INSTRUCTION MANUAL
Mini-M™
RTD TRANSMITTER
(1 mA sensing current) MODEL M2RS1
BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.
■PACKAGE INCLUDES:
Signal conditioner (body + base socket).............................(1)
■MODEL NO.
Confirm Model No. marking on the product to be exactly
what you ordered.
■INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
POINTS OF CAUTION
■CONFORMITY WITH EU DIRECTIVES
• This equipment is suitable for Pollution Degree 2 and In-
stallation Category II (transient voltage 2500V). Rein-
forced insulation (signal input or output to power input:
300V) and basic insulation (signal input to output: 300V)
are maintained. Prior to installation, check that the insu-
lation class of this unit satisfies the system requirements.
• The equipment must be mounted inside a suitable fire
enclosure.
• Altitude up to 2000 meters.
• The equipment must be installed such that appropriate
clearance and creepage distances are maintained to con-
form to CE requirements. Failure to observe these re-
quirements may invalidate the CE conformance.
• The actual installation environments such as panel con-
figurations, connected devices, connected wires, may affect
the protection level of this unit when it is integrated in
a panel system. The user may have to review the CE re-
quirements in regard to the whole system and employ ad-
ditional protective measures to ensure the CE conformity.
• Install lightning surge protectors for those wires connect-
ed to remote locations.
■POWER INPUT RATING & OPERATIONAL RANGE
• Locate the power input rating marked on the product and
confirm its operational range as indicated below:
100 – 240V rating: 85 – 264V, 47 – 66 Hz, approx. 3 – 5VA
24V DC rating: 24V ±10%, approx. 3W
11 – 27V DC rating: 11 – 27V, approx. 3W
110V DC rating: 85 – 150V, approx. 3W
■GENERAL PRECAUTIONS
• Before you remove the unit from its base socket or mount
it, turn off the power supply and input signal for safety.
■ENVIRONMENT
• Indoor use.
• When heavy dust or metal particles are present in the
air, install the unit inside proper housing with sufficient
ventilation.
• Do not install the unit where it is subjected to continuous
vibration. Do not subject the unit to physical impact.
• Environmental temperature must be within -5 to +55°C
(23 to 131°F) with relative humidity within 30 to 90% RH
in order to ensure adequate life span and operation.
• Be sure that the ventilation slits are not covered with ca-
bles, etc.
■WIRING
• Do not install cables close to noise sources (relay drive
cable, high frequency line, etc.).
• Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
■AND ....
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data
sheet.
• With voltage output, do not leave the output terminals
shortcircuited for a long time. The unit is designed to
endure it without breakdown, however, it may shorten ap-
propriate life duration.
COMPONENT IDENTIFICATION
Model
SPAN
ZERO
Zero Adj.
Span Adj.
Body Base Socket
Connection
Diagram
Specifications
Fixing Screw
INSTALLATION
Loosen the fixing screw at the front of the unit in order to
separate the body from the base socket.
DIN Rail
35mm wide
Spring Loaded
DIN Rail Adaptor
■DIN RAIL MOUNTING
Set the base socket so that
its DIN rail adaptor is at
the bottom. Position the
upper hook at the rear side
of base socket on the DIN
rail and push in the lower.
When removing the socket,
push down the DIN rail
adaptor utilizing a minus
screwdriver and pull.
■WALL MOUNTING
Refer to “EXTERNAL DIMENSIONS.”
M2RS1
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-5127 Rev.3 P. 2 / 2
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the connection diagram on the side of the unit.
■EXTERNAL DIMENSIONS unit: mm (inch)
5
9
21.5 (.85)
70.5 (2.78)
84 (3.31)10 (.39) 15 (.59)
6 (.23)
59 (2.32)
DIN RAIL
35mm wide
[4 (.16)]
8–M3
SCREW
72 (2.83)
23 (.91)
2–4.2x5
(.17x.20)
MTG HOLE
6 (.24) deep
114 (4.49)
• When mounting, no extra space is needed between units.
4
8
12
14 13
1
4 (.16)
■CONNECTION DIAGRAM
1
4
5
+
–
OUTPUT
9
12
U(+)
V(–)
POWER
13
14
A
B
B
RTD
WIRING INSTRUCTIONS
■SCREW TERMINAL
Torque: 0.8 N·m
■SOLDERLESS TERMINAL
Refer to the drawing below for recommended ring tongue
terminal size. Spade tongue type is also applicable.
Applicable wire size: 0.25 to 1.65 mm2(AWG 22 to 16)
Recommended manufacturer: Japan Solderless Terminal
MFG.Co.Ltd, Nichifu Co.,ltd
6 (.24) max
3.3 (.13) max
mm (inch)
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage across the terminal
13 – 14 with a multimeter.
3) Input: Check voltage across the terminal 1 – 4 with a
sensitive voltmeter (with 20°C or 68°F, approx. 110mV
with Pt 100, approx. 55mV with Pt 50Ω).
If RTD wires are broken, the output goes over 100%
(below 0% with downscale) due to the burnout function.
Check leadwires in such a case.
4) Output: Check that the load resistance meets the de-
scribed specifications.
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any cali-
bration.
The output signal can be finely adjusted to match it to a re-
ceiving instrument, or to compensate input wire resistance
when the unit is combined with a zenor barrier. Follow the
regular calibration procedure explained below:
■HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 10 minutes.
1) ZERO: Apply 0% input and adjust output to 0%.
2) SPAN: Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 0% input.
4) When ZERO value is changed, repeat the above proce-
dure 1) – 3).
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 10 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.

Other M-system Transmitter manuals

M-system M3LT User manual

M-system

M-system M3LT User manual

M-system Mini-M M2EXT User manual

M-system

M-system Mini-M M2EXT User manual

M-system Mini-M M2RTS User manual

M-system

M-system Mini-M M2RTS User manual

M-system Mini-M M2TT User manual

M-system

M-system Mini-M M2TT User manual

M-system W2XR User manual

M-system

M-system W2XR User manual

M-system W5 Series User manual

M-system

M-system W5 Series User manual

M-system M5XREB User manual

M-system

M-system M5XREB User manual

M-system W2TS User manual

M-system

M-system W2TS User manual

M-system 10TG User manual

M-system

M-system 10TG User manual

M-system M2XRP2 User manual

M-system

M-system M2XRP2 User manual

M-system 6AC User manual

M-system

M-system 6AC User manual

M-system M5VF User manual

M-system

M-system M5VF User manual

M-system 10RS User manual

M-system

M-system 10RS User manual

M-system M3SXV User manual

M-system

M-system M3SXV User manual

M-system Mini-M M2LCS User manual

M-system

M-system Mini-M M2LCS User manual

M-system M3LM User manual

M-system

M-system M3LM User manual

M-system JPA2 User manual

M-system

M-system JPA2 User manual

M-system 10VK User manual

M-system

M-system 10VK User manual

M-system M5MS-5-R User manual

M-system

M-system M5MS-5-R User manual

M-system M2EXS User manual

M-system

M-system M2EXS User manual

M-system Mini-M M2XU User manual

M-system

M-system Mini-M M2XU User manual

M-system 10VS User manual

M-system

M-system 10VS User manual

M-system Mini-M M2RT User manual

M-system

M-system Mini-M M2RT User manual

M-system MLV User manual

M-system

M-system MLV User manual

Popular Transmitter manuals by other brands

Acromag IntelliPack 895M Series user manual

Acromag

Acromag IntelliPack 895M Series user manual

API DuoPak APD 2053 instructions

API

API DuoPak APD 2053 instructions

SPRIANO SPRIANO Series operating manual

SPRIANO

SPRIANO SPRIANO Series operating manual

Sonnet Podfreq user guide

Sonnet

Sonnet Podfreq user guide

GESTRA LRGT 16-3 Original Installation & Operating Manual

GESTRA

GESTRA LRGT 16-3 Original Installation & Operating Manual

Powercast Powercaster TX91501B user manual

Powercast

Powercast Powercaster TX91501B user manual

IDTOLIGHT NASHVILLE installation instructions

IDTOLIGHT

IDTOLIGHT NASHVILLE installation instructions

Key Digital KD-IP1080Tx operating instructions

Key Digital

Key Digital KD-IP1080Tx operating instructions

Primex UK SATELLITE TRANSMITTER Installation & user guide

Primex

Primex UK SATELLITE TRANSMITTER Installation & user guide

PR electronics 5333 Series manual

PR electronics

PR electronics 5333 Series manual

Altinex TP115-110 user guide

Altinex

Altinex TP115-110 user guide

Gigalink PYROMATE Programming instructions

Gigalink

Gigalink PYROMATE Programming instructions

Amana DT01* installation instructions

Amana

Amana DT01* installation instructions

Altinex DA1930CT user guide

Altinex

Altinex DA1930CT user guide

Absolute Process Instruments APD 7010 quick start guide

Absolute Process Instruments

Absolute Process Instruments APD 7010 quick start guide

NOSHOK KING-GAGE ES2 5501 Installation and operation manual

NOSHOK

NOSHOK KING-GAGE ES2 5501 Installation and operation manual

Sennheiser SI 406 HDI 406 user guide

Sennheiser

Sennheiser SI 406 HDI 406 user guide

Sony DCC-FMT50U operating instructions

Sony

Sony DCC-FMT50U operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.