Inficon TC3000L Quick guide

Translation of the original operating instructions
TC3000L
Large rigid test chamber for ELT3000
mimc95en1-01-(2112)
Catalog No.
600-101
From software version
--

INFICON GmbH
Bonner Strasse 498
50968 Cologne, Germany

INFICON Table of contents
TC3000L-Operating-instructions-mimc95en1-01-(2112) iii
Table of contents
1 About this manual ..........................................................................................................................................4
1.1 Target groups ...........................................................................................................................................4
1.2 Warnings...................................................................................................................................................4
1.3 Definition of terms.....................................................................................................................................5
2 Safety...............................................................................................................................................................6
2.1 Intended use .............................................................................................................................................6
2.2 Duties of the operator ...............................................................................................................................8
2.3 Owner requirements .................................................................................................................................8
2.4 Dangers ....................................................................................................................................................9
3 Scope of delivery, transport ........................................................................................................................11
4 Description....................................................................................................................................................12
4.1 Function ..................................................................................................................................................13
4.2 Markings on the device...........................................................................................................................13
4.3 Technical data ........................................................................................................................................14
5 Installation.....................................................................................................................................................15
6 Operation.......................................................................................................................................................18
6.1 Carrying out leak testing .........................................................................................................................20
7 Cleaning and maintenance ..........................................................................................................................21
8 Sending in the device...................................................................................................................................23
8.1 Disposal ..................................................................................................................................................23
9 Declaration of Contamination .....................................................................................................................24
10 Declarations of Conformity .........................................................................................................................25

1 | About this manual INFICON
4 / 28 TC3000L-Operating-instructions-mimc95en1-01-(2112)
1 About this manual
1.1 Target groups
This instruction manual is intended for operators and technically qualified personnel
with experience in leak detection technology and the integration of leak detectors in
leak detection systems. In addition, the installation and use of the device require
knowledge of electronic interfaces.
1.2 Warnings
DANGER
Imminent hazard resulting in death or serious injuries
WARNING
Hazardous situation resulting in potential death or serious injuries
CAUTION
Hazardous situation resulting in minor injuries
NOTICE
Hazardous situation resulting in damage to property or the environment

INFICON About this manual | 1
TC3000L-Operating-instructions-mimc95en1-01-(2112) 5 / 28
1.3 Definition of terms
Minimum detectable leak rate
The minimum detectable leak rate that can be detected by the leak detector under
ideal conditions (< 1 x 10-6 mbar l/s*).
* Helium equivalent leak rate for DMC with a pressure difference of 1000 mbar versus
0 mbar.
GCU
Gas Control Unit (basic unit, operating unit)
GDU
Gas Detection Unit
DMC
Dimethyl carbonate, typical solvent in battery electrolyte. CAS No. 616-38-6
MSDS
Material Safety Data Sheet

2 | Safety INFICON
6 / 28 TC3000L-Operating-instructions-mimc95en1-01-(2112)
2 Safety
2.1 Intended use
The test chamber is designed for leak testing of Li-ion cells and Li-ion batteries.
Uncharged cells/batteries can be tested safely. For testing charged cells/batteries,
further safety measures must be taken to avoid personal injury and damage to
property.
• For validation of the test, cells prepared as leaky are placed in the chamber. The
preparation of the cells, which creates further hazards, is performed by the
operator and is not an integral part of the test or the device. Testing of leaking
cells will result in soiling of the test chamber and possibly "contamination". Cells to
be tested with the test chamber must be pre-qualified. For this purpose,
uncharged cells are examined, for example, to determine whether they can
withstand the stresses of the leak testing. Suitable measures for the safety of man
and machine must be taken by the operator.
• The test objects must be filled with an electrolyte in which one solvent component
ideally consists of dimethyl carbonate (DMC, CAS No. 616-38-6).
• Depending on the internal structure of the test objects and the external
geometries, local mechanical stresses can occur on the test objects. This can
damage the test objects themselves, but also other components located in the test
chamber.
• Make sure that the test chamber and the outer wall of the test objects are clean.
Leak testing with charged test objects represents an additional safety risk and
should only be carried out by appropriately trained personnel and with the
installation of further safety measures.
• To avoid short circuits between the test object and chamber wall, use the insulator
included in the scope of delivery.
• An external pump may optionally be connected to the ISO-KF16 connection
according to the specifications in the documentation. The pump is not included in
the scope of delivery. The pump capacity must be at least 40 L/min up to 1000 L/
min. The optional external pump must be connected via an electrically switchable
valve, a corrugated hose and, if required, a suitable adapter.
Incorrect usage Avoid the following unintended uses:
• Testing of partially or fully charged cells and batteries without further safety
measures.
• Use outside the technical specifications, see "Technical Data".
• Testing of non-vacuum Li-ion cells, batteries or other test objects.

INFICON Safety | 2
TC3000L-Operating-instructions-mimc95en1-01-(2112) 7 / 28
• Testing of Li-Ion cells, batteries or other test objects that cannot withstand the
stresses that occur during testing. Depending on the internal structure of the test
objects and the external geometries, local mechanical stresses can occur on the
test objects. This can damage the test objects themselves, but also other
components in the test chamber.
• Testing of test objects whose current collectors can be short-circuited via the test
chamber (e.g. via the lid or other conductive parts).
• Testing of test objects that come into contact with the sealing lips of the chamber.
• Testing of wet or moist test objects.
• Testing of test objects with significant temperature differences to the environment.
• Testing of damaged test objects, batteries or other test objects.
• Testing of test objects without insulator.
• Testing of components or substances other than lithium-ion batteries.
• Testing of contaminated test objects.
• Operation of a contaminated test chamber.
• Installation and operation in potentially explosive atmospheres.
• Installation and operation in locations with very low humidity.
• Operation of the chamber by inadequately trained personnel.
• Insufficient distance between the test objects in the test chamber.
• Use beyond the technical specifications.
• Use in radioactive areas.
• Closing the test chamber while your fingers are in the rotation range of the test
chamber.
• Use of unauthorized accessories or spare parts.
• Mounting by untrained or unauthorized personnel. Only trained personnel or
INFICON employees are permitted to carry out the assembly.
• Interchanging the exhaust air ("INLET") and supply air ("VENT") lines on the
ELT3000.
• Pumping out liquids or vapors that can condense.
• Using an improperly dimensioned optional external pump.
• Use of the optional pump connection for abrupt venting.
• Testing of test objects that are too small or too light and may move uncontrollably
during venting.
• Use of tools that may damage the sealing surfaces of the test chamber during
mechanical cleaning.

2 | Safety INFICON
8 / 28 TC3000L-Operating-instructions-mimc95en1-01-(2112)
This device is not intended to be used in residential areas and cannot ensure
adequate protection of radio reception in such environments. The battery leak detector
does not perform a safety function. In the event of strong electromagnetic interference,
measured values could be falsified. It is recommended to check the function of the test
chamber regularly (e.g. with a calibration leak).
2.2 Duties of the operator
• Read, observe, and follow the information in this manual and in the work
instructions provided by the owner. This concerns in particular the safety and
warning instructions.
• Always observe the complete operating instructions for all work.
• If you have any questions about operation or maintenance that are not answered
in this manual, contact customer service.
2.3 Owner requirements
The following notes are for companies or any person who is responsible for the safety
and effective use of the product by the user, employees or third parties.
Safety-conscious operation
• Only operate the test chamber and the battery leak detector if it is in technically
perfect condition and shows no signs of damage.
• Only operate the test chamber and the battery leak detector as intended, in a
safety-conscious and hazard-conscious manner and in compliance with these
operating instructions.
• Provide ambient conditions that are suitable for operating staff, the test chamber
and the test specimen.
• Adhere to the following regulations and observe their compliance:
– Intended use
– General applicable safety and accident prevention regulations
– International, national and local standards and guidelines
– Additional device-related provisions and regulations
• Only use original parts or parts approved by the manufacturer.
• Keep this instruction manual available on site.
Personnel qualifications
• Only allow instructed personnel to work with the test chamber and the battery leak
detector. The instructed personnel must have received appropriate training. This
includes knowledge of the dangers posed by leaking electrolyte/solvent.

INFICON Safety | 2
TC3000L-Operating-instructions-mimc95en1-01-(2112) 9 / 28
• Make sure that authorized personnel have read and understood the instruction
manual and all other applicable documents.
2.4 Dangers
The measuring instrument was built according to the state-of-the-art and the
recognized safety regulations. Nevertheless, improper use may result in risk to life and
limb on the part of the user or third parties, or damage to the unit or other property
may occur.
WARNING
Danger to health due to hazardous materials and substances
Test specimens are usually filled with substances that are hazardous to health. If
these substances leak during the test, they are freely accessible to the operator after
the test.
► Wear appropriate protective clothing, especially gloves, gowns, and face shields.
► Ensure sufficient ventilation at the installation location.
► Avoid contact with skin, eyes, or clothing.
► Avoid inhaling these substances.
► Only test specimens for leak tightness that do not show any damage or smell of
electrolyte or solvent after initial inspection.
► Before removing the test specimen (visual check and smell test), check whether
any contents of the test specimen have leaked out.
► Pay attention to the risks posed by released electrolyte components and their
reaction products.
► Defective test specimens can develop gross leaks during the leak testing. In the
case of test specimens with gross leaks, observe the company's internal regulations
for handling electrolyte and the safety instructions in the safety data sheets.
► Do not pump out toxic or corrosive gases.
► Clean the device regularly and keep it clean at all times.
► Observe the safety instructions in the safety data sheets for the test objects.
► Operate the device only with a connected exhaust air connection and in well-
ventilated rooms. Alternatively, the device may be used in rooms where hazardous
substances under test are monitored.
► When nitrogen or argon is used as a purge gas, it can cause asphyxiation at
dangerous room concentrations. Suitable measures must be taken. The pressure in
the gas line to the purge gas connection must not exceed 100 mbar over
atmospheric pressure. An exhaust gas line must be connected.

2 | Safety INFICON
10 / 28 TC3000L-Operating-instructions-mimc95en1-01-(2112)
WARNING
Fire and explosion hazard
Reaction products during combustion can lead to further health risks.
► Do not operate the device unattended.
► Only operate the device with the exhaust hose connected.
► Do not pump off explosive gases.
CAUTION
Warning about hand injuries
Danger of crushing by opening the flap if the clearance is too small at the top and rear.
Danger of crushing when closing the test chamber lid in the gap between the test
chamber lid and the test chamber, or between the test chamber rings.
► Ensure that there is sufficient space where the devices are located, see also
"Setup".
► Only open and close the test chamber when your fingers are outside the test
chamber halves and outside the pivoting range of the test chamber.
► Do not touch the hinge when closing the test chamber.
Measurement inaccuracies due to dirty test chamber
Escaping electrolyte can contaminate the test chamber.
► After detecting leakage, check the test chamber for contamination caused by
escaping electrolyte.
► Avoid inhaling harmful gases or vapors.
► Keep the gasket of the test chamber clean. Do not use grease or lubricants.
► Remove coarse soiling with a dust-free cloth. This soiling can falsify the
measurement results. The device features a purging function which can be carried
out in case of minor soiling, see ELT3000 instructions "Purging the device". Use
personal protective equipment.

INFICON Scope of delivery, transport | 3
TC3000L-Operating-instructions-mimc95en1-01-(2112) 11 / 28
3 Scope of delivery, transport
Scope of delivery Item Quantity
Test chamber TC3000L 1
Operating instructions 1
Insulator 1
Inline Filter 1
"VENT" hose 1
"INLET" hose 1
Angle clips 12
►Check the scope of delivery after receipt of the product to make sure it is complete
and it has no visible damage.
Transport
NOTICE
Damage caused by transport
Transport in unsuitable packaging material can damage the device.
► Keep the original packaging.
► Only transport the device in its original packaging.

4 | Description INFICON
12 / 28 TC3000L-Operating-instructions-mimc95en1-01-(2112)
4 Description
TC3000L overview
1 2 3
5
8 7 6
4
Fig.1: TC3000L
1 "INLET" connection 5 Connection for optional auxiliary pump
2 "VENT" connection 6 Hinge (spring-mounted)
3 Proximity switch with red-green display
(for automatic measurement start)
7 Hinge (dampened)
4 Insulator 8 Hinge (spring-mounted)

INFICON Description | 4
TC3000L-Operating-instructions-mimc95en1-01-(2112) 13 / 28
4.1 Function
The test chamber, in conjunction with the Gas Control Unit of the ELT3000, makes it
possible to perform leak testing on test objects. Test objects that are not vacuum-tight
can thus be tested for leak tightness.
Place the test object in the test chamber. By pumping out the air from the test
chamber, a pressure gradient between test object and test chamber is generated.
Due to this pressure gradient, gas flows through leakages out of the test object and
into the test chamber. This gas is sent to the GDU (Gas Detection Unit) for analysis.
After the analysis, the result is compared with the setpoint. A distinguishable leak-
proof/leaky signal is output.
4.2 Markings on the device
The markings on the device have the following meanings:
Device cannot be disposed of as normal domestic waste.
Note: Only put objects that are obviously undamaged and vacuum-tight in
the device.
Warning about hand injuries
Read operating instructions

4 | Description INFICON
14 / 28 TC3000L-Operating-instructions-mimc95en1-01-(2112)
4.3 Technical data
Mechanical data
Dimensions (L × W × H) 420mm x 550mm x 200mm
Installation depth 600mm
Weight 17.0 kg
Electrical data
Operating voltage 24V DC
Power consumption 5VA
Physical data
Pressure range 1080 hPa to 1 hPa
Ambient conditions
Temperature range (°C) 10 °C to 40 °C
Relative humidity (%) 80 % at 30 °C, linear decrease to 50% at
40 °C
Height above sea level (m) 2000m
Degree of contamination II

INFICON Installation | 5
TC3000L-Operating-instructions-mimc95en1-01-(2112) 15 / 28
5 Installation
WARNING
Risk of injury from escaping electrolytes
► Make sure that you do not incorrectly swap the vent line and the inlet line when
connecting the lines on the ELT3000 battery tester.
CAUTION
Risk of injury from falling or tipping device
If the device slips off its surface, it can fall down and crush your feet.
► Only place the test chamber on a clean, horizontal, non-slip and vibration-free
surface.
► The recommended installation height of the chamber is 27.6 - 35.4 in. (70 - 90 cm).
The choice of the best ergonomic installation height is the responsibility of the
owner.
► Do not place the test chamber on the gas control unit.
► Use non-slip rubber bumpers for the device feet.
NOTICE
Property damage due to improper installation
It is recommended to have the installation carried out by INFICON or by trained
personnel.

5 | Installation INFICON
16 / 28 TC3000L-Operating-instructions-mimc95en1-01-(2112)
Installation scheme
1
2
3
4
1 "INLET" connection 3 External pump connection
(optional)
2 "VENT" connection 4 Proximity switch connection
Connecting an external pump (optional)
CAUTION
Uncontrolled movements of chamber lid or test objects
An incorrectly selected or incorrectly connected external pump can lead to hazards
from the lid or test objects, which can move uncontrollably due to excessive air flows.
► Only use an external pump that complies with the permitted technical data.
► When installing and commissioning an external pump, be sure to follow the
instructions below.
To increase the pumping capacity, you can connect an external pump to the ISO-
KF16 connection. The ISO-KF16 connection is closed with a blind flange when
delivered. To do this, connect an electrically switchable valve with a minimum nominal
diameter of DN4 to the ISO-KF16 connection. The valve must be suitable for a
pressure range of at least 0 to 1.5 bar (absolute). This enables the external pump to
be switched on and off later. When using an IO1000 module connected to the
ELT3000, the valve can be switched via one of the PLC outputs. Please note that the
valve is switched with 24V, with a maximum current consumption of 0.75 A. You can
connect a suitable pump to the valve via a corrugated hose and adapter. For this

INFICON Installation | 5
TC3000L-Operating-instructions-mimc95en1-01-(2112) 17 / 28
purpose, we recommend the use of a single-phase, dry vacuum pump with a pumping
capacity of at least 40 L/min (up to 1000 L/min). The external pump is not operated via
the ELT3000, but must be permanently operated externally.

6 | Operation INFICON
18 / 28 TC3000L-Operating-instructions-mimc95en1-01-(2112)
6 Operation
WARNING
Risk of injury from explosion
If both battery poles come into contact with the electrically conductive test chamber, a
short circuit and thus high temperatures can occur in the tested battery.
► Do not perform a battery test without the original insulator.
► Make sure that the battery connections do not touch the test chamber cover.
► Only test vacuum-tight and obviously undamaged battery cells in this test chamber.
► Make sure that there is no material in the test chamber that could damage the
battery (e.g. pointy, sharp objects).
► Leave batteries in the chamber only for the test period, and not permanently.
► Keep the test chamber clean and clean it regularly.
► No smoking.
► Keep ignition sources away from the test chamber.
WARNING
Respiratory tract irritation
The solvents in lithium-ion batteries can cause respiratory irritation and loss of
consciousness if they leak from the battery.
► To avoid irritating your airways, avoid contact with electrolytes.
► Only place vacuum-tight and obviously undamaged battery cells in this test
chamber.
WARNING
Damage to battery cells due to pointed foreign objects
Pointed foreign bodies can penetrate battery cells during the test procedure and
trigger a short circuit.
► Make sure that the test chamber is always free of pointed foreign bodies.
CAUTION
Warning about hand injuries
► Only open and close the test chamber when your fingers are outside the pivoting
range of the test chamber.

INFICON Operation | 6
TC3000L-Operating-instructions-mimc95en1-01-(2112) 19 / 28
CAUTION
Danger due to physical stress / ergonomics
Continuous opening and closing of the test chamber cover can result in fatigue of the
arm muscles.
Incorrect setup of the test chamber can impair its moving parts.
► Plan enough breaks to avoid fatigue.
► Set up the test chamber so that there is no impairment of its moving parts.
ðWhen setting up the test chamber, pay attention to the height and distance
from the operator.
ðPay attention to the arrangement of the test object trays.
► Do not lean on the handle of the test chamber.
► If a sudden change in opening or closing forces is observed while operating the
chamber, check all hinges immediately and replace if necessary.
NOTICE
Property damage due to accumulation or blockage in detachable parts
with solvents
Accumulation or blockage in detachable parts due to solvents lead to functional
problems.
► Replace the particle filter as needed or annually.
► Replace all gaskets and hoses as necessary, otherwise annually.

6 | Operation INFICON
20 / 28 TC3000L-Operating-instructions-mimc95en1-01-(2112)
6.1 Carrying out leak testing
This test chamber is used to perform leak tests on vacuum-tight and undamaged test
objects (lithium-ion batteries).
üThe test chamber and Gas Control Unit are properly connected, see also
"Installation".
üThe Gas Control Unit is switched on. The two LEDs flash simultaneously during
startup.
1Place the test object in the test chamber.
2By closing the cover you actuate the proximity switch and the measurement is
started.
ðThe air is pumped out of the test chamber.
ðAfter the analysis, the result is compared with the setpoint and shown on the
display of the Gas Control Unit.
ðThe red LED lights up when a leak greater than the configured setpoint has
been detected.
ðThe green LED lights up when a leak less than the configured setpoint has
been detected.
3You can now open the lid and remove the test object. Skin contact with
electrolytes when cleaning the test chamber or removing leak-tested test
objects should be avoided.
Table of contents
Other Inficon Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

IKA
IKA IKAMAG Maxi MR 1 operating instructions

Velp Scientifica
Velp Scientifica ARE instruction manual

Renfert
Renfert Basic Master Translation of the original instructions for use

Chopin
Chopin Alveo PC graph Quick installation and first test

Burg Wächter
Burg Wächter BWnet-10120 quick guide

Yamato
Yamato RE300 instruction manual

DDS Calorimeters
DDS Calorimeters CAL3K-F installation guide

HTL
HTL SWIFTPET manual

IKA
IKA Lab disc manual

Four E's Scientific
Four E's Scientific Ironman III operating instructions

Thermo Scientific
Thermo Scientific iCAP RQ ICP-MS Pre-installation requirements guide

SP Scientific
SP Scientific VirTis BenchTop Pro Series Operator's manual