manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Millipore
  6. •
  7. Laboratory Equipment
  8. •
  9. Millipore SNAP i.d User manual

Millipore SNAP i.d User manual

Millipore, Milli-Q and Immobilon are registered trademarks of
Millipore Corporation.
SNAP i.d. and the M mark are trademarks of Millipore Corporation.
Tween is a registered trademark of ICI Americas Inc.
Apply appropriate volume of secondary antibody (see12. OPTIMIZATION
GUIDELINES on reverse side) evenly across the blot holder surface.
Incubate for 10 minutes at room temperature. Again, solution will be ab-
sorbed into the blot holder and surface may appear dry.Apply vacuum.
IMPORTANT: Do not apply vacuum until after the 10-minute
incubation.
With vacuum on, wash 3 times with buffer. See13. OPTIMIZATION
GUIDELINES on reverse side for volumes.
TURN VACUUM OFF.
Open lid of system and place8. blot holder in chamber, aligning
blot holder tabs with notches of
chamber. Close and latch lid.
Add volume of blocking solution9. as indicated under OPTIMIZA-
TION GUIDELINES on reverse
side. Using knobs on the system,
apply vacuum until well(s) are
completely empty.
TURN VACUUM OFF.
Add volume of primary anti-10. body as indicated under
OPTIMIZATION GUIDE-
LINES on reverse side.
Antibody solution must evenly
cover entire blot holder
surface.
Incubate for 10 minutes at
room temperature. Solution
will be absorbed into the blot
holder and surface may ap-
pear dry.Apply vacuum.
IMPORTANT: Do not ap-
ply vacuum until after the
10-minute incubation.
++
With vacuum running11. continuously, wash 3 times
with wash buffer. See
OPTIMIZATION GUIDE-
LINES on reverse side for
volumes.
TURN VACUUM OFF.
3X
Remove blot and incubate with the14. appropriate detection reagent such
as Immobilon®HRP, or, if using Milli-
pore uorescently labeled antibod-
ies, visualize.
Squeeze rmly at base of tab 7. area to secure lid.
Close the blot holder lid.6.
Thoroughly wet the white2. surface of the blot holder with
Milli-Q®water.
NOTE: If using only one
well of a double or triple
well blot holder, the unused
well(s) must also be wet.
Place the spacer (wetting not5. necessary) on top of the blot
membrane and roll again to
ensure contact of spacer with
blot membrane.
Roll blot membrane gently to4. remove air bubbles.
Open the blot holder lid, taking1. care not to damage the inner
white surface.
OVERVIEW OF PROCEDURE
Before using the SNAP i.d. Protein Detection System, please read the User Guide completely.
Place the pre-wet blot in the3. center of the blot holder with
the protein side down. The blot
membrane should not exceed
size specied in the User Guide.
00103870 Rev. A, 01/08
AVID QS 021108.indd 1 2/11/2008 6:40:46 PM
Antibody Volume and Concentration
Most users will be able to use the same amount of antibody, but in 1/3 to 1/5 the■volume at 3–5 fold higher concentration.
Standard
Immunodetection SNAP i.d.
Immunodetection
Mass of antibody
required 1 µg 1 µg
Stock solution 1 mg/mL 1 mg/mL
Diluted stock 1:10,000 (0.1 µg/mL) 1:3,333 (0.33 µg/mL)
Volume required for assay 10 mL 3 mL
Antibody used 0.1 µg/mL × 10 mL = 1 µg 0.33 µg/mL × 3 mL = 1 µg
This guideline is intended as a starting point to develop the nal antibody concentration necessary for desired
performance. Because each antibody is different, it may be necessary to adjust the blot exposure time, antigen
load or both.
Blot Blocking Concentration
The use of non-fat/low fat dry milk at concentrations higher than 0.5% is not recom-■
mended, as this will result in clogging of the blot holder and prevent reagent ow.
Blocking agents should be prepared in tris or phosphate buffered saline solutions contain-■ing 0.1% Tween20 surfactant, to reduce surface tension and ensure even distribution of
blocking agent across the blot holder surface.
The SNAP i.d. system is compatible with the most commonly used blocking agents. Refer■
to User Guide for complete list with recommended concentrations.
In order to insure optimal ow through the blot holder, it is essential that blocking solu-■tions be completely solubilized and free of all particulate matter. In some cases, it may be
necessary to reduce the concentration of the blocking agent to achieve the required ow.
Blocking, Antibody and Wash Recommended Volumes
Single well Double well Triple well
Blocking solution volume 30 mL/well 15 mL/well 10 mL/well
Antibody volume 3 mL/well 1.5 mL/well 1 mL/well
Wash buffer* volume 30 mL/well 15 mL/well 10 mL/well
It is not necessary to use all the wells of double and triple well blot holders, but unused wells must
be wet out with Milli-Q water.
* Tris or phosphate buffered saline solutions, supplemented with 0.1% Tween® 20 surfactant.
OPTIMIZATION GUIDELINES
AVID QS 021108.indd 2 2/11/2008 6:40:46 PM

Other Millipore Laboratory Equipment manuals

Millipore Elix 35 User manual

Millipore

Millipore Elix 35 User manual

Millipore PR00SIA1 Use and care manual

Millipore

Millipore PR00SIA1 Use and care manual

Millipore Steritest Symbio User manual

Millipore

Millipore Steritest Symbio User manual

Millipore XX11 000 00 User manual

Millipore

Millipore XX11 000 00 User manual

Millipore Amicon Ultra-2 User manual

Millipore

Millipore Amicon Ultra-2 User manual

Millipore Amicon 6028 User manual

Millipore

Millipore Amicon 6028 User manual

Millipore Millicell MDCI10000 User manual

Millipore

Millipore Millicell MDCI10000 User manual

Millipore Samplicity User manual

Millipore

Millipore Samplicity User manual

Millipore Milliflex Oasis User manual

Millipore

Millipore Milliflex Oasis User manual

Millipore Amicon Ultra-4 Series User manual

Millipore

Millipore Amicon Ultra-4 Series User manual

Millipore APC ErgoTouch Pro 2 User manual

Millipore

Millipore APC ErgoTouch Pro 2 User manual

Millipore Labscale User manual

Millipore

Millipore Labscale User manual

Millipore Viresolve Pro VPMCPDKNB9 User manual

Millipore

Millipore Viresolve Pro VPMCPDKNB9 User manual

Millipore Spectroquant Multy User manual

Millipore

Millipore Spectroquant Multy User manual

Millipore M Air T User manual

Millipore

Millipore M Air T User manual

Millipore XX4004700 User manual

Millipore

Millipore XX4004700 User manual

Millipore milli-q User manual

Millipore

Millipore milli-q User manual

Millipore Millicell ERS-2 User manual

Millipore

Millipore Millicell ERS-2 User manual

Millipore Direct Detect User manual

Millipore

Millipore Direct Detect User manual

Millipore MERCK EZ-Pak User manual

Millipore

Millipore MERCK EZ-Pak User manual

Millipore 8003 User manual

Millipore

Millipore 8003 User manual

Millipore ReadyStream User manual

Millipore

Millipore ReadyStream User manual

Millipore Helicon S10 User manual

Millipore

Millipore Helicon S10 User manual

Millipore Milli-Q Advantage A10 User manual

Millipore

Millipore Milli-Q Advantage A10 User manual

Popular Laboratory Equipment manuals by other brands

Drucker Diagnostics Horizon 24 Service manual

Drucker Diagnostics

Drucker Diagnostics Horizon 24 Service manual

Cole Parmer pH 550 Operation manual

Cole Parmer

Cole Parmer pH 550 Operation manual

AS QC2 Series operating instructions

AS

AS QC2 Series operating instructions

Molecular Devices SkanWasher 300 user guide

Molecular Devices

Molecular Devices SkanWasher 300 user guide

Aison STE16L-C instruction manual

Aison

Aison STE16L-C instruction manual

IKA ROTAVISC manual

IKA

IKA ROTAVISC manual

Buchi Inert Loop S-395 Operation manual

Buchi

Buchi Inert Loop S-395 Operation manual

Bruker Ascend 400'54 user manual

Bruker

Bruker Ascend 400'54 user manual

GHM Delta OHM PMsense-M operating manual

GHM

GHM Delta OHM PMsense-M operating manual

Nederman HD82 instruction manual

Nederman

Nederman HD82 instruction manual

Sigma 3-30K operating manual

Sigma

Sigma 3-30K operating manual

Molecular Devices QPix 450 user guide

Molecular Devices

Molecular Devices QPix 450 user guide

Olink Signature Q100 user manual

Olink

Olink Signature Q100 user manual

VistaLab Ovation 9060-1011 Replacement instructions

VistaLab

VistaLab Ovation 9060-1011 Replacement instructions

Thermo Scientific Dionex Easion Operator's manual

Thermo Scientific

Thermo Scientific Dionex Easion Operator's manual

Yamato BK300 instruction manual

Yamato

Yamato BK300 instruction manual

FMS ZipThaw 202 user manual

FMS

FMS ZipThaw 202 user manual

NanoEnTek EVE PLUS user manual

NanoEnTek

NanoEnTek EVE PLUS user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.