25
9. Specifications
Product Anemomaster
Model A031, A041, A034, A044
Measuring Object Clean Air Flow
Air Velocity 0.10 to 30.0 m/s (0.00 to 9.99m/s: 0.01m/s, 10.0 to 30.0m/s: 0.1m/s)
Air Temperature -20.0 to 60.0 °C (0.1°C)
Measuring
Range
(Resolution) Pressure*1-5.00 to +5.00 kPa (0.01kPa)
Duct Size Range 0 to 2550mm (0 to 255inch)
Air Velocity ± (3% of the reading +0.1 ) m/s
Air Temperature ± 0.5 °C
Accuracy
Pressure*1± (3% of the reading +0.01) kPa
Air Velocity Approx.1sec. (at Air Velocity: 1 m/s, 90% Response)
Air Temperature Approx 30sec. (at Air Velocity: 1 m/s, 90% Response)
Response
Time Pressure*1Approx 1 sec.
Temperature Compensation
Accuracy (Velocity) ± (5% of the reading +0.1)m/s in temperature range of 5 to 60°C
Sampling Functions
- Hold the reading
- Statistical calculation (Average, Maximum, Minimum)
- Time constant setting (1, 5, 10, 20 sec)
- Remaining battery level indicator (4 levels)
- Selection of sampling units (Air velocity: m/s or ft/min; Flow rate: m3/h, m3/min
or ft3/min; Air temperature: °C or °F )
- Data storage: Max. 800 data records.
- Duct shape setting : Rectangular or Circular / Duct size unit: mm or inch
Output
Digital output: RS232C (Baud rate: 4800, 9600, 19200, 38400bps) for outputting
to printer and PC.
Analog output*2: DC 0 to 3V (Only for air velocity output)
Power Manganese AA Batteries × 6 (Alkaline or Ni-Cd batteries can be used as well)
AC Adapter: AC 100 to 240V (50/60Hz)
Battery Life Approx. 10 hours (when, Air velocity: 5m/s, Air temp: 20°C, and using Alkaline
batteries)
Main Unit 5 to 40°C
Probe -20 to 60°C
Operating
Environment Storage
Temperature 5 to 40°C
Weight Approx. 500g (Including Batteries)
Standard Accessories Carrying case, Operation manual, Manganese AA Batteries × 6, RS232C cable,
Software (for Windows), and AC adapter
Optional Accessories Analog output, Printer, AC adapter for the printer
*1: Pressure measurement is only available for A041 and A044
*2: Optional