manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. NeuLog
  6. •
  7. Accessories
  8. •
  9. NeuLog NUL-220 User manual

NeuLog NUL-220 User manual

NEULOG CO2 LOGGER SENSOR GUIDE
NeuLog CO2logger sensor
NUL-220
The CO2Sensor can be used for any science experiment or activity where CO2levels
are required. Some fields of study include Chemistry, Physiology, Exercise Science,
Biochemistry, Biology, Earth Science, Environmental Science, etc.
This sensor is practical for testing CO2levels both in laboratories and in the field. An
electromechanical cell in the sensor reacts with carbon dioxide to produce a voltage
which is directly interpreted to give a CO2 concentration reading.
Just a few of the thousands of possible experiments that can be done with the NUL-
202 sensor are: Monitoring of ecological systems, combustion reactions, CO2
consumption rates, studying photosynthesis, testing chemical solutions, hydrogen
peroxide decomposition, and many more.
The NeuLog CO2 sensor collects data in the following units of measurement:
Parts per million (ppm): The total number of carbon dioxide molecules per
million molecules sampled.
Sensor calibration:
Prior to each use the CO2 needs to warm up for roughly half an hour. To do this, plug
the CO2 sensor into a USB module or a battery pack and allow the sensor to sit for half
an hour prior to collecting data for the best results.
Open air mode:
The CO2sensor needs to be calibrated before use. Calibration of the probe assumes
a standard atmospheric CO2concentration of 20.9% (assuming stable levels in the
Earth’s atmosphere at sea level).
Materials Needed:
NUL-220 CO2sensor
1 of the following (depending on experiment type)
oUSB-200 USB Module (For On-line experiments)
oBAT-200 Battery (For Off-line experiments)
oVIEW-101 Viewing Module and BAT-200 Battery (For use with
NeuLog Viewer)
o
Procedure:
1. The CO2sensor needs a power supply to properly calibrate so connect the
CO2sensor to:
a. For On-line experiments: USB-200 USB Module.
b. For Off-line experiments: BAT-200 Battery
c. For use with the NeuLog Viewer: VIEW-101 Viewing Module
and BAT-200 Battery.
2. The sensor now needs to be calibrated:
a. Connect the CO2sensor to the left side of a USB-200 Module
(which is connected to a computer via the USB to mini USB
cable).
b. Open the NeuLog software.
c. Once the CO2sensor is detected click the “Module setup” button.
d. Hold the probe vertically so the opening is facing the ground.
e. Press and hold the blue “Start/Stop” button the faceplate of the
sensor for 3 seconds.
3. The CO2Sensor is now calibrated for use.
Quick start procedure:
PC or Mac Computer
Materials needed:
NUL-220 CO2Sensor
USB-200 USB Module
A USB to mini USB cable (which comes with the USB-200)
Prior to initial use the NeuLog CO2sensor needs to be calibrated, please review the
“Sensor calibration” section of this document on beginning on page 2 for detailed
instructions. Your CO2sensor needs to be connected to a USB-200 module. The
USB-200 module then connects to a computer via a USB to mini-USB cable. Please
note that you cannot plug the CO2sensor directly into the computer.
NEULOG CO2 LOGGER SENSOR GUIDE
Resident PC software and browser based software can be downloaded for free at
www.NeuLog.com/dowload as well as a full software user guide. Note: Make sure not
to download and install both types of software, they will conflict on the computer.
Procedure:
1. Calibrate the CO2sensor. (Follow the ‘Sensor Calibration” guide in this
document)
2. Install the NeuLog software
3. Connect the USB-200 module to the PC or Mac
4. Connect the CO2sensor to the USB-200 module (they directly plug
together).
5. Open the NeuLog software.
6. Once a CO2sensor logo appears on the left side of the screen the probe
has been automatically identified and you can begin experimentation.
7. If the CO2sensor is not automatically identified then click the “Search for
sensors” icon to find the sensor.
8. Select the “On-line experiment” button; this will open a graph below.
9. Click on the “Module setup” button located on the CO2sensor icon in the
module window to change the sensor settings if need be.
10. Click on the experiment set up button to change the experiment settings if
need be (experiment duration for example).
11. The CO2sensor will give a live reading in the box to the left of the screen
while plugged in.
12. To run an experiment and collect data click “Run experiment”.
13. To end data collection early, click “Stop experiment”.
Tablet, smart phone device
Materials needed:
NUL-220 CO2Sensor
WIFI-201 WIFI module
BAT-200 Battery
Prior to initial use the NeuLog CO2sensor needs to be calibrated, please review the
“Sensor calibration” section of this document on beginning on page 2 for detailed
instructions. Your CO2sensor needs to be connected to a WIFI-201 module. The
WIFI-201 module will create a closed NeuLog wifi network which will stream the
NeuLog data to a device of your choosing. Once your device is wirelessly connected
to the NeuLog network you can run experiments and collect data through a browser of
your choosing.
Procedure:
1. Calibrate the CO2sensor. (Follow the ‘Sensor Calibration” guide in this
document)
2. Connect the CO2sensor directly to the left side of a WIFI-201 module (no
wires required).
3. Connect a BAT-200 module to the right side of the WIFI-201 module.
4. Although not required, we recommend plugging the BAT-200 to an outlet
using a USB to micro USB charger (such as a typical cell phone charger).
The WIFI-201 module will run for 15-75 minutes (depending on the sensor)
without being plugged in.
5. For further WIFI-201 instructions or the WIFI-201 quick start guide please
visit www.NeuLog.com/download.
6. The WIFI-201 indicator lights will flash; take no action until the LED to the
far left turns blue.
7. Take your tablet or smart phone and go to the Wi-Fi settings and select the
NeuLog network which matches the WIFI module ID found on the back of
the WIFI-201 device. (NeuLogXXXX for example)
8. Give your device 1-2 minutes to connect to the WIFI-201 network.
9. Once the device is connected go to your browser and type in the website
wifi201.com into the URL bar, then wait for 30-60 seconds.
10. You will see a “Control mode” icon in the browser, click on this icon.
11. The browser will then load a new screen and begin to auto detect the
sensors, this can take a few minutes.
12. If the browser does not auto detect, select “Search for sensors”
13. Once the sensor is found you will see an icon on the left side of the screen
for the CO2sensor. The icon will display data in real time.
14. To set the CO2sensor’s settings click on “Module setup” button located on
the CO2sensor icon on the left side of your screen.
15. To change the experimental settings click the “Experiment setup” button
(experiment duration for example).
16. To run an experiment and collect data click “Run experiment”.
NEULOG CO2 LOGGER SENSOR GUIDE
Operation with NeuLog viewer
Materials needed:
NUL-220 CO2 Sensor
VIEW-101 Viewing Module
BAT-200 Battery
Procedure:
1. Calibrate the CO2sensor. (Follow the ‘Sensor Calibration” guide in this
document).
2. Connect the CO2 sensor to the left side of the viewer.
3. Connect the BAT-200 module to the right side of the viewer.
4. The VIEW-101 will auto detect the sensor and you will see it appear on the
left side of the screen.
5. Once the sensor appears it will be monitoring data in real time.
6. To run an experiment and collect data click the run button (little green
person).
Off-line experiments
(Off-line experiments are for when you do not have a sensor connected directly to a
computer, tablet, smartphone, or NeuLog viewer).
Materials needed:
NUL-220 CO2Sensor
BAT-200 Battery
Materials needed to configure your offline experiment:
USB-200 USB Module or WIFI-201 WIFI module or VIEW-101 Viewing Module
A USB to mini USB cable (which comes with the USB-200)
Procedure:
1. Calibrate the CO2sensor. (Follow the ‘Sensor Calibration” guide in this
document)
2. Connect the CO2sensor directly to the left side of a charged BAT-200
module.
3. When ready to collect data press the “Start/Stop” button on the CO2sensor,
a red light will turn on for the duration of the experiment.
4. To change the experiment settings first connect to your chosen device (PC,
Mac, tablet, smart device) and select the “Module setup” button. (For more
information on how to change the experiment settings view quick start
procedure section for your chosen device in this document.)
5. After the experiment has concluded (when the “Start/Stop” button is pressed
again or the data collection period ends) connect the sensor to your chosen
device. (full instructions in each device’s quick start procedure section)
6. Open NeuLog software.
7. Click the “Off-line experiment” button.
8. Click the “Load data from sensors” button.
9. Select which experimental data to upload. (5 experimental runs can be
stored at one time on the NUL-220 CO2sensor)
Restoring sensor’s factory default settings:
Procedure:
1. Connect the USB-200 to a PC, Mac, tablet, or smart device.
2. Connect the CO2sensor to a USB-200 module (they directly plug together).
3. Open the NeuLog software.
4. Click the “Tools” icon.
5. Click “Restore sensor’s factory defaults” icon.
6. If prompted to clear the graph, click either the “Off-line experiment” or “On-
line experiment” (whichever you are using).
7. Click “Clear experiment results”.
8. Resume from step 4.
Included with sensor:
NeuLog CO2sensor probe which is attached to the sensor body by a
durable rubber-coated wire.
NeuLog NUL-220 CO2Sensor Guide (This Document)
NEULOG CO2 LOGGER SENSOR GUIDE
Sensor specifications:
Range and operation modes
350-10,000 ppm
ADC resolution
10 bit
Resolution
1 ppm
Max sample rate (S/sec)
100
Sensor features:
Fully digital data
Rugged plastic ergonomic case
Push button switch for Start/Stop experiments in off line mode
LED indicator of experiment status (blinks while collecting data)
Attached probe cased in durable plastic
Note: NeuLog products are intended for educational use.
Videos and experiment examples:
Videos, literature and other probes can be found at www.NeuLog.com
http://www.NeuLog.com/Products/CO2loggerSensor.aspx
Photosynthesis -
http://www.neulog.com/Portals/3/Users/Experiment B-10 Photosynthesis Ver
3.2.5.pdf
Technical background:
The philosophy behind NeuLog’s plug and play technology is based on each sensor’s
ability to store its own data due to an internal flash memory chip and micro-controller
in each plastic NeuLog body. This technology allows the sensor to collect and then
store the digital data in the correct scientific units (oC, oF, Lux, %, ppm, for example).
The sensor needs to be calibrated prior to initial use, please review the “Sensor
Calibration” guide in the beginning of this document for detailed instructions.
Inside the probe a small incandescent light bulb emits infrared radiation which
interacts and is absorbed by carbon dioxide molecules. An infrared sensor detects
how much infrared radiation was not affected by carbon dioxide molecules and in turn
will heat up from the radiation it is collecting. This heat produced is inversely related
to the amount of carbon dioxide present. More carbon dioxide present will produce
less heat and vice versa. The heat acts as a resistor in the wires and based on how
much electrical resistance there is a carbon dioxide reading can be determined
through calculation.
Maintenance and storage:
Never submerge the NeuLog sensor’s plastic body in any liquid.
Do not allow liquid into the NeuLog sensor’s plastic body.
After using the probe wipe off all excess material, liquid or residue from the
CO2probe body.
Store in a box at room temperature out of direct sunlight.
Do not put the plastic case, or cord of the CO2probe in an acids or bases.
Warranty:
We promise to deliver our sensor free of defects in materials and workmanship for a
period of 3 years from the date of purchase. Our warranty does not cover damage of
the product caused by improper use, abuse, or incorrect storage. Sensors with a shelf
life such as ion selective proves have a warranty of 1 year. Should you need to act
upon the warranty please contact your distributor. Your sensor will be repaired or
replaced.
Thank you for using NeuLog!
Flexible, simple, fast, forward thinking.
W: www.neulog.com
E: info@neulog.com
A: 850 St Paul Street, Suite 15, Rochester, NY 14605
P: 1.866.553.8536
V04302013

Other NeuLog Accessories manuals

NeuLog NUL-207 User manual

NeuLog

NeuLog NUL-207 User manual

NeuLog NUL-242 User manual

NeuLog

NeuLog NUL-242 User manual

NeuLog NUL-245 User manual

NeuLog

NeuLog NUL-245 User manual

NeuLog NUL-215 User manual

NeuLog

NeuLog NUL-215 User manual

NeuLog NUL-236 User manual

NeuLog

NeuLog NUL-236 User manual

NeuLog NUL-232 User manual

NeuLog

NeuLog NUL-232 User manual

NeuLog NUL-210 User manual

NeuLog

NeuLog NUL-210 User manual

NeuLog NUL-201 User manual

NeuLog

NeuLog NUL-201 User manual

NeuLog NUL-233 User manual

NeuLog

NeuLog NUL-233 User manual

NeuLog NUL-227 User manual

NeuLog

NeuLog NUL-227 User manual

NeuLog NUL-236 User manual

NeuLog

NeuLog NUL-236 User manual

NeuLog NUL-221 User manual

NeuLog

NeuLog NUL-221 User manual

NeuLog NUL-233 User manual

NeuLog

NeuLog NUL-233 User manual

NeuLog NUL-202 User manual

NeuLog

NeuLog NUL-202 User manual

NeuLog NUL-223 User manual

NeuLog

NeuLog NUL-223 User manual

NeuLog NUL-235 User manual

NeuLog

NeuLog NUL-235 User manual

NeuLog NUL-205 User manual

NeuLog

NeuLog NUL-205 User manual

NeuLog NUL-231 User manual

NeuLog

NeuLog NUL-231 User manual

NeuLog NUL-211 User manual

NeuLog

NeuLog NUL-211 User manual

NeuLog NUL-206 User manual

NeuLog

NeuLog NUL-206 User manual

NeuLog NUL-202 User manual

NeuLog

NeuLog NUL-202 User manual

NeuLog NUL-208 User manual

NeuLog

NeuLog NUL-208 User manual

NeuLog NUL-231 User manual

NeuLog

NeuLog NUL-231 User manual

NeuLog NUL-206 User manual

NeuLog

NeuLog NUL-206 User manual

Popular Accessories manuals by other brands

Tattletale Glass Break Detector Guide

Tattletale

Tattletale Glass Break Detector Guide

Hanger Clinic Migo user guide

Hanger Clinic

Hanger Clinic Migo user guide

Berker RolloTec Operation and assembly instructions

Berker

Berker RolloTec Operation and assembly instructions

NATURE & DECOUVERTES GOLDEN MOON DIFFUSER manual

NATURE & DECOUVERTES

NATURE & DECOUVERTES GOLDEN MOON DIFFUSER manual

Avaya M-ACCF/SF user guide

Avaya

Avaya M-ACCF/SF user guide

TMT Automation ChowBELL user manual

TMT Automation

TMT Automation ChowBELL user manual

Novital Covatutto 16L user manual

Novital

Novital Covatutto 16L user manual

PROZIS LUMINY user manual

PROZIS

PROZIS LUMINY user manual

Rega Planar RP78 instructions

Rega

Rega Planar RP78 instructions

Atek AWP820 user manual

Atek

Atek AWP820 user manual

Weka Holzbau 609.3260.50.00 Assembly instructions

Weka Holzbau

Weka Holzbau 609.3260.50.00 Assembly instructions

Sensopart FT 50 RLA-70 Mounting and operating instructions

Sensopart

Sensopart FT 50 RLA-70 Mounting and operating instructions

Dolby Laboratories Program Optimizer DP600 manual

Dolby Laboratories

Dolby Laboratories Program Optimizer DP600 manual

nedis PEBL140CWT quick start guide

nedis

nedis PEBL140CWT quick start guide

Balluff BAE SA-OH-059-YI-DP02 user guide

Balluff

Balluff BAE SA-OH-059-YI-DP02 user guide

Telco Sensors SpaceMaster Series user manual

Telco Sensors

Telco Sensors SpaceMaster Series user manual

Axis F2105-RE installation guide

Axis

Axis F2105-RE installation guide

Omron ZN-DPS1- *** -S Series instruction sheet

Omron

Omron ZN-DPS1- *** -S Series instruction sheet

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.