Functions available from main screen
How to change battery
1. Remove the front part at the back of the
instrument.
2. Remove old battery.
3. Insert the new battery in accordance with proper
polarity drew inside the housing.
4. Replace the front.
Velocity
activation 1x
Select air
velocity unit
Press to activate or deactivate Validate with
Press to scroll down units Validate with
Ambient
Temp. 6x Enter value with Validate with
Atmo pheric
Pre ure 7x Enter value with Validate with
Select
pre ure unit Press to scroll down units Validate with
Automatic
hut-off Press to scroll down units Validate with
Beep Press to activate or deactivate Validate with
2x
3x
4x
5x
Select pre ure
unit Press to scroll down units Validate with
1x
Automatic
hut-off Press to scroll down minutes Validate with
2x
Beep Press to activate or deactivate Validate with
3x
Self-calibration 1x Start of self calibration countdown*
Functions available from main screen (only for MP120 instrument)
Self-calibration 1x Start self calibration countdown*
*self calibration is performed by disconnecting the two pressure inlets of the sensor, then by pressing AutoZ key.
Using Pitot tube
1. Connect the Pitot tube to MP120 instrument.
2. The probe must be placed at the center of the duct and
parallel to the air flow.
3. Press then to activate velocity.
Dynamic pressure = PT PS
+
-
(+) Static
Pressure PS ( ) Total
Pressure PT