SERVICE INSTRUCTIONS
Disasse bly
Ensure that the fluid suppl to the meter is
disconnected, and the line pressure is released
before disassembl , with the exception for repair or
maintenance to the LC Displa or PCB where
there is no necessit to isolate the meter from flow.
Refer to the exploded parts diagram on subsequent
pages for item numbers.
1a) Pulse Caps odels: Undo the conduit
connector, remove pulse cap (item 9) and
remove the wires from the pulse terminal board
(item 5).
1b) Standard LC Display: Mark the displa
orientation with a marking pen, unscrew the four
large screws (Item 26) on top of the LC Displa .
Carefull separate the LC Displa from the
plastic housing and disconnect the wires from
the pulse terminal block. (Refer to additional
Standard LC Displa
instruction manual).
1c) Deluxe LC Display: Mark the displa
orientation with a marking pen, remove the four
retaining screws on the displa face (Item 16).
Lift off the displa unit and remove the 9 pin
connector at the back of the displa unit. (Refer
to additional Deluxe LC Displa instruction
manual).
2) Remove the mounting adaptor plate and gasket
(Item 14).
3) Loosen the cap head screws (Item 7) that hold
down the meter cap (Item 4), remove the screws,
washers and lift off the cap.
4) Remove the o’ring (Item 2) from the o’ring groove
in the meter cap (Item 4).
5) Remove rotors (Item 3). Note the position of
the timing marks.
Reasse bly
3) Replace the rotors (Item 3) onto the shafts at 90dgr
to each other (refer Fig. 5). Check the operation of the
rotors b turning either b hand. If the rotors are
not in mesh correctl or do not move freel ,
remove one of the rotors and replace correctl at
90dgr to the other rotor.
Re-check the operation of the rotors.
4) Replace the o’ring (Item 2) into groove in the
meter cap, if the o’ring has grown or is damaged in an
wa replace it with a new part.
5) Replace the meter cap making sure that the locating
pin in the bod lines up with the hole in the meter cap.
Insert the cap head screws (Item 7) and tighten in
a diagonal sequence 1, 3, 2, 4, etc.
6) The replacement of cables and connectors are a
reversal of the disassembl procedure, replace
conduit fitting if required.
When replacing the Standard LC Displa or the Deluxe
LC Displa , confirm the orientation marks made on
disassembl are aligned then screw the register into
place.
7) Test the meter b turning the rotors with a finger
or b appl ing ver low air pressure (no more than a
good breath) to one end of the meter, before
returning the meter to the line.
Pulse Circuit Board (PCB) Notes:
The pulse PCB (Item 5) is fitted with (A) two reed
switches; (B) hall effect sensors; or (C) one reed
switch and one hall effect sensor.
The PCB board is fastened to the meter cap
(Item 4) b two screws and stand off’s.
All care and caution should be taken when
removing or handling the PCB as both the reed switch
and hall effect sensor are fragile.
Reed switch or hall effect sensors are not available
as individual replacement parts and are onl
available with the complete PCB (Item 5).
1) Before reassembling check the condition of
the rotors (replace if necessar ).
2) Check that the plug side of the rotors is facing
ou when inserting the rotors, the plug side of
the rotor is the magnet side.
There is no difference between rotor one or
rotor two.