13
Inspection, care and maintenance
Care of the lift cassette
The cassette should be recharged every night.
When necessary the cassette can be cleansed with a damp cloth with dishwashing
liquid.
Do NOT use any strong detergents.
If disinfecting the product DuPontTM Virkon®is to be used.
Never flush the cassette with water or other liquids.
The main switch (10) should always be OFF.
Turn the main switch (10) if the lift will not be used in a few days. Otherwise, the
battery power runs out.
Annual inspection
The lift must be inspected at least once a year and be performed by Human Care
authorized personnel. Pay particular attention to parts that show wear. Installation, service
and maintenance are to be performed by Human Care authorized personnel using original
spare parts.
Daily inspections
Check all parts of the sling before use. Loose seams and other signs of wear
on lift belts, belts and fabric may not exist. Change the sling in that case.
Check the lift cassette for any sign of visible damage
Monthly inspections
Feed out the lift belt as far as possible to check for visible damage. If the
belt is damaged, contact Human Care.
Service agreements
Human Care invites you to sign a service agreement for regular maintenance and testing.
Transportation and storage
During transport, or when the lift is not to be used for some time, the main switch shall
be off or the emergency button pushed in, otherwise the battery will discharge.
The climatic conditions should be as follows, an ambient temperature from 0 °C to 50 °C,
relative humidity from 30 % to 80 % and air pressure from 790 hPa to 1060 hPa.
Expected lifetime
The lift is expected to function for 10 000 lifting cycles with maximum load or for 10 years
with correct service and maintenance intervals, whichever comes first.
If the lift is dismantled from the rail system and then used again, it shall be cleaned,
disinfected and undergo service.
Recycling
Rejected lifts can be returned to Human Care for recycling (NOTE! contains lead (acid)
batteries). For information concerning accessories and technical documentation we refer
to our homepage (www.humancare.se).