1-4 TOS6200/6210
7. Contact check function
The testeris provided with acontactcheck function that verifies the connection of
the DUT (by currentdetection) before initiatingtesting.
8. Voltage judgment function (TOS6210 only)
This function allows eitheraresistance value or avoltage drop value (voltage
between the SAMPLING terminals) to be selected as the reference value for judg-
mentcriteria.
It alsoallows voltage drop value-based testingto be conducted,which is required by
safety standards such as UL 60950-1 and IEC 60950-1.
9. Simple operations
Testerfunctions are fastand easily controlled,with an intuitive control method
that's easy to pick up,even for first-time users.
For example,testconditions are setsimply by selectingan item displayed on the
LCD with the cursor keys and turningthe rotary knob to set/selectavalue.
Any items you wish to setup can be assigned to the function keys.
10. Stores up to 100 types of test conditions
The testerallows you to store and name up to 100 testconditions involvingparame-
ters such as testcurrent, determined resistance value,and testtime.For example,
you can store testconditions for aspecific safety standard underthe name of the
standard,or store testconditions underthe name of the destination of the DUT.
When testconditions need to be modified due to changes in the destination of the
productor revised safety standards,you can recall asetof testconditions simply by
enteringamemory number, makingnecessary modifications to the pre-existing
standard.
Assigningspecific names allows testconditions to called up by name.This function
is available through both the frontpanel and the remote control.
11. Programming of test conditions
Acombination of stored testconditions allows automatic execution of tests consist-
ingof several programs of up to 100 steps each.
Although the total numberof steps is limited to 500, 100 types of programs can be
stored and recalled from the frontpanel or the remote control.
12. GPIB and RS-232Cinterfaces
The testeris provided with standard GPIB and RS-232C interfaces.It is therefore
not necessary to buy additional GPIB and RS-232C boards.
An interface cable and PC or sequencerallows remote control of testconditions
such as testcurrent, determined resistance value,and testingtime.Measured values
and testresults can alsobe read back.
13. Supplied test leads
The testercomes with alligator clip testleads,lettingyou start testingimmediately.
14. Memo function
The testerhas amemofunction that can store up to 20 characters perline on 3lines,
which can be used to store serial numbers,calibration dates,and/or comments.