KERN DLB-A01 User manual

KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Betriebsanleitung Temperatur-Kalibrierset
Operating instruction temperature calibrating set
Mode d’emploi kit de calibrage de température
KERN DLB_A01
Version 1.1
10/2011
DLB_A01-BA-def-1111

D KERN DLB_A01
Version 1.1 10/2011
Betriebsanleitung
Temperatur-Kalibrierset
1 Technische Daten
Celsius [°C]
Fahrenheit [°F]
Einheiten
Kelvin [K]
-50°C – 1300°C
-58°F - 2000°F
Messbereich
223K -2000K
Ablesbarkeit umschaltbar 1°C / 1°F 0,1 °C / 0,1 °C / 1K
BereichMessgenauigkeit
-50°C - 0°C
-58°F – 32°F
100 - 200°C
200 - 400°C
± 2°C
± 4°F
± 1,5°C
± 3,0°C
Batterien 9 V Blockbatterie
Gewicht 210 g
Abmessungen 162mm (H) x 76mm (B) x 38.5mm (T)
Umgebungstemperatur
0°C - 50°C / 32°F - 122°FC
Umgebungsbedingungen
Relative Luftfeuchtigkeit
0% - 80% (0°C - 35°C) (32°F - 95°F)
0% - 70% (35°C - 50°C) (95°F - 122°F)
2 DLB_A01-BA-def-1110

2 Geräteübersicht
1. Anzeige „Anzeigegerät“
2. Tastatur „Anzeigegerät“
3. Temperatursensorplatte
2.1 Anzeigenübersicht
Anzeige Beschreibung
MAX Kennzeichnet den Maximalwert einer Messung
HOLD Kennzeichnet einen festgehaltenen Anzeigenwert
1Messbereich überschritten
Kapazität der Batterie erschöpft
DLB_A01-BA-def-1110 3

2.2 Tastaturübersicht
Beschreibung
1 Gerät ein- und ausschalten
2 Nach Tastendruck wird die Anzeige für ca. 5 s hinterleuchtet.
3 Nicht dokumentiert
4 Einheit auswählen, °C oder °F oder K
0.1°
1°C
Ablesbarkeit umschalten 1°C ⇔0.1°C / 1°F ⇔0.1°F.
MAX Nach Tastendruck wird der höchste Messwert angezeigt, ge-
kennzeichnet durch den Indikator „MAX“
5
HOLD Nach Tastendruck wird der aktuelle Anzeigewert festgehalten,
gekennzeichnet durch den Indikator „HOLD“. Er wird solange an-
gezeigt, bis er durch erneuten Tastendruck gelöscht wird.
1°
Tc
C
° K
F°
BACK-
LIGHT
10. °
MAX
+
C
°
F
°
K
1
2
5
3
4
4 DLB_A01-BA-def-1110

3 Batterie wechseln
Wenn die Batteriespannung nicht mehr ausreicht, erscheint das Batteriesymbol in der
Anzeige.
+Gerät ausschalten und Kabel zur Temperatursensorplatte entfernen.
+Deckel des Batteriefachs durch Lösen der Schraube öffnen.
+Batterie tauschen.
Auf die richtige Polung achten.
+Den Deckel wieder verschließen.
4 Temperaturwert des KERN DLB-A justieren/kalibrieren
4.1 Vorbereitung
Wir empfehlen den Temperaturwert des DLB-A gelegentlich mit dem optionalen Tem-
peratur-Kalibrierset DLB-A01 zu prüfen. Zuvor sollte sich das Gerät mindestens 3
Stunden nach der letzten Heizphase abgekühlt haben.
ÖDie einzelnen Bauteile „Probenschale“der Reihenfolge nach entfernen
1.
2.
3.
4.
DLB_A01-BA-def-1110 5

ÖTemperatur-Kalibrierset lt. Abb. installieren.
ÖHeizaufsatz an der Rückseite einschalten.
ÖDeckel des Heizaufsatzes schließen.
ÖTemperatur-Kalibrierset einschalten und folgende Einstellungen vornehmen,
s. Kap. 2.2.
Einheit °C wählen,
Ablesbarkeit 1°C
Servicefunktion aufrufen:
ÖIm Wägemodus MENU -Taste drücken, der Menüpunkt „Therm“ wird angezeigt.
ÖMit PRINT-Taste bestätigen. MENU -Taste wiederholt drücken bis „Service“ an-
gezeigt wird.
ÖPRINT-Taste 2 sec gedrückt halten. Mit der MENU-Taste gewünschte Einstellung
auswählen.
Temperaturkalibrierung
Ð
Temperaturjustierung
Ð
Einstellungen Lampe
6 DLB_A01-BA-def-1110

4.2 Temperatur kalibrieren
Bei der Temperaturkalibrierung findet nur eine Überprüfung statt, d.h. es werden keine
Werte verändert.
ÖServicefunktion „Tmp test“ aufrufen, s. Kap. 4.1.
ÖMit PRINT-Taste bestätigen. Die aktuelle Einstellung wird angezeigt. Mit den
Navigationstasten ÐÏTemperatur auswählen, bei der die Überprüfung statt fin-
den soll, wählbar 35- 160°C.
ÖMit PRINT-Taste bestätigen, das Gerät heizt auf die eingestellte Temperatur auf.
Nach ca. 15 Minuten den angezeigten Temperaturwert am Thermometer mit dem
des Feuchtebestimmers vergleichen. Stimmen die beiden Werten nicht überein,
empfehlen wir eine Temperaturjustierung, s. Kap. 3.
ÖMit der ON/OFF-Taste wird die Kalibirung beendet.
ÖMit ON/OFF-Taste zurück in den Wägemodus.
DLB_A01-BA-def-1110 7

4.3 Temperatur justieren
Die Temperatur wird an zwei wählbaren Punkten gemessen und kann an diesen korri-
giert werden.
ÖServicefunktion „Tmp Adj“ aufrufen, s. Kap. 4.1.
ÖMit PRINT-Taste bestätigen. Die aktuelle Einstellung für den ersten Temperatur-
punkt wird angezeigt. Entweder auf Werkseinstellung 80°C lassen oder mit den
Navigationstasten ÐÏgewünschte Temperatur auswählen (wählbar 50-130°C).
ÖMit PRINT-Taste bestätigen, die erste Aufheizphase wird gestartet. Die
verbleibende Zeit wird angezeigt.
ÖNach 15 Min ist die Temperaturkalibrierung des 1. Punktes beendet. Den
angezeigten Temperaturwert am Thermometer mit dem des Feuchtebestimmers
vergleichen. Stimmen die beiden Werten nicht überein, mit den Navigationstasten
ÐÏkorrigieren und mit PRINT-Taste bestätigen
.
ÖDie aktuelle Einstellung für den zweiten Temperaturpunkt wird angezeigt. Entwe-
der auf Werkseinstellung 150°C lassen oder mit den Navigationstasten ÐÏge-
wünschte Temperatur auswählen. Der zweite Temperaturpunkt muss mindestens
30°C über dem ersten liegen, max. 160°C.
ÖMit PRINT-Taste bestätigen, die zweite Aufheizphase wird gestartet. Die
verbleibende Zeit wird angezeigt.
ÖNach 15 Min ist die Temperaturkalibrierung des 2. Punktes beendet. Den
angezeigten Temperaturwert am Thermometer mit dem des Feuchtebestimmers
vergleichen. Stimmen die beiden Werten nicht überein, mit den Navigationstasten
ÐÏkorrigieren.
ÖMit PRINT-Taste bestätigen. Die Justierung ist beendet, das Gerät kehrt zurück
ins Menü. Mit ON/OFF-Taste zurück in den Wägemodus.
Durch die AUTO-PPOWER OFF-Funktion schaltet sich das Anzeigegerät des
Temperatur-Kalibriersets während der Justierung automatisch aus. Gerät erneut
einschalten, die laufendeTemperatur-Justierung wird dadurch nicht gestört.
8 DLB_A01-BA-def-1110

GB KERN DLB-A01
Version 1.1 10/2011
Operating Instructions
Temperature calibration set
1 Technical data
Celsius [°C]
Fahrenheit [°F]
Units
Kelvin [K]
-50°C – 1300°C
-58°F - 2000°F
Measuring range
223K -2000K
Readability commutable 1°C / 1°F 0,1 °C / 0,1 °C / 1K
Display exactness ± 0,2°C
RangeSensor measuring exact-
ness -50°C - 0°C
-58°F – 32°F
100 - 200°C
200 - 400°C
± 2°C
± 4°F
± 1,5°C
± 3,0°C
Batteries 9 V compound battery
Battery life 150 hrs.
Weight 210 g
Dimensions 162mm(H) x 76mm(W) x 38.5mm(D)
Ambient temperature
0°C - 50°C / 32°F - 122°FC
Environmental conditions
Relative air humidity
0% - 80% (0°C - 35°C) (32°F - 95°F)
0% - 70% (35°C - 50°C) (95°F - 122°F)
DLB_A01-BA-def-1110 9

2 Appliance overview
1. Display „Display unit“
2. Keyboard „Display unit“
3. Temperature sensor plate
2.1 Overview of display
Display Description
MAX Means the maximum value of a measurement
HOLD Means a recorded display value
1Measuring range exceeded
Capacity of battery exhausted
10 DLB_A01-BA-def-1110

2.2 Keyboard overview
Description
1 Switch unit on and off
2 Press button, the display background will be illuminated for approx. 5 s.
3 Not documented
4 Select unit, °C or °F or K
0.1°
1°C
Switch-over readability 1°C ⇔0.1°C / 1°F ⇔0.1°F.
MAX Press button, the highest measured value will be displayed, iden-
tified by the indicator „MAX“
5
HOLD Press the button to record the current displayed value, identified
by the indicator „HOLD“. It remains displayed until it is deleted by
pressing button again.
1°
Tc
C
°K
F°
BACK-
LIGHT
1
°
0.
+
MAX
F
°
K
C
°
1
2
3
5
4
DLB_A01-BA-def-1110 11

3 Change battery
If the battery voltage is not more sufficient, the battery symbol will appear in the dis-
play.
+Switch off appliance and remove cable from temperature sensor plate.
+Open the lid of the battery compartment loosening the screw.
+Replace battery.
Ensure correct polarisation.
+Close again the lid.
4 Adjust/calibrate the temperature value of the KERN DLB-A
4.1 Preparation
We recommend sometimes to check the temperature value of the device using the op-
tional temperature calibrating set DLB-A01. Before you do this, allow the device to cool
down for at least 3 hours after the last heating phase.
ÖRemove the separate component parts „sample dish“ according to their sequence
1.
2.
3.
4.
12 DLB_A01-BA-def-1110

ÖInstall the temperature-calibration set acc. to fig.
ÖSwitch-on the heating top on the rear side.
ÖClose cover of the heating top
ÖSwitch-on the temperature calibration set and carry out the following adjustments,
see chap. 2.2. Select unit °C, readability 1°C
Invoke service function:
ÖIn weighing mode press the MENU button, the menu item „Therm“ will be dis-
played.
ÖConfirm by pressing the PRINT button. Press repeatedly the MENU button until
„Service“ will be displayed.
ÖKeep pressed the PRINT button for 2 sec. Select the desired settings by pressing
the MENU button
Temperature calibration
Ð
Temperature adjustment
Ð
Settings of lamp
DLB_A01BA-def-1110 13

4.2 Calibrate temperature
During temperature calibration only a check is carried out, i.e. no values are changed.
ÖInvoke service function „Tmp test“, see chap. 4.1.
ÖConfirm by pressing the PRINT button. The current setting will be displayed. Use
the navigation buttons ÐÏto select the temperature during the check, available
35- 160°C.
ÖConfirm using the PRINT button, the appliance heats up to the set temperature.
After approx. 15 minutes compare the temperature displayed on the thermometer
with that of the moisture analyzer. If the two values do not match, we recommend
a temperature adjustment, see chap. 3.
ÖUse the ON/OFF button to finish the calibration.
ÖUse the ON/OFF button to return into weighing mode.
14 DLB_A01-BA-def-1110

4.3 Adjust temperature
The temperature is measured at two selectable points and it is possible to correct it
there.
ÖInvoke service function „Tmp Adj“, see chap. 4.1.
ÖConfirm by pressing the PRINT button. The current setting for the first tempera-
ture point is displayed. Or leave it on factory setting 80°C or select a desired tem-
perature with the help of the navigation buttons ÐÏ(available 50-130°C).
ÖConfirm with the PRINT button, the first heat-up phase is started. The remaining
time is displayed.
ÖTemperature calibration of first point will take 15 min. Compare the temperature
displayed on the thermometer with that of the moisture analyzer. If the two values
do not match, correct them using the navigation buttons ÐÏand confirm with the
PRINT button
ÖThe current setting for the second temperature point is displayed. Or leave it on
factory setting 150°C or select the desired temperature with the help of the navi-
gation buttons ÐÏ. The second temperature point has to be at least 30°C over
the first one, max. 160°C.
ÖConfirm with the PRINT button, the second heat-up phase is started. The remain-
ing time is displayed.
ÖTemperature calibration of the second point will take 15 min. Compare the tem-
perature displayed on the thermometer with that of the moisture analyzer. If the
two values do not match, correct with the help of the navigation buttons ÐÏ.
ÖConfirm by pressing the PRINT button. The adjustment is completed, the appli-
ance returns into the menu. Use the ON/OFF button to return into weighing mode.
Due to the AUTO-POWER OFF function the display unit of the temperature
calibration set switches off automatically during adjustment. Turn-on appli-
ance again, the current temperature adjustment will not be disturbed.
DLB_A01-BA-def-1110 15

F KERN DLB-A01
Version 1.0 04/2011
Mode d’emploi
Kit de calibrage de température
1 Caractéristiques techniques
Celsius [°C]
Fahrenheit [°F]
Unités
Kelvin [K]
-50°C – 1300°C
-58°F - 2000°F
Plage de mesure:
223K -2000K
Lisibilité commutable 1°C / 1°F 0,1 °C / 0,1 °C / 1K
Précision d'affichage ± 0,2°C
PlagePrécision de mesure du
capteur -50°C - 0°C
-58°F – 32°F
100 - 200°C
200 - 400°C
± 2°C
± 4°F
± 1,5°C
± 3,0°C
Batteries Batterie monobloc de 9 V
Durée des batteries 150 h
Poids 210 g
Dimensions 162mm(H) x 76mm(B) x 38.5mm(T)
Température ambiante
0°C - 50°C / 32°F - 122°FC
Conditions d'environne-
ment
Humidité de l’air rélative
0% - 80% (0°C - 35°C) (32°F - 95°F)
0% - 70% (35°C - 50°C) (95°F - 122°F)
16 DLB_A01-BA-def-1110

2 Aperçu de l’appareil
1. Affichage „Appareil d’affichage“
2. Clavier „Appareil d’affichage“
3. Plaque capteur de température
2.1 Vue d’ensemble des affichages
Afficheur Description
MAX Signifie la valeur maximale d’un affichage
HOLD Signifie une valeur d’affichage enregistrée
1Plage de mesure dépassée
Capacité de la batterie épuisée.
DLB_A01-BA-def-1110 17

2.2 Vue d’ensemble du clavier
Description
1 Mise en marche et arrêt de l’appareil
2 Après avoir enfoncée la touche le fond de l’affichage est éclairé env. 5 sec.
3 Non documenté
4 Choisir l’unité, °C ou °F ou K
0.1°
1°C
Commuter la lisibilité 1°C ⇔0.1°C / 1°F ⇔0.1°F.
MAX Enfoncer la touche, la valeur mesurée plus haute est affichée,
identifiée par l’indicateur „MAX“
5
HOLD Enfoncer la touche pour enregistrer la valeur affichée actuelle,
caractérisée par l’indicateur „HOLD“. Elle continue d’être affichée
jusqu’à ce que soit effacée en appuyant autre fois sur la touche.
1°
Tc
C
° K
F°
BACK-
LIGHT
10. °
+
MAX
F
°
K
C
°
1
2
3
5
4
18 DLB_A01-BA-def-1110

3 Changement de la pile
Si la tension de la pile ne suffit plus, le symbole de batterie apparaît à l’affichage.
+Mettre l’appareil à l’arrêt et enlever le câble vers la plaque du capteur de tempé-
rature.
+Ouvrir le couvercle du compartiment à piles en desserrant la vis.
+Remplacer la pile.
Veillez à la polarité correcte.
+Refermer le couvercle.
4 Ajuster/calibrer la valeur de température du KERN DLB-A
4.1 Préparation
Nous recommandons de vérifier occasionnellement la valeur de température de
l’appareil à l’aide du jeu de calibrage de la température en option DLB-A03. Aupara-
vant l’appareil devrait s'être refroidi au moins 3 heures après la dernière phase de
chauffage.
ÖAssembler les différentes pièces constitutives du „plateau porte-échantillon“ dans
l’ordre de succession indiqué.
1.
2.
3.
4.
DLB_A01-BA-def-1110 19

ÖInstaller le kit de calibrage de la température conformément à la figure.
ÖMettre en marche la rehausse de chauffage.
ÖFermer le couvercle de la rehausse de chauffage
ÖMettre en marche le kit de calibrage de température et faire les réglages suivants,
voir au chap. 2.2.
Choisir l’unité °C,
lisibilité 1°C
Appel de la fonction du SAV:
ÖAppuyer sur la touche du MENU, le point de menu „Therm“ est affiché.
ÖConfirmez sur la touche PRINT. Répéter la pression sur la touche MENU jusqu’à
ce que „Service“ apparaisse.
ÖMaintenir la touche PRINT pendant 2 sec enclenchée. Sélectionner sur la touche
MENU votre réglage voulu
Calibrage de la température
Ð
Ajustage de température
Ð
Réglages de la lampe
20 DLB_A01-BA-def-1110
Table of contents
Languages:
Other KERN Measuring Instrument manuals

KERN
KERN ABP User manual

KERN
KERN MLS 50-3...N series User manual

KERN
KERN DLB 160-3A User manual

KERN
KERN MLS 50-3C User manual

KERN
KERN TMBA-A01-A User manual

KERN
KERN KFE-TM User manual

KERN
KERN DAB-BA-e-1511 User manual

KERN
KERN DBS User manual

KERN
KERN ORD 45BM, ORD 92BM, ORD 85BM, ORD 92HM, ORD 1SM, ORD 3SM, ORD 2WM, ORD 1PM, ORD 2UM, ORD 5UM, ORD 6 US, ORD... User manual

KERN
KERN DLB A01N User manual