
T4 EEG Amplifier User Manual
16
2.4 Battery Operation Time
USB Powerbank Li-Polymer battery pack
When fully charged the battery pack will typically power the unit for 36 hours depending on the number of
channels, sample rate and wireless usage.
The typical service life is 2 years.
Internal Li-Ion backup battery
The internal backup battery will enable the unit to continue operating for a short period of time (15 mins ap-
prox.) to allow the main battery pack to be replaced. It is recharged automatically, whenever the main battery
pack is connected, with acquire off. The state of charge is displayed, as described in section 3.4, whenever the
unit is internally powered from the backup battery. To charge manually, operate the pushbutton several times
to activate the backlight and if the reading drops below approximately 70%, charge the battery for about 10
minutes by connecting the main battery pack or connecting to a USB port.
The typical service life is 500 charge-discharge cycles.
2.5 Use in the home environment
Where the equipment is intended to be used in the home, the unit and its battery should be operated in its bag
where it is protected against ingress of solid objects and water to a degree of IP22.
The laptop PC is optional in the home environment and may be used for video recordings. There is no connec-
tion between the PC and the T4 Amplifier unit, as all communication is accomplished wirelessly.
Keep the equipment away from sources of heat.
Do not use mobile phones.
Do not allow pets or children to interfere with the sensor cables.
When the equipment is operated with or without its Bluetooth or WiFi on, other devices in the vicinity should
be moved away or turned off to reduce the likelihood of interference to the equipment or by the equipment.
The T4 may have internal radios fitted. These are approved industry-standard Bluetooth and Wi-Fi types which
present minimal risk of reciprocal interference with other equipment.
2.6 Use with other equipment
Defibrillators and HF surgical equipment
The equipment is not defibrillator proof and should not be used in situations where a defibrillator is likely to be
used.
The equipment should not be used with, or in the presence of, high frequency surgical equipment.
Other patient-connected equipment
When used simultaneously with other patient-connected equipment, for example a cardiac pacemaker or
other electrical stimulator, it is unlikely that a safety hazard will arise. However always consult the documenta-
tion supplied with the other patient-connected equipment to ensure that all hazards, warnings and cautions
are considered before the equipment is used together.
Leakage current
This system is designed to comply with IEC 60601-1, the international standard for medical electronic equip-
ment, which specifies the permissible levels of leakage current. A potential hazard exists in the summation of
leakage currents caused by connecting several pieces of equipment together. Because this system can be used
in conjunction with standard electronic devices, the total leakage current should be tested whenever the sys-
tem is modified.
There should be no electrical connections between the system equipment, which is powered via a medical
grade power supply, and any other equipment powered from another mains supply.
2.7 Interference
The T4 will continue to operate in the presence of radio frequency magnetic fields (RF) and the effects of elec-
trostatic discharges (ESD) and other interference, in accordance with the requirements of EN60601-1-2. How-
ever, the T4 amplifier records signals of very low amplitude, and these signals themselves are not immune to