TEST CONNECTION
Adapter Box Connections
To increase operator safety, you may elect to use an adapter box for products
terminating in either a two-prong or three-prong line cord.
Connect the black ground return lead (P/N 2100A-13) to the front panel
Return terminal and connect the other end of the lead to the dead metal
on the chassis of the DUT (Device Under Test). Check to ensure a solid
connection is made between the DUT and the return clip.
Plug the black lead from the adapter box (P/N 36544) into the CONT.
CHECK terminal located on the front panel.
Plug the white lead from the adapter box into the High Voltage terminal
located on the front panel.
Plug the line cord of the DUT into the adapter box receptacle.
WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.
Note: Be sure to enable the Ground Continuity test before moving forward. See page 4.
CONT.
CHECK
RESET
TEST
Hypot® Series
DIELECTRIC WITHSTAND TESTER
RETURN
HIGH VOLTAGE
My Menu / Select
POWER DATA
BARCODE
CAUTION
HIGHVOLTAGE
5KVAC MA X.
6KVDCMAX .
CAUTION
HIGH VOLTAGE
5KVACMAX.
6KVDC MAX.
NL
Find parts for Instrument Setup in Section 2 of your Accessory Box.
Did You Know?
The Data port on the front of the instrument can be used to
download test results directly to a USB thumb drive.
The Barcode port is compatible with a USB barcode scanner
so you can automatically scan product barcodes and run test
sequences.