manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. NeuLog
  6. •
  7. Accessories
  8. •
  9. NeuLog NUL-219 User manual

NeuLog NUL-219 User manual

NEULOG COLORIMETER LOGGER SENSOR GUIDE
NeuLog colorimeter logger sensor
NUL-219
The colorimeter sensor can be used for any science experiment where liquid samples
of different colors and concentrations are utilized such as in the fields of Chemistry,
Biology, Pharmacology, Biochemistry, Environmental Science, Physics, or even just to
study the behavior of light waves.
Colorimeters are used to determine absorbance or transmittance of given wavelengths
(generally red, green, and blue) in a liquid solution. Absorbance and transmittance are
directly related to the concentration of known solutes in the solution which is
calculated using the Beer-Lambert law. One parameter can be calculated from the
other.
The sensor comes pre-calibrated so you can start experimentation right out of the box
using this guide.
Hundreds of possible experiments that can be performed with the NUL-219 sensor
are: Analyzing chemical reactions, studying Beer’s Law, water quality testing,
monitoring of ecological systems, study of photosynthesis, studying light waves. The
colorimeter could take the place of a spectrophotometer in many experiments.
The NeuLog colorimeter sensor's measurement units are:
% Transmittance: The amount of light that passes through a sample
Absorbance: The amount of light which is absorbed by a sample
Cuvette usage and handling:
It is important to maintain a standard for cuvette usage as their quality will directly
impact the results of your experiments. If the cuvette has fingerprints, smudges, or
cracks the sensor will detect them and give off an improper colorimeter reading.
Smudges left from handling can easily be wiped off with tissue paper.
Proper usage:
1. Carefully remove the cuvettes from their packaging.
2. Load your liquid sample into the cuvette before placing it into the sensor to
prevent spilling and damaging the sensor.
3. Plug the cap into the cuvette and gently invert a few times. Shaking will create
air bubbles which will be detected by the sensor and give an incorrect reading.
4. Before placing the cuvette in the sensor; hold it as far towards the top as you
can and wipe off the bottom with a tissue. Do not touch the bottom of the cuvette
again as this will leave a residue which could affect the results of the experiment.
5. If you have a cuvette with four clear sides, place it in the colorimeter sensor.
If you have a cuvette with two opaque sides:
a. Determine which sides of the cuvette are clear and which sides have an
opaque finish.
b. The clear opposing sides must be positioned facing the front and back of the
sensor (they won’t fit the other way) while the glossy sides face either side.
c. If the cuvette does not fit with a gentle push, rotate 90° in either direction
and try again.
6. After experimentation, empty your samples, gently rinse out the cuvettes with
distilled water and let them air-dry.
Selecting a wavelength:
Selecting the correct wavelength is important to get the best data and can easily be
determined using one of two different methods. There is no incorrect wavelength, just
some which are more informative than others. Lab experiments often come with a
predetermined wavelength to use. If your experiment does not, follow one of these
methods for best results:
Knowing that the color we see solutions is based on the color reflected from them
(and not absorbed): Choose a color you know will be absorbed. For example, if
the solution is red try the blue LED option.
Once the sensor is connected and the NeuLog software is open (follow one of
the guides below for more instruction) click the “Module setup” button in the
module box and try the different LED color options to determine which has the
highest absorbance (it will appear in real time on the sensor’s module box).
Select the color which has the highest absorbance reading.
NEULOG COLORIMETER LOGGER SENSOR GUIDE
Quick start procedure:
PC or Mac Computer
Materials needed:
NUL-219 Colorimeter Sensor
Cuvettes + Cuvette Lids
USB-200 USB Module
USB to mini USB cable (included with the USB-200 module)
Your colorimeter sensor needs to be connected to a USB-200 module. The USB-200
module then connects to a computer via a USB to mini-USB cable. Please note that
you cannot plug the colorimeter sensor directly into the computer.
The browser based application can be downloaded for free at www.NeuLog.com as
well as a full application user guide.
Choose "Downloads" on the main menu and then choose "Software and Application".
Procedure:
1. Install the NeuLog application.
2. Connect the USB-200 module to the PC or Mac.
3. Connect the colorimeter sensor to the USB-200 module (they directly plug
together). Please note that no calibration is required for this sensor.
4. Open the NeuLog application by clicking on the shortcut on the screen.
5. Once a colorimeter sensor module box appears on the left side of the screen the
probe has been automatically identified and you can begin experimentation.
6. If the colorimeter sensor is not automatically identified then click the “Search for
sensors” icon to find the sensor.
7. Select the “On-line experiment” icon; this will open a graph below.
8. Click on the “Module setup” button located in the colorimeter sensor module box
to change the sensor's settings if need be.
9. Click on the "Experiment setup" icon to change the experiment settings if need
be (experiment duration for example).
10. The colorimeter sensor will give a live reading in the module box to the left of the
screen while plugged in.
11. To run an experiment and collect data click the “Run experiment” icon.
12. To end data collection early, click the “Stop experiment” icon.
Tablet, smart phone device
Materials needed:
NUL-219 Colorimeter Sensor
Cuvettes + Cuvette Lids
WIFI-201 WIFI Module
BAT-200 Battery Module
Your colorimeter sensor needs to be connected to a WIFI-201 module. The WIFI-201
module will create a closed NeuLog WiFi network which will stream the NeuLog data
to a device of your choosing. Once your device is wirelessly connected to the NeuLog
network, you can run experiments and collect data through a browser of your
choosing.
Procedure:
1. Connect the colorimeter sensor directly to the left side of a WIFI-201 module (no
wires required).
2. Connect a BAT-200 module to the right side of the WIFI-201 module.
3. Although not required, we recommend plugging the BAT-200 to an outlet using a
USB to mini USB charger (such as a typical cell phone charger). The WIFI-201
module will run for 60-90 minutes (depending on the sensor) without being
plugged in.
4. The WIFI-201 can be powered directly using a mini to USB cord and plugging it
into your computer or a wall charger. Please note this function is only available
on WIFI-201 modules with the USB icon on the cover.
5. For further WIFI-201 instructions or the WIFI-201 quick start guide please visit:
www.NeuLog.com; Choose "Products" on the main menu,
then choose "WiFi Communication module".
6. The WIFI-201 indicator lights will flash; take no action until the LED to the far left
turns blue; this can take up to one minute.
7. Take your tablet or smart phone, go to the WiFi settings and select the NeuLog
network (NeuLog0184 for example) which matches the ID found on the back of
the WIFI-201 module (0184 for example).
8. Give your device about 20 seconds to connect to the WIFI-201.
9. Once the device is connected, go to your browser and type the website
wifi201.com into the URL bar, then wait for 30-60 seconds.
10. You will see a “Control mode” icon in the application, click on it.
11. The application will then load a new screen and begin to auto detect the
sensor(s); this can take a minute. (You can stop the search when the sensor is
found.)
NEULOG COLORIMETER LOGGER SENSOR GUIDE
12. If the application does not auto detect the sensor(s), select “Search for sensors”.
13. Once the sensor is found you will see a module box on the left side of the screen
for the colorimeter sensor. In this box, real time data will be displayed.
14. Click on the “Module setup” button located in the colorimeter sensor module box
to change the colorimeter probe settings if need be.
15. Click on the “Experiment setup” icon to change the experiment settings if need
be (Experiment duration for example).
16. To run an experiment and collect data, click the “On-Line” experiment icon, then
click the "Run experiment" icon.
17. If you have a newer generation WIFI-201 module (with the USB icon on the
cover) this can be used as a USB-200, meaning you can tether the probes
directly to the computer using the WIFI-201. To engage operation in this fashion,
press 3 times on the panel button. Repeating this operation will return the
module to WiFi mode.
Operation with NeuLog Graphic Display Module (viewer)
Materials needed:
NUL-219 Colorimeter Sensor
Cuvettes + Cuvette Lids
VIEW-101 Graphic Display Module
BAT-200 Battery Module
Procedure:
1. Connect the colorimeter sensor to the left side of the viewer.
2. Connect the BAT-200 module to the right side of the viewer.
3. The VIEW-101 will auto detect the sensor and you will see it appear on the left
side of the screen.
4. Once the sensor appears it will be monitoring data in real time.
5. To run an experiment and collect data press the run icon (little green person).
Off-line experiments
(Off-line experiments are when you do not have a sensor connected directly to a
computer, tablet, smartphone, or NeuLog viewer).
Materials needed:
NUL-219 Colorimeter Sensor
Cuvettes + Cuvette Lids
BAT-200 Battery Module
Materials needed to configure your offline experiment:
USB-200 Module, WIFI-201 module or VIEW-101 Module
USB to mini USB cable (included with the USB-200 module)
Procedure:
1. To determine the experiment settings, first connect to your chosen device (PC,
Mac, tablet, smart device) and click on the “Off-line experiment” icon after the
sensor search has ended. (For more information on how to change the
experiment settings, view the quick start procedure section for your chosen
device in this document.)
2. Connect the colorimeter sensor directly to the left side of a charged BAT-200
module.
3. When ready to collect data press the “Start/Stop” button on the colorimeter
sensor, a red light will turn on for the duration of the experiment.
4. After the experiment has concluded (the “Start/Stop” button is pressed again or
the data collection period ends) connect the sensor to your chosen device (you
can find full instructions in each device’s quick start procedure section).
5. Open NeuLog application.
6. Click the “Off-line experiment” icon.
7. Click the “Load data from sensors” icon.
8. Select which experimental data to upload (5 experimental runs can be stored at
one time on the NUL-219 colorimeter sensor).
Restoring sensor’s factory default settings:
Procedure:
1. Connect the USB-200 to a PC, Mac, tablet, or smart device.
2. Connect the colorimeter sensor to a USB-200 module (they directly plug
together).
3. Open the NeuLog application.
4. Click the “Tools” icon.
5. Click the “Restore sensor’s factory defaults” on the menu.
NEULOG COLORIMETER LOGGER SENSOR GUIDE
Included with the sensor:
Three plastic cuvettes with three lids
NeuLog NUL-219 Colorimeter Sensor Guide (This Document)
Sensor's specifications
Transmittance
Absorbance
Range and
operation modes
0-100%
Red, green, blue,
and orange
0-4
Red, green, blue,
and orange
ADC resolution
14 bit
Resolution
0.02%T
0.01 abs.
Max sample rate
(S/sec)
100
Experiment Duration: 1 second to 31 days.
Sensor's features:
Fully digital data
Rugged plastic ergonomic case
Push button switch for Start/Stop experiments in off line mode
Red, green, blue and orange LEDs for four different wavelength tests.
Testing in both transmittance and absorbance
Internal photodiode for detecting light

Note: NeuLog products are intended for educational use.
Videos and experiment examples:
Videos, literature and other probes can be found at www.NeuLog.com.
In order to access the colorimeter sensor's page, choose "Products" on the main
menu and then "Colorimeter logger sensor".
In order to find the science experiments, choose "Experiments" on the main
menu of the NeuLog website, then choose "Physics", "Chemistry", "Biology" or
"Environmental Science".
Chemistry Experiments:
oBeer-Lambert law (C-28)
oColor Absorption (C-31)
Technical background:
The philosophy behind NeuLog’s plug and play technology is based on each sensor’s
ability to store its own data due to an internal flash memory chip and micro-controller
in each plastic NeuLog body. This technology allows the sensor to collect and then
store the digital data in the correct scientific units (°C, °F, Lux, %, ppm, for example).
The sensor is pre-calibrated at the factory. The built-in software in the logger can be
upgraded for free at any time using the provided firmware update.
Inside the sensor’s plastic body there are four LED lights (red, green, blue and
orange) to test various wavelengths.
Light is transmitted from four different colored LEDs inside the sensor's body, one
color at a time, passed through the cuvette containing a solution and then detected at
the other end to determine the absorbance or transmittance of light at a specific
wavelength.
The sensor works off of Beer’s Law which is explained in the section “Beer’s Law”
below.
The four wavelengths emitted are:
Red: 634 - 662 nm
Blue: 452 - 476 nm
Green: 500 - 530 nm
Orange: 595 - 625 nm
Beer’s Law:
Beer’s Law is an extremely useful concept for solution concentration testing in the
fields of Chemistry and Biology.
Standard curves are created by using solutions of known concentrations which are
analyzed in a colorimeter to determine the absorbance and then plot a graph. The
absorption for unknown samples can easily be tested and plotted on the standard
curve to determine their concentration.
NEULOG COLORIMETER LOGGER SENSOR GUIDE
The law states that there is a logarithmic dependence between the transmission T of
light through a substance, the product of the absorption coefficient of the substance α,
and the distance the light travels through the material l.
For liquids, these relations are usually written as:
T = I/I0= 10-αl
where I0and I are the intensity of the incident light and the transmitted light,
respectively.
The transmission (or transmittance) is expressed in terms of absorbance which, for
liquids, is defined as
A = -log10(I/I0) = -log10(T)
This implies that the absorbance becomes linear with the concentration (or number
density of absorbers) according to A=αl
Maintenance and storage:
Never submerge the NeuLog plastic body in any liquid.
Never place anything in the colorimeter sensor’s opening on the faceplate except
the approved cuvettes.
Do not allow liquid into the NeuLog plastic body.
After using the probe wipe off all excess material, liquid or residue from the
colorimeter probe.
Store in a box at room temperature out of direct sunlight.
The cuvette handling and storage guide can be found at the beginning of this
document.
Warranty:
We promise to deliver our sensor free of defects in materials and workmanship. The
warranty is for a period of 3 years from the date of purchase and does not cover
damage of the product caused by improper use, abuse, or incorrect storage. Sensors
with a shelf life such as ion selective probes have a warranty of 1 year. Should you
need to act upon the warranty, please contact your distributor. Your sensor will be
repaired or replaced.
Thank you for using NeuLog!
Flexible, simple, fast, forward thinking.
W: www.neulog.com
E: [email protected]m
A: 850 St Paul Street, Suite 15, Rochester, NY 14605
P: 1.866.553.8536
V11042014

Other NeuLog Accessories manuals

NeuLog NUL-215 User manual

NeuLog

NeuLog NUL-215 User manual

NeuLog NUL-216 User manual

NeuLog

NeuLog NUL-216 User manual

NeuLog NUL-224 User manual

NeuLog

NeuLog NUL-224 User manual

NeuLog NUL-209 User manual

NeuLog

NeuLog NUL-209 User manual

NeuLog NUL-233 User manual

NeuLog

NeuLog NUL-233 User manual

NeuLog NUL-211 User manual

NeuLog

NeuLog NUL-211 User manual

NeuLog NUL-213 User manual

NeuLog

NeuLog NUL-213 User manual

NeuLog NUL-219 User manual

NeuLog

NeuLog NUL-219 User manual

NeuLog NUL-214 User manual

NeuLog

NeuLog NUL-214 User manual

NeuLog NUL-236 User manual

NeuLog

NeuLog NUL-236 User manual

NeuLog NUL-233 User manual

NeuLog

NeuLog NUL-233 User manual

NeuLog NUL-233 User manual

NeuLog

NeuLog NUL-233 User manual

NeuLog NUL-215 User manual

NeuLog

NeuLog NUL-215 User manual

NeuLog NUL-214 User manual

NeuLog

NeuLog NUL-214 User manual

NeuLog NUL-232 User manual

NeuLog

NeuLog NUL-232 User manual

NeuLog NUL-227 User manual

NeuLog

NeuLog NUL-227 User manual

NeuLog NUL-204 User manual

NeuLog

NeuLog NUL-204 User manual

NeuLog NUL-218 User manual

NeuLog

NeuLog NUL-218 User manual

NeuLog NUL-235 User manual

NeuLog

NeuLog NUL-235 User manual

NeuLog NUL-229 User manual

NeuLog

NeuLog NUL-229 User manual

NeuLog NUL-230 User manual

NeuLog

NeuLog NUL-230 User manual

NeuLog NUL-211 User manual

NeuLog

NeuLog NUL-211 User manual

NeuLog NUL-212 User manual

NeuLog

NeuLog NUL-212 User manual

NeuLog NUL-246 User manual

NeuLog

NeuLog NUL-246 User manual

Popular Accessories manuals by other brands

Makita CJ105D instruction manual

Makita

Makita CJ105D instruction manual

WIKA MH-3 operating instructions

WIKA

WIKA MH-3 operating instructions

Hydronix Hydro-Probe SE user guide

Hydronix

Hydronix Hydro-Probe SE user guide

Aginova Sentinel PRO user guide

Aginova

Aginova Sentinel PRO user guide

Petzl OWALL M41 quick start guide

Petzl

Petzl OWALL M41 quick start guide

NOVAK PRO SC Setup guide

NOVAK

NOVAK PRO SC Setup guide

Kathrein ESD 64 manual

Kathrein

Kathrein ESD 64 manual

Spector T1039 user manual

Spector

Spector T1039 user manual

HB Products HBX Vapor Quality sensor Installation and configuration guide

HB Products

HB Products HBX Vapor Quality sensor Installation and configuration guide

Sunbeam Sleep warm. Sleep well. user manual

Sunbeam

Sunbeam Sleep warm. Sleep well. user manual

SICK WSE26 operating instructions

SICK

SICK WSE26 operating instructions

ETCR 016 user manual

ETCR

ETCR 016 user manual

Cardinal 210 Installation and technical manual

Cardinal

Cardinal 210 Installation and technical manual

EDS i1 Series user manual

EDS

EDS i1 Series user manual

IFM Electronic PA32xx Series installation instructions

IFM Electronic

IFM Electronic PA32xx Series installation instructions

SICK W4F operating instructions

SICK

SICK W4F operating instructions

LuxaFlex Base Lite Instruction manual for use and maintenance

LuxaFlex

LuxaFlex Base Lite Instruction manual for use and maintenance

X10 HawkEye MS13A Set up and operating instructions

X10

X10 HawkEye MS13A Set up and operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.