2HP01A-TOS/HP02A-TOS
WARNING
The test probe uses a dangerously high voltage capable of causing electric
shocks resulting in death unless handled correctly. To prevent accidents,
handle the probe with extreme care, abiding by all instructions given below.
Ensure the safety of test environments.
• Insulate the DUT and other high-voltage areas1 with an electrical insulation
rubber plate or sheet to prevent electric shock even if the operator contacts the
DUT or other high-voltage areas. If electrical insulation rubber plates or sheets
are not available, contact your Kikusui distributor or agent.
• Implement measures so that only authorized operators can approach high-volt-
age areas1.
Never touch the contact pin. Contact will result in electric shock.
• High voltage is delivered to the contact pin.
Never touch the contact pin.
• The DUT is also charged to high test volt-
ages during testing. Never touch high-volt-
age areas1.
• Never expose the contact pin except when
testing.
• Never touch the red area around the contact pin; this area is also capable of
inflicting electric shock.
Modification may cause electric shock.
• Never fasten the slide lever to the grip such that the contact pin is exposed.
• Do not lock start switch 1 and/or start switch 2 to the on state.
• Do not fasten the high voltage cable to the high voltage terminal of the Tester. If
the cable catches against an object, it may be disconnected from the Tester.
Do not operate the probe with a single hand.
For safety reasons, this test probe is designed to be operated using both hands.
Never attempt to operate the probe using only one hand.
Be sure to connect the remote control cable.
To prevent electric shock, never perform testing using the Tester START (TEST)
switch without the remote control cable.
Never use the test probe with Testers from other manufacturers.
The test probe is designed specifically for use with a KIKUSUI Tester. It cannot
be used with testers from other manufacturers, even if the connector appears to
be compatible.
Never use the product at voltages above 4 kV ac rms or 5 kV dc.
The rated maximum voltage2 of the probe is 4 kV ac rms or 5 kV dc. Use at
higher voltages may result in electric shock.
Handle the test probe attentively. Do not swing it around.
Do not yank the test probe cable or drape it over any part of your body.