10 11
Familiarizing Yourself with the Nasco Life/form
®
Child Blood Pressure Simulator Control Panel
Under the display window are three buttons: Menu,
Gap, and Calibrate. (See figure 5.) The systolic
pressure is set by pressing the Menu key once.
The pressure is adjusted up or down using the up
and down arrow keys. (See figure 6.) The diastolic
pressure is set by pressing the Menu key a second
time. Adjust the setting up and down with the arrow
up and down keys. (See figure 7.) The heart rate
is set by pressing the Menu key a third time and
adjusting the rate with the arrow up and down keys.
The pulse rate can be set from 0 beats per minute
to 300 beats per minute. (See figure 8.) The palpa-
tion can be set to either on or pulseless. When the
pulseless setting is used, the diastolic and systolic
pressures will automatically be set to 0.
Figure 5
Figure 6
Figure 7
Figure 8
PALPABLE PULSE FEATURE
The Nasco Life/form
®
Child Blood Pressure
Simulator also incorporates a palpable pulse at the
radial location. (See figure 9.) Palpations can be
felt upon start-up of the unit or after blood pressure
settings have been made. Press the Menu key
repeatedly until “Set PALPATION” menu appears.
“Pulse ON” is defaulted and enables the palpation
feature. The pulse in the wrist will begin when the
electronic box is turned on. During the actual blood
pressure reading, the palpable pulse will turn off
when the cuff is inflated and surpasses the systolic
set point. It will turn on again when the cuff is
deflated 20 mmHg BELOW the diastolic set point.
Pulse Location
Figure 10 Figure 11
Note: The pulse feature is turned off during this time to allow students to clearly hear the Korotkoff
sounds through the stethoscope without interference from the background noise of the pulse feature in
the wrist. The palpable pulse is delicate and should be palpated lightly, as you would with a real
patient. Pressing too hard can damage the pulse feature.
Make sure the cable from the blood pressure arm is properly connected to the electronic box and the
palpation feature is in the “PULSE ON” mode.
A pulseless condition can be simulated by switching your unit to the “pulseless” mode. Press the Menu
key repeatedly until “Set PALPATION” menu appears. By pressing the down arrow at this point, palpa-
tions can be disabled, causing the simulator to be pulseless. When in the “pulseless” mode, all settings
are automatically reset to 0 and all blood pressure sounds are disabled.
The pulseless setting will also turn off the sounds in the arm. The pulse will always be on unless the
pulseless feature is activated or if the systolic or heart rate levels are set to zero. To do this, press the
Menu key four times. The down arrow key will set the pulse to pulseless. Press the arrow up key to turn
the pulse back on. (See figure 10.) Located to the right of the Menu key is the auscultatory Gap key. (See
figure 11.) This key is included to simulate the auscultatory gap that is sometimes present between phase
1 and phase 2 sounds in which no audible sound is noted during this portion of auscultation. This control
function is included so that the trainee can become familiar with this phenomenon. Pressing the Gap
key simply turns the gap function off or on. When the key is pressed, a message will briefly appear that
the auscultatory gap is enabled or disabled. Also the main display will show (at the bottom right of the
display) either AGap:Y (for on) or AGap:N (for off).
The arrow up and down keys also control the volume of the sounds that are present in the arm. From the
main menu, press the up arrow key to increase the volume, press the down arrow key to decrease the vol-
ume. The volume levels can be adjusted from level 1 (the lowest volume) to level 7 (the highest volume).
USING THE NASCO
Life/form®
CHILD BLOOD PRESSURE SIMULATOR
First, verify that the pressure line tubing from the sphygmomanometer and the audio line coming from
the simulated arm assembly are properly connected to the blood pressure simulator unit, as previously
described in the setup procedures. Apply the sphygmomanometer cuff and gauge to the simulated arm
assembly in the usual manner. Apply the stethoscope to the simulated arm, also in the usual manner.
Set the systolic and diastolic controls to the desired levels. Select the auscultatory gap if desired. Finally,
adjust the pulse rate control to the desired setting.
To proceed with the simulated blood pressure measurement, first close the valve on the sphygmoma-
nometer bulb tightly and begin pumping air into the cuff until the gauge reads higher than the preset
systolic level chosen. Once this point is reached, loosen the valve on the bulb slightly to allow the gauge
pressure reading to decrease slowly. While monitoring the arm assembly with the stethoscope, note the
point on the sphygmomanometer gauge when the first Korotkoff sound is heard. This will be the systolic
blood pressure. Allow the pressure in the cuff to continue to decrease until the point at which the last
pulse is heard, noting the reading on the gauge. This is the diastolic blood pressure.