KERN ORA 3 HA User manual

1
2
ORA_3HB_6HB-BA-d-1410
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Tel: +49-[0]7433-9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Betriebsanleitung
Analoges Refraktometer
KERN ORA 3 HBBaumé, Brix,
Feuchtigkeit Honig
ORA 6 HBHonig Refraktometer
Version 1.0
12/2014
CONTENTS
1 Technical data.................................................................................................1
2 Description......................................................................................................2
3 General information.......................................................................................3
3.1 Bestimmungsgemäße Verwendung .............................................................3
3.2 Warranty..........................................................................................................3
4 Basic safety information ...............................................................................4
4.1 Follow the instructions in the operating manual .......................................4
4.2 Warning ....................................................................................................... 4-5
5 Supplied items................................................................................................5
6 Before the first use ........................................................................................6
7 Use/measurement.........................................................................................6
7.1 Zero point calibration ................................................................................ 7-8
7.2 Additional advice............................................................................................9
7.3 Measuring procedure .............................................................................. 9-10
8 Cleaning and maintenance......................................................................... 11
9 Storage ......................................................................................................... 11
10 Service.......................................................................................................... 12
11 Disposal........................................................................................................ 12
12 Additional information................................................................................ 12
13 Brix to refractive index (nD)conversion table ......................................... 13
14 Annex............................................................................................................ 14
3 ORA_3HB_6HB-BA-e-1510
1Technical data
Models
KERN
Measuring
range
and
Scales
Scalegradua
tions
Accuracy
Dimensions
Product
Net weight
ORA3HB
5892%Brix
3843°Bé
1227%water
0.5%Brix
0.5°Bé
1%water
160x40x40mm
0.140kg
ORA6HB
1230%water
contentin
honey
0.1%
Watercon
tentinhoney
168x40x40mm
0.140kg
Example scale on the ORA 3 HB
1. Technical data
Models
KERN
Measuring
range and
Scales
Scale
graduations
Accuracy
Dimensions
Product Net weight
ORA 3 HA/HB
58-92% Brix
38-43°Bé
12-27% water
0,5% Brix
0,5°Bé
1 % water 160x40x40mm 0,140kg
ORA 6 HA/HB
12-30%
water content
in honey
0,1%
water content
in honey 168x40x40mm 0,140kg
Operating manual
Analogue refractometer
KERN & Sohn GmbH
KERN ORA 3 HA/HB Baumé, Brix,
honey moisture
ORA 6 HA/HB Honey refractometer
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Tel: +49-[0]7433-9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
GB
2. Description
1. Prism cover
2. Prism surface
3. Adjustment screw
4. Eyepiece with rubber eyeshade
5. Diopter adjustment ring
6. Optical tubes with rubber grip
Example scale on
ORA 3 HB
ORA_3HB_6HB-BA-d-1410 4
2Gerätebeschreibung
1. Prismenabdeckung
2. Prismenfläche
3. Justierschraube
4. Okular mit Gummi-Augenmuschel
5. Einstellring des Diopters
6. Optische Röhre mit Gummigriff
7. Aufbewahrungsbox
8. Pipette
9. Refraktometer
10. Destilliertes Wasser / Kalibrierflüssigkeit
11. Justierwerkzeug
12. Reinigungstuch
1
6
3
5
4
2
7
10
9
11
12
8
ORA_3HB_6HB-BA-d-1410 4
2Gerätebeschreibung
1. Prismenabdeckung
2. Prismenfläche
3. Justierschraube
4. Okular mit Gummi-Augenmuschel
5. Einstellring des Diopters
6. Optische Röhre mit Gummigriff
7. Aufbewahrungsbox
8. Pipette
9. Refraktometer
10. Destilliertes Wasser / Kalibrierflüssigkeit
11. Justierwerkzeug
12. Reinigungstuch
1
6
3
5
4
2
7
10
9
11
12
8
Version 1.1 05 / 2016
7. Storage box
8. Pipette
9. Refractometer
10. Contact liquid
(Clove oil)
11. Adjustment tool
12. Cleaning cloth
6
5
3
4
3. General information
3.1 Intended use
The refractometer is a measuring instrument for determining the refractive index of
transparent substances in the liquid or the solid state. It is used to observe the
behaviour of light as it passes from a prism with known properties to the substance
being tested.
Use of the refractometer for other purposes is contrary to its intended use and may
be hazardous. The manufacturer shall not be liable for any damages caused
by improper use.
3.2 Warranty
The warranty shall be void in the event of:
▸ Failure to observe the instructions in the operating manual
▸ Use for purposes other than those described
▸ Modifications or opening the device housing
▸ Mechanical damage and/or damage resulting from media, liquids,
natural wear and tear
4. Basic safety information
4.1 Follow the instructions in the operating manual
▸ Carefully read through the operating manual even if you have prior
experience with KERN refractometers.
▸ Every language version includes a non-authoritative translation.
The original German document is the definitive version.
4.2 Warning
▸ Do not let acids come into contact with skin or eyes. If acid comes into contact
with skin, flush with copious amounts of water. Shower if larger areas of skin are
affected.
▸ If acid comes into contact with eyes, keep the eyelid open and flush the eye with
running lukewarm water from the outer corner to the inner corner. Flush eyes for
at least 15 minutes. Then consult a doctor or ophthalmologist immediately.
▸ Thoroughly clean the refractometer aer each use.
▸ The refractometer must not be exposed to extreme temperatures, high
mechanical stresses, strong direct sunlight or high humidity.
▸ This refractometer is not a toy. Keep out of reach of children.
▸ Make sure that you will not be hit by anything else while you are using the
refractometer, as this could cause serious eye injuries
▸ The rubber eyeshade may cause irritation when in prolonged contact with the
skin. If this happens, consult your doctor.
▸ Do not touch the lenses with your fingers.
6. Before the first use
Remove the protective film (if present)from the prism surface [2]and check that
the rubber eye-cup [4]is fitted correctly.
7. Use/measurement
The refractometer can be used to quickly and accurately determine the refractive
index of transparent substances, liquid or solid ones. To ensure correct measure-
ment, the measuring device should be adjusted before measurements are carried
out. Please make sure your hands are dry before handling the measuring device.
5. Supplied items
Aer unpacking and before using the device for the first time, check that all
listed parts have been supplied. Replace damaged or faulty parts immediately
and do not put them into operation.
▸ Refractometer
▸ Storage box
▸ Pipette
▸ Adjustment tool
▸ Cleaning cloth
▸ Calibration block + Contact liquid (Clove oil)
!
6
Avoid bringing the supplied contact liquid (clove oil) into contact with the prism
cover [1]. Even the smallest amounts of clove oil will damage the material of the
cover and lead to its opacification.
Therefore, pay big attention to this aspect while carrying out the adjustment of the refractometer
described below.
If still there is accidentally contact between clove oil and the prism cover, it would not affect
subsequent measurement operations even when high opacification occurs.
Important!

7
14
8
9
10
12
11
13
7.1 Zero point calibration
▸ Hold up the refractometer against a sufficiently bright light source and look through
the eyepiece [5], holding the rubber eyeshade [4]closely against your eye/glasses.
▸ Rotate the eyepiece [5]to adjust it for your eyesight, until you can see the scale
sharply.
▸ Open the prism cover [1].
▸ Thoroughly clean the prism [2]and the underside of the prism cover [1]using a so
cloth or soft paper (with alcohol if necessary)and wipe dry.
▸ Moisten the polished surface of the calibration block with the contact fluid. Now put the
block with the moistened surface downward onto the prism surface [2].
▸ Close the prism cover [1]. The amount of fluid should be sufficient to moisten most of the
block surface. There should not be any air bubbles between the measuring prism [2] and
calibration block.
▸ Wait about 30 seconds to allow the temperatures of the fluid and prism to equalise.
▸ Look through the eyepiece [4]while pointing the refractometer‘s prism surface [2]
at a bright light source.
▸ Through the eyepiece [4], you will see a bright and a blue field. The boundary line
between them shows the measured value on the scale which is also visible
through the eyepiece [4].
▸ Use the supplied adjustment tool [11]to turn the adjustment screw [3]behind
the prism surface [2](under the rubber cap), and adjust the scale so that the
measuring device is optimally set up by moving the boundary line upwards or downwards.
Calibration value ORA 3HA/3HB: 78.8 % (Brix) (See the calibration line in the scale)
Calibration value ORA 3HA/3HB: 19,6 % (Water content) (See the calibration line in the scale)
▸ Repeat step 4 (cleaning).
Important!
The ambient/room temperature and the sample temperature influ-ence
the refractometer measuring result. The scales of the refractometer
models, which have "HB" in its name, are designed for an ambient
7.2 Additional advice
It is important that the samples being measured are representative samples. For
samples that evaporate easily, measurements should be carried out quickly. The
samples should be at the same temperature as the measuring instrument in order
to achieve an accurate result.
7.3 Measuring procedure
▸ Make sure your hands are dry before handling the refractometer.
▸ Open the prism cover [1]and use the supplied pipette [8]to apply a few drops of
the sample liquid [8]onto the prism surface [2]. Close the prism cover [1]. Spread
the liquid evenly by pressing down on the prism cover [1]and elimi-nate any air
bubbles present.⇨
▸ Hold the device horizontally and wait about 30 seconds (for optimal tempera-ture
equalisation between the sample and device).
▸ View the measurement scale through the eyepiece [4]. Point the prism surface [2]
of the refractometer at a bright light source while doing this.
▸ Rotate the adjustment ring [5]on the eyepiece [4]to adjust the focus.
▸ The boundary line will move on the measurement scale depending on the concen-
tration. This bright/dark boundary line shows the result on the scale.
▸ If the temperature deviates from +20 °C and a refractometer without ATC is used,
correct the measured result using the corresponding value from the temperature
correction table [13].
▸ Carefully clean the supplied pipette [8]and the refractometer aer carrying out
the measurement.
Important!
Aer every measurement, use a lint-free, absorbent cloth to remove
the fluids from the prism surface [2]and prism cover [1]. Then
carefully clean the prism and prism cover using a cloth moistened
with water or if necessary alcohol, and dry both parts using a so,
dry and lint-free cloth. Avoid rubbing the prism [2].
9
11 ORA_1SB_2SB_3SB-BA-e-1510
13 Brix to refractive index (nD) conversion table
BRIX
Refractive
indexBRIX
Refractive
indexBRIX
Refractive
index
%
nD
%
nD
%
nD
0 1.33299301.38115601.44193
1
1.33442
31
1.38296
61
1.44420
2 1.33586321.38478621.44650
3 1.33732331.38661631.44881
4 1.33879341.38846641.45113
5 1.34026351.39032651.45348
6 1.34175361.39220661.45584
7 1.34325371.39409671.45822
8 1.34476381.39600681.46061
9 1.34629391.39792691.46303
101.34782401.39986701.46546
111.34937411.40181711.46792
121.35093421.40378721.47037
131.35250431.40576731.47285
141.35408441.40776741.47535
151.35568451.40978751.47787
161.35729461.4118176 1.48040
171.35891471.41385771.48295
181.36054481.41592781.48552
191.36218491.41799791.4881
201.36384501.42009801.49071
211.36551511.42220811.49333
221.36720521.42432821.49597
23
1.36889
53
1.42647
83
1.49862
241.37060541.42862841.50129
251.37233551.43080851.50398
261.37406561.43299
271.37582571.43520
281.37758581.43743
291.37936591.43967
Data from "ICUMSA" International Commission for Uniform Methods
of Sugar Analysis, at 20 °C and 589 nm wavelength.
!
8. Cleaning and maintenance
Clean the refractometer using a so, lint-free cloth moistened with either water
or, if necessary, alcohol. Do not use any aggressive or abrasive cleaning agents.
Never immerse the device in water or hold it under running water. Never handle the
device with wet or damp hands.
Never touch the measuring prism [2]with hard tools made from plastic, wood,
rubber, metal, glass etc. Hard objects can quickly damage the relatively so prism
glass, resulting in measurement errors.
The refractometer is maintenance-free.
Cleaning should be carried out immediately before and aer each use of the refrac-
tometer to maximise its life and optimise measurement results.
9. Storage
Store the refractometer in a dry, non-corrosive environment, preferably
between 10 °C and 30 °C.
10. Service
Aer reading this operating manual, if you have any questions about setting up
or using the refractometer, or if any unexpected problem occurs, please contact
your dealer. The device housing may only be opened by trained service technicians
authorised by KERN.
11. Disposal
The packaging consists of environmentally friendly materials which can be disposed
of via local recycling facilities.
The device and storage box should be disposed of by the operator in accordance
with applicable national or regional regulations at the place of use.
12. Additional information
The product may differ slightly from the illustrations. Avoid exposing the refracto-
meter to direct sunlight. Never bring the refractometer into contact with solvents.
13. Brix to refractive index (nD) conversion table
Data from „ICUMSA“ International Commission for Uniform Methods
of Sugar Analysis, at 20 °C and 589 nm wavelength.
!
ORA_1SB_2SB_3SB-BA-e-1510
14 Annex
Table 1: International Temperature Correction Table for°Brix (% sugargradient)
Correcttheresultbythefollowingvalues(refractometermustbecorrectlyadjustedat20°C)
% Brix
reading
0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 55.0 60.0 65.0 70.0 75.0 80.0 85.0
Temperature °C
10.0 -0.53 -0.56 -0.59 -0.62 -0.65 -0.67 -0.69 -0.71 -0.72 -0.73 -0.74 -0.75 -0.75 -0.75 -0.75 -0.75 -0.74 -0.73
11.0 -0.49 -0.52 -0.54 -0.57 -0.59 -0.61 -0.63 -0.64 -0.65 -0.66 -0.67 -0.68 -0.68 -0.68 -0.68 -0.67 -0.67 -0.66
12.0 -0.44 -0.47 -0.49 -0.51 -0.53 -0.55 -0.56 -0.57 -0.58 -0.59 -0.60 -0.60 -0.61 -0.61 -0.60 -0.60 -0.60 -0.59
13.0 -0.40 -0.41 -0.43 -0.45 -0.47 -0.48 -0.50 -0.51 -0.52 -0.52 -0.53 -0.53 -0.53 -0.53 -0.53 -0.53 -0.52 -0.52
14.0 -0.34 -0.36 -0.38 -0.39 -0.40 -0.42 -0.43 -0.44 -0.44 -0.45 -0.45 -0.46 -0.46 -0.46 -0.46 -0.45 -0.45 -0.44
15.0 -0.29 -0.31 -0.32 -0.33 -0.34 -0.35 -0.36 -0.37 -0.37 -0.38 -0.38 -0.38 -0.38 -0.38 -0.38 -0.38 -0.37 -0.37
16.0 -0.24 -0.25 -0.26 -0.27 -0.28 -0.28 -0.29 -0.30 -0.30 -0.30 -0.31 -0.31 -0.31 -0.31 -0.31 -0.30 -0.30 -0.30
17.0 -0.18 -0.19 -0.20 -0.20 -0.21 -0.21 -0.22 -0.22 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.22
18.0 -0.12 -0.13 -0.13 -0.14 -0.14 -0.14 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15
19.0 -0.06 -0.06 -0.07 -0.07 -0.07 -0.07 -0.07 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.07
20.0 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00
21.0 0.06 0.07 0.07 0.07 0.07 0.07 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.07
22.0 0.13 0.14 0.14 0.14 0.15 0.15 0.15 0.15 0.16 0.16 0.16 0.16 0.16 0.16 0.15 0.15 0.15 0.15
23.0 0.20 0.21 0.21 0.22 0.22 0.23 0.23 0.23 0.23 0.24 0.24 0.24 0.24 0.23 0.23 0.23 0.23 0.22
24.0 0.27 0.28 0.29 0.29 0.30 0.30 0.31 0.31 0.31 0.32 0.32 0.32 0.32 0.31 0.31 0.31 0.30 0.30
25.0 0.34 0.35 0.36 0.37 0.38 0.38 0.39 0.39 0.40 0.40 0.40 0.40 0.40 0.39 0.39 0.39 0.38 0.37
26.0 0.42 0.43 0.44 0.45 0.46 0.46 0.47 0.47 0.48 0.48 0.48 0.48 0.48 0.47 0.47 0.46 0.46 0.46
27.0 0.50 0.51 0.52 0.53 0.54 0.55 0.55 0.56 0.56 0.56 0.56 0.56 0.56 0.55 0.55 0.54 0.53 0.52
28.0 0.58 0.59 0.60 0.61 0.62 0.63 0.64 0.64 0.64 0.65 0.65 0.64 0.64 0.64 0.63 0.62 0.61 0.60
29.0 0.66 0.67 0.68 0.69 0.70 0.71 0.72 0.73 0.73 0.73 0.73 0.73 0.72 0.72 0.71 0.70 0.69 0.68
30.0 0.74 0.75 0.77 0.78 0.79 0.80 0.81 0.81 0.81 0.82 0.81 0.81 0.81 0.80 0.79 0.78 0.77 0.75
14. Annex
Table 1: International Temperature Correction Table for °Brix (% sugar gradient)
Correct the result by the following values (refractometer must be correctly
adjusted at 20 °C).
temperature of +20 °C! If the measurements are carried out at a temperature other than +20 °C,
the results must be correspondingly corrected. A correction table can be found in the annex,
Point 14. Refractometer models, which have "HA" in its name, are equipped with an automatic
temperature compensation (ATC). Between +10°C and +30°C measurement differences due to
temperature changes are automatically compensated.
This manual suits for next models
3
Other KERN Measuring Instrument manuals

KERN
KERN DLB 160-3A User manual

KERN
KERN KKP V40 Series User manual

KERN
KERN ORA 3AA/AB Beer User manual

KERN
KERN ABP User manual

KERN
KERN ORD 45BM User manual

KERN
KERN MBA-A01 User manual

KERN
KERN YDB-02 User manual

KERN
KERN DBS 60-3 User manual

KERN
KERN DAB-SH-e-1610 User manual

KERN
KERN TMBA-A01-A User manual