manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. MadgeTech
  6. •
  7. Data Logger
  8. •
  9. MadgeTech Cryo-Temp Specification sheet

MadgeTech Cryo-Temp Specification sheet

Other MadgeTech Data Logger manuals

MadgeTech RFCO2RHTemp2000A Specification sheet

MadgeTech

MadgeTech RFCO2RHTemp2000A Specification sheet

MadgeTech HiTemp140-FP Specification sheet

MadgeTech

MadgeTech HiTemp140-FP Specification sheet

MadgeTech RHTemp101A Specification sheet

MadgeTech

MadgeTech RHTemp101A Specification sheet

MadgeTech Event101A User manual

MadgeTech

MadgeTech Event101A User manual

MadgeTech TSR101-EB User manual

MadgeTech

MadgeTech TSR101-EB User manual

MadgeTech MadgeNET RFTCTemp2000A Specification sheet

MadgeTech

MadgeTech MadgeNET RFTCTemp2000A Specification sheet

MadgeTech RTDTempX Series Specification sheet

MadgeTech

MadgeTech RTDTempX Series Specification sheet

MadgeTech SHOCK300 Specification sheet

MadgeTech

MadgeTech SHOCK300 Specification sheet

MadgeTech LF101A Specification sheet

MadgeTech

MadgeTech LF101A Specification sheet

MadgeTech DHS Specification sheet

MadgeTech

MadgeTech DHS Specification sheet

MadgeTech RFC1000 User manual

MadgeTech

MadgeTech RFC1000 User manual

MadgeTech pHTemp2000 User manual

MadgeTech

MadgeTech pHTemp2000 User manual

MadgeTech Shock101-5 User manual

MadgeTech

MadgeTech Shock101-5 User manual

MadgeTech Titan S8 Specification sheet

MadgeTech

MadgeTech Titan S8 Specification sheet

MadgeTech PRHTemp101 User manual

MadgeTech

MadgeTech PRHTemp101 User manual

MadgeTech X Series Specification sheet

MadgeTech

MadgeTech X Series Specification sheet

MadgeTech Process101A User manual

MadgeTech

MadgeTech Process101A User manual

MadgeTech pHTemp101 User manual

MadgeTech

MadgeTech pHTemp101 User manual

MadgeTech QuadTemp2000 User manual

MadgeTech

MadgeTech QuadTemp2000 User manual

MadgeTech PR140 User manual

MadgeTech

MadgeTech PR140 User manual

MadgeTech TC101A User manual

MadgeTech

MadgeTech TC101A User manual

MadgeTech HiTemp1400 User manual

MadgeTech

MadgeTech HiTemp1400 User manual

MadgeTech RTDTemp101A User manual

MadgeTech

MadgeTech RTDTemp101A User manual

MadgeTech DHS User manual

MadgeTech

MadgeTech DHS User manual

Popular Data Logger manuals by other brands

UGCS SkyHub user manual

UGCS

UGCS SkyHub user manual

Avisaro 4.0 Series user manual

Avisaro

Avisaro 4.0 Series user manual

DDB C3 manual

DDB

DDB C3 manual

TESTO 176 instruction manual

TESTO

TESTO 176 instruction manual

HBM SOMAT XR CX22B-R operating manual

HBM

HBM SOMAT XR CX22B-R operating manual

Agtrix 380i manual

Agtrix

Agtrix 380i manual

Beijing ART Tech PCI2320 user manual

Beijing ART Tech

Beijing ART Tech PCI2320 user manual

Athena GET SL1 user manual

Athena

Athena GET SL1 user manual

Motec Systems C185 user manual

Motec Systems

Motec Systems C185 user manual

Beijing ART Tech USB2832 user manual

Beijing ART Tech

Beijing ART Tech USB2832 user manual

HT Italia XL421 user manual

HT Italia

HT Italia XL421 user manual

Racelogic VBOX LapTimer quick start guide

Racelogic

Racelogic VBOX LapTimer quick start guide

Edu-Logger 900-240 Guide

Edu-Logger

Edu-Logger 900-240 Guide

Elma DT173 manual

Elma

Elma DT173 manual

Comet U0121M user guide

Comet

Comet U0121M user guide

pico Technology PT-104 user guide

pico Technology

pico Technology PT-104 user guide

Sensus DIAVASO user manual

Sensus

Sensus DIAVASO user manual

Soil Instruments GTecLink user manual

Soil Instruments

Soil Instruments GTecLink user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Cryo-Temp
To view the full MadgeTech product line,
visit our website at madgetech.com.
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE,
FORCE OVER DISCHARGE, DISASSEMBLE, CRUSH, PENETRATE OR INCINERATE. BATTERY MAY
LEAK OR EXPLODE IF HEATED ABOVE 100 °C (212 °F).
Specifications subject to change.
See MadgeTech’s terms and conditions at madgetech.com
MadgeTech, Inc.
6 Warner Road lWarner, NH 03278
Phone (603) 456.2011 lFax (603) 456.2012
madgetech.com l[email protected]
Cryo-Temp
Ultra Low Temperature Data Logger
Description Cryo-Temp
Temperature Sensor NTC Thermistor
Temperature Range -86 °C to +35 °C (-122.80 °F to +95 °F) recommended,
may be used up to +55 °C (131 °F)
Temperature Resolution 0.1 °C (0.18 °F)
Calibrated Accuracy +1.0 °C
Memory 32,767 readings; software configurable memory wrap
Reading Rate 1 reading every 5 seconds up to 1 reading every 30 minutes
LED Indicator Red, Green & Yellow
Manual Start Yes
Required Interface Package IFC300
Baud Rate 38,400
Battery Type 3.6V lithium battery (LTC-7PN);
Battery Life 2 years at +25 °C, 1 min. reading intervals
1 year at -80 °C, 1 min. reading intervals
Operating Environment -86 °C to +35 °C (-122.80 °F to +95 °F), 0 %RH to 100 %RH
Submergible Splash Resistant (IP64)
Material ABS plastic, white, polycarbonate labeling
Weight 1.4 oz (40 g)
Dimensions 3.3 in x 2.1 in x 0.4 in (83 mm x 54 mm x 11 mm)
Approvals CE
Product User Guide
DOC-1001035-00 REV 8 2019.08.26
Product Notes
Manual Start
To start the data logger with the
Manual Start option, use the
magnetic wand provided (in the
IFC300) to touch the start/mark switch. (manual start is defaulted to a 10
minute reading rate) The reading rate can be modified in the MadgeTech 4
Software. Connect the data logger, touch the wand to the start/mark switch,
the green LED (OK) will blink 5 times followed by the yellow LED (WARN) and
the red LED (ALARM), indicating the device is recording. The logger must be
manually stopped through the MadgeTech 4 software.
Marking Feature
A date and time stamp may be placed in the recorded data, by touching
the magnetic wand to the start/mark switch. The marking feature may be
configured to clear the alarm and warn indicators, while the logger is still active.
Submergibility
The Cryo-Temp is splash resistant, rated IP64.
LEDs
Green: Blinks to indicate safe conditions during acquisition.
Yellow: Blinks to indicate warning limits have been exceeded.
Red: Blinks to indicate alarm criteria (limits/delay) have been exceeded
Alarm Settings
To change the settings for the temperature alarm;
- Select Alarm Settings from the Device Menu in the MadgeTech Software. A window
will appear allowing to set the high and low temperature alarms and warning alarms.
- Press Change to edit the values.
- Check Enable Alarm Settings to enable the feature and check each high and low, warn
and alarm box to activate it. The values can be entered in the field manually or by using
the scroll bars.
- Click Save to save the changes. To clear an active alarm or warn, press the Clear Alarm
or Clear Warn button.
- To set an alarm delay, enter the duration of time into the Alarm Delay box in which the
readings can be outside of the alarm parameters.
Set Password
To password protect the device so that others cannot start, stop or reset the device;
- In the Connected Devices panel, click the device desired.
- On the Device Tab, in the Information Group, click Properties. Or, right-click the device
and select Properties in the context menu.
- On the General Tab, click Set Password.
- Enter and confirm the password in the box that appears, then select OK.
Cryo-Temp Product User Guide
Installation Guide
Installing the Interface cable
- IFC300
Refer to the “Quick Start Guide” included in the package.
Installing the software
The Software can be downloaded from the MadgeTech website at madgetech.com.
Follow the instructions provided in the Installation Wizard.
Device Operation
Connecting and Starting the data logger
- Once the software is installed and running, insert the Cryo-Temp into the IFC300 docking
station.
- Connect the USB end of the interface cable into an open USB port on the computer.
- The device will appear in the Connected Devices list, highlight the desired data logger.
- For most applications, select “Custom Start” from the menu bar and choose the desired
start method, reading rate and other parameters appropriate for the data logging
application and click “Start”. (“Quick Start” applies the most recent custom start options,
“Batch Start” is used for managing multiple loggers at once, “Real Time Start” stores the
dataset as it records while connected to the logger.)
- The status of the device will change to “Running”, “Waiting to Start” or “Waiting to
Manual Start”, depending upon your start method.
- Disconnect the data logger from the docking station and place it in the environment to
measure.
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device
cannot be restarted until it has been re-armed by the computer.
Downloading data from a data logger
- Connect the logger to the interface cable.
- Highlight the data logger in the Connected Devices list. Click “Stop” on the menu bar.
- Once the data logger is stopped, with the logger highlighted, click “Download”. You will
be prompted to name your report.
- Downloading will offload and save all the recorded data to the PC.
Device Maintenance
Battery Replacement & Calibration
When battery life is expended or device calibration is required, the Cryo-Temp data logger
must be returned to MadgeTech to participate in the product exchange program.
Please call a MadgeTech sales representative for details.
To start the device, hold
the magnetic wand over
the start/mark switch. (Also
indicated below)
Close-up of the
start/mark switch