POSITIONING OF THE SENSOR
The first step is to place the Medpage-MP5x in a safe place next to the bed that it will be used
in. The sensor is then to be placed underneath the mattress of the bed to be monitored. The
sensor should be positioned so that the person laying on it has maximum coverage over the
sensor, around the waist/torso area underneath the mattress is the best position. As you
progress through this instruction manual you will gain a better understanding for the best
place for the sensor. The key point is that the sensor reliably detects the movements that
occur, this will be tested later.
CHECK POWER
Now that the sensor is positioned correctly, plug the Medpage-MP5x power adaptor into a
mains power socket and turn on the power socket. To turn the unit on, push in the On/Off
switch on the left hand side of the front panel. The POWER LED will indicate the status of
the mains power supply, a GREEN LED means that the monitor is connected to the mains
power supply, a RED LED indicates a power failure or that the monitor is unplugged. Please
ensure that the POWER LED is GREEN before continuing. Should there be a power failure
during operation, an internal backup battery will provide 12 hours of use. To turn the monitor
off, push the on/off button and wait until the POWER and MOVEMENT LEDs turn off.
If the POWER LED doesnot show GREEN when plugged into a
mainssocket that isswitched on, please do not use the monitor as
the backup battery will not charge.
PREPARING THE RADIO PAGERS FOR USE
NOTE: The pagersare supplied with the batteriesfitted. (The
following prodcedureisonly required for replacing batteriesin the
future)
A) Undo the screw on the belt clip. Install the two AA batteries observing the polarity markings.
Replace the cover (batteriesare usually factory installed). Do not touch the button
between the batteries. NOTE: Batteriesrequire replacement when the LED flashesred
repeatedly at a slow rate.
B) Slide the on/off switch on the side of the pager to on. The pager will vibrate and beep, (if beep
is selected by alarm option switch), and the LED will flash green occasionally to indicate
pager on. If the pager does not vibrate or beep when switched on do not use the system. Check
that the batteries are correctly inserted. If the pager fails to respond contact your distributor
for further advice.
C) Select your choice of alert using the selector switch on front of the pager.
Position: Vib only = vibration with no sound. Vib + vol = vibration with low volume beeper. Vib + Vol
= vibration with high volume beeper.
MP5x User Instruction REV 01-12-17 Helpline 01536 264869 www.medpage-ltd.com