and on the instrument:
CAUTION: observe the instructions given in this manual; an improper use could
damage the instrument or its components.
High voltage danger: electrical shock hazard.
Double-insulated meter
AC voltage or current
DC voltage or current
Connection to earth
1.1. PRELIMINARY INSTRUCTIONS
This clamp has been designed for use in environments of pollution degree 2.
It can be used for CURRENT and VOLTAGE measurements on installations with
measurement category CAT IV 600V and CAT III 1000V. For a definition of
measurement categories, see § 1.4.
We recommend to follow the standard safety rules devised by the procedures for
carrying out operations on live systems and using the prescribed PPE to protect the
user against dangerous currents and the instrument against incorrect use.
Only the leads supplied with the instrument guarantee compliance with the safety
standards. They must be under good conditions and replaced with identical models,
when necessary.
Do not test circuits exceeding the specified current and voltage limits.
Check that the battery is correctly inserted.
Before connecting the test leads to the circuit under test, make sure that the switch is
correctly set.
Make sure that the LCD display and the switch indicate the same function.

HT9020
EN - 3
1.2. DURING USE
Please carefully read the following recommendations and instructions:
CAUTION
Failure to comply with the Caution notes and/or Instructions may damage
the instrument and/or its components or be a source of danger for the
operator.
Before activating the switch, remove the conductor from the clamp jaw or disconnect
the test leads from the circuit under test.
When the instrument is connected to the circuit, do not touch any unused terminal.
Keep your hands always under the hand protection. This protection is always located in
a suitable position to guarantee a correct safety distance from possible exposed or live
parts (see Fig. 3)
Avoid measuring resistance if external voltages are present. Even if the instrument is
protected, excessive voltage could cause a malfunction of the clamp.
During current measurement, any other current near the clamp may affect
measurement accuracy.
When measuring current, always put the conductor as close as possible to the middle
of the clamp jaw, to obtain the most accurate reading.
While measuring, if the value or the sign of the quantity under test remain unchanged,
check if the HOLD function is enabled.
1.3. AFTER USE
When measurement is complete, switch OFF the instrument.
If the instrument is not to be used for a long time, remove the batteries
1.4. DEFINITION OF MEASUREMENT (OVERVOLTAGE) CATEGORY
Standard IEC/EN61010-1: “Safety requirements for electrical equipment for measurement,
control and laboratory use, Part 1: General requirements” defines measurement category,
commonly called overvoltage category. In § 6.7.4: Measured circuits, circuits are divided
into the following measurement categories:
(OMISSIS)
Measurement category IV is for measurements performed at the source of the low-
voltage installation.
Examples are electricity meters and measurements on primary overcurrent protection devices
and ripple control units.
Measurement category III is for measurements performed on installations inside
buildings.
Examples are measurements on distribution boards, circuit breakers, wiring, including cables,
bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for
industrial use and some other equipment, for example, stationary motors with permanent
connection to fixed installations.
Measurement category II is for measurements performed on circuits directly
connected to the low-voltage installation.
Examples are measurements on household appliances, portable tools and similar equipment.
Measurement category I is for measurements performed on circuits not directly
connected to MAINS.
Examples are measurements on circuits not derived from MAINS, and specially protected
(internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that
reason, the standard requires that the transient withstand capability of the equipment is made
known to the user.

HT9020
EN - 4
2. GENERAL DESCRIPTION
The HT9020 instrument carries out the following measurements:
DC and AC+DC TRMS voltage
DC and AC+CD TRMS current
Phase sequence and conformity test
AC powers and power factor on single-phase and/or balanced three phase systems
AC energies on single-phase and/or balanced three-phase systems
DC power
AC voltage harmonics up to 25° order and THD%
AC current harmonics up to 25° order and THD%
Frequency on voltage (with test leads) and current (with clamp jaw)
Resistance and continuity test
Electric motor starting currents (Dynamic Inrush)
Detection of AC voltage with and without contact with built-in sensor
Each of these functions can be selected using the 6-position selector switch, including
OFF position. Keys F1, F2, F3, F4/OK and H / ESC / are also provided. For their use,
please refer to § 4.2.
2.1. MEASURING AVERAGE VALUES AND TRMS VALUES
Measuring instruments of alternating quantities are divided into two big families:
AVERAGE-VALUE meters: instruments measuring the value of the single wave signals
TRMS (True Root Mean Square) VALUE meters: instruments measuring the TRMS
value of the quantity being tested
In the presence of a perfectly sinusoidal wave, both families of instruments provide
identical results. In the presence of distorted waves, on the other hand, the readings shall
differ. Average-value meters provide the RMS value of the sole fundamental wave, TRSM
meters, instead, provide the RMS value of the whole wave, including harmonics (within the
instrument’s bandwidth)
2.2. DEFINITION OF TRUE ROOT MEAN SQUARE VALUE AND CREST FACTOR
The root mean square value of current is defined as follows: "In a time equal to a period,
an alternating current with a root mean square value of the intensity of 1A, circulating on a
resistor, dissipates the same energy as that which would have been dissipated by a direct
current with the intensity of 1 A during the same time ". This definition results in the
numeric expression:
G=
Tt
t
dttg
T
0
0
)(
12The root mean square value is indicated with the acronym RMS. The
Crest Factor is defined as the relationship between the Peak Value of a signal and its RMS
value: CF (G)=
RMS
p
G
GThis value changes with the signal waveform, for a purely sinusoidal
wave it is 2=1.41. In case of distortion, the Crest Factor takes higher values as wave
distortion increases.

HT9020
EN - 5
3. PREPARATION FOR USE
3.1. INITIAL CHECKS
Before shipping, the instrument has been checked from an electric as well as mechanical
point of view. All possible precautions have been taken so that the instrument is delivered
undamaged.
However, we recommend generally checking the instrument in order to detect possible
damage suffered during transport. In case anomalies are found, immediately contact the
forwarding agent.
We also recommend to check whether the package contains all components indicated in
§ 7.3. In case of discrepancy, please contact the Dealer.
In case the instrument should be replaced, please carefully follow the instructions given in
chapter 8.2.
3.2. INSTRUMENT POWER SUPPLY
The instrument is supplied by 2x1.5V LR03 AAA batteries. Replace them following the
instructions in § 5.2.
3.3. CALIBRATION
The instrument has the technical specifications described in this manual. The instrument’s
performance is guaranteed for one year.
3.4. STORAGE
In order to guarantee accurate measurements, after a long storage time under extreme
environmental conditions, wait for the instrument to come back to normal condition (see §
7.2.1)

HT9020
EN - 6
4. OPERATING INSTRUCTIONS
4.1. INSTRUMENT DESCRIPTION
4.1.1. Description of the controls
CAPTION:
1. Inductive clamp jaw
2. AC voltage indicator LED
3. Jaw trigger
4. Rotary selector switch
5. H/ESC/ key
6. F1,F2,F3,F4/OK function
keys
7. LCD display
8. Input terminal V
9. Input terminal COM
Fig. 1: Instrument description
4.1.2. Alignment marks
Put the conductor as close as possible to the middle of the jaws on the intersection of the
indicated marks (see Fig. 2) in order to meet the meter accuracy specifications.
CAPTION:
1. Alignment marks
2. Conductor
Fig. 2: Alignment marks

HT9020
EN - 7
4.1.3. Hand protection
CAPTION:
1. Hand protection
2. Safe area
Fig. 3: hand protection
Always keep your hands under the hand protection. This protection is always located in a
suitable position to guarantee a correct safety distance from possible exposed or live parts
(see Fig. 3)
4.1.4. Indication of the conventional direction of Current
The Fig. 4 shows an arrow which indicates the conventional direction of current
Fig. 4: Current direction arrow

HT9020
EN - 8
4.2. FUNCTION KEYS DESCRIPTION
4.2.1. F1, F2, F3, F4/OK keys
The F1, F2, F3, F4/OK keys perform different functions according to the measurement set
(for detailed information, see the single functions).
4.2.2. H/ESC/ key
A single press activates the Data HOLD function and the value of the measurement
quantity is frozen at display. The symbol "H" is displayed when this function is enabled.
This operating mode is disabled when “H” key is pressed again or the switch is operated.
To improve the readability of the values measured in dark places, the display has been
provided with a backlight function which is turned on and off by long-pressing “H” key. If
the feature is set in MAN mode (see § 5.1) the backlight deactivates after approximately
30 seconds after its activation, in order to save battery life.
The same key identify the ESC (Exit) functionality inside the different modes of the
instrument.
4.3. INITIAL SCREEN
When switching on the instrument, the initial screen appears for
a few seconds. It shows:
The instrument’s model
The instrument’s serial number
The instrument’s firmware version
HT9020
Sn 15120020
V. 2.00
CAUTION
Please note down this information, especially the firmware version, in case
it should be necessary to contact the service department.
After a few seconds, the instrument switches to the selected function.

HT9020
EN - 9
5. OPERATING INSTRUCTIONS
5.1. INSTRUMENT SETTINGS
By positioning the selector switch to “Settings”, the screen aside
will appear, containing the possible settings of the instrument.
Press F1 (Sel) key to see the different selections. Press F2, F3
(, ) keys to modify the settings of the selected items and F4
(OK) key to confirm the selections.
Sel OK
Language:
English
Auto-Off:
OFF
Backlight:
MAN
Language
In the “Language” section it is possible to set the system language. Press F2, F3 (, )
keys for the selection of the available languages and F4 (OK) key to confirm the selected
item. The “Saved data” message is displayed for a while in the bottom part of display.
Auto - Off
In the “Auto - Off” section it is possible to activate/deactivate the auto power off feature.
Press F2, F3 (, ) keys for the selection of the “ON” or “OFF” options and F4 (OK) key
to confirm the selected item. The “Saved data” message is displayed for a while in the
bottom part of display.
The “ ”symbol is displayed with auto power off feature activated and the instrument
switches off after approx.. 5 minutes of idleness.
Backlight
In the “Backlight” section it is possible to select the activation mode of display backlight.
Press F2, F3 (, ) keys for the selection of the “MAN” option (backlight manually
activated by pressing “H” key and disable after approx. 30 seconds) or “ON” (backlight
always active) and F4 (OK) key to confirm the selected item. The “Saved data” message is
displayed for a while in the bottom part of display.
The “ON” option can result a significant reduction of the battery life.
5.2. AC VOLTAGE DETECTION
With the selector switch set to “V” by taking the end of the clamp jaw near an AC source,
the red LED at the base of the clamp jaw will turn on (see Fig. 1 – part 2), which indicates
that voltage is present.
CAUTION
Phase detection is active only when the clamp selector switch is set to “V”
position

HT9020
EN - 10
5.3. DC VOLTAGE MEASUREMENT
CAUTION
The maximum DC or AC+DC input voltage is 1000V. When the display
shows “> 999.9V”, it means that the maximum value that clamp is capable
of measuring has been exceeded. Exceeding these limits could result in
electrical shocks to the user and damage to the instrument.
Fig. 5: DC Voltage measurement
1. By positioning the selector switch to “V”, the screen aside
will appear
Mod Har Fnc
AC <42.5 Hz
---- V
2. Press F1 (Mod) key to open the drop-down menu shown on
the screen nearby and select the “DC” option with the same
key
3. Press the F4 (OK) to confirm
Mod Har Fnc OK
AC <42.5 Hz
---- V
AC
DC
Ph Seq
Help

HT9020
EN - 11
4. Connect red cable to input lead Vand black cable to input lead COM then position
the leads at the desired points of the circuit under test (see Fig. 5)
5. The screen shows an example of DC Voltage measurement. Mod Fnc
DC
9.1 V
6. Press F3 (Fnc) key to open the drop-down menu shown on
the screen nearby. At each subsequent pressure of F3 key,
the cursor will scroll through the available items, as follows:
Max: it constantly displays the maximum value of the
measured DC Voltage
Min: it constantly displays the minimum value of the
measured DC Voltage
Cr+: it constantly displays the maximum positive crest
value
Cr-: it constantly displays the minimum negative crest
value
RST: (RESET) it deletes all stored Max, Min, Cr+ and Cr-
values and re-start with a new measure
ESC: it goes back from Max/Min/Cr+/Cr- and return to
normal measuring mode
Mod Fnc OK
DC
9. V
7. By pressing F4 (OK), the selected item is confirmed. Nearby
an example of measurement with active Max function. The
“Max” symbol indicates the active function
Mod Fnc
Max DC
12.0 V
8. For the use of HOLD and backlight feature see § 5.1
CAUTION
The measurement of the 4 Max, Min, Cr+ and Cr- values is simultaneous,
regardless of the displayed value.
Max
Min
Cr+
Cr-
RST
Esc

HT9020
EN - 12
5.4. AC/AC + DC VOLTAGE MEASUREMENT
CAUTION
The maximum AC/AC+DC input voltage is 1000V. When the display shows
“> 999.9V”, it means that the maximum value which clamp is capable of
measuring has been exceeded. Exceeding these limits could result in
electrical shocks to the user and damage to the instrument.
Fig. 6: AC/AC + DC voltage measurement
1. By positioning the selector switch to “V”, the screen nearby
will appear
Mod Har Fnc
AC <42.5 Hz
---- V
2. Press F1 (Mod) to open the drop-down menu shown on the
screen aside and select the “AC” option with the same key
3. Press F4 (OK) to confirm
Modd Har Fnc OK
AC <42.5 Hz
---- V
AC
DC
Ph Seq
Help

HT9020
EN - 13
4. Connect red cable to input lead Vand black cable to input lead COM then position
the leads to the desired points of the circuit under test (see Fig. 6)
5. The screen shows an example of AC voltage measurement.
The instrument allows the evaluation of possible DC
components overlapped on a generic alternate waveform
(AC+DC) signal and this can be very useful for the
measurements on impulsive signals typically of no-linear
loads (e.g: welders, electric ovens, etc)
Mod Har Fnc
AC 50.0 Hz
230.1 V
6. Press F3 (Fnc) to open the drop-down menu shown on the
screen aside. At each subsequent pressure of key F3, the
cursor will scroll through the available items, as follows:
Max: it constantly displays the maximum value of the
measured AC+DC Voltage
Min: it constantly displays the minimum value of the
measured AC+DC Voltage
Cr+: it constantly displays the maximum positive crest
value
Cr-: it constantly displays the minimum negative crest
value
RST: (RESET) it deletes all stored Max, Min, Cr+ and Cr-
values and re-start with a new measure
Esc: it goes back from Max/Min/Cr+/Cr- and return to
normal measuring mode
Mod Har Fnc OK
AC
23. V
CAUTION
Note: the measurement of the 4 Max, Min, Cr+ and Cr- values is
simultaneous, regardless of the displayed value.
7. By pressing F4 (OK), the selected item is confirmed. Nearby,
an example of measurement with active Max function.
Mod Har Fnc
Max AC 50.0 Hz
231.5 V
8. For the use of HOLD and backlight features see § 5.1
Max
Min
Cr+
Cr-
RST
Esc

HT9020
EN - 14
5.4.1. Voltage Harmonics measurement
1. Press F2 (Har) key to select the screen of voltage harmonics
as shown nearby. Press again F2 (RMS) to go back to
voltage measurement screen
Mod Har Fnc OK
AC 50.0 Hz
220.5 V
2. By pressing F1 (◄) or F4 (►), it is possible to move the
cursor over the graph and select the harmonic to be
measured. The correspondent absolute or percentage value
of harmonic voltage is shown. It is possible to measure up to
the 25th harmonic
◄RMS Fnc ►
H01 215.0 V
ThdV 10.0 %
3. While measuring Voltage Harmonics, press F3 (Fnc) to open
the drop-down menu shown on the screen aside. At each
subsequent pressure of F3, the cursor will scroll through the
available items, as follows:
Max: it constantly displays the maximum RMS value of
the selected current harmonic
Min: it constantly displays the minimum RMS value of the
selected current harmonic
Abs: it displays the absolute value of the harmonics in
Volts
%: it displays the value of the harmonics as percentage
value with respect to the fundamental
RST (RESET) it deletes all stored Max, Min values and
re-start with a new measure
Esc: it goes back to a normal measuring mode
◄Par Fnc
H01 215.0 V
ThdV 10.0 %
CAUTION
Since the menu contains functions with a different meaning (Max-Min and
Abs-%), it is necessary to enter the menu twice: first for displaying Abs or
% values and second time to enable the Max or Min functions
4. By pressing F4 (OK), the selected item is confirmed. Nearby,
an example of measurement with active Max function. The
display shows the active function.
◄RMS Fnc ►
H01 215.0 V
ThdV 10.0 %
Max
5. For the use of HOLD and backlight features see § 5.1
Max
Min
Abs
%
RST
Esc

HT9020
EN - 15
5.4.2. Phase Sequence and Phase Conformity
CAUTION
While measuring, the instrument must be held in the operator’s hand.
Phase sequence test
Fig. 7: Verification of phase sequence
1. Press F1 (Mod) to open the drop-down menu shown on the
screen nearby and select the “Ph Seq” option with the same
key
2. Press F4 (OK) to confirm
Mod OK
AC <42.5 Hz
---- V
3. The instrument shows the “PH1” message and waits for the
detection of L1 phase
4. Connect red cable to input lead Vand black cable to
input lead COM then position the leads respectively to the L1
phase and the ground reference PE of the circuit under test
(see Fig. 7).
Mod
Ph Seq
PH1
Wait
AC
DC
Ph Seq
Help

HT9020
EN - 16
CAUTION
If the frequency of the measured voltage is lower than 42.5Hz or higher than
69Hz, the display shows the message “F<42.5 Hz” or “F>69 Hz” and phase
detection does not start.
5. When a voltage higher than or equal to 100V is detected, the
instrument emits a sound signal (buzzer) and the message
“Meas” is displayed. Do not press any key and keep the test
lead connected to L1 phase cable.
Mod
Ph Seq
PH1
Meas
6. Once phase L1 acquisition is complete, the instrument stops
the acoustic signal and the “Discon.” Message is shown.
Disconnect the test lead from phase L1 cable.
Mod
Ph Seq
Discon.
Wait
7. The message “PH2“ is shown and the instrument waits for
the detection of L2 phase. Connect the test lead to phase L2
cable (see Fig. 7)
Mod
Ph Seq
PH2
Wait
CAUTION
If more than 3 seconds elapse before detecting phase L2, the instrument
displays the message “Time Out”. It is necessary to repeat the measuring
cycle from the beginning, by pressing F3 (New) and starting again from point
3.
8. When a voltage higher than or equal to 100V is detected, the
instrument emits a sound signal (buzzer) and the message
“Meas” is displayed. Do not press any key and keep the test
lead connected to L2 phase cable
Mod
Ph Seq
PH2
Meas

HT9020
EN - 17
9. If the two phases, to which the test lead has been connected,
are in the correct sequence, the message “123”is displayed.
If the phase sequence is incorrect, the message “132” is
displayed
10.To start a new measurement, press F3 (New)
Mod New
Ph Seq
123
Phase conformity test
CAUTION
While measuring, the instrument must be held in the operator’s hand.
Fig. 8: Verification of phase conformity
1. The instrument shows the screen nearby, and waits for the
detection of L1 phase of the first system
2. Connect the red cable to the input lead Vand the black
cable to the input lead COM then position the leads
respectively to the L1 phase of the first system and the
ground reference PE of the circuit under test (see Fig. 8)
Mod
Ph Seq
PH1
Wait

HT9020
EN - 18
3. When a voltage higher than or equal to 100V is detected, the
instrument emits a sound signal (buzzer) and the message
“Meas” is displayed. Do not press any key and keep the test
lead connected to L1 phase cable of the first system
Mod
Ph Seq
PH1
Meas
4. Once the voltage of L1 phase acquisition is complete, the
instrument stops the sound signal and the “Discon.”
Message is displayed. Disconnect the test lead from L1
phase of the first system.
Mod
Ph Seq
Discon.
Wait
5. The message “PH2“ is displayed and the instrument waits
for the detection of L1 phase of the second system. Connect
the test lead to L1 phase of the second system.
Mod
Ph Seq
PH2
Wait
CAUTION
If more than 3 seconds elapse before detecting the phase L1 of the second
sequence, the instrument displays the message “Time Out”. It is necessary
to repeat the measuring cycle from the beginning, by pressing F3 (New) key
and starting again from step 1.
6. When a voltage higher than or equal to 100V is detected, the
instrument emits a sound signal (buzzer) and the message
“Meas” is displayed. Do not press any key and keep the test
lead connected to L1 phase cable of the second system
Mod
Ph Seq
PH2
Meas
7. If there is correct conformity between the two phases, to
which the test lead has been connected, the message “11-“
is displayed. If not, the messages “123” or “132” are
displayed.
To start a new measurement, press F3 (New).
Mod New
Ph Seq
11-
Other manuals for HT9020
1
Table of contents
Other HT Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Toshiba
Toshiba RemotEye 4 Procedure for Installing

CS Instruments
CS Instruments LD 500 instruction manual

Kessel
Kessel SonicControl Installation, operation and maintenance instructions

Nortek
Nortek DVL 1000 300m manual

LOVATO ELECTRIC
LOVATO ELECTRIC EXP10 11 instruction manual

Pitney Bowes
Pitney Bowes SendPro C Operator's guide