Labex COMFORT EL User manual

Labex Comfort User Manual, Date: 20 April 2020, # 1
LABEX™COMFORT EL
USER MANUAL

Labex Comfort User Manual, Date: 20 April 2020, # 2
OVERVIEW
1. What is Labex Comfort
Labex Comfort is a digital electrolarynx device. It is specially designed to
help laryngectomees talk. Its main feature is the two configurable talk
buttons.
Labex Comfort is a CLASS I medical device.
General principle of operation for all models electrolarynges manufactured
by Labex Trade Ltd.:
The digital electrolarynx produces mechanical vibrations with a low sound
frequency between 60-195 Hz. Applied to the neck, these vibrations create
trembling in the oral cavity similar to the human voice.
The combination of starting and stopping sound vibrations and oral
articulation creates comprehensible human speech.
2. In the box
In the box, you will find everything you need to start using the device:
Labex Comfort electrolarynx device;
2 batteries 9v rechargeable Li-Io;
Two socket charger;
„Labex“lanyard;
User manual;
Warranty card.

Labex Comfort User Manual, Date: 20 April 2020, # 3
3. Parts and buttons
Labex Comfort consists of several main
components:
1) Top cap –The part which you put on your
neck to receive the sound.
2) TALK button 1 –The upper talk button, can
be configured with its own volume and pitch.
3) TALK button 2 - The lower talk button, can be
configured with its own volume and pitch.
4) SETTINGS button –Used for changing
settings.
5) Indicator LED –Lights up when one of the
TALK buttons is pressed.
6) Battery indicator LED –Lights up when the
battery needs charging.
7) Battery cover –Used to secure the battery
inside the device
NOTE: The actuator buttons are sensitive. They
require about 160 grams of force to make contact. Please avoid pushing
them with extreme force. Doing so could damage your device.

Labex Comfort User Manual, Date: 20 April 2020, # 4
GETTING STARTED
4. Basic use
You must insert the battery into the device in order to turn it
on. Take one of the two, included in the package, batteries
and carefully remove the transparencies. Swivel the bottom
cap of the device. Insert the battery by ensuring that the "+"
and "-" signs of the battery and the device match. Swivel the
bottom cap back.
NOTE: Inserting the battery incorrectly will not damage the
device, but it will not work.
When the Top cap of the device touches the throat skin and
you press the Speech Button, the device starts to emit sound.
This allows the proper function of the product.
5. Finding the sweet spot
A sweet spot is a place where you feel the sound from the electrolarynx
transmits through your neck tissue and your oral cavity with the most
clarity. The easiest way to find your sweet spot is to place the electrolarynx
on your neck in different positions and actuate the sound while you are
pushing the START button. What you are looking for is the place on your
neck, which transmits sound the best. While looking for a sweet spot, it is
crucial that you press it firmly against your neck tissue, so the sound
transmits through your tissue and out your oral cavity. Take your time
finding your sweet spot –this is the most crucial part of learning to use an
electrolarynx. Finding just the right place will set you up for success.
6. First Words
After you have pinpointed the desired spot, you can start making some
basic sounds. Focus on single vowels first. Over-articulate every sound with
your mouth until you feel comfortable with all the vowels. After that, you
can start practicing short words. Slowly work your way up to sentences and
conversation.

Labex Comfort User Manual, Date: 20 April 2020, # 5
For a more detailed guide, please check our article:
https://labextrade.com/tips-for-using-electrolarynx/
7. Expectations
Sometimes learning to speak with an electrolarynx device is not easy. Think
of it as learning to play a musical instrument. It will take time, practice, and
patience.
In some cases, talking can be achieved in a month, in other cases it can
take only 3 days.
SETTINGS
8. How to change the volume
You can adjust the volume of each button individually. To do so, follow this
procedure:
Press and hold the talk button you wish to adjust the volume of
(TALK button 1 or TALK button 2).
While holding your desired talk button, press and hold the
SETTINGS button.
Keep both buttons pressed. The sound will stop, and the red LED
will start blinking.
After a few seconds, the device will start buzzing on its own.
Release all buttons. The device is now in “Volume setup mode.”
You can now use TALK button 1 to increase the volume, or TALK
button 2 to decrease the volume.
After you reached your desired volume, press the SETTINGS button
to confirm, the device will automatically save your desired volume.
NOTE: You can choose different volume for each TALK button.
9. How to change the pitch
You can adjust the pitch of each button individually. To do so, follow this
procedure:

Labex Comfort User Manual, Date: 20 April 2020, # 6
●Press and hold the SETTINGS button.
●While holding the SETTINGS button, press and hold the TALK
button, you wish to change the pitch of (TALK button 1 or TALK
button 2).
●Keep both buttons pressed. The sound will stop, and the red LED
will start blinking.
●After a few seconds, the device will start buzzing on its own.
Release all buttons. The device is now in “Pitch setup mode.”
●You can now use TALK button 1 to increase the pitch, or TALK
button 2 to decrease the pitch.
●After you reached your desired volume, press the SETTINGS button
to confirm, the device will automatically save your desired pitch.
NOTE: You can choose a different pitch for each TALK button.
10. How to revert to the factory settings
If you feel you want to revert the device to its factory default volume &
pitch follow these steps.
1) Press and hold all the buttons at the same time (TALK button 1,
TALK button 2 and SETTINGS buttons).
2) The red Battery Indicator LED will start blinking. Continue to hold all
the buttons until the Battery Indicator LED stops blinking and lights
up in solid red.
3) When the Battery Indicator LED turns solid red, release all the
buttons and wait 3-4 seconds. All the settings have been reset to
factory default.
Check our website for video tutorials.
BATTERY
11. Low battery
Labex Comfort has a built-in battery indicator, which will warn you when
the battery level gets low. When the battery goes down to 20%, the

Labex Comfort User Manual, Date: 20 April 2020, # 7
Battery Indicator LED will start blinking after every release of a speech
button. When the battery drains down to 10%, the Battery Indicator LED
will light up in solid red. At this point, you should replace your battery with
a fully charged one.
12. Different batteries
There are two high capacity Li-Io batteries included with your Labex Digital.
This will allow you to have always a fully charged battery on hand.
However, Labex digital is compatible with all 9V batteries on the market –
rechargeable or non-rechargeable.
13. Charging the battery
After you pull out your drained battery, you should charge it with the
included intelligent charger:
1) Put the battery inside the charger while minding the “+” and the “–“.
2) Put the charger into an electrical socket.
3) The corresponding red LED will light up, meaning the battery is
currently charging.
4) The red LED will turn off when the battery is fully charged.
Note: The included intelligent charger will NEVER overcharge your
batteries. Each battery life is more than 500 recharge cycles.
WARNING: Do not try to charge ANY other kind of battery with the
charger! The charger is designed ONLY for Li-Io (Lithium-Ion) batteries!
Trying to charge any other kind of battery will result in permanent damage
to the battery and the charger!
Note: If you find that the battery in your device can not hold a charge for more
than one hour of use, there is a possibility that the battery you are using is near
the end of its life cycle. If this happens we advise you to recycle this battery and
get a new one.

Labex Comfort User Manual, Date: 20 April 2020, # 8
CARING FOR YOUR DEVICE
14. Cleaning and maintenance
Since the device is in constant contact with the skin and other surfaces, it
should be cleaned at least once every 24 hours. For the best hygiene
practices, the following procedure should be followed:
Clean the outside surfaces of the device with a slightly wet soft
microfiber cloth.
After cleaning all the surfaces, you can disinfect the device by
applying a small amount of disinfectant gel or alcohol. Alternatively,
alcohol wipes can be used.
Wipe the device with a soft microfiber cloth again.
15.Water resistance
Labex Comfort can withstand being used in the rain or cleaned with a wet
cloth. However, submerging the unit in water will damage the device and
will void your warranty.
16.Contact information
LABEX TRADE LTD.
59 Sankt Peterburg Blvd., Floor 6, Office 604
4017 Plovdiv, Bulgaria, European Union
T: +359 32 54 9090, M: +359 878 616 711
For more information visit https://labextrade.com

Labex Comfort User Manual, Date: 20 April 2020, # 9
17. Technical specifications
General:
Adjustable tone
Yes
Adjustable volume
Yes
Speech button
2
Settings buttons
1
Frequency range
55 - 155 Hz
Volume range
60-83 dB
Speech time
Up to 4 h. per charge
Physical:
Diameter x Length
Ø34 x 98 mm
Weight
72g without battery
Dust / Moisture resistive
Yes
Battery:
Capacity
600 mAh
Battery type
Rechargeable 9V 6F22 Li-Ion
Batteries in the box
2 pcs
Battery life
More than 500 recharge cycles
Chargins station in the box
2 slots 9V Li-Ion
Charging time from 0 to 100%
4h 30 min with Labex charger
Protection
Smart electronics

Labex Comfort User Manual, Date: 20 April 2020, # 10
INFORMATION
18. Warnings
Do not submerge the device underwater. Do not wash the device
under running water. This will potentially damage the device.
Do not try to disassemble the device under any circumstances. This
will permanently void your warranty.
It is recommended not use any third-party chargers or rechargeable
batteries. Use the original ones provided in the box. Using any
other charging equipment could potentially damage the battery or
the device.
Do not insert any part of the electrolarynx or any accessories into
the stoma.
Do not burn or physically damage the battery.
Do not short-circuit a cell or battery.
The device contains a strong magnet that may interfere with
pacemakers or other implantable devices. Consult with your
physician before use.
Please note that due to the nature of the electrolarynges and the principle
of their use, the device is not intended to be used by multiple people,
shared use, and second-hand use. Due to the contact of the device to the
user's neck, stoma, or other cavities - the device is for personal use only!
Once a person has used the device, it should not be used by anyone else.
Because of this fact, the included warranty provided by Labex Trade Ltd. is
non-transferable.
19. Safe storage:
Store in a dark and dry place.
Store the device and its accessories in its original box.
Store in temperatures between -20°C to +45°C.

Labex Comfort User Manual, Date: 20 April 2020, # 11
20.Warranty
All Labex devices come with a limited 2-year warranty from the date of
purchase. The warranty period of a replacement electrolarynx or part lasts
as long as the remaining warranty of the original electrolarynx.
This Limited Warranty covers any defects in material or workmanship
under normal use according to the User Manual during the Warranty
Period.
Trying to open or repair the device; causing extensive shock damage
(dropping it, hitting it against other objects, etc.); or liquid damaging it
(submerging in water or other chemicals) will void your warranty.
During the Warranty Period, Labex Trade Ltd. will repair or replace, at no
charge, products or parts of a product that prove defective because of
improper material or workmanship, under normal use and maintenance.
This Limited Warranty is not transferable.
This Limited Warranty does not cover any problem that is caused by:
Conditions, malfunctions or damage not resulting from defects in material
or workmanship.
Conditions, malfunctions or damage resulting from negligence, improper
maintenance or modification.
Damaged or destroyed by natural causes including but not limited to
lightning, flood, or another natural disaster - Theft or loss of the EL.
Shipping charges as well as the risk of loss or damage to the electrolarynx
while in transit to Labex Trade Ltd are taken care of by the customer. The
back shipping is covered by the Company.
Please read the warranty card. To learn more about the Labex warranty,
please visit https://labextrade.com/warranty-policy/
If you have any problems with your device, contact your distributor or
contact us directly https://labextrade.com/contacts/

Labex Comfort User Manual, Date: 20 April 2020, # 12
21.Information on Disposal for Users of WEEE:
This symbol on the product box and/or accompanying documents means
that used electrical and electronic equipment (WEEE) should not be mixed
with general household waste. For proper treatment, recovery, and
recycling, please take this product(s) to designated collection points where
it will be accepted free of charge. Alternatively, in some countries, you may
be able to return your products to your local retailer upon the purchase of
an equivalent new product. Disposing of this product correctly will help
save valuable resources and prevent any potential negative effects on
human health and the environment, which could otherwise arise from
inappropriate waste handling. Please contact your local authority for
further details of your nearest designated collection point. Penalties may
be applicable for incorrect disposal of this waste, in accordance with your
national legislation.
For professional users in the European Union, if you wish to discard
electrical and electronic equipment (EEE), please contact your dealer or
supplier for further information.
For disposal in countries outside of the European Union: This symbol is
only valid in the European Union (EU). If you wish to discard this product,
please contact your local authorities or dealer and ask for the correct
method of disposal.
Table of contents
Other Labex Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Wissner-Bosserhoff
Wissner-Bosserhoff Eleganza 1 User manual and technical description

Clinton
Clinton 6198 Assembly instructions

Devon
Devon extriCARE 2400 operating manual

Innovasis
Innovasis Excella III-D Instructions for use

Organ recovery systems
Organ recovery systems LifePort Kidney Transporter Operator's manual

ResMed