5. PRECAUTION & PREPARATIONS FOR
MEASUREMENT
* Please check carefully if the meter's power supply is
AC 110 V or AC 220 V before operating the meter.
There is a label at the rear of the meter to show the
proper power source outlet AC 110 V or AC 220 V.
* Please don't input voltage to input terminal ( Lo, Rx1,
Rx2, Hi ) to prevent any internal circuit damage.
6. MEASURING PROCEDURES
1) Rotate the " Power off/Range Rotary Switch " ( 3-5, Fig. 1 )
to the convenient & right range 200 m ohm, 2000 m ohm,
20 ohm, 200 ohm, 2000 ohm.
2) After select the above right range, then should make
the following " ZERO ADJ. PROCEDURE " :
a. Short the 2 clips.
b. Rotate the " Zero Adjust Knob " (3-4, Fig. 1) until
the display reading values show the zero values.
Consideration :
* Zero adjustment is necessary for range 200 m
ohm, 2000 m ohm, 20 ohm only.
* Recommend user select 200 m ohm range to
make "ZERO ADJ. PROCEDURE"
3) Connect the 2 clips as following Fig. 3 to measure
the unknown resistance.
5