Operating Instructions ULTRAHEAT®XS
Ultrasonic Heatmeter ULTRAHEAT®XS Subject to change without prior notice UH 304 – 101l 2/2
Fault Code Fault Measures
FL nEG Wrong flow direction Check / correct flow or
mounting direction
Eventually changing
with
DIFF nEG Negative temperature
difference Check / exchange mount-
ing position of sensors
Eventually changing
with
F0 No flow rate can be
measured
Air in the measuring
unit/pipe, vent pipe (as-
delivered state)
F1 Interruption in the
supply sensor Contact service
F2 Interruption in the
return sensor Contact service
F3 Electronics for tem-
perature evaluation
defective Contact service
F4 Battery empty Contact service
F5 Short-circuit in the
supply sensor Contact service
F6 Short-circuit in the
return sensor. Contact service
F7 Fault in internal
storage operation Contact service
F8
Fault F1, F2, or F3 or
F5, F6 pending for
longer than 8 hours,
detection of fraud
attempts. No more
measurements are
performed.
This F8 error message
must be reset by service.
F9 Fault in the electron-
ics Contact service
If the response thresholds are exceeded and the flowrate and
temperature are positive, the heat quantity and volume will be
summated. The segment test displays all display segments for
test purposes.
On the yearly set day, the meter readings for heat quantity, vol-
ume, and missing time are placed in a previous year memory
each year.
The flowrate, heat power, and temperature difference are re-
corded signed. Values below the response threshold are pre-
ceded by a "u". The current temperatures are displayed together
as integer °C values on one display line.
The 8-digit customer number (secondary address for M-bus op-
eration) can be set in parameterization mode. The most signifi-
cant digit is suppressed on the display. The device number is
assigned by the manufacturer.
The operating hours are counted from initial connection of the
power source. Missing hours are summated if a fault is pending
that prevents the heat meter from measuring. The date is incre-
mented daily.
The firmware version number is assigned by the manufacturer.
Technical Data:
Technical Data Electronic Unit:
Ambient temperature 5 - 55°C
Power supply Battery for 6 or 11 years
or 24V AC/DC external (special
version)
Temperature sensor Pt 500
Communication Optical interface standard, M-bus
or pulse output optional
Protocol IEC870, 300 baud in Nb
Splittability Always removable,
cable length 1 m
Technical Data of the Flowrate Measuring Unit
Type
qpDimensions
0.6 110mm (3/4“) 190mm(1“)
1.0 110mm (3/4“) 190mm(1“)
1.5 110mm (3/4“) 190mm(1“)
2.5 130mm (1“) 190mm(1“)
Temperature range 15 - 105°C
Nom. pressure 1.6 MPa (PN 16)
Overload qs = 2 x qp, permanent
Mounting position Horizontal or vertical
Measurement range 1:100 approved, calibrated
1:50
Measurement accuracy: EN 1434 cl. 3
Technical Data Temperature Sensor:
Temperature sensor Pt 500 acc. to EN 60751, not
removable
Connection 2-wire, permanent.
Type DS direct short, M10 x
27.5mm acc. to EN1434 or
rod sensor 45 x 5.2 dia. mm
Cable length 1.5m standard, 5 m optional
Max. temperature 105 °C
Installation supply sensor Mounting element for DS 1/2“
x M10, ball valves for DS,
brass pocket 1/2“ for rod sen-
sor
Installation return sensor Integrated (when meter is for
installation in return)
Notes:
Observe national calibration laws when replacing the battery.
Batteries must not be opened, come into contact with water, or be
exposed to temperatures above 80 °C. Dispose of used batteries
at suitable collection centers.
The transport of the heat meter is permitted only in the original
package. If the meter needs to be sent by air freight then the
battery must be removed prior to shipping!
Latest information about our heat meters you may also find in the
internet: www.landisgyr.com
EC Declaration of conformity
Landis+Gyr herewith declares that this product complies
with the relevant requirements of the following directives:
−2004/22/EC Measuring instruments directive
−89/336/EEC Electromagnetic compatibility
−73/23/EEC Low voltage directive
EC type-examination certificate
DE-06-MI004-PTB007
Landis+Gyr GmbH
Humboldtstr. 64
D-90459 Nürnberg
Germany